05 — Light in tissue

Light in tissue.

+ The questionWhy does light from one laser stop within a hundredth of a millimeter of the surface, while light from another reaches millimeters deep?

Absorbers and scatterers

Tissue is about three quarters water, and the light of a carbon dioxide laser is absorbed within about 16 µm of its surface, yet the light of a diode laser at 980 nm enters the same tissue and spreads over millimeters before it is absorbed. The difference lies in which parts of the tissue absorb each wavelength and which mostly scatter it. Some molecules absorb light, taking each photon's energy and turning it into heat. These absorbers are called chromophores, and three matter most in surgery: water, which is everywhere, hemoglobin in the blood and melanin in pigmented skin. Cell membranes, fibers, organelles and the rest of the tissue mostly scatter light, bending it in new directions without absorbing it.

How strongly a chromophore absorbs depends on the wavelength, and the dependence is often enormous. Water is almost perfectly transparent to green light and almost opaque to much of the infrared, while hemoglobin absorbs green light strongly and near-infrared light weakly. A laser emits one wavelength, so choosing a laser means choosing which chromophore will do the absorbing, and therefore where the heat will appear.

The Beer–Lambert rule

The Beer–Lambert rule describes light that is absorbed and not scattered: its intensity falls by the same fraction over every equal step of depth. How fast it falls is set by the absorption coefficient, written µa, which is the probability of absorption per centimeter. At a depth of one divided by the absorption coefficient, the intensity has fallen to about 37 % of its value at the surface, and that depth is the absorption depth, within which most of the light's energy, and so most of the heat, is deposited.

Across the surgical wavelengths, published absorption coefficients for water span about seven and a half orders of magnitude, so the strongest absorption is tens of millions of times the weakest. A carbon dioxide laser emits at 10.6 µm, where water absorbs about 855 per centimeter, so the absorption depth in water is about 12 µm. A KTP laser emits green light at 532 nm, where water absorbs only about 0.00044 per centimeter, which gives an absorption depth of more than twenty meters.

14 — What absorbs each surgical laser
0.0001 0.01 1 100 10,000 µa, per cm 0.5 1 2 5 10 µm (log scale) WAVELENGTH KTP 532 diode 980 Nd:YAG 1064 diode 1470 thulium fiber 1940 Ho:YAG 2100 Er:YAG 2940 CO₂ 10,600 laser lines, nm water water: schematic between the marked values water data: Hale & Querry 1973; Pope & Fry 1997 at 532 nm; Segelstein 1981 at 1.47 µm hemoglobin in whole blood ≈ 235 0.00044 at 532 nm blood absorbs, water hardly at all
OpticalFocal detail
Plate 14 — Water's absorption of these wavelengths spans about seven and a half orders of magnitude. The lasers beyond 1.4 µm are absorbed by water within micrometers to a fraction of a millimeter; the near-infrared lines pass through water and are absorbed weakly; at 532 nm water is almost transparent and hemoglobin does the absorbing.

Water rules the infrared beyond 1.4 µm

Most surgical lasers work in the infrared, where water absorbs strongly, so the depth at which their light is absorbed follows almost directly from water's absorption coefficient. Tissue is about 75 % water, so it absorbs these wavelengths about three quarters as strongly as pure water, and each depth in tissue is about a third longer than in pure water.

LaserWavelengthWater absorption (per cm)Absorption depth in tissue
Er:YAG2.94 µmabout 12,500about 1 µm
CO₂10.6 µmabout 855about 16 µm
Thulium fiber1.94 µmabout 120about 0.11 mm
Diode1.47 µmabout 29about 0.47 mm
Ho:YAG2.1 µmabout 27about 0.5 mm
15 — How deep each laser's energy lands
ABSORPTION DEPTH IN 75 %-WATER TISSUE Er:YAG 2.94 µm ≈ 1 µm CO₂ 10.6 µm ≈ 16 µm thulium fiber 1.94 µm ≈ 0.11 mm Ho:YAG 2.1 µm ≈ 0.5 mm diode 1.47 µm ≈ 0.47 mm KTP 532 nm set by hemoglobin; about 43 µm in whole blood diode 980 nm Nd:YAG 1064 nm millimeters, set by scattering 0.11 mm against 0.5 mm: four to five times shallower 1 µm 10 µm 0.1 mm 1 mm 1 cm 10 cm absorption only, 1/(0.75 µa); real depth is shorter where scattering matters
OpticalFocal detail
Plate 15 — Where water is the absorber, the depth follows directly from the absorption coefficient: about a micrometer for Er:YAG, 16 µm for CO₂, a tenth of a millimeter for thulium and half a millimeter for holmium. The near-infrared lines reach millimeters because water barely absorbs them, and scattering, not absorption, sets how far they go.

Each laser's character follows from its absorption depth. An Er:YAG laser deposits its energy within about a micrometer of the surface, so it removes tissue in thin layers and leaves very little heat behind. A CO₂ laser deposits its energy in a layer about 16 µm deep and vaporizes that layer, which makes a precise incision but seals only the smallest vessels, because very little tissue below the cut is heated. The light of holmium and thulium lasers is absorbed within about half a millimeter and a tenth of a millimeter, and both lasers cut and coagulate, but because water absorbs thulium's light about four to five times more strongly than holmium's, thulium puts the same energy into a shallower layer.

Thermal damage reaches below the absorption depth

The depths in the table are optical: they show where the light is absorbed. The thermal damage beneath an incision usually reaches deeper, because heat conducts outward from the absorbing layer during and after the pulse. In a 2012 position paper for the Academy of Laser Dentistry on lasers in oral soft tissue, Pang and colleagues give thermally affected layers of 10 to 50 µm for Er:YAG, ten to fifty times its optical depth. How much deeper the damage reaches depends on the length of the pulse, which the last section of this chapter takes up.

Hemoglobin and scattering at shorter wavelengths

Toward the visible end of the spectrum, water absorbs very little, and two other effects take over: absorption by hemoglobin, and scattering. At 532 nm, oxygenated whole blood absorbs about 235 per centimeter, an absorption depth of about 43 µm, while the water around it is almost transparent. Green light therefore passes through water or saline irrigation and is absorbed by tissue filled with blood. That difference is the basis of KTP laser vaporization of the prostate, a procedure that runs under irrigation. Blood absorbs roughly 40 to 200 times more strongly at 532 nm than near 1,000 nm, depending on how much oxygen it carries.

At 980 and 1,064 nm neither water nor hemoglobin absorbs strongly, so scattering dominates, and the light bounces around inside the tissue and spreads over millimeters before it is absorbed. Diode lasers at 980 nm and Nd:YAG lasers at 1,064 nm therefore coagulate deeply rather than cut precisely. Scattering sets their depth, and because scattering varies with the tissue, published depths for these lines are less consistent than those for the lines that water absorbs.

Time: confining the heat

Chapter 2 showed that heat needs time to move. A layer that has absorbed light cools by conducting heat into its surroundings, and the time this cooling takes is the thermal relaxation time. It is roughly the layer's thickness squared divided by four times the thermal diffusivity, which for water and most soft tissues is about 0.14 square millimeters per second.

Solved example: the thermal relaxation times of three layers

Four times the thermal diffusivity of tissue is 4 × 0.14, or 0.56 square millimeters per second. The layer that absorbs CO₂ light is about 16 µm, or 0.016 mm, thick. Its thickness squared is 0.000256 square millimeters, and dividing that by 0.56 gives about 0.00046 seconds, a relaxation time of about half a millisecond. The layer that absorbs Er:YAG light is about 1 µm thick, sixteen times thinner, so its relaxation time is sixteen squared, or 256 times, shorter: about 2 microseconds. A layer 0.1 mm thick, the size of a small blood vessel, has a thickness squared of 0.01 square millimeters, and dividing by 0.56 gives about 18 milliseconds. The relaxation time therefore grows with the square of the thickness, so a layer sixteen times thinner loses its heat in a 256th of the time.

When the energy arrives in a pulse shorter than the relaxation time, the absorbing layer heats and vaporizes before it can pass heat to its neighbors, and little tissue beneath the crater is damaged. This is thermal confinement. When the pulse is longer, heat spreads while energy is still arriving, and more tissue is coagulated. Neither choice is right in general, because confinement gives precision while spreading gives hemostasis, the stopping of bleeding.

16 — Confining heat with a short pulse
PULSE SHORTER THAN THE RELAXATION TIME PULSE LONGER THAN THE RELAXATION TIME laser pulse crater: layer vaporized shallower crater thick coagulated layer heat spread during the pulse thin heated layer heat had no time to leave relaxation time ≈ L² / 4α L: layer thickness; α: diffusivity CO₂ layer 16 µm: about 0.5 ms Er:YAG layer 1 µm: about 2 µs a 0.1 mm blood vessel: about 18 ms same 1 J: in 300 µs the peak power is about 3.3 kW; in 50 ms it is 20 W pulse glyphs: equal area, not to scale
OpticalHeatFocal detail
Plate 16 — Heat needs time to leave the layer that absorbed it. A pulse shorter than that time, about 0.5 ms for the 16 µm layer that absorbs CO₂ light, leaves the energy in place and the layer vaporizes with little damage below. A longer pulse gives heat time to spread and coagulates more tissue: useful for hemostasis, costly for precision.

Power is the rate at which energy arrives, so the length of a pulse also sets its peak power. One joule delivered in 300 µs is a peak of about 3.3 kW, while the same joule spread over 50 ms is 20 W. At very high peak power, especially in liquid, the absorbed energy boils water so fast that the vapor expands explosively and generates pressure waves. The effect then becomes partly mechanical as well as thermal, and Chapter 12 shows how lithotripsy lasers use it to break stones.

In practice

Read a surgical laser's specification as a set of choices about where the heat goes. Start with the wavelength, because it decides which chromophore absorbs the light and so how deep the energy is deposited, from about a micrometer for Er:YAG to millimeters for the lines at 980 and 1,064 nm. Then set the pulse length against the thermal relaxation time of the absorbing layer, because a shorter pulse confines the heat and a longer one coagulates the tissue around it. Check the peak power as well as the pulse energy, because at very high peak power the effect becomes partly mechanical. In verification, measure the thermally affected depth in tissue rather than quoting the optical depth, because heat conducts beyond the layer that absorbed the light.

+ What this chapter established
  • Tissue absorbs light through chromophores, mainly water and hemoglobin, and scatters it off everything else.
  • Without scattering, light loses the same fraction of its intensity over each equal step of depth, and water's absorption spans about seven and a half orders of magnitude across surgical wavelengths, from a depth of about 1 µm for Er:YAG to meters at 532 nm.
  • Water sets the depth of the infrared lasers beyond 1.4 µm, hemoglobin that of the 532 nm line, and scattering that of the 980 and 1,064 nm lines, which therefore coagulate deeply.
  • A pulse shorter than the thermal relaxation time confines the heat and gives precision, while a longer pulse spreads the heat and gives hemostasis.