+ Part IV · Sound and light

Energy without current.

The blade of an ultrasonic shear has no sharp edge. It sits at the end of a titanium rod made a whole number of half-wavelengths long, and at the 55.5 kHz of Ethicon's Harmonic instruments each half-wavelength is about 46 mm. The blade moves 50 to 100 µm back and forth 55,500 times a second. Ultrasonic instruments and lasers both deliver their energy with no current through the patient. One vibrates a blade at a mechanical resonance, and the other turns electricity into light of a single wavelength. This part follows each of them from the power supply to the tissue, and shows how its physics sets what it does well. Each also brings hazards of its own: Chapter 11 follows the heat an ultrasonic blade keeps after activation, and Chapter 19 takes up the hazards of laser light.

11 — Ultrasonic and hybrid

Ultrasonic and hybrid instruments.

+ The questionHow does a blade that is never sharp cut tissue and seal vessels without passing any current through the patient?

From voltage to motion

Ethicon's original Harmonic transducer was a stack of piezoelectric elements held between two aluminum cylinders. Elements of this kind turn electrical power into vibration through the piezoelectric effect: certain crystals and ceramics change shape when a voltage is applied across them, and produce a voltage when they are squeezed. The movement of a single piezoelectric ring is tiny, well under a nanometer per volt, so instruments stack several rings and drive them at their mechanical resonance, where small forces build up large motion.

The standard arrangement is called a Langevin stack. A bolt through the middle clamps a few piezoelectric rings between a heavy back mass and a front mass, and it keeps the rings in compression, because ceramic cracks easily under tension. The front mass tapers into a horn, which amplifies the motion toward its narrow end.

A waveguide cut to the wavelength

From the horn, the vibration travels down a slender titanium-alloy waveguide to the blade. Vibration in a rod is a sound wave, and at the operating frequency the rod carries a standing wave. A standing wave has points that never move, called nodes, half a wavelength apart, and between them points of maximum motion, called antinodes. Patents describe waveguides whose length is a whole number of half-wavelengths. The blade tip sits at an antinode, where its motion is largest, and the supports that hold the waveguide inside its sheath sit at nodes, where there is no motion to lose.

31 — The ultrasonic stack and its standing wave
NODE ANTINODE λ/2 ≈ 46 mm at 55.5 kHz titanium alloy, thin-rod speed about 5,090 m/s; 54 mm at 47 kHz PIEZOELECTRIC RINGS BACK MASS FRONT MASS HORN the taper amplifies the motion TITANIUM WAVEGUIDE mounts at nodes: no motion BLADE bolt keeps the rings in compression tip excursion 50–100 µm the tip sits where motion is largest
ElectricalAcousticFocal detail
Plate 31 — Piezoelectric rings clamped between two masses convert voltage into motion, a tapered horn amplifies it, and a titanium waveguide cut to whole half-wavelengths carries it to the blade. The waveguide is held at nodes, where nothing moves, and the blade tip sits at an antinode, where it moves 50–100 µm tens of thousands of times a second.

The half-wavelength follows from the speed of the wave and its frequency, with one subtlety. A slender rod vibrates in an extensional mode, whose speed is set by the material's stiffness and density. For titanium and its alloys that speed is about 5,000 to 5,100 m/s, which is lower than the bulk speed of sound that tables quote. Half a wavelength therefore comes to about 46 mm at 55.5 kHz, the operating frequency of Ethicon's Harmonic instruments, and to about 54 mm at 47 kHz, the frequency Olympus lists for its current ultrasonic generator. The bulk speed of about 6,100 m/s would overestimate both by about 20 %.

Solved example: the length of a half-wave titanium waveguide

A titanium waveguide is to vibrate at 55.5 kHz. A slender rod carries the vibration at its extensional speed, about 5,090 m/s in titanium. A wavelength is the distance the wave travels during one cycle, so it is the speed divided by the frequency: 5,090 m/s divided by 55,500 cycles a second is about 0.092 m, or 92 mm. Half a wavelength is therefore about 46 mm, and the waveguide is made a whole number of these half-wavelengths long. At the 47 kHz Olympus lists, the same division gives about 108 mm, and half of it is about 54 mm. The bulk speed of about 6,100 m/s from a table would give about 55 mm at 55.5 kHz, some 20 % too long. So at 55.5 kHz each half-wave section of a titanium waveguide is about 46 mm long.

How far the blade travels in each cycle decides how hard it works the tissue. Homma's 2022 review of the energy devices used in lung cancer surgery puts this excursion at 50 to 100 µm, and Ethicon's own patents give about 40 to 100 µm. Neither says whether the figure is the full stroke, from one end of the motion to the other, or half of it. Chapter 4 took 100 µm as the full stroke. On that reading the tip at 55.5 kHz reaches about 17 m/s and accelerates at more than 600,000 times gravity, and pressed against tissue, that motion becomes heat.

The transducer as a circuit

Electrically, a piezoelectric stack looks like a capacitor in parallel with a resonant circuit. The capacitor is the ceramic's own capacitance, called C0. The resonant branch has an inductance that stands for the moving mass, a capacitance that stands for the stiffness, and a resistance that stands for the mechanical losses, which include the tissue pressed against the blade. This equivalent circuit is the Butterworth–Van Dyke model, and patents on ultrasonic generators describe the transducer in its terms.

32 — The transducer as a circuit
C0 the ceramic's own capacitance Lm moving mass Cm stiffness Rm losses and the tissue load MOTIONAL BRANCH the generator tracks zero phase and holds the current in the motional branch, which sets blade velocity 0° series resonance parallel resonance |Z| (LOG) PHASE FREQUENCY lock here tissue load and temperature move the resonance schematic
ElectricalReadoutFocal detail
Plate 32 — Electrically, the stack looks like a capacitor in parallel with a resonant branch whose inductance, capacitance and resistance stand for mass, stiffness and losses. The generator hunts for the frequency where the phase crosses zero and regulates the current in the resonant branch, because that current is proportional to how fast the blade moves.

The model explains how the generator works. The current in the resonant branch is called the motional current, and it is proportional to the velocity of the blade. The generator therefore regulates that current to hold the blade's motion constant, and it compensates for the current that flows through C0 and does no mechanical work.

The generator also tracks the resonance, because the tissue load and the temperature shift it. The original Harmonic generator used a phase-locked loop, a circuit that adjusted the generator's frequency up or down to stay at resonance, and later patents describe tracking the frequency at which the phase between voltage and current crosses zero. Because the stack works at resonance, the drive it needs is modest. A reseller's copy of Ethicon's specification for its GEN11 generator lists the ultrasonic output as 150 V RMS maximum and 35 W continuous, about half the voltage and a ninth of the power of the 300 W cut in Chapter 10's solved example.

How it cuts and seals

Pressed against tissue, the blade heats it by friction and by internal losses, as Chapter 4 described. Proteins unfold into a sticky coagulum that seals small vessels, and with enough pressure the tissue parts along the blade. Sealing depends on compression, heat and time, the same variables as bipolar sealing (Chapter 9). Standard ultrasonic shears are indicated for vessels up to 5 mm. In October 2013, FDA cleared Ethicon's Harmonic ACE+7 for vessels up to 7 mm, in a mode that modulates the delivery of energy during the seal, and Ethicon called it the first ultrasonic device indicated for vessels that large. Olympus's THUNDERBEAT, which delivers bipolar current together with ultrasonic energy, had already been cleared for 7 mm vessels in January 2012.

Because no current flows through the patient, ultrasonic instruments avoid the stray-current hazards of Chapter 18, and that matters most near structures such as bowel. They carry a different hazard instead: the heat retained in the blade.

Heat left in the blade

Ultrasonic cutting is often described as a low-temperature technique. For the tissue, the maker agrees: Houser's patent for Ethicon, filed in 2008, states that the coagulum seals small vessels at temperatures below 100 °C. The blade is another matter. Govekar and colleagues at the University of Colorado measured it in a bench study, reported as an abstract to a meeting of the Society of American Gastrointestinal and Endoscopic Surgeons. With thermal imaging, they recorded how far the temperature rose at the tips of laparoscopic energy devices fired four times for five seconds each. The rise averaged 172 ± 63 °C for ultrasonic devices, against 81 °C for monopolar and 46 °C for bipolar devices. Bovine liver that the monopolar and bipolar devices touched after activation was within 20 °C of its starting temperature by five seconds, while liver that the ultrasonic devices touched stayed more than 20 °C above it for the full 20 seconds observed.

33 — Heat left in an ultrasonic blade
ACTIVATION 0 50 100 150 200 250 0 5 10 15 20 25 °C RISE TIME, SECONDS 20 °C rise ultrasonic: 172 ± 63 °C monopolar: 81 ± 18 °C bipolar: 46 ± 19 °C still hot when the next tissue is touched fired with no tissue in the jaws, a blade has nothing to absorb its energy and runs hotter still peak rise measured at the device tips; after activation, the rise in tissue each device touched (bench study, Govekar, SAGES abstract) schematic curves through the published values
HeatFocal detail
Plate 33 — A vibrating blade makes heat by friction at the tissue and keeps it after activation stops. In one bench study the temperature rise measured at an ultrasonic device averaged 172 °C, and tissue it touched afterward stayed hot for the full 20 seconds observed, while tissue touched by bipolar and monopolar devices fell back within five. The ultrasonic hazard is the residual heat, not current.

Both descriptions are true, because they refer to different things. The tissue at the cut stays moderate because its water holds the temperature near 100 °C and the tissue is removed. The metal has no water and a small heat capacity, and it keeps its heat after the vibration stops. A blade fired with no tissue in its jaws has nothing to absorb its energy, so it runs hotter still. A hot blade that touches bowel on its way to the next bite causes a burn as surely as a current would.

An ultrasonic blade's hazard is the heat it keeps

The absence of current through the patient removes one class of hazard and leaves another. For ultrasonic instruments, the thermal hazard sits in the blade after activation, not in the tissue during it. Instructions for use, training and jaw design all address it, and so should any comparison of devices.

Hybrid instruments and aspirators

A hybrid instrument combines ultrasonic and bipolar energy in one jaw. Olympus launched THUNDERBEAT in March 2012, after FDA cleared it in January, and describes it as delivering ultrasonic and bipolar energy to the tissue at the same time. The two energies share the work. Bipolar current heats the tissue from within, while friction at the probe heats it from the surface and divides it, so one bite can seal and cut.

34 — One jaw, two energies
electrode electrode pad UPPER JAW frictional heat at the probe ULTRASONIC PROBE bipolar current two sources of heat in one bite Seal & Cut: bipolar and ultrasonic together Seal: bipolar only Ultrasonic only (THUNDERBEAT II): vessels to 3 mm, against 7 mm in the sealing modes ultrasonic 47 kHz; bipolar 380 kHz (Olympus's summary, 2025)
ElectricalAcousticHeatFocal detail
Plate 34 — A hybrid instrument puts an ultrasonic probe and bipolar electrodes in the same jaw. Bipolar current heats the tissue from within while friction at the probe heats it from the surface and divides it, so one bite can seal and cut. The summary Olympus filed for the newest version lists 47 kHz ultrasonic and 380 kHz bipolar output.

The current version illustrates how such devices evolve. FDA cleared THUNDERBEAT II on 7 October 2025. The summary Olympus filed for it lists ultrasonic output at 47 kHz and bipolar output at 380 kHz. It describes three modes: Seal & Cut with both energies, Seal with bipolar only, and an ultrasonic-only mode indicated for vessels up to 3 mm, against 7 mm for the two sealing modes. The generator of THUNDERBEAT II, the USG-410, was first cleared in 2021 and is listed again in the 2025 submission. Olympus calls the new transducer cordless, which here means that the cable has moved to the base of the handle, not that the device runs on a battery. Olympus also describes a thermal shield on the jaw that slows the transfer of heat to its outer surface.

A second family of ultrasonic instruments, the aspirators, does not seal at all. Vieira-Coimbra and Ferreira describe how they work in a prospective case series, published in 2025, on the laparoscopic excision of endometriosis with the CUSA aspirator. Integra LifeSciences, the maker of CUSA, supported its publication. An aspirator vibrates a hollow tip at 23 or 36 kHz and fragments the tissue just in front of it, mainly by the cavitation of Chapter 4 together with direct mechanical impact, while irrigation and suction carry the fragments away. Water-rich, weakly cohesive tissue such as liver parenchyma or tumor breaks up, while collagen-rich vessels, ducts and nerves survive to be clipped or sealed. The selectivity comes from the mechanics of the tissue rather than from heat.

In practice

Specify and verify an ultrasonic instrument against three properties from this chapter. First, size the waveguide from the extensional speed of the alloy actually used, because the bulk speed in a table makes every half-wavelength about 20 % too long. Second, measure the blade's temperature after activation and the time it takes to cool, because that retained heat is the hazard an instrument without current through the patient still carries. Third, state the vessel size for each mode separately, because one instrument can carry different limits, as THUNDERBEAT II does with 7 mm in its sealing modes and 3 mm in its ultrasonic-only mode.

+ What this chapter established
  • A Langevin stack of piezoelectric rings drives a titanium waveguide at resonance, and the waveguide is a whole number of half-wavelengths, about 46 mm each at 55.5 kHz, with the blade tip at an antinode.
  • The generator regulates the motional current, which is proportional to the blade's velocity, and tracks the resonance as the tissue load and the temperature shift it.
  • The blade heats tissue by friction and internal losses, seals vessels up to 5 mm, or 7 mm for newer designs, and parts the tissue along its edge with no current through the patient, but it keeps its heat after activation and can burn tissue it touches afterward.
  • Hybrid instruments combine bipolar and ultrasonic energy in one jaw, and aspirators fragment tissue mechanically, mainly by cavitation, while sparing vessels.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.