To calibrate each Alcotest breath-test machine in New Jersey, an officer warmed solutions of known alcohol content to about 34 °C and confirmed their temperature to within 0.2 °C with a thermometer traceable to NIST, the US national metrology institute. One coordinator was charged with skipping that check, and the machines he had calibrated had taken evidential breath samples from 20,667 people. A retired judge who heard the evidence for the state's Supreme Court accepted testimony that a solution one degree off could make a sample of 0.075 % alcohol read as 0.082 %. On 13 November 2018 the court ruled that breath-test results from machines calibrated without a NIST-traceable thermometer could not be used as evidence. This part follows such checks upward: what a calibration hands back, what makes a chain of calibrations traceable, why a stranger's certificate is believed, and how long a calibration lasts.

14 — Calibration

What calibration hands back.

+ The questionWhen a laboratory calibrates an instrument, what exactly does it return?

Two steps, and neither adjusts

Calibration is an operation in two steps, as the VIM, the vocabulary of metrology of Chapter 1, defines it, and neither step adjusts the instrument. In the first step a laboratory relates the values that its measurement standards provide to the indications of the instrument, each with its uncertainty. In the second step it uses that relation to obtain a measurement result from an indication. The first step therefore produces a record of pairs: at each test point, what the standard said and what the instrument showed. The second step turns that record into a rule for reading the instrument afterward.

A laboratory may express the relation as a statement, a calibration function, a diagram, a curve or a table, the forms the VIM lists. In some cases the relation is a correction of the indication with its uncertainty, either added to each indication or multiplying it. The first step alone is often taken for the whole calibration, according to a note in the VIM. A laboratory that returns only a table of errors has done only the first step and left the second to the user. In the monitor of the earlier chapters the firmware performs the second step. The coefficients of Chapter 9, which turn the converter's counts into millimeters of mercury, are the monitor's rule for obtaining a result from an indication.

Two other operations are often called calibration, and the VIM defines each of them separately. Adjustment of a measuring system is a set of operations that makes the system give prescribed indications for given values of the quantity, and its kinds include zero, offset and span adjustment, by the VIM's list. Verification is the provision of objective evidence that an item fulfills specified requirements, such as a maker's specification. Calibration measures the error, adjustment changes it, and verification judges it against a limit.

Calibration is "a prerequisite for adjustment", in the VIM's words, and after an adjustment the system "must usually be recalibrated". A calibration made before an adjustment and one made after it therefore describe two different instruments, and a record that keeps only the second has lost the evidence of how far the instrument had drifted before it was adjusted.

The monitor on the bench

For the monitor, the first step takes place on a bench. The tester replaces the cuff with a rigid metal vessel of 500 ml, within 25 ml, as OIML's test procedure for automated sphygmomanometers requires. The vessel holds a steady pressure because its volume cannot change, while a cuff wrapped round an arm moves with every pulse (Chapter 4). T-shaped connectors, called T-pieces, join the vessel to the monitor's pressure port, a hand pump and a reference manometer. The tester raises the pressure in steps of at most 50 mmHg, from zero to the top of the monitor's scale, and then lowers it through the same steps. By OIML's procedure, the result at each step is the monitor's reading minus the reference's.

An automated monitor does not normally let anyone else set its pressure. In use it pumps the cuff up by itself and lets the air out through a valve, reading the pressure all the while (Chapter 7). For the test the monitor enters a service mode, which OIML's rules allow only service personnel to use. In that mode the monitor closes its vent valve, so that the air from the hand pump stays in the vessel, as Fluke Biomedical, a maker of test equipment, describes it. Rossmax, a monitor maker, states how its own monitors enter the mode: the power is drained, the power button is pressed three times, and the display shows "00". In that state the monitor no longer controls the pressure, and its display shows only what its sensor reads.

The same set-up also tests for leaks, because a pressure that falls while the two instruments are read gives a pair of readings that describes no single pressure. Fluke advises waiting at least 60 seconds at each step before reading, and measuring the loss of pressure over five minutes. Fluke adds that most monitor makers recommend a leak of less than 8 mmHg per minute, while OIML allows an automated monitor at most 6 mmHg per minute. A leak between the two figures meets the makers' usual recommendation and fails OIML's limit.

36 — The monitor on the calibration bench
HAND PUMP with release valve T-piece RIGID VESSEL 500 ml ± 25 ml, in place of the cuff T-piece 150.0 mmHg REFERENCE MANOMETER the reference: maximum error ±0.8 mmHg MONITOR, CUFF REMOVED 150 mmHg service mode: vent valve closed PRESSURE IN THE VESSEL, mmHg time 0 50 100 150 200 250 300 monitor reference rising: 50 mmHg steps to the top falling through the same steps result at each step = monitor's reading − reference's reading Apparatus and steps: OIML R 149-2:2020. Service mode: Fluke Biomedical; Rossmax (manufacturers). Top of scale drawn at 300 mmHg: schematic.
The quantity being measuredReferences and standardsReadingFocal detail
Plate 36 — On the bench a rigid vessel replaces the cuff, the monitor's valve stays shut, and the reference manometer's reading serves as the reference value. At every step, rising and falling, the result is the monitor's reading minus the reference's.

The reference manometer

The reference manometer is the pressure gauge whose reading serves as the reference value at each step, and the whole calibration rests on it. A reference manometer is one whose own error is known from its own calibration and is small beside the error being measured. OIML allows the reference a maximum permissible error of ±0.8 mmHg, or 0.1 kPa, and India sets the same limit in its rules for automated sphygmomanometers. Mindray, a maker of patient monitors, checks the pressure channel of its VS-900 against ±3 mmHg and asks in its service manual for a manometer "calibrated with accuracy equal to or better than 0.75 mmHg". That figure is a quarter of the ±3 mmHg, since 0.75 divided by 3 is 0.25. A reference must be several times better than the limit it serves, because its own error raises the risk of a wrong pass or fail (Chapter 12).

A reference used in practice can be far better than these limits. In a 2021 bench study, Ramírez-Barrera and colleagues calibrated four non-automated sphygmomanometers from the General Hospital of Medellín in the metrology laboratory of the city's Instituto Tecnológico Metropolitano, which funded the work. Their steps followed the pattern of OIML's procedure: each test point lay at most 50 mmHg from the next, from zero to the top of each scale, and the authors ran two rising and two falling series on every instrument. Their reference was a digital manometer with a resolution of 0.01 mmHg, an accuracy of ±0.025 % of the reading and, by its own certificate, an expanded uncertainty of ±0.062 mmHg.

A reference stays a reference only while something better checks it in turn. Rossmax sells a calibration system for its own monitors, the CalPro, and states its accuracy as ±0.8 mmHg. Rossmax asks in the manual that the CalPro be compared with a reference at a metrology center every two years, and that it differ from that reference by no more than ±0.2 mmHg. The reference manometer is checked in its turn by a better standard, and that standard by a better one still (Chapter 15). Perkin-Elmer, in the opening story, relied on a reference that no one had measured independently. It did not verify its null corrector's dimensions after assembly, and a lens about 1.3 mm out of place guided the polishing of the whole mirror.

Laboratories also keep their best references out of daily use. In a procedure revised in 2019, FDA's own testing laboratories use their reference masses and thermometers "for calibration verification only and not as working standards", store them in a protected area and have the masses calibrated every five years.

Adjustment, and the record before and after it

Adjustment follows only after the first step has shown an error worth removing. Rossmax's CalPro runs the sequence in order. Over a data link it first reads the monitor at 11 pressures, in four falling groups that each start from zero: 295, 270 and 240 mmHg; 210, 180 and 150; 120, 90 and 60; and 30 mmHg. The monitor passes if it reads within ±3 mmHg at every point. If it fails, the CalPro programs it at 0, 300, 200 and 100 mmHg, which is the adjustment, and then reads the 11 points again. The display shows PASS or FAIL, and Rossmax tells its partners to return a monitor that still fails to the maker. Rossmax presents the CalPro as a system for its partners' service work in the field, not as its production station.

A shorter sequence adjusts at two points, because two pressures far apart fix a straight line, and therefore both the offset and the span of Chapter 11. Welch Allyn, a maker of patient monitors, describes one in its service manual for the Ambulo 2400, the ambulatory monitor of Chapter 9. The technician sets the main processor's pressure channel at two reference points, 250 and 15 mmHg, zeroes it with all pressure drained, and then tests the result against ±3 mmHg.

37 — Calibration, adjustment, calibration again, verification
CALIBRATION: AS FOUND VIM 2.39 −6 −3 0 +3 +6 0 295 270 240 0 210 180 150 0 120 90 60 0 30 check points in reading order, mmHg ERROR, mmHg OIML (dashed): 3 mmHg or 2 % of the reading, whichever is greater Rossmax (solid): ±3 mmHg at every point keep the as-found record: it shows how the monitor was reading in use ADJUSTMENT VIM 3.11 MEM mmHg 0 300 200 100 programmed at 0/300/200/100 mmHg CALIBRATION AGAIN: AS LEFT VIM 3.11, Note 3 −6 −3 0 +3 +6 same points and bands as panel 1; every point inside ±3 mmHg VERIFICATION VIM 2.44 PASS a monitor that still fails goes back to the maker Sources: points, limit and adjustment pressures, Rossmax CalPro manual (manufacturer); OIML band, R 149-1:2020, 5.1. Error dots schematic.
The signal in the chainLimits and decisionsReadingFocal detail
Plate 37 — Calibration measures, adjustment changes, verification judges, and a second calibration shows what the adjustment did. Rossmax's flat ±3 mmHg is tighter than OIML's limit at every check point above 150 mmHg.
A monitor that adjusts itself has still not been calibrated

Adjustment is "often mistakenly called 'self-calibration'", in the words of the VIM's note on calibration. A monitor that zeroes its sensor at power-on (Chapter 17) adjusts itself: it changes its own indication to match a value it assumes, the pressure of the open air. Nothing outside the monitor takes part, so no relation to a measurement standard is established, and the zero is only as good as the assumption that the sensor feels no pressure at that moment. Only a comparison with a reference whose value and uncertainty are known establishes that relation. A built-in routine can hold a calibrated monitor near its calibration between visits to the bench, but it cannot replace a visit.

A laboratory that adjusts an instrument calibrates it first and, usually, again afterward, and the two calibrations are known in the trade as as found and as left. ISO/IEC 17025 is the standard for the competence of testing and calibration laboratories (Chapter 16). A calibration certificate under that standard adds four things to an ordinary test report, as NIST and two bodies that accredit laboratories, UKAS in the United Kingdom and SADCAS in southern Africa, describe the standard. The certificate gives the uncertainty of each result, the conditions that influence the results and a statement of how the results are traceable, and, where available, the results before and after any adjustment or repair. The as-found values show whether the monitor had been outside its limit while in use, and the maker must then decide what to do about the readings it gave (Chapter 26). The as-left values describe the monitor that leaves the bench.

As of October 2026, open sources have yielded no monitor maker's account of its own production calibration station, with the pressures it applies and the coefficients it writes, and the makers' patents describe designs instead. Health and Life, a Taiwanese maker, patented a factory tool with which "more than one electronic sphygmomanometer can be set and calibrated at the same time". The tool works from two states, atmospheric pressure and a maximum, with a straight-line proportion between them. Shenzhen CHK Technology, a Chinese monitor maker, states in its 2010 patent application that "each sensor must be calibrated", and in its method a program sets the gain and offset of the monitor's amplifier (Chapter 9). Each inventor described a proposal rather than a factory's present practice, so a production test rests on the published test methods rather than on any one maker's practice (Chapters 22 and 23).

Verification asks whether a limit is met

Verification needs a limit, and the limit belongs to whoever sets the requirement. Under OIML's rules, verifying an automated monitor, new or in service, covers at least the cuff-pressure limit, the repeatability and the leak, and needs no climatic chamber (Chapter 30). OIML sets the cuff-pressure limit at ±3 mmHg or 2 % of the reading, whichever is greater. Because 2 % of 150 mmHg is 3 mmHg, OIML allows more than 3 mmHg above 150 mmHg: 2 % of 200 mmHg is 4.0 mmHg, and 2 % of 295 mmHg is 5.9 mmHg.

A maker may set a limit of its own. Mindray verifies the VS-900's pressure channel at 0, 50 and 200 mmHg against ±3 mmHg, and Rossmax allows ±3 mmHg at every point of its check. At five of Rossmax's 11 points, from 180 to 295 mmHg, the maker's limit is tighter than OIML's, and at 295 mmHg it is about half, since 3 divided by 5.9 is 0.51. A monitor that passes Rossmax's check therefore meets OIML's limit at every point, and a monitor that meets OIML's limit may still fail Rossmax's.

A calibration hands back an error with an uncertainty, and a verification has to decide what to do with both. Ramírez-Barrera and colleagues reported errors and uncertainties but did not state whether any instrument passed at any point. Their figures leave the decision to whoever applies a limit, and at one point the decision is not obvious.

Solved example: what one calibration point hands back, and what it leaves undecided

For the fourth of their instruments at 120 mmHg, Ramírez-Barrera and colleagues report:

  • an error of −2.43 mmHg;
  • five standard uncertainties: repeatability 0.28 mmHg, resolution 0.29 mmHg, hysteresis 0.30 mmHg, the reference's accuracy 0.11 mmHg and the reference's certificate 0.031 mmHg;
  • an expanded uncertainty of ±1.1 mmHg.

Squared, the five rows are 0.0784, 0.0841, 0.0900, 0.0121 and 0.0010 mmHg², and they add to 0.2656 mmHg². The square root, 0.515 mmHg, is the combined standard uncertainty of Chapter 11, and it matches the authors' 0.514 mmHg within the rounding of the printed rows. Twice 0.514 mmHg, for a coverage factor of 2, would be 1.03 mmHg. The authors' ±1.1 mmHg is slightly larger because they take the coverage factor from the effective degrees of freedom, which puts it a little above 2, as Chapter 12 showed for their first gauge. The reference's two rows supply 0.0121 plus 0.0010, which is 0.0131 of the 0.2656 mmHg², about 5 % of the total of squares, so the instrument itself supplies about 95 %.

The error and its expanded uncertainty describe an interval from −2.43 minus 1.1, which is −3.53 mmHg, to −2.43 plus 1.1, which is −1.33 mmHg (Chapter 2). Under OIML's current recommendation for non-automated sphygmomanometers, R 148-1 of 2020, the limit is ±3 mmHg at any point of the scale. The measured error, 2.43 mmHg in size, lies inside that limit, but 0.53 mmHg of the interval's 2.2 mmHg lies beyond −3 mmHg. The authors calibrated according to R 148-1's predecessor, R 16-1, whose limit for a device in use was ±4 mmHg, and under that limit the whole interval lies inside. The calibration is complete, because it has stated the error and how well the error is known. Whether the instrument passes at 120 mmHg is a separate decision, made by a limit and a decision rule chosen in advance (Chapter 13). Against ±3 mmHg it passes under a rule that accepts any measured error inside the limit, and it cannot be passed under a rule that requires the whole interval to lie inside.

38 — What one calibration point hands back
limit ±3 mmHg (OIML R 148-1:2020, non-automated) −4 −3 −2 −1 0 +1 +2 +3 +4 ERROR AT 120 mmHg, mmHg error −2.43 mmHg expanded uncertainty ±1.1 mmHg, k a little above 2 0.53 mmHg of the interval lies beyond the limit rule 1, measured error inside the limit: passes rule 2, whole interval inside the limit: cannot be passed the rule is chosen in advance (Chapter 13) repeatability 0.0784 · 29.5 % resolution 0.0841 · 31.7 % hysteresis 0.0900 · 33.9 % the instrument itself: about 95 % the reference: about 5 % reference's accuracy 0.0121 · 4.6 % reference's certificate 0.0010 · 0.4 % total of squares 0.2656 mmHg², root 0.515 mmHg Sources: Ramírez-Barrera et al. 2021, instrument 4 at 120 mmHg; OIML R 148-1:2020. Interval and shares worked from the figures above.
References and standardsNoise and errorUncertaintyLimits and decisionsReadingFocal detail
Plate 38 — A calibration hands back an error and how well it is known. Here the error lies inside the limit but its interval does not, so whether the instrument passes depends on a rule chosen before the result was seen.

The reading, not the estimate

Calibration, adjustment and verification all concern the cuff-pressure reading, and none of them reaches the blood-pressure estimate. The reference manometer and the vessel hold a steady pressure, and a steady pressure carries no oscillations, so the algorithm that derives systolic and diastolic values from the oscillations (Chapter 9) never runs.

The estimate can be adjusted in the field, but only against another estimate. In a patent granted in 2004, Microlife, a monitor maker, describes a calibrating device that recalibrates a wrist monitor's estimate in the field. The device writes new constants into the monitor's memory after three paired readings against an upper-arm monitor, and accepts ±20 mmHg for the systolic value and ±12 mmHg for the diastolic. The procedure adjusts one estimate against another, so it is not a calibration of the cuff pressure in the VIM's sense. A clinical study validates the estimate instead, comparing it with a reference measured on patients (Chapter 24).

Writing calibration into the monitor's procedures
  • Write the procedure with its points, both directions, the vessel, the waiting time and the reference's limit, and name the service mode the monitor must enter.
  • Record the as-found values before any adjustment, calibrate again after it, and file both with the coefficients written.
  • Keep three records apart: the calibration (pairs of readings with their uncertainty), the adjustment (what was changed) and the verification (which limit, which rule, which result).
  • State which limit a verification used, the maker's ±3 mmHg at every point or OIML's greater of 3 mmHg and 2 % of the reading, because the two differ above 150 mmHg.
  • Keep the reference manometer's certificate with every record it produced, so that a later fault in the reference can be traced to the monitors it passed (Chapter 26).
+ What this chapter established
  • Calibration is two steps, as the VIM defines it: a relation between a standard's values and an instrument's indications, each with its uncertainty, and then the use of that relation to obtain results; adjustment and verification are separate operations.
  • The monitor is calibrated in a service mode, with a rigid vessel in place of the cuff and a reference manometer whose maximum permissible error OIML and India each set at ±0.8 mmHg, in steps of at most 50 mmHg, rising and falling.
  • Adjustment follows a calibration and is followed by another, and a certificate under ISO/IEC 17025, as NIST and accreditation bodies describe the standard, gives the results before and after it, the as-found and as-left values, where they are available.
  • A calibration hands back an error and its uncertainty, a verification judges them against a limit by a decision rule chosen in advance, and both concern the cuff-pressure reading, never the blood-pressure estimate.