§ Glossary
Glossary.
Terms are defined as they are used in this guide.
- 510(k)
- A US premarket notification showing that a device is substantially equivalent to a legally marketed predicate device; FDA's decision on it is a clearance (Chapters 9 and 26).
- Ablation
- Destroying tissue in place without removing it, usually a tumor, or lines of heart muscle whose scar treats an irregular heart rhythm; it reaches its target through a needle, a catheter or the intact skin (Before we start).
- Absorption coefficient (µa)
- The probability per centimeter that light is absorbed; at a depth of 1/µa, the absorption depth, the intensity has fallen to about 37 % of its value at the surface (Chapter 5).
- Acoustic impedance
- A material's density multiplied by its speed of sound, measured in rayls; sound crosses a boundary well only when the impedances on its two sides match, and a mismatch reflects it (Chapter 4).
- Acoustic window
- A path to a target that is free of gas and bone, which focused ultrasound needs (Chapter 4).
- Active electrode
- The electrode through which monopolar current enters the patient, such as the pencil, hook or scissors in the surgeon's hand (Chapter 8).
- Active electrode monitoring
- A conductive shield around an instrument's insulated conductor, connected back to the generator's return and covered by a second insulation, which collects current that escapes through a defect or by capacitance and returns it; a monitor tells fault current, in phase with the voltage, from normal capacitive current, a quarter of a cycle apart, and stops the generator on a fault (Chapter 18).
- Advanced bipolar sealer
- A bipolar instrument that seals vessels by combining high, controlled jaw pressure, high current at low voltage and a closed loop that ends energy delivery by itself; many are indicated for vessels up to 7 mm (Chapter 9).
- Antinode
- A point of maximum motion in a standing wave; the blade tip of an ultrasonic instrument sits at one (Chapter 11).
- Applied part (B, BF, CF)
- A part of a device that contacts the patient, classified as type B, BF or CF from the least to the most isolated; type CF is required for anything applied directly to the heart (IEC 60601-1; Chapter 20).
- Arrhenius model
- A description of thermal damage as a chemical reaction whose rate rises exponentially with temperature, the damage accumulating for as long as the temperature is held (Chapter 1).
- Basic safety
- Freedom from unacceptable risk caused directly by physical hazards, such as electric shock, fire, mechanical injury and excessive temperature, when equipment is used in normal condition or with a single fault (IEC 60601-1; Chapter 20).
- Bioheat equation
- The heat balance for living tissue that Harry Pennes published in 1948: the temperature of a small volume rises by the heat deposited in it, minus the heat conducted to its neighbors, minus the heat carried off by blood; metabolic heat is also in it but is negligible during surgery (Chapter 2).
- Bipolar
- Electrosurgery with an instrument that carries both electrodes in its tips, usually the two blades of a forceps, so that current flows only through the tissue held between them and no return pad is needed (Chapter 9).
- Blocking capacitor
- A capacitor in series with each lead of a generator output that passes high-frequency current and blocks direct and low-frequency current, which could otherwise stimulate nerves and muscles (Chapter 8).
- Burst pressure
- The pressure at which a seal fails when the sealed vessel is filled with liquid and the pressure raised; the bench test of a seal's strength (Chapter 9).
- Butterworth–Van Dyke model
- The equivalent circuit of a piezoelectric transducer: the ceramic's own capacitance in parallel with a resonant branch whose inductance, capacitance and resistance stand for the moving mass, the stiffness and the mechanical losses, which include the tissue pressed against the blade (Chapter 11).
- Capacitive coupling
- Transfer of alternating current across an insulator without contact, for example from an instrument shaft to a metal cannula (Chapter 7).
- Carbonization
- Charring of tissue once it has dried out, with an onset put at about 150 to 200 °C, well above the 100 °C at which wet tissue stalls; char wastes energy, releases smoke and conducts current poorly (Chapter 1).
- Cavitation
- The growth and motion of gas bubbles in a sound field: in stable cavitation a bubble oscillates gently in step with the wave, and in inertial cavitation it grows to many times its size and collapses violently, and the collapse damages tissue (Chapter 4).
- CDSCO
- The Central Drugs Standard Control Organisation, India's device regulator under the Medical Devices Rules, 2017, which publishes classification lists by specialty and, through its Central Licensing Authority, licenses class C and D devices (Chapter 27).
- CE marking
- The mark a manufacturer affixes to a device for the EU market once its conformity assessment is complete; for a class IIb device, after a notified body has issued its certificates (Chapters 16 and 27).
- CEM43
- Cumulative equivalent minutes at 43 °C, the thermal dose used in hyperthermia; on this measure each degree above 43 °C halves the time needed for the same effect (Chapter 1).
- Chromophore
- A molecule that absorbs light and turns its energy into heat; in surgery, mainly water, hemoglobin and melanin (Chapter 5).
- Cleared
- Of a device in the United States: allowed onto the market by the Food and Drug Administration through a 510(k), as substantially equivalent to a device already legally marketed (Before we start; Chapter 26).
- Coagulation
- Irreversible unfolding and setting of proteins by heat, which kills cells and stops small vessels bleeding; published onsets range from about 50 to 60 °C, depending on the tissue, species and exposure time. To a surgeon the word means this change caused by heat, and the clotting by which blood closes a small wound on its own is a different process (Chapter 1).
- Collagen
- The protein that gives vessel walls, skin and tendons their strength, three chains wound around each other like the strands of a rope; in measurements on rat tissues, heat unwinds its helix between about 60 and 90 °C, depending on the tissue and the test, the fiber shrinks as it unwinds, and unwound layers pressed together as they cool fuse, which is how a vessel seal works (Chapter 1).
- Collateral standard
- A standard in the IEC 60601-1 family that adds requirements many kinds of equipment share, such as IEC 60601-1-2 for electromagnetic compatibility, IEC 60601-1-6 for usability and IEC 60601-1-8 for alarm systems (Chapters 20 and 25).
- Contact quality monitoring
- Monitoring of a split return pad's contact by the impedance between its two halves, which stops the output when the pad lifts; since edition 6.0 of 2017, IEC 60601-2-2 has required adult return electrodes of this kind (Chapters 8 and 25).
- Crest factor
- Peak voltage divided by RMS voltage: 1.41 for a continuous sine wave and higher the lower the duty cycle, so it is low for cutting waveforms and high for fulguration and spray (Chapter 8).
- Critical task
- In FDA's human factors guidance, a user task that, if performed incorrectly or not performed at all, would or could cause serious harm (Chapter 22).
- Cryoablation
- Ablation by freezing: a probe colder than tissue draws heat out, and ice grows outward from it (Chapters 6 and 14).
- Cutting
- Dividing tissue with current or a laser by delivering energy so quickly into a thin line that the water in its cells boils explosively and the tissue parts; no sharp edge is involved (Before we start; Chapter 1).
- Cyber device
- Under section 524B of the US Federal Food, Drug, and Cosmetic Act, a device that includes software validated, installed or authorized by its sponsor, can connect to the internet and has characteristics that could be vulnerable to cybersecurity threats; submissions for one made on or after 29 March 2023 must include a vulnerability plan, update processes and a software bill of materials (Chapters 20 and 26).
- Damping
- The decay of an oscillation. In Bovie's apparatus, as his patent filed in 1928 describes it, the less damped and steadier current severed tissue, and the heavily damped one, whose bursts of oscillation died away quickly, dried tissue and stopped bleeding (the opening story).
- De Novo
- The US route for a new kind of device of low or moderate risk that has no predicate: when FDA grants the request it creates a new regulation and product code that later devices can follow by 510(k) (Chapters 14 and 26).
- Desiccation
- Slow drying of tissue: when energy arrives slowly, the water leaves gradually and the tissue dries and shrinks into a firm coagulum, which is how most contact coagulation works (Chapter 1).
- Design and development file
- The record that holds the trace from each claim to its requirements, tests and label line, which ISO 13485 requires for each device type or family (Chapter 22).
- Design controls
- The process that turns an intended use and its claims into design inputs, design outputs, verification and validation; FDA's guidance of 1997 draws it as a waterfall, and ISO 13485 sets out the same steps (Chapter 22).
- Design input
- A requirement the design must meet, written so that a test can fail it, such as seals of vessels up to 7 mm, measured rounded and perfused, holding at least 360 mmHg in each mode (Chapter 22).
- Design output
- What the design process produces to define the device, which must meet the input requirements; for a sealer, its jaw pressure, sealing mode, end-point algorithm and instructions for use (Chapter 22).
- Dielectric heating
- Heating by an alternating electromagnetic field that turns water molecules, which lag behind it, and drives ions back and forth; both motions dissipate energy as heat throughout a volume, with no conducting path to an electrode, as in microwave ablation (Chapters 6 and 13).
- Diffusion length
- An estimate of how far heat spreads in a given time: the square root of four times the thermal diffusivity times the time, about 0.75 mm in one second and 2.4 mm in ten for soft tissue (Chapter 2).
- Direct coupling
- Current passing from an active electrode into another conductor it touches, such as a metal grasper or the laparoscope, and from there into tissue at that conductor's point of contact (Chapter 18).
- Duty cycle
- The fraction of time an output is on. The waveform duty cycle of Chapter 8 is the share of each cycle in which the carrier is on, 100 % for pure cut and 6.25 % for the FT10's fulguration. The activation duty cycle of Chapter 10 is the share of time a generator's output is switched on, as in a rating of 10 seconds on and 30 off, and it is the sense of IEC 60601-2-2's 45 % rule (Chapters 8 and 10).
- Electromagnetic compatibility (EMC)
- The interface between a device and the other equipment in the room through emissions and immunity; for a surgical generator, set by IEC 60601-1-2 as modified by IEC 60601-2-2, with emissions tested in the idle state (Chapters 19 and 21).
- Electroporation
- The opening of pores in cell membranes by the voltage an external electric field induces across them, reported above about 0.5 to 1 V; reversible when the pores reseal and the cell survives, irreversible when the cell cannot repair them and dies (Chapter 15).
- Essential performance
- The performance a device must maintain for the patient to be safe, as identified by the manufacturer's risk analysis; for a generator, an output that matches its setting within its declared limits, appears only when commanded and stops when commanded (IEC 60601-1; Chapter 20).
- EUDAMED
- The EU's register for devices; four of its modules, for actor registration, the registration of devices and their unique device identifiers, notified bodies and certificates, and market surveillance, became mandatory on 28 May 2026, and vigilance was not among them (Chapter 27).
- Formative evaluation
- In FDA's human factors guidance, a test of a device's user interface while the interface is still changing, before human factors validation testing on the final design (Chapter 22).
- Fulguration
- Coagulation of a tissue surface by bursts of current at kilovolts that jump an air gap and dry and char the surface over a broad area, with no contact; also called spray coagulation (Chapter 8).
- General standard
- IEC 60601-1, the standard for the basic safety and essential performance of all medical electrical equipment, at edition 3.2 of 2020 as of October 2026 (Chapters 20 and 25).
- Harm
- In ISO 14971:2019, injury or damage to the health of people, or damage to property or the environment, such as a skin burn under a lifted pad (Chapter 22).
- Hazard
- In ISO 14971:2019, a potential source of harm, such as the monopolar current that leaves the patient through the return pad (Chapter 22).
- Hazardous situation
- In ISO 14971:2019, circumstances in which people, property or the environment are exposed to a hazard, such as a pad partly lifted during activation (Chapter 22).
- Heat sink
- Cooling of the tissue around a large vessel by the blood flowing through it; the "3 mm rule" of ablation practice treats a vessel 3 mm or more across beside a tumor as a heat sink (Chapter 2).
- Heating factor
- Current squared multiplied by time; under IEC 60601-2-2, a monopolar mode is a high current mode when its intended use, at maximum current and maximum activation duty cycle, gives more than 30 A²s in any 60 seconds (Chapter 10).
- HIFU
- High-intensity focused ultrasound, in which sound absorbed at a small focus deep inside the body heats it during exposures that last seconds (Chapter 16).
- High current mode
- Under IEC 60601-2-2, a monopolar mode whose intended use, at maximum current and maximum activation duty cycle, gives more than 30 A²s in any 60 seconds; such a mode brings stricter requirements for the return electrode (Chapter 10).
- High-frequency leakage current
- High-frequency current that escapes from an electrosurgical output by paths other than the intended return, through stray capacitance to earth; IEC 60601-2-2 limits it to 150 mA for monopolar outputs, measured through a 200 Ω resistor that stands in for a person (Chapter 20).
- Histotripsy
- Mechanical destruction of tissue by focused ultrasound pulses lasting microseconds, at a duty cycle below 1 %, whose inertial cavitation breaks cells apart while the tissue barely warms (Chapter 16).
- Human factors evaluation
- A study of how users work with a device, such as Intuitive's study of how users work with its E-200 generator in its integrated, backup and standalone set-ups (Chapter 21).
- Human factors validation testing
- FDA's term for the test at the end of development, on the final user interface, with in general at least 15 participants for each distinct user population; the Johner Institute notes that IEC 62366-1 calls it summative evaluation (Chapter 22).
- Hybrid cannula
- A metal cannula held in the abdominal wall by a plastic anchor that insulates it from the wall, so that current coupled into the metal can discharge into whatever touches it, which may be bowel (Chapter 18).
- Hybrid instrument
- An instrument that combines two energies in one jaw, such as bipolar current and ultrasonic vibration (Chapter 7).
- Impedance
- The opposition a circuit offers to an alternating current, measured in ohms like a resistance (Before we start).
- Impedance roll-off
- The point in RF ablation at which tissue at the needle dries and chars, the impedance rises sharply and the generator can no longer deliver power, so the zone stops growing (Chapter 13).
- Indication
- The use a device's maker claims for it, which a clearance or approval covers as the label states it, such as sealing vessels up to and including 7 mm in diameter; a general soft-tissue indication covers soft tissue in general rather than one procedure or organ (Chapters 2 and 15).
- Information for safety
- The third option for controlling a risk under ISO 14971, used where design and protective measures reduce the risk too little, such as the instruction not to cut or fold a pad; it ranks last because it works only if people act on it (Chapter 22).
- Inherently safe design
- The first and preferred option for controlling a risk under ISO 14971, in BSI's wording to eliminate the risk by the design itself, such as a return pad large enough to stay below the threshold of necrosis when two thirds of it lifts (Chapter 22).
- Interface
- A place where two parts of an energy system meet, or where a part meets a person: high-frequency power, signals, data or procedures cross it, and the party that writes its specification owns it (Chapter 21).
- Irreversible electroporation (IRE)
- Tumor ablation by electric pulses that open membrane pores the cells cannot repair (Chapter 15).
- Isolated output
- A generator output that floats electrically instead of being referenced to earth, so that current returns through the generator's own return lead rather than through earthed contacts on the patient; high-frequency current that still escapes to earth through stray capacitance is limited as leakage (Chapters 8 and 20).
- Joule heating
- Heating by current flowing through a resistance; in tissue, the heat released in each unit of volume is the square of the current density divided by the conductivity (Chapter 3).
- Joule–Thomson effect
- The temperature change of a real gas under high pressure expanding through a narrow orifice without doing work; argon and nitrogen cool, while helium, at room temperature, warms (Chapter 14).
- Langevin stack
- A few piezoelectric rings clamped between a heavy back mass and a front mass by a bolt through the middle, which keeps the rings in compression; the standard arrangement of an ultrasonic instrument's transducer (Chapter 11).
- Lateral thermal spread
- Thermal damage beyond the tissue a device was meant to affect, measured after the fact by cutting and staining the tissue and measuring the damaged zone under magnification (Chapter 2).
- Lethal isotherm
- The surface of equal temperature at the temperature that kills tissue, and so the edge of an ablation zone: about 60 °C in RF ablation, and below about −20 to −40 °C in cryoablation, several millimeters inside the visible edge of the ice (Chapters 13 and 14).
- Load
- The tissue as a generator sees it: an electrical impedance that changes as the tissue heats and dries, from about 20 Ω in a sealer's jaws to 2 kΩ or more across an arc (Before we start; Chapter 10).
- Maximum output voltage
- IEC 60601-2-2's figure for a generator: for each mode, the highest peak voltage that can appear between its patient connections; FDA asks for it in the generator's labeling (Chapters 10 and 21).
- Means of protection
- The barriers that keep a device's mains side apart from the patient and the operator, even after a fault: insulation, clearances through air and creepage distances along surfaces, each rated for a test voltage, with patient protection and operator protection rated separately (IEC 60601-1; Chapter 20).
- Medical device report
- A report to FDA of a death, serious injury or malfunction involving a device, which FDA counts by product code; FDA counted 46,673 under the ultrasonic-instrument code LFL from 2002 to 2021, 95 % of them malfunctions (Chapter 26).
- Monopolar
- Electrosurgery in which current flows from an active electrode through the patient to a return pad (Chapter 8).
- Motional current
- The current in the resonant branch of a piezoelectric transducer, proportional to the velocity of the blade; the generator regulates it to hold the blade's motion constant (Chapter 11).
- Neutral electrode
- IEC 60601-2-2's term for the return electrode, or pad; in the standard's heating test a pad carries a current set by the weight category of patient it is rated for, and the skin beneath it must not warm by more than 6 °C (Chapter 25).
- Node
- A point of no motion in a standing wave, half a wavelength from the next; the supports of an ultrasonic waveguide sit at nodes (Chapter 11).
- Nominal ocular hazard distance
- The distance along the direct beam over which a laser's exposure stays above the eye's maximum permissible exposure; reflections and focusing optics define a wider nominal hazard zone (Chapter 19).
- Notified body
- An organization designated under the EU's Medical Device Regulation to assess the conformity of devices; for a class IIb device, under the usual route, it audits the maker's quality management system, assesses the technical documentation of at least one representative device in each generic device group and issues certificates (Chapters 22 and 27).
- Particular standard
- A standard for one type of medical electrical equipment that modifies the general standard and overrides it where the two differ, such as IEC 60601-2-2 for electrosurgery (Chapter 25).
- Piezoelectric effect
- The change of shape of certain crystals and ceramics when a voltage is applied across them, and the voltage they produce when they are squeezed (Chapter 11).
- Population inversion
- More atoms in the upper laser level than in the level below it, the state a pump creates so that light passing through the gain medium is amplified by stimulated emission rather than absorbed (Chapter 12).
- Predicate device
- A legally marketed device to which a new device is shown to be substantially equivalent in a 510(k) (Chapter 9).
- Premarket approval (PMA)
- FDA's route for class III devices, the class of highest risk, by which a device is approved rather than cleared; the pulsed field ablation systems approved in the US, four between December 2023 and November 2024 and two more in 2025, reached patients this way (Chapters 16, 22 and 26).
- Process standard
- A standard that governs how a device is designed and made rather than what it does, such as ISO 13485 for the quality management system, ISO 14971 for risk management, IEC 62304 for the software life cycle and IEC 62366-1 for usability engineering (Chapter 25).
- Product code
- The three-letter code under which FDA files every record for one type of device, such as GEI for electrosurgical devices; a code can hold devices beyond the surgical ones, such as aesthetic RF devices under GEI (Chapter 23).
- Product standard
- A standard for how a kind of product's output is measured or how it is classed, such as IEC 61847 for ultrasonic surgical systems and IEC 60825-1 for laser hazard classes (Chapter 25).
- Protective measure
- The second option for controlling a risk under ISO 14971, after inherently safe design; on a lifting pad, the split pad and its monitor, which stops the generator when the impedance between the halves rises (Chapter 22).
- Pulsed field ablation (PFA)
- Irreversible electroporation used in the heart, to make the lines of scar that treat atrial fibrillation, usually with short biphasic pulses (Chapter 15).
- Quality Management System Regulation (QMSR)
- The US rule for device makers' quality systems, 21 CFR Part 820, which since 2 February 2026 incorporates ISO 13485:2016 by reference and replaces the older Quality System Regulation (Chapter 26).
- Rated accessory voltage
- The highest peak voltage an active accessory is designed to withstand, stated in its instructions for use; its insulation is tested at 120 % of it for 30 seconds, and the generator's peak in every mode used with it must stay at or below it (Chapters 18 and 21).
- Rated load
- The impedance at which a maker specifies a generator mode's power and waveform, such as 20 Ω for the FT10's sealing and 300 Ω for its monopolar cut (Chapter 10).
- Recall class
- FDA's grading of a correction or removal of a marketed device, on a scale of its own, separate from the device classes. FDA calls Class I "the most serious type of recall", and such a recall need not remove anything, as with the 2025 Varipulse correction of the catheter's instructions; less serious actions fall in Class II (Chapter 26).
- Recognized consensus standard
- A standard, in a stated edition, that FDA lists as one to which a maker may declare conformity in a premarket submission; FDA entered edition 6.1 of IEC 60601-2-2 on 25 May 2026 and accepts declarations to edition 6.0 until 1 July 2029 (Chapter 25).
- Recurrent test
- A safety test of medical electrical equipment in service, made at intervals and after repair under IEC 62353, whose abstract describes safety tests, such as leakage, and no test of output; a generator's output is checked by its maker's own procedure (Chapter 21).
- Return electrode
- The large pad on the patient's skin through which monopolar current leaves the body and returns to the generator; IEC 60601-2-2 calls it the neutral electrode (Before we start; Chapter 8).
- Rule 9
- The EU Medical Device Regulation's classification rule for active therapeutic devices that administer or exchange energy: class IIa unless they may do so in a potentially hazardous way, then class IIb, judged by the nature, density and site of application of the energy (Chapter 27).
- Sealing
- Closing a vessel held shut under pressure by heating it until the collagen of its walls fuses into a band strong enough to replace a suture, so that it stays shut when it is divided (Before we start; Chapter 1).
- Spark gap
- A small air gap across which each spark sets off a burst of high-frequency oscillation; Bovie's apparatus made both its currents this way (the opening story; Chapter 8).
- Special controls
- In the US, requirements written for one class II device type, such as particular tests and labeling, in addition to the rules for every device (Chapter 26).
- Split pad
- A return pad in two halves: the generator passes a small interrogating current from one half to the other through the skin, and stops when the impedance between them leaves its window, as it does when the pad lifts (Chapter 8).
- Summative evaluation
- IEC 62366-1's name, as the Johner Institute gives it, for human factors validation testing, the test of the final user interface at the end of development (Chapter 22).
- Thermal relaxation time
- The time a heated layer takes to lose its heat by conduction, roughly its thickness squared divided by four times the thermal diffusivity: about half a millisecond for the 16 µm layer that absorbs CO₂ light (Chapter 5).
- Traceability
- The links from each claim to the requirements that serve it, the tests that verify them and the label line that limits the claim, kept so that a change shows which tests must be repeated (Chapter 22).
- Transient ischemic attack
- A brief stroke-like episode; with stroke, the event that 4 of 132 patients in the Varipulse catheter's US evaluation had, about 3 % against an expected 1 % or less (Chapters 22 and 24).
- Use error
- In FDA's human factors guidance, a user action or lack of action that was different from what the manufacturer expected (Chapter 22).
- Use life
- The number of times a reusable instrument can be reused, which FDA's reprocessing guidance asks the maker to state from testing, with a way for users to tell when it is exceeded; Intuitive gives 10 uses for its da Vinci monopolar scissors and Covidien 50 for its reusable pencil (Chapter 21).
- Validation
- The assessment of whether devices made to the design meet user needs and intended uses, for a sealer by a chronic animal study, simulated use and human factors validation testing (Chapter 22).
- Vaporization
- Explosive boiling of the water in cells, which bursts them when energy arrives in a thin layer faster than steam can escape; in electrosurgery it begins once the water in the cells reaches about 100 °C, and a moving line of bursting cells is an incision (Chapter 1).
- Verification
- The check of each design output against its design input, such as a zero-failure burst test of seals (Chapter 22).
Let's Talk