27 — The EU and India

Same evidence, other assessors.

+ The questionHow do the European Union and India classify and assess the same devices, and what does each add?

Rule 9 in its own words

Sixteen words in the European Union's Medical Device Regulation decide whether an energy device is class IIa or IIb. Regulation (EU) 2017/745 classifies devices by the rules of its Annex VIII, and Rule 9 covers active therapeutic devices intended to administer or exchange energy. Such devices are class IIa unless they "may administer energy to or exchange energy with the human body in a potentially hazardous way", in which case they are class IIb. The rule names what decides the hazard: "the nature, the density and site of application of the energy". Those are the variables of this book. Chapter 6 showed that the nature of the carrier sets where its energy is deposited, Chapter 3 that current heats where its density is highest, and Chapters 2 and 13 that the site changes what the same dose does.

The Medical Device Coordination Group's guidance on classification, MDCG 2021-24, was revised as rev. 1, published by the Commission on 20 April 2026. The revision restructured the guidance's section on Rule 9 "to better match the rule" and kept the examples. As class IIb under Rule 9 it names "high-frequency electrosurgical generators, and electrocautery equipment, including their electrodes", "surgical lasers", "lithotriptors, surgical ultrasound devices" and "high-intensity focused ultrasound (HIFU)". By contrast, its class IIa examples include muscle stimulators and phototherapy. It places the cryo-ablation of the heart or spine under another rule, Rule 6, in class III, and it does not mention pulsed field ablation under Rule 9. The guidance does not bind; the rule in Annex VIII does.

71 — Rule 9: class IIa or class IIb
active therapeutic device administering or exchanging energy MDR Annex VIII, Rule 9 may it administer or exchange energy in a potentially hazardous way? NATURE DENSITY SITE OF APPLICATION the rule weighs nature, density and site of application yes no class IIb high-frequency electrosurgical generators, and electrocautery equipment, including their electrodes surgical lasers lithotriptors, surgical ultrasound devices high-intensity focused ultrasound (HIFU) class IIa muscle stimulators, phototherapy cryo-ablation of the heart or spine: Rule 6, class III pulsed field ablation: not named under Rule 9 examples and wording as in MDCG 2021-24 rev. 1 (April 2026)
ElectricalAcousticOpticalColdFocal detail
Plate 71 — Rule 9 asks one question of an energy device, and the three things it tells the assessor to weigh are the variables of this book: what carries the energy, how concentrated it is, and where it is applied. The coordination group's examples of class IIb are the electrosurgical, laser and ultrasonic devices of the earlier chapters; cryo-ablation of the heart or spine falls under another rule, in class III.

A notified body, a certificate and a register

A class IIb device needs a notified body, an organization designated under the regulation to assess conformity. Under the usual route, in Annex IX, the notified body audits the manufacturer's whole quality management system and assesses the technical documentation of at least one representative device in each generic device group; a route by type examination is also available. The notified body then issues its certificates, and the manufacturer affixes the CE marking.

What happens when a certified device changes depends on what the notified body certified. For a device whose technical documentation a notified body has certified, such as a class III device, Annex IX requires the notified body's approval for any change that "could affect the safety and performance of the device or the conditions prescribed for use of the device". A generator's documentation is assessed through a representative device of its group rather than certified device by device. The maker of a generator therefore informs its notified body of any planned substantial change to its quality management system or to the devices it covers, and any approval is issued as a supplement to the quality management system certificate. The EU's test for a certified device is the one Chapter 26 met in the US, asked without the word "significantly".

The EU's register for devices, EUDAMED, became partly mandatory on 28 May 2026. Four of its modules apply from that day: actor registration, device registration with unique device identifiers, notified bodies and certificates, and market surveillance. Ireland's Health Products Regulatory Authority sets out the transition: devices placed on the market before that date are to be registered by 28 November 2026, and certificates issued before mandatory use by 28 May 2027. Vigilance, the reporting of serious incidents after sale, was not among the four modules.

A transition and a proposal

Devices certified under the older directives are still moving across to the regulation. Regulation (EU) 2023/607, adopted on 15 March 2023, came into force on its publication on 20 March 2023 and extended the validity of certificates issued under the directives. For class III devices and implantable class IIb devices, with listed exceptions, the extension runs to 31 December 2027. For other class IIb devices, and for class IIa and sterile or measuring class I devices, it runs to 31 December 2028. It applies only where, among other conditions, the manufacturer had put a quality management system in place and lodged a formal application with a notified body by 26 May 2024, and signed a written agreement with it by 26 September 2024.

Solved example: how long a directive certificate for a generator can run

An electrosurgical generator is a non-implantable class IIb device under Rule 9, so the amended transition puts it in the later group, with a deadline of 31 December 2028 rather than 31 December 2027. Its directive certificate stays valid that long only if its maker met the conditions of the transition, among them an application by 26 May 2024 and an agreement by 26 September 2024. From the later of those dates to the deadline is four years, three months and five days, or about 51 months. From the end of October 2026, the month this book is current to, the deadline is two years and two months away, or 26 months. A generator still sold on a directive certificate in October 2026 therefore has a little over two years left on it, unless a later act moves the date.

On 16 December 2025 the Commission proposed a targeted revision of the regulation, to simplify it and reduce its burden. Its explanatory memorandum states that the proposal would remove the five-year maximum validity of certificates in favor of risk-proportionate periodic reviews, let notified bodies replace on-site audits with remote ones, and set surveillance audits every two years where there are no safety issues. The proposal would also adapt some classification rules to lower classes, naming reusable surgical instruments among others, and the memorandum does not name Rule 9. As of October 2026 the proposal has not been adopted. The Council's progress report of 5 June 2026 records that the member states had not reached a general approach, the Council's agreed position for negotiating the text with the Parliament. The report states that the European Parliament aims to vote on its position in early 2027, an aim rather than a scheduled date.

72 — The EU transition and the proposal, 2023 to 2028
2023 2024 2025 2026 2027 2028 IN FORCE 20 Mar 2023 Regulation 2023/607 in force 26 May 2024 application to a notified body due 26 Sep 2024 written agreement due MDCG 2021-24 rev. 1 20 Apr 2026 EUDAMED: four modules mandatory 28 May 2026 register devices placed before 28 May 2026 28 Nov 2026 register certificates issued before 28 May 2026 28 May 2027 directive certificates end: class III and implantable IIb 31 Dec 2027 directive certificates end: other IIb, 31 Dec 2028 IIa and sterile or measuring I, a generator among them PROPOSED, NOT ADOPTED 16 Dec 2025 Commission proposal 5 Jun 2026 Council progress report: no general approach Parliament's aim: vote on its position, early 2027 this guide 26 months from the end of October 2026 dates from Regulation (EU) 2023/607, HPRA (EUDAMED) and the Council's report
Focal detail
Plate 72 — The transition applies only where, among other conditions, the maker lodged an application by 26 May 2024 and signed an agreement by 26 September 2024. Directive certificates then end on 31 December 2027 or 31 December 2028 by class. A non-implantable class IIb generator falls in the later group. The proposal of December 2025, below the line, had not been adopted in October 2026.

India: the Medical Devices Rules

India regulates devices under the Medical Devices Rules, 2017, which set four risk classes, A to D. Its regulator, the Central Drugs Standard Control Organisation, or CDSCO, publishes classification lists by specialty. Its 2021 lists for operation theatre devices and for dermatology and plastic surgery place general-purpose electrosurgical units and laser surgical instruments in class C, the class for moderate to high risk.

CDSCO began a new cycle of lists with a draft for comment in January 2025, and issued its final oncology list on 2 January 2026. The list has 77 entries, and the energy devices it names are all in class C: cryoablation systems, electroporation therapy systems, microwave ablation systems, a microwave or electrosurgical system generator, HIFU ablation systems and ultrasonic ablation systems. Not every device in this book could be placed in the new lists: no CDSCO class for vessel sealers, ultrasonic shears or cardiac pulsed field catheters was found in any list issued since late 2025.

The class sets who licenses the device. CDSCO's FAQ on the Rules states that the Central Licensing Authority licenses class C and D devices, while lower classes are licensed by the states or only registered, so a generator or a surgical laser is licensed centrally. An importer applies on Form MD-14 and receives its import license on Form MD-15. The FAQ gives timelines: an inspection of a class C or D manufacturing site within 60 days of the application, an import license within nine months, and approval of a major change within 45 days for a manufacturer or 60 days for an importer. A license stays valid until it is suspended or cancelled, against a retention fee every five years.

A device must meet the standards of the Bureau of Indian Standards where they exist, then ISO or IEC standards, then pharmacopoeial or validated manufacturer's standards, and after sale the license holder must submit post-marketing surveillance data. For a generator, the IEC tests of Chapter 25 therefore apply wherever no Indian standard covers the same ground.

One evidence core, filed three ways

The three systems ask in different words and through different assessors, but they draw on one body of evidence: a quality management system to ISO 13485, a risk management file to ISO 14971, IEC 60601 test reports, software documentation to IEC 62304, usability files, and the tissue and animal testing of Chapter 25. An electrosurgical generator shows how the forms differ. In the US the generator is class II under product code GEI, FDA clears it by 510(k), and its summary becomes public, as the VIO 3n's did. In the EU it is class IIb under Rule 9, a notified body audits its maker and assesses its technical documentation, and the maker affixes the CE marking. In India it is class C, licensed by the Central Licensing Authority against Indian standards first. A maker that builds the core once and keeps it current can present it in the form each assessor asks for, while one that builds it separately for each market repeats most of the work.

Planning one dossier for three markets

Classify the device in each target market before planning the evidence: the US product code and route, the EU rule and class, and the CDSCO list and class, noting where a list gives no class. Build the quality system, the risk file and the test reports once, to editions that each regulator accepts on the planned dates of submission. Keep one register of changes, and record for each change the US, EU and Indian answer to whether it needs a new submission or approval.

From Bovie's knobs to a public record

In the autumn of 1926, at the Peter Bent Brigham Hospital, William Bovie adjusted the knobs of his apparatus while Harvey Cushing operated. No standard governed the apparatus and no regulator had seen it. Its current went where the circuit allowed. By the account in Li and Isaacson's 2024 history of the collaboration, a shock once passed through a metal retractor to Cushing's arm, and the first unit lacked an effective return electrode. Each of those hazards now has a test. The current that escapes a generator by any path other than its return lead is measured through a 200 Ω resistor, and in the standard's test a return pad must keep the skin beneath it within 6 °C of its starting temperature (Chapter 25). A new generator may be sold in the United States only after FDA clears it, on evidence that includes test reports against IEC 60601-2-2 and the tissue tests of FDA's guidance. Whereas no regulator had seen Bovie's apparatus, the summary for Erbe's VIO 3n generator, cleared on 29 August 2025, is open to anyone.

The operating room of 1926 and the public record of 2025 together hold the answer this book has built. Where energy goes when it enters tissue is set by physics: current heats where it is densest, light where its wavelength is absorbed, focused sound at its focus, and the heat then spreads about 0.75 mm in one second and about 2.4 mm in ten (Chapters 2, 3, 5 and 16). Surgery needs eleven kinds of energy because its tasks need different geometries and times: a cut along a line, a seal across a vessel's wall, a zone of dead tissue around a needle, a lesion deep inside a closed body, and a kill that spares the vessels and ducts running through it. Keeping each energy where the surgeon intends takes three things more: a design that confines it, such as an isolated output, a monitored pad and a safety channel; tests that show the confinement holds; and an assessment by someone other than the maker, whether FDA, a notified body or CDSCO.

The physics of 1926 still holds. Bovie's patent, filed in 1928, separated two currents: a steady one that cuts, and a heavily damped one at about 4,000 V that dries tissue and stops bleeding. Nearly a century later, Medtronic declares the same distinction for each mode of its FT10 generator as a duty cycle and a crest factor (Chapter 8). The years since have added an account of where that current goes: the path by which it returns, the limits it must stay within, and a record of what was tested before it reached a patient.

+ What this chapter established
  • Under the EU's Rule 9, an energy device is class IIb when it may administer or exchange energy in a potentially hazardous way, judged by the nature, density and site of the energy, and the MDCG guidance of April 2026 keeps electrosurgical generators, surgical lasers, surgical ultrasound and HIFU as its examples.
  • A class IIb device needs a notified body's assessment, directive certificates for non-implantable class IIb devices can run to 31 December 2028 under 2023/607, and the Commission's December 2025 proposal has not been adopted as of October 2026.
  • Under its 2021 lists and its oncology list of 2 January 2026, India places electrosurgical units, surgical lasers and ablation systems in class C, licenses them centrally and applies Indian standards first.
  • One core of quality, risk and test evidence serves all three jurisdictions; physics decides where energy goes, the tasks of surgery explain the eleven kinds, and design, test and assessment keep each where the surgeon intends.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.