+ Part II · The signal chain
Where error enters.
Between a physical quantity and the number on a display lie six stages, each named by what it does: sense, condition, digitize, compute, connect, decide. In the blood-pressure monitor, the sensor turns the air pressure in the cuff into a few millivolts, the electronics amplify and filter them, a converter turns them into counts, and firmware turns the counts into millimeters of mercury on the display and in the stored record. Each stage can add an error, and noise is not a stage of its own but something that every stage adds. This part follows the cuff-pressure channel through the first five stages: sensing in Chapter 4, noise in Chapter 5, conditioning in Chapter 6, sampling and quantization in Chapters 7 and 8, and computing and connecting in Chapter 9. The sixth stage, the decision, comes in Chapter 13, once Part III has stated how sure each reading is.
From a quantity to a signal.
+ The questionHow does a pressure, a temperature or a light level become a voltage, and what else comes in with it?
A pressure becomes a few millivolts
Between an empty cuff and a cuff at 40 kilopascals, which is 300 mmHg, the sensor in Texas Instruments' reference design for a blood-pressure monitor changes its output by only about 30 millivolts (Chapter 1). A millivolt is a thousandth of a volt. The sensor is Omron's 2SMPP-03, and TI drives it with a constant current of 100 microamperes, written 100 µA, a ten-thousandth of an ampere. Omron gives the sensor's bridge of resistors a resistance of 20 kilohms, written 20 kΩ. A voltage is the current times the resistance, so 100 µA through 20 kΩ sets 2 V across the bridge. The 30 mV is about fifteen thousandths of those 2 V, since 30 divided by 2,000 is 0.015. Every later stage of the monitor works with this small change, and every error the sensor adds travels with it to the display.
Like every sensor, the 2SMPP-03 is a transducer, a device that turns one physical quantity into another, here a pressure into a voltage. The transducer sets how large the signal is, and also which other quantities enter the measurement alongside the one intended. The force sensor of a weighing instrument is a transducer too: OIML, the intergovernmental body that writes model rules for measuring instruments, defines a load cell as a "measuring transducer that will produce an output in response to an applied load".
The cuff is part of the sensor
The sensor never touches the blood, so the cuff, its tube and the air inside them belong to the sensing system as much as the chip does. A pump fills the cuff around the upper arm with air, and a tube carries the pressure of that air to the sensor inside the monitor (Chapter 1). The 2SMPP-03 measures gauge pressure: it compares the pressure on one side of its chip with the pressure of the surrounding air on the other, so its reading is the cuff's pressure above the pressure of the room. Omron rates the sensor for use with air only, so air must carry the pressure from the cuff to the chip, and the chip can respond only to what the air brings it.
The air also carries a much smaller signal at the pace of the heart. The artery under the cuff swells a little with every heartbeat and pushes on the cuff, so the air pressure rises and falls slightly with each beat. These small rises and falls are the oscillations. They ride on the much larger, slowly falling pressure of the cuff, and together the two make the oscillometric signal. Typical oscillations measure less than 1 mmHg to 3 mmHg, by the figures that Chua and Siew, two engineers at the chip maker Freescale, give in a 2005 application note on a digital blood-pressure meter. The cuff pressure is the monitor's reading, and the systolic and diastolic values are the estimate that its firmware derives from the oscillations (Chapter 9).
A bridge of four resistors
The chip turns pressure into a voltage through piezoresistance, the change in a material's electrical resistance when the material is strained, that is, stretched or compressed. The 2SMPP-03 is a MEMS sensor, short for microelectromechanical system, in which the moving part and the resistors are made in the same piece of silicon. Pressure strains the silicon, and the strain changes the resistances. Four of the resistors are joined in a square called a bridge. The current enters at one corner and leaves at the opposite one, so it divides between two paths of two resistors each, and the other two corners are the outputs. When the four resistances are equal, each path divides the voltage across the bridge in half, so both outputs sit at the same voltage and the voltage between them is zero.
A bridge responds to a difference between its resistances and ignores a change that all four share. Pressure changes the resistances unequally, the bridge loses its balance, and a small voltage appears between the outputs. A change that moves all four resistances by the same fraction leaves each path dividing the voltage in half, so the bridge stays balanced and that change adds nothing to the difference. Both outputs stay close to 1 V, half of the 2 V across the bridge, while the signal is only their difference of a few millivolts. The amplifier that follows must measure that difference of millivolts while ignoring the 1 V that both outputs share (Chapter 6).
The current or voltage that powers a bridge is its excitation, and Omron states every output figure for the 2SMPP-03 at an excitation of 100 µA, the current that TI's design supplies. The offset is the output with no pressure applied, typically −2.5 mV, and one sensor's offset may lie anywhere within 4.0 mV of that value. The span is the output at full scale minus the output at zero, typically 42.0 mV for the sensor's full scale of 50 kPa, which is about 375 mmHg. A sensor's sensitivity is the change in its output for a given change in the quantity it measures, and it differs from a diagnostic test's sensitivity, which Chapter 24 explains. For the 2SMPP-03 it is the span divided by the range, 42.0 mV over 375 mmHg, or 112 µV for each millimeter of mercury, where a microvolt is a millionth of a volt.
The span is 42.0 mV over 50 kPa. One millimeter of mercury is 133.322 Pa, so 50,000 Pa divided by 133.322 Pa is 375.0 mmHg, and the sensitivity is 42.0 mV divided by 375.0 mmHg, which is 0.112 mV, or 112 µV, for each mmHg.
- At a cuff pressure of 150 mmHg the pressure adds 150 times 112 µV, which is 16.8 mV, to the output. With the typical offset of −2.5 mV, the bridge's output is 16.8 minus 2.5, about 14.3 mV.
- Over the range that Chua and Siew give, an oscillation of 1 mmHg adds 112 µV to the output, and one of 3 mmHg adds 3 times 112 µV, which is 336 µV.
- TI expects oscillations of 100 to 300 µV at its amplifier's input, measured from peak to peak, that is, from the lowest to the highest point of each oscillation. Divided by 112 µV for each mmHg, that is about 0.9 to 2.7 mmHg.
Freescale's application note of 2005 and TI's design of 2023, written 18 years apart about different sensors, therefore agree on the size of the oscillations. At 150 mmHg, oscillations of 1 to 3 mmHg are 1 part in 150 to 1 part in 50 of the steady pressure, about 0.7 % to 2 %. The electronics must resolve a change of about a hundred microvolts riding on about 14 mV, so the noise they add must stay well below a hundred microvolts (Chapter 5). For the same reason the monitor sends the steady pressure and the oscillations along two separate paths (Chapter 6).
How other transducers compare
A bridge can be powered by a current or by a voltage, and its maker states its output per unit of whichever excitation it uses. The 2SMPP-03 takes a current, so Omron states its span at 100 µA. TDK's C32/2 takes a voltage. The C32/2 is a die, a bare silicon chip that carries a bridge, and TDK sells it for medical devices among other uses. Its sensitivity is ratiometric to the supply, as TDK states it, which means that its output grows in proportion to the supply voltage (Chapter 6). TDK gives that output per bar, a unit of pressure equal to 100 kPa. HBK, a maker of load cells, states the rated output of its PW6D in millivolts for each volt of excitation, because the PW6D takes a voltage too. The table adds two transducers of other kinds, a platinum resistance thermometer and a photodiode, with figures from TI's guide to such thermometers and from an article on low-light measurement by Analog Devices, a maker of amplifiers and converters.
| Transducer | What changes | A maker's figure | Share of the excitation at full scale |
|---|---|---|---|
| Silicon pressure bridge, Omron 2SMPP-03 | resistance, under strain | 42.0 mV over 50 kPa at 100 µA | about 21 thousandths of 2 V |
| Silicon pressure die, TDK C32/2 | resistance, under strain | 215 mV per bar at 5 V | about 17 thousandths over 0.4 bar |
| Strain-gauge load cell, HBK PW6D | resistance, under strain | 2.0 mV per volt at full load | 2 thousandths, 10 mV at 5 V |
| Platinum thermometer, Pt100 | resistance, with temperature | 100 Ω at 0 °C | read with a current |
| Photodiode | light makes a current | 0.5 A per watt of light | read with an amplifier |
The two silicon sensors, from different makers, put out a similar share of their excitation at full scale. Omron's span of 42.0 mV is about 21 thousandths of the 2 V across its bridge, and TDK's 215 mV per bar over a range of 0.4 bar is 86 mV, about 17 thousandths of its 5 V supply. HBK's 2.0 mV for each volt is 2 thousandths, or 10 mV at 5 V. The silicon sensors' share is therefore about 8.5 to 10.5 times the load cell's at its full load, since 17 divided by 2 is 8.5 and 21 divided by 2 is 10.5, so for the same excitation these silicon bridges give the amplifier the larger signal. Not every monitor uses a bridge. In one Omron Healthcare patent, granted in 2013, the pressure sensor's capacitance changes with pressure, and an oscillator circuit turns that capacitance into a frequency for the processor to read.
The thermometer and the photodiode each need a circuit to turn what they sense into a voltage. The Pt100 is a platinum sensor of 100 Ω at 0 °C. Its resistance rises by about 0.39 Ω for each degree: the standard IEC 60751 sets a coefficient of 3.9083 thousandths for each degree, as TI reports it, and 3.9083 thousandths of 100 Ω is about 0.39 Ω. A current turns that resistance into a voltage, and the current also heats the sensor (Chapter 6). In a photodiode, light generates a current, and an amplifier turns the current into a voltage through a resistor whose value sets the gain. In an example by Analog Devices, a photocurrent of up to 125 nanoamperes, billionths of an ampere, passes through a resistor of 80 megohms, written 80 MΩ, and gives up to 10 V, since 125 nA times 80 MΩ is 10 V.
What else comes in with the pressure
A sensor responds to more than the quantity it is meant to measure, and any other quantity that changes its output is an influence quantity. For the 2SMPP-03, a silicon bridge, the largest in Omron's figures is temperature. Omron names another: the board the sensor is mounted on may influence its output, so mounting is an influence quantity too.
The 2SMPP-03's offset can move with temperature by several times the accuracy that a monitor must meet. Omron states how far the offset and the span move between 0 and 85 °C, relative to 23 °C, in percent of full scale, written %FS: a share of the whole 375 mmHg range rather than of the reading. At 85 °C, where the change is widest, a typical sensor's offset has moved by 2.0 %FS, and any one sensor's offset may have moved by anywhere from −3.0 to 7.0 %FS, which Omron sums up as ±5.0 %FS about the typical change. The same sensor's span may have moved by −2.0 to 4.0 %FS, which Omron sums up as ±3.0 %FS about a typical change of 1.0 %FS. Omron does not state that the sensor compensates for temperature. The largest offset change in Omron's figures, 7.0 %FS, is 7.0 % of 375 mmHg, about 26 mmHg, more than eight times the ±3 mmHg to which monitor makers such as Microlife and A&D state their cuff-pressure accuracy (Chapter 8). Even Omron's summary figure of 5.0 %FS is about 19 mmHg, so a monitor built on this sensor must compensate for temperature or correct for it.
A warming of 10 K can move a reading by a monitor's whole accuracy limit, as TDK's figures for its die show. TDK gives the C32/2's span a typical temperature coefficient of −2.1 thousandths per kelvin, where a kelvin is the same step as a degree Celsius. Over 10 K the span therefore falls by 10 times 2.1 thousandths, which is 21 thousandths, or about 2.1 %. In a monitor that used the die without compensation, the reading at a cuff pressure of 150 mmHg would fall by 2.1 % of 150 mmHg, about 3 mmHg, the whole of a ±3 mmHg limit.
The effect of temperature can be compensated on the sensor's chip, corrected in firmware or, for the offset, removed by zeroing. Where the sensor is not compensated on its chip, the monitor's firmware can correct the reading from a measured temperature (Chapter 9). Zeroing the sensor just before each measurement removes the offset's share of the effect, as long as the sensor's temperature holds steady while the measurement lasts.
Omron's typical span of 42.0 mV is the center of a band from 36.5 to 47.5 mV, 5.5 mV or about 13 % either side, since 5.5 divided by 42.0 is about 0.13. A monitor whose firmware assumed the typical span would read correctly only with a sensor that happened to be typical, and with any other it could read up to 13 % high or low at every pressure. The maker therefore calibrates each monitor in its factory, and the monitor stores coefficients of its own (Chapters 9 and 14). A design review checks every calculation against the minimum and maximum, not against the typical value alone.
Even at a steady temperature, the 2SMPP-03's output does not follow the pressure exactly, and Omron gives two figures for how far it departs. The output departs from a straight line by up to 1.3 %FS, which is 1.3 % of 375 mmHg, about 4.9 mmHg, and that departure is the sensor's nonlinearity. At any one pressure, the output with the pressure rising and the output with it falling differ by up to 0.2 %FS, about 0.75 mmHg, and that difference is the sensor's hysteresis.
Omron gives no figure for the 2SMPP-03's repeatability, and none for its long-term stability, the drift of the output over months of use. TDK does give a long-term figure for its die. The long-term stability of the C32/2's offset is typically 0.15 % and at most 0.3 % of the die's normalized full-scale output of 120 mV, measured to TDK's own test standard. The typical 0.15 % of 120 mV is 0.18 mV. A bar, 100 kPa, is about 750 mmHg, so the die's typical sensitivity of 215 mV per bar is about 0.29 mV for each mmHg, and 0.18 mV is about 0.6 mmHg. An engineer who chooses Omron's sensor has no such figure from its maker and must establish one by testing. During the monitor's life, periodic checks against a reference must cover the drift that Omron does not state (Chapter 17).
- Convert every %FS figure into the display's unit over the full-scale range, and set each beside the accuracy the product must meet.
- Check whether the sensor is driven by a current or a voltage before computing its sensitivity, because its maker states the output per unit of that excitation.
- Look for a statement of temperature compensation on the data sheet, and where none appears, plan the correction in hardware or in firmware and give the error that remains after it a row of its own in the uncertainty budget that Chapter 11 describes.
- List what the data sheet does not give, such as repeatability or long-term stability, and plan a test or a second sensor's figures for each.
- Record each figure as the maker's, with the document's revision, in the design and development file.
- A transducer turns the measured quantity into another: the monitor's silicon bridge, driven by a constant 100 µA, turns the cuff's air pressure into about 112 µV for each millimeter of mercury, while a platinum thermometer and a photodiode need a circuit to turn a resistance or a current into a voltage.
- The cuff, its tube and its air belong to the sensing system, and the oscillometric signal they deliver carries oscillations of about 1 to 3 mmHg riding on the cuff pressure.
- A bridge puts out only the difference between its two outputs, so a change common to all four resistors cancels, while pressure appears as a few millivolts.
- Temperature is the sensor's largest influence quantity: uncompensated, the 2SMPP-03's offset may move by as much as 7.0 %FS, about 26 mmHg, at 85 °C, so the design must compensate or correct it.
Let's Talk