17 — Intervals

How long a calibration lasts.

+ The questionHow long does a calibration stay true, and who decides?

A date the owner writes

The due date on a calibration label is the owner's prediction, not a result of the calibration. The laboratory measures the instrument on one day, and its results describe the instrument on that day only. The laboratory cannot know how the instrument will be used, stored or carried afterward, so the owner, who does know, chooses the interval. Choosing it means predicting how long the instrument will stay within its limits, or, in OIML and ILAC's words, determining "the maximum period that should be permitted between successive calibrations".

An accredited laboratory accordingly leaves the interval to its customer. Under a clause of ISO/IEC 17025, as NIST, SADCAS and UKAS each describe it, a certificate or label carries no recommendation on the calibration interval unless the customer agreed it or a legal regulation requires one. UKAS asks a laboratory that does print a due date to show whose date it is, with a phrase such as "Customer requested calibration due date."

A US device maker sets its own calibration intervals, now under ISO 13485, and until 2026 it also had to show each due date by law. FDA's former rule on inspection, measuring and test equipment required records that included "the next calibration date", displayed on or near each piece of equipment or readily available. FDA replaced that rule with its Quality Management System Regulation, the QMSR, which took effect on 2 February 2026, brings in ISO 13485:2016 by reference and so makes the standard's requirements part of the regulation.

What changes between calibrations

A calibration stays true only as long as the instrument does not change, and every part of the monitor's chain can change. Semiconductor pressure sensors drift strongly with temperature, as Omron's engineers wrote in a patent filed in 1980, so the monitor in that patent zeroes its sensor before each measurement. The span can drift too, and so can the tightness of the valves and tubes, and the reference manometer itself.

Drift also continues with time, and use wears the monitor. The zero of a pressure sensor "may initially change 0.1% annually and can become progressively worse over time", as engineers at Baker Hughes wrote in a 2022 patent on industrial sensors, not blood-pressure monitors. OIML asks that an automated monitor's cuff-pressure reading change by no more than 3 mmHg after 10,000 simulated measurement cycles, a test of wear rather than of time.

Four makers recommend intervals for their own instruments, and none of them states how it chose its interval. Microlife recommends that its BP B3 Basic home monitor "is tested for accuracy every 2 years or after mechanical impact", through its own service. Welch Allyn recommends calibrating its Ambulo 2400 at least once a year, Mindray recommends checking the VS-900's blood-pressure channel at least once a year, and Rossmax asks for its CalPro to be recalibrated every two years (Chapter 14). The owner therefore cannot tell what evidence each figure rests on. For some instruments, India's rules fix the interval by law (Chapter 30).

How an interval is chosen

In the international guidance, the owner chooses an interval by judging risk, not by applying a fixed rule. OIML and ILAC publish that guidance jointly as OIML D 10 and ILAC G24, in an edition of 2022. Staff with the technical competence for it make the initial decision, which "is based mainly on a risk assessment analysis". The factors the staff weigh, as OIML and ILAC list them, include the "risk of the measuring equipment exceeding the predetermined limits", the uncertainty needed, the maker's recommendation, wear and drift, how heavily and where the equipment is used, and the results of checks between calibrations. The staff also weigh cost against uncertainty, because frequent recalibration costs more, and "these costs need to be balanced against increased measurement uncertainties". Fixed intervals, in OIML and ILAC's words, "are not recommended unless the interval has been specified in a normative document", such as a standard or a regulation.

Once results accumulate, the laboratory reviews the interval by one of five methods that OIML and ILAC describe. In the staircase method, the laboratory extends or keeps the next interval when the deviation found at calibration lies within a defined share of the maximum permissible error, and shortens it otherwise. A control chart, a plot of each calibration's result against the date, follows the results over calendar time, and that method suits only equipment that drifts. In the other methods, the laboratory counts hours of use, checks critical parameters often, "once a day" in one example, or applies other statistics. OIML and ILAC give no step sizes and no percentages, and they leave each laboratory free to use any of the methods or none.

NIST does give numbers, for mass standards, in GMP 11, its good-practice procedure for US legal-metrology laboratories, which check instruments that the law regulates. A laboratory's working standards that it calibrates itself, and its balances, start at 12 months, and pairs of reference kilograms calibrated at NIST start at 48 months. No interval may exceed 10 years without exceptional analysis of the laboratory's data. The laboratory's technical manager changes an interval on documented evidence, and an initial interval ends early when a proficiency test fails, or when results on a control chart fall outside its warning limits, the lines drawn on either side of the expected value, with no correctable cause. In NIST's scheme as in OIML's, the interval is a hypothesis that each calibration tests.

The monitor's own zero

Between calibrations, the monitor corrects the largest part of its sensor's error every time it starts. Before measuring, the firmware reads the sensor with the cuff at the pressure of the open air and takes that reading as its zero, a step called auto-zero. In the method of Omron's patent filed in 1980, the monitor leaves its sensor "exposed to the atmosphere", stores the reading as a correction value and subtracts it from every later reading. The capture ends "After a certain period of time has elapsed", and no duration is stated. The same method follows slow drift during use. The monitor sets a second value about 1 mmHg above the stored zero, and if the reading rests between the two for a set time, 10 seconds in the patent's example, the higher value becomes the new zero. Because the second value lies above the zero, the method corrects only upward drift.

In Omron's later patents the monitor zeroes its sensor at power-on, and India now requires the step by law. In an Omron patent granted in 2015, pressing the power switch starts "the 0 mmHg correction of the pressure sensors", and in one of 2005 the monitor completes "zero setting, which is to say initial reset" at the third step of its flowchart. In Shenzhen CHK's 2010 patent application, the monitor likewise performs "the zero point offset correction" when the sensor is turned on. Under India's rules for automated sphygmomanometers, notified on 28 July 2025, as under OIML's recommendation, the monitor "shall be capable of automatic zero adjustment", at least when it is powered on. It must zero again or switch itself off when the transducer's output drifts by one scale interval, 1 mmHg, or more. None of these documents states how long the zero takes to settle, or how large an offset the firmware accepts before it reports a fault.

Auto-zero is an adjustment in the VIM's sense (Chapter 14), and its strength and its limit come from the same fact: it compares the sensor with one pressure only, that of the open air. An offset is the same at every pressure, so one reading at zero measures it completely. An error of span grows with the reading and vanishes at zero, so the zero cannot see it at all.

43 — What the monitor's zero can and cannot see
cuff at the pressure of the open air time zero at least at power-on (India's rules, 2025; OIML R 149-1) power-on inflation begins settling time: not stated in any source found raw sensor output reading = output − stored zero stored as the zero stored zero AN OFFSET 0 150 300 ERROR, mmHg CUFF PRESSURE, mmHg offset: the same at every pressure the zero sees it all A CHANGE OF SPAN 1 %, illustrative 0 150 250 300 ERROR, mmHg CUFF PRESSURE, mmHg 1.5 mmHg 2.5 mmHg at zero a span error is nothing: only a known pressure finds it bare sensor at 150 mmHg: about 26 mmHg; after auto-zero: 11.9 mmHg, the span about 90 % of its total of squares (worked from Chapter 11's rows) Sources: Omron patent JPS5834159B2 (filed 1980, published 1983; manufacturer); G.S.R. 498(E), 28 Jul 2025; OIML R 149-1:2020. Time line and span line schematic, 1 % illustrative.
The quantity being measuredThe signal in the chainNoise and errorReadingFocal detail
Plate 43 — Auto-zero measures the offset completely, because an offset is the same at every pressure. A change of span vanishes at zero, so the zero never sees it, and only a comparison at a known pressure can.
Solved example: what the zero removes, and what only a known pressure can find

The bare sensor's six rows at 150 mmHg, as standard uncertainties from Chapter 11, are the offset 20.6 mmHg, the temperature effect on the offset 10.8 mmHg, the span 11.3 mmHg, the nonlinearity 2.8 mmHg, the temperature effect on the span 2.6 mmHg and the hysteresis 0.43 mmHg.

  • Squared, they are 424.4, 116.6, 127.7, 7.8, 6.8 and 0.2 mmHg², a total of 683.5 mmHg², whose square root is about 26 mmHg.
  • Auto-zero removes the two offset rows, 424.4 and 116.6 mmHg², which add to 541.0 mmHg², 79 % of the total, since 541.0 divided by 683.5 is 0.79. The four rows left add to 142.5 mmHg², and the square root is 11.9 mmHg.
  • The span supplies 127.7 of the 142.5, about 90 % of what is left, since 127.7 divided by 142.5 is 0.90.
  • Suppose, as an illustration, that the span changes by 1 % after a calibration. At zero the error stays 0 mmHg, so the next auto-zero finds nothing to correct. At 150 mmHg the reading moves by 1 % of 150, or 1.5 mmHg, and at 250 mmHg by 2.5 mmHg.

A monitor that zeroes itself at every power-on therefore still needs a comparison at a known pressure, by calibration or by a check between calibrations. After the zero, almost all of the remaining error is span, and only a known pressure reveals it, so the interval of a self-zeroing monitor has to be chosen for its span, which no zero can detect.

Checks between calibrations

Between full calibrations, a laboratory or a factory checks that nothing has moved. Intermediate checks are comparisons made between two calibrations to confirm that an instrument still works and keeps its calibration status. A laboratory "should have an appropriate system of intermediate checks", in OIML and ILAC's words, and the results of those checks should fall within predetermined limits before the equipment is approved for further use. Under ISO/IEC 17025, as SADCAS describes it, such checks follow a procedure where they are needed. A reference manometer can be compared with a second gauge at one pressure. For the monitor, a check at zero and at one pressure tests both the offset and the span, and in India's test procedure for automated sphygmomanometers, a reading at 50 mmHg taken after switch-on is checked for drift.

A failed check calls into question every result since the last good one. An air-quality laboratory at the University of California, Davis, applies that rule with numbers, although it checks balances rather than pressure. Under its procedure of July 2022, the laboratory weighs reference weights at the start of work and every four hours, with an acceptance limit of ±0.003 mg, and processes no samples until the check passes. When a check fails, the laboratory weighs again every filter it weighed since the last check that passed.

44 — Between two calibrations
LIMIT ±3 mmHg −4 −3 −2 −1 0 +1 +2 +3 +4 ERROR AT 150 mmHg, mmHg 0 6 12 18 24 MONTHS SINCE CALIBRATION drift of the span calibration (as left) calibration (as found) Microlife recommends a test every 2 years or after mechanical impact check at zero: reads 0, sees nothing check at one pressure: sees the drift the check that catches the span before the calibration does a failed check calls into question every result since the last good one (Chapter 26) Interval: Microlife BP B3 Basic manual (manufacturer). Drift line and check results schematic.
References and standardsNoise and errorLimits and decisionsReadingFocal detail
Plate 44 — The interval is a prediction, and checks between calibrations test it. A check at zero cannot see a drifting span; a check at one pressure can, months before the next calibration.

What the quality system asks

Under ISO 13485, the quality-system standard for device makers, a maker calibrates or verifies its monitoring and measuring equipment at specified intervals, or before use, against standards traceable to international or national measurement standards, as FDA's training for auditors and other open sources describe the standard's clause. Where no such standard exists, the maker records the basis it used. Japan's quality-system rule for device makers, Ministerial Ordinance 169, set out the same two points in its Article 53: equipment is "calibrated or verified at the specified intervals", and "in case where no such standards exist, the basis used for the calibration or verification shall be recorded". That English text, quoted in a trade article in 2012, is of the version that matched the standard's 2003 edition. Japan revised the ordinance's second chapter, which holds Article 53, in March 2021 to align it with ISO 13485:2016.

Software used for monitoring and measurement is also validated for its intended use, according to FDA's training. When equipment is found not to conform, the maker assesses and records the validity of the earlier results and acts on the equipment and on any product affected (Chapter 26).

Setting and defending an interval
  • Set each instrument's interval from risk: how close its errors run to the limit, how it drifts, how it is used and what its maker recommends, and record the reasons.
  • Start from the maker's figure where one exists, such as 2 years for Microlife's home monitor or at least once a year for Welch Allyn's and Mindray's monitors, and adjust it from the as-found results.
  • Write the due date on the label in-house and record whose date it is, since an accredited laboratory does not print one unless asked.
  • Check the monitor between calibrations at zero and at one pressure, because auto-zero removes the offset but cannot see a change of span.
  • When a check fails, list every result since the last good check before deciding what to do with them (Chapter 26).
+ What this chapter established
  • An accredited laboratory's certificate and label carry no due date unless the customer agrees one or a regulation requires one, so the interval is the owner's prediction of how long the instrument stays within its limits.
  • Under OIML D 10 and ILAC G24, competent staff set intervals by judging risk, fixed intervals are not recommended unless a normative document sets one, and five review methods are described without numbers.
  • The monitor zeroes its sensor against the open air at least at power-on, and India's and OIML's rules require it to zero again or switch off when the drift reaches 1 mmHg.
  • Auto-zero removes the two offset rows, 79 % of the bare sensor's total of squares, but the span supplies about 90 % of the total of squares that remains, whose square root is 11.9 mmHg, and only a comparison at a known pressure can find it.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.