+ Part V · The top of the chain

From an object to a constant.

For 130 years the kilogram was a cylinder of platinum and iridium, about 39 millimeters high and 39 across, kept in a safe at the International Bureau of Weights and Measures near Paris. Every calibrated mass in the world led back to that one object, and so did every cuff pressure calibrated through the masses of a pressure balance. Chapter 28 shows that the links at the top of the chain add almost nothing to a monitor's uncertainty budget, but a maker's claim that its cuff pressures are traceable depends in part on them. An error at this level, such as the one the BIPM found in 2014, has to be traced through every certificate that rests on it. This part follows the top of the chain through its change in 2019. Chapter 18 shows what the cylinder could and could not tell the people who kept it, and Chapter 19 how a fixed value of the Planck constant, a constant of quantum physics, came to define the kilogram. Chapter 20 describes the instruments that measure a mass from that constant, and Chapter 21 follows the kilogram they realize down to the balances of every country, India's among them.

18 — The artifact

An object that could not be wrong.

+ The questionWhat goes wrong when a unit is defined by a single object?

One cylinder, right by definition

Until 2019 the kilogram was the mass of one metal cylinder, which Johnson Matthey, in England, had made from an alloy of 90 % platinum and 10 % iridium. Its height and its diameter were each about 39 mm, according to a 2004 review by Schwitz and colleagues at METAS, Switzerland's national metrology institute. The BIPM, the International Bureau of Weights and Measures, kept the cylinder in a safe at Sèvres near Paris. In 1889 the first General Conference on Weights and Measures, the CGPM of Chapter 3, sanctioned it as the international prototype of the kilogram and declared that the prototype "shall henceforth be considered as the unit of mass".

A unit defined by an object has a property that no other standard shares: its value cannot be wrong. Every other kilogram was a copy whose mass someone had measured, and every such measurement carried an uncertainty. The international prototype was 1 kg by declaration, exactly, and no measurement could give it an uncertainty, because nothing stood above it to compare it with. For the same reason, no measurement could reveal a change in it. If the cylinder lost a microgram, a millionth of a gram, the kilogram lost a microgram with it, and every copy measured against the cylinder would seem to have gained one. Until 2019 the chain of Chapter 15, from the monitor's reference manometer through a pressure balance and its masses, ended at this cylinder.

Copies, compared again and again

The prototype came out of its safe only on rare occasions, so laboratories around the world measured mass against copies of it. Johnson Matthey delivered 40 copies in 1884, and in 1889, 34 of them went to the states that had signed the Metre Convention of 1875 and asked for one, by the account of a 2001 review by Jabbour and Yaniv of NIST. These copies became the states' national prototypes. The BIPM kept official copies beside the prototype, two at first and six after one was replaced and four were added. In its declaration of 1889 the CGPM recorded that the national kilograms differed from the international one by no more than 1 mg, and only a comparison with the prototype could show whether a copy still agreed that closely.

A copy is checked against the prototype on a balance, and a balance used this way reports only the difference between two masses. The differences that mattered were small: a kilogram holds a billion micrograms, so a difference of one microgram between two kilograms is a difference of one part per billion. The BIPM compared the national prototypes with the international one in three periodic verifications, from 1899 to 1911, from 1947 to 1954 and from 1988 to 1992, the dates that Schwitz and colleagues give. Georges Girard, the head of the BIPM's mass section, led the third. In that verification each national prototype was measured against the international one with a standard uncertainty of 2.3 µg, or 2.3 parts per billion.

What the verifications found

The verifications found that the copies and the prototype were drifting apart, and that the copies tended to grow heavier relative to the prototype. In the first, two of the 25 prototypes verified had changed by as much as 50 µg. The seventeen national prototypes that Schwitz and colleagues plot across the verifications drifted by an average of 50 µg in 100 years, which is 50 parts per billion a century. In the second and third verifications many of the BIPM's official copies had also increased in mass with respect to the prototype, as Stock and colleagues at the BIPM report in their 2015 account of a later calibration campaign.

The surface of every cylinder, the prototype's included, gathers contamination between cleanings, and the contamination adds mass. During the third verification the BIPM found that the official copies gained between 0.6 and 1.3 µg a year, and the international prototype itself 1.5 µg a year. The typical rate that the BIPM observed for national prototypes is 1 µg a year. The long-term instability is attributed to surface effects "that are not yet fully understood", in the words of Jabbour and Yaniv. A cause that no one could model was also a cause that no one could correct, because a correction needs a rate that holds for each object, and the rates differed from one cylinder to the next.

45 — The copies drift apart from the prototype
DIFFERENCE FROM THE PROTOTYPE, µg 1 µg in 1 kg = 1 part per billion FIRST 1899–1911 SECOND 1947–1954 THIRD 1988–1992 25 50 75 0 1889 1900 1925 1950 1975 2000 2014 2.3 µg standard uncertainty per prototype average drift 50 µg in 100 years (17 national prototypes) international prototype: 1 kg by definition ? its own change cannot be drawn: every point is a difference Sources: Schwitz et al. 2004 (METAS review) for the average drift, the prototype count, the uncertainty and the dates, as they give them. The thin lines are schematic.
References and standardsNoise and errorUncertaintyFocal detail
Plate 45 — The verifications measured only differences from a cylinder fixed at 1 kg by definition. They could see the copies move apart by about 50 µg a century, but not whether the copies had gained mass or the prototype had lost it.

Cleaning before weighing

Because contamination adds mass between cleanings, the BIPM cleaned every prototype by hand before comparing it. In the method that Girard described in a BIPM monograph of 1990, a metrologist rubbed the whole surface of the cylinder fairly vigorously with chamois leather that had soaked for 48 hours in equal parts of alcohol and ether. A jet of steam from a flask of double-distilled water then washed each surface for fifteen to twenty minutes, about fifty minutes in all. The method "relies on the human touch and therefore can be highly irreproducible", in Jabbour and Yaniv's words, so two careful metrologists could leave the same cylinder at slightly different masses.

A cleaned prototype starts to gain mass again at once. Girard followed prototypes 7, 67 and 73 after cleaning and washing, as the same monograph reports, and each gained about 1 µg a month during the first three to four months. The mass of a prototype therefore depended on how long ago it had been cleaned, and on the hands that had cleaned it. At about 1 µg a month, a prototype gained about 3 µg in the three months after a cleaning, more than the 2.3 µg standard uncertainty with which the third verification compared it.

Nobody could say which had moved

The verifications showed a growing difference between the prototype and its copies, and a difference cannot show which of two masses changed. If the copies gained 50 µg, the difference grows by 50 µg; if the prototype lost 50 µg, the difference grows by exactly the same amount, and the balance reads the same numbers in both cases. A third object would not settle the question, because it too could be compared only with the prototype, whose mass was fixed by declaration rather than measured. The prototype's own mass "is subject to small and not easily predictable changes", in the words of Schwitz and colleagues, and no comparison could say how large those changes were.

The CIPM, the International Committee for Weights and Measures of Chapter 3, drew that conclusion in 1993. In a recommendation after the third verification, it noted the significant drifts found in national prototypes and "the fundamental impossibility of drawing definite conclusions from these results". It recommended work toward "a new definition of the unit of mass based upon fundamental or atomic constants". The experiments that followed led to the definition of 2019 (Chapters 19 and 20).

A copy that gains mass and a prototype that loses it give the same comparison

A drawing of the verifications shows the copies' masses rising against a flat line for the prototype, and a reader can take that line to mean that the prototype held steady while its copies grew heavier. The flat line is a convention. The prototype's mass is 1 kg by definition, so a drawing has no way to show a change in it, and every point is a difference. A rise of 50 µg in a century fits copies that gained 50 µg, a prototype that lost 50 µg, or any mixture of the two.

The weighing of December 2013

To prepare a new definition of the kilogram, based on a fixed value of the Planck constant (Chapter 19), the BIPM began the calibration campaign by weighing the international prototype between 11 and 27 December 2013. It was the prototype's first weighing since the third verification, and in the meantime the prototype and its six official copies had stood in a safe, unused. In 2013 the CCM, the consultative committee that advises the CIPM on mass, had set conditions for the new definition. Among them, it asked that the BIPM's standards and the mass standards used in the experiments that measured the Planck constant be compared "as directly as possible with the international prototype".

The BIPM then cleaned and washed the prototype and its copies between 23 December 2013 and 17 February 2014. Weighed before and after, the official copies lost an average of 15 µg in the cleaning, with a standard deviation of 2 µg, as Stock of the BIPM reported to the CCM in February 2015. The BIPM followed the mass that the objects regained for about ten months, until November 2014. At first they gained 0.01 to 0.03 µg a day, which over 30 days is 0.3 to 0.9 µg a month, close to Girard's figure of about 1 µg a month for prototypes 7, 67 and 73.

The prototype and its copies still stood in the same relation to one another as in the third verification, within the uncertainty of the comparisons. The differences between the prototype and the official copies had changed by an average of 1 µg since then, against an uncertainty of 3 µg for each comparison, so the set had "behaved as a consistent set", in the words of Stock and colleagues. A copy kept away from the BIPM gave a further check. Prototype No. 34, which belongs to the French Academy of Sciences and had stayed unused in Paris since the third verification, came to the BIPM in September 2014. Cleaning removed 19.6 µg from it after 22 years, which is 19.6 ÷ 22, about 0.9 µg a year, close to the typical 1 µg a year. Its mass relative to the international prototype came out 1 µg lower than in the third verification, a change of the same size as the official copies' average.

The BIPM did find an error, but one level below the prototype, in the masses it used every day. Its working standards, the kilograms against which it calibrated other countries' standards, had all lost between 18 and 88 µg against the prototype since 1992, and the BIPM later attributed the loss to wear. The BIPM's unit was still traceable to the prototype through the values those standards had been given in the third verification. A standard that has lost mass but keeps its old value gives every kilogram compared with it a value that is too high. The as-maintained mass unit, the kilogram as the BIPM actually kept it in those standards, was therefore found in 2014 to be "offset by 35 µg with respect to the IPK", the international prototype of the kilogram. The CCM put the change at 35 µg over 22 years.

46 — The weighing of 2013–14 finds the fault one step down
11–27 Dec 2013 weighed · 23 Dec 2013 to 17 Feb 2014 cleaned and washed · regain of mass followed to Nov 2014 THE PROTOTYPE AND ITS SIX OFFICIAL COPIES differences changed by an average of 1 µg since the third verification, against 3 µg per comparison Stock and colleagues: the set "behaved as a consistent set" INTERNATIONAL PROTOTYPE SIX OFFICIAL COPIES 0 20 40 60 80 copies: 1 µg change µg from the prototype working standards: 18 to 88 µg +35 µg: the as-maintained unit in 2014 BIPM WORKING STANDARDS lost 18 to 88 µg against the prototype since 1992, attributed to wear; used to calibrate other kilograms 1 kg certificates issued 1 Jan 2003 to 31 Dec 2013 updated NIST's kilogram K85: 2012 value lowered by about 35 µg h Planck-constant results corrected down by 35 parts per billion Sources: Stock et al. 2015 (BIPM); Stock, CCM slides, Feb 2015; CCM recommendation 2015; Schlamminger et al. 2015 (NIST preprint).
References and standardsNoise and errorFocal detail
Plate 46 — The prototype and its copies still agreed with one another; the working standards used every day did not. The BIPM's kilogram as maintained had moved 35 µg from the prototype, and the certificates and Planck-constant results built on it were corrected.

The offset passed into the results that rested on the BIPM's calibrations of those years. By NIST's account the maintained unit was 35 µg larger than the prototype, so NIST had to lower by about 35 µg the value that the BIPM had given NIST's own kilogram K85 in 2012. The BIPM updated the certificates for calibrations of 1 kg standards that it had issued between 1 January 2003 and 31 December 2013. The experiments on the Planck constant that used masses calibrated by the BIPM carried the same offset, and these were the experiments that the campaign was meant to support. In 2015 the CIPM recommended that the determinations of the Planck constant, which "will provide the basis for the future definition of the kilogram", be updated with the corrections that the BIPM gave the institutes concerned. NIST lowered its own result for the Planck constant by 35 parts per billion, the same fraction of its value as 35 µg is of a kilogram. NRC, Canada's national metrology institute, applied the same correction, and both did so before the value of the Planck constant was fixed (Chapter 20).

Solved example: the prototypes' drift and the 2014 offset, in parts per billion

A kilogram holds a billion micrograms, so 1 µg in a kilogram is 1 part per billion.

  • The national prototypes' average drift: 50 µg in 100 years is 50 ÷ 100 = 0.5 µg a year, or 50 parts per billion a century.
  • The third verification's standard uncertainty was 2.3 µg for each national prototype. At 0.5 µg a year, a prototype drifts by that amount in 2.3 ÷ 0.5 = 4.6 years.
  • The second verification ran from 1947 to 1954 and the third from 1988 to 1992, about 40 years apart. In that time the average drift adds 0.5 × 40 = 20 µg, and 20 ÷ 2.3 = 8.7, about nine times the 2.3 µg of a comparison.
  • The 2014 offset: 35 µg in a kilogram is 35 parts per billion. Built up over 22 years, it grew by 35 ÷ 22 = 1.6 µg a year, and 1.6 ÷ 0.5 = 3.2, about three times the national prototypes' average drift.
  • In 2013 the CCM asked that at least one experiment measure the Planck constant with a standard uncertainty no larger than 20 parts per billion of its value (Chapter 19). The offset of 35 parts per billion was 35 ÷ 20 = 1.75 times that target.

Over forty years the drift grew to about nine times the uncertainty of a comparison, so the verifications detected it. They could not assign it to the copies or to the prototype, because each of their numbers was a difference from a mass fixed by declaration. The offset found in 2014 was far too small to matter on a factory's balance. At the top of the chain, though, it was larger than the uncertainty that the CCM required of the best measurement of the Planck constant, which is why the CIPM asked for the determinations of the Planck constant to be corrected for it.

Keeping a reference that nothing above it checks
  • Keep at least one reference that is rarely used beside the references used every day. The BIPM's working standards lost 18 to 88 µg through use, while the differences between the rarely handled prototype and its copies changed by an average of only 1 µg.
  • Compare a set of references with one another at fixed intervals and record each difference, not only a pass. A difference that grows over several checks shows that something in the set moves, even when it cannot show which member.
  • Record how each reference was handled and cleaned before it was used. A freshly cleaned prototype gained about 1 µg a month, so the time since cleaning was part of its value.
  • When a reference turns out to have been wrong, find every certificate and result that rests on it, as the BIPM did for the 1 kg certificates it had issued from 2003 to 2013 (Chapter 26).
+ What this chapter established
  • From 1889 to 2019 the kilogram was the mass of one platinum-iridium cylinder at the BIPM, exact by declaration, so no measurement could give it an uncertainty or reveal a change in it.
  • Three periodic verifications compared the national prototypes with it and found an average drift of 50 µg a century, which a comparison uncertainty of 2.3 µg could detect but could not assign.
  • Every comparison was a difference, so no one could say whether the copies had gained mass or the prototype had lost it, and in 1993 the CIPM recommended a definition based on constants.
  • In the campaign of 2014 the differences between the prototype and its copies had changed by about 1 µg since the third verification, while the BIPM's worn working standards had shifted its mass unit by 35 µg, and the determinations of the Planck constant were corrected for that offset.