14 — Cryoablation

Cryoablation.

+ The questionEvery energy so far adds heat. How does taking heat away kill tissue, and why is the edge you can see not the edge that dies?

Cooling by expansion

A cryoablation probe is a needle a couple of millimeters wide, and its tip must get far colder than anything else in the operating room without freezing the needle's shaft or the skin it passes through. Most systems do this with a property of real gases. When a gas under high pressure expands through a narrow orifice without doing work on anything, its temperature changes, and this is the Joule–Thomson effect. In most gases at room temperature, including argon and nitrogen, attraction between the molecules dominates, so pulling them apart as the gas expands takes energy from their motion, and the gas cools. In helium at room temperature repulsion dominates, and the gas warms.

In an argon cryoprobe, gas from a high-pressure supply flows down a thin capillary inside the needle to its tip. There it expands through an orifice into the closed chamber of the tip and cools sharply. The cold gas absorbs heat from the surrounding tissue through the wall of the tip, then flows back along the space between the capillary and the outer wall and vents outside the patient. Argon boils at about −186 °C, which sets the floor. Hinshaw and colleagues report that argon systems cool the probe to −160 °C or colder. Their 2014 review compares microwave, RF and cryoablation tools, and three of its authors held shares in NeuWave Medical, the original maker of the NEUWAVE microwave system (Chapter 13).

41 — Cooling by expansion
TISSUE ICE BALL high-pressure argon supply cold gas returns along the annulus and vents JOULE–THOMSON EXPANSION: THE GAS COOLS all the cooling happens here argon systems: probe at −160 °C or colder; argon boils at about −186 °C, the floor to thaw: helium, which warms when it expands liquid-nitrogen systems circulate pressurized liquid instead; it boils at −196 °C at atmospheric pressure, a little warmer under pressure
ColdFocal detail
Plate 41 — A real gas cools when it expands through a small orifice without doing work, the Joule–Thomson effect. Argon under high pressure, released inside the probe tip, drops toward its boiling point and freezes the tissue outside. Helium does the opposite at room temperature and warms, which is how some systems thaw.

The same effect can also run the other way. Because helium warms when it expands, some systems thaw the tissue by switching the probe from argon to helium. Liquid nitrogen gives a second way of cooling, one that avoids high-pressure gas altogether. Liquid-nitrogen systems circulate the liquid under pressure to the tip, where it boils, at −196 °C at atmospheric pressure and somewhat warmer under pressure. IceCure makes one such system, ProSense, and FDA's decision summary for it gives a minimum of −170 °C at the center of the probe's cooling zone.

How cold kills

Freezing kills cells in several ways at once. As tissue cools, ice forms first in the fluid between the cells. The ice excludes salt, so the remaining fluid becomes more concentrated and draws water out of the cells by osmosis, and the cells shrink and their contents are damaged. If the cooling is fast, water has no time to leave, and ice forms inside the cells themselves. This intracellular ice is far more damaging than the extracellular ice between the cells, and at the probe, where cooling is fastest, it is the main cause of death.

Farther from the probe, cooling is slower and the temperature higher, and cells may survive. There, injury depends on what happens next. During thawing, small ice crystals recrystallize into larger, more damaging ones, and water rushes back into the shrunken cells. The small vessels of the frozen zone are injured, and in the hours after the procedure the tissue loses its blood supply. A 2010 review of cryoablation by Erinjeri and Clark lists four factors that together decide the injury: the cooling rate, the lowest temperature reached, the time held there, and the thawing rate.

The ice ball and its isotherms

A frozen zone has one great advantage over RF and microwave zones: it can be seen. Ice shows on ultrasound, CT and MRI, so the operator watches the zone grow in real time. The difficulty is that the visible edge of the ice ball is a surface of equal temperature, the 0 °C isotherm, and cells at 0 °C survive. Both reviews place reliable cell death below about −20 to −40 °C. Hinshaw and colleagues estimate the lethal temperature at −20 to −35 °C and suggest that cell death may not occur until about 8 mm inside the edge of the visible ice ball.

42 — The ice ball and the lethal isotherm
frozen, may survive TUMOR 0 °C: the edge seen on ultrasound, CT and MRI −20 °C: cell death may begin only about 8 mm inside the visible edge (one review) −40 °C: reliable cell death what you see is not what dies the −40 °C isotherm spans 44–60 % of the visible ice-ball width, depending on the system (single needles) breast (ProSense) freeze 9 min thaw 8 min freeze 9 min
ColdFocal detail
Plate 42 — The ice ball's visible edge is the 0 °C line, but cells die reliably only several millimeters farther in, below −20 to −40 °C. So the visible ice must extend beyond the tumor by that margin. Two freeze cycles with a thaw between them extend the lethal zone toward the edge.
The visible edge of the ice marks 0 °C

The ice ball's edge is the easiest thing in ablation to see and the easiest to misread. It marks 0 °C, not death. A plan that covers the tumor with visible ice, but not with the lethal isotherm of Chapter 13, leaves a rim of living tumor inside what looked like a successful freeze.

A 2025 study measured the −40 °C isotherm of current needles against the visible ice. With a single needle, the isotherm spanned 44 to 60 % of the width of the visible ice ball, depending on the system, and clustering several needles raised the fraction. The practical rule follows: the visible ice must extend well beyond the tumor, by at least the distance between the visible edge and the lethal isotherm, if the whole tumor is to die.

Solved example: the visible ice needed to kill a 15 mm tumor

A tumor 15 mm across, the largest breast cancer ProSense is authorized to treat, is to lie inside the −40 °C isotherm, before any margin of normal tissue is added. If that isotherm spans 60 % of the visible width, the best case the 2025 study found for single needles, the visible ice must be 15 divided by 0.60, or 25 mm wide. At 44 %, the worst case, it must be 15 divided by 0.44, or about 34 mm wide. Hinshaw and colleagues' figure of about 8 mm inside the visible edge gives a similar width by another route: 8 mm on each side of the tumor makes 15 plus 16, or 31 mm. Either way the visible ice must be roughly twice as wide as the tumor. The ProSense breast protocol targets more than 35 mm after the first freeze and more than 40 mm after the second.

Protocols usually freeze twice with a thaw between, because the second freeze meets tissue already damaged by the first and extends the lethal zone toward the edge of the ice. The whole sequence takes time: Hinshaw and colleagues give typical ablation times of about 25 to 30 minutes. For the breast, FDA's decision summary for ProSense gives a default of nine minutes of freezing, eight of thawing and nine more of freezing, 26 minutes in all, with target ice-ball widths of more than 35 mm after the first freeze and more than 40 mm after the second.

What cold gives and takes

Cryoablation's strengths follow from its physics: the zone is visible as it forms, there is no current through the patient and no return pad, and several probes can be combined to shape the ice around an irregular tumor. Its weaknesses follow too. Cold does not coagulate blood or cauterize injured vessels, so frozen tissue has no built-in hemostasis, the stopping of bleeding, and Erinjeri and Clark note that cryoablation can therefore worsen bleeding. RF ablation, by contrast, coagulates damaged vessels as it goes. Blood flow also works against cold as it works against heat. Large vessels warm the ice near them, acting as a heat source rather than as the heat sink of Chapter 2, so they can change the size and shape of the ice ball. Argon systems also need high-pressure cylinders and gas lines, which the makers of liquid-nitrogen systems cite as part of the case for their own.

Cryoablation reached a new indication in the US in 2025. On 3 October 2025, FDA granted IceCure's ProSense a De Novo authorization for the local treatment of biologically low-risk breast cancers of 1.5 cm or less in patients aged 70 or over who receive adjuvant endocrine therapy. A De Novo is the route FDA uses for a new kind of device of low or moderate risk when no device already on the market can serve as its comparison.

The supporting trial was IceCure's own ICE3 study, a single-arm study with no comparison group. FDA's analysis of it reported ipsilateral recurrence rates of 2.3 to 8.7 %, depending on the analysis population, while IceCure reports a five-year estimate of 3.1 %. An ipsilateral recurrence is the cancer's return in the same breast. The figures differ, and each is correct for its own population and method.

In practice

Ask of every zone size in a cryoablation system's specification which isotherm it refers to, because a "4 cm ice ball" and a "4 cm lethal zone" are very different claims. Plan from zone charts that give the −20 °C and −40 °C isotherms, measured for single and for clustered needles, because a chart of the visible ice alone does not show which tissue will die. On the bench, measure temperatures inside the ice as well as at its edge, so that a stated lethal zone rests on a measured isotherm.

+ What this chapter established
  • Cryoprobes cool by the Joule–Thomson expansion of high-pressure argon or by circulating liquid nitrogen, and helium can thaw the tissue.
  • Freezing kills by extracellular ice and osmotic injury, by intracellular ice where cooling is fast, and by injury to the small vessels after thawing.
  • The visible edge of the ice is at 0 °C and reliable cell death lies below about −20 to −40 °C, several millimeters inside it, so the ice must extend beyond the tumor by at least that distance.
  • Cryoablation gives a visible zone and no current through the patient, but no hemostasis, and its newest US indication is low-risk breast cancer in patients aged 70 or over.