Classification decides the route.
+ The questionWhat must a surgical energy device show to reach US patients, and what happens when it changes?
A regulation, a product code and a class
On 26 and 27 October 2022, an FDA advisory panel met to consider how ultrasonic surgical instruments should be classified, although the device type had been sold in the United States since before the 1976 amendments that created the country's classification system. FDA's briefing for the panel recommended class II with special controls, requirements written for that device type, such as particular tests and labeling, in addition to the rules for every device. As of October 2026, no proposed rule has been found, and FDA's product-code page still lists the instruments as unclassified, a pre-amendment device type that reaches the market by 510(k). Ultrasonic instruments are therefore an exception to the order in which the US system works: a device type is normally placed first in a regulation and a class, and the class decides what its maker must show.
FDA describes most device types in a classification regulation in Title 21 of the Code of Federal Regulations, which assigns each a class, and FDA files the records for each type under one or more three-letter product codes (Chapter 23). Most class II devices reach the market by a 510(k), a premarket notification showing that the device is substantially equivalent to a legally marketed predicate device, and FDA's decision on it is a clearance (Chapter 9). Devices in class III, the class of highest risk, need premarket approval, or PMA, and FDA's decision is an approval.
Electrosurgery sits in 21 CFR 878.4400, which describes a device "intended to remove tissue" and "control bleeding by use of high-frequency electrical current". The regulation places it in class II, FDA files these devices under product code GEI, and GEI devices are eligible for FDA's program in which accredited third parties review 510(k)s, while ultrasonic instruments are excluded from it. The table gives the codes that FDA's records show for devices in this book, as of October 2026. Where it shows a dash, the records cited give no regulation.
| Device type | Product code | Regulation, 21 CFR | Class | Route |
|---|---|---|---|---|
| Electrosurgical cutting and coagulation, with accessories | GEI | 878.4400 | II | 510(k) |
| Microwave ablation | NEY | 878.4400 | II | 510(k) |
| Powered laser surgical instrument | GEX | 878.4810 | II | 510(k) |
| Ultrasonic surgical instrument | LFL | — | unclassified | 510(k) |
| HIFU for prostate tissue ablation | PLP | 876.4340 | II | De Novo (2015), then 510(k) |
| Focused ultrasound for non-thermal, mechanical tissue ablation (histotripsy) | QGM | 878.4405 | II | De Novo (2023) |
| Cryoablation for low-risk breast cancer | QXW | 878.4355 | II | De Novo (2025) |
| MR-guided HIFU ablation | NRZ | — | III | PMA |
| Cardiac ablation catheter, pulsed field | QZI | — | III | PMA |
Hybrid instruments that deliver bipolar current and ultrasonic energy through one jaw have been cleared under the electrosurgical regulation with both product codes (Chapter 11). Other cryosurgical devices reach the market by 510(k) under product code GEH, and AngioDynamics states that its NanoKnife IRE system is cleared by 510(k) for the surgical ablation of soft tissue (Chapter 15).
When there is no predicate
The first device of a new kind has no predicate, and so no 510(k). If its risk is low or moderate, its maker can request a De Novo classification, and when FDA grants the request it creates a new regulation and product code that later devices can follow. The ablation chapters met three such grants: Sonablate's HIFU system for prostate tissue on 9 October 2015, HistoSonics' Edison system for liver tumors on 6 October 2023 (Chapter 16), and IceCure's ProSense cryoablation system for low-risk breast cancer on 3 October 2025 (Chapter 14). The route then works as designed. Chapter 16 tells how a competitor withdrew its own De Novo request after Sonablate's grant and was cleared by 510(k) on 6 November 2015, four weeks later, with Sonablate as its predicate.
An announced request is not yet a grant. As of 20 October 2026, no FDA decision had been found on the De Novo request for kidney tumors that HistoSonics announced on 11 May 2026 (Chapter 16). Devices of the highest risk go to premarket approval instead of a De Novo. FDA approved the cardiac pulsed field ablation systems of Chapter 15 by PMA, the first on 13 December 2023, and its first approval of a focused-ultrasound device was a PMA of 30 June 1988, for a glaucoma treatment (Chapter 16).
A 510(k) device is cleared, a De Novo request is granted and a PMA device is approved. A request that has been submitted or announced is an application, and only FDA's decision makes a device legally marketable for the use it names. That use is the indication on the label: an indication that names a tissue, as NanoKnife's names prostate tissue, makes no claim about a disease. Companies' own releases sometimes use the wrong verb. Medtronic's release of 30 September 2025 on launching its FT10 generators in India called them "USFDA-approved", although FDA clears such generators by 510(k), and HistoSonics' release of 22 June 2026 called the liver authorization a "De Novo clearance". In the EU, a notified body issues certificates and the manufacturer affixes the CE marking (Chapter 27). Voluntary ISO 13485 certificates come from certification bodies, and ISO itself issues none.
The quality system, software and cybersecurity
Whatever the route, the maker must run a quality system, and on 2 February 2026 the US rule for it changed its basis. The Quality Management System Regulation took effect that day. A final rule published on 2 February 2024 created it by amending 21 CFR Part 820. The new regulation incorporates ISO 13485:2016 by reference and replaces the older Quality System Regulation. Chapter 22 described design controls as the process that turns an intended use and its claims into requirements, risk controls, verification and validation, a trace from claim to test report, and a loop from the field back into the design. Under the new regulation, a US maker runs and records that process to ISO 13485's requirements for design and development. The VIO 3n's summary, prepared on 7 April 2025, shows the older form: it cites its design controls under 21 CFR 820.30, a section of the regulation then in force.
Chapter 20 placed software and connectivity in a generator's architecture, and in a US submission they carry obligations of their own. An opening notice in FDA's electrosurgical guidance says that general-surgery electrosurgical devices should generally address the Enhanced documentation level of FDA's 2023 software guidance (Chapter 20). The body of the electrosurgical guidance has not been revised since 2020 and still speaks of a "major" level of concern, a term from the framework that the 2023 guidance replaced (Chapter 25). The electrosurgical guidance therefore uses the terms of both frameworks, the newer in its notice and the older in its body.
Connectivity adds obligations that come from the statute itself. For premarket submissions made on or after 29 March 2023, section 524B of the Federal Food, Drug, and Cosmetic Act sets requirements for the sponsor of a cyber device. The section defines a cyber device by three criteria, which FDA's frequently asked questions restate: it includes software validated, installed or authorized by the sponsor, it can connect to the internet, and it has characteristics that could be vulnerable to cybersecurity threats. Among the requirements, the sponsor must provide a plan to monitor, identify and address vulnerabilities and exploits after sale, processes that give reasonable assurance that the device and related systems are cybersecure, including updates and patches, and a software bill of materials. FDA reissued its premarket cybersecurity guidance on 3 February 2026, superseding the edition of 27 June 2025, with "Quality Management System Considerations" now in its title.
When the device changes
A device is cleared as it was described, and every later change, to its software, components, labeling or materials, has to be assessed against that description. FDA's guidance on deciding when to submit a 510(k) for a change to an existing device, issued on 25 October 2017, walks through the assessment with flowcharts for labeling changes, for changes in technology, engineering or performance, and for changes in materials. Each flowchart serves the question that 21 CFR 807.81(a)(3) asks: whether the change "could significantly affect the safety or effectiveness of the device", or is a "major change or modification in the intended use of the device". If either holds, a new 510(k) is needed. If neither does, the guidance asks the maker to "document your analysis and file it for future reference". A companion guidance of the same month covers software changes.
Chapter 22 placed the change assessment inside the design loop, and these guidances set the US threshold for it. Both records of Chapter 25 are new models cleared against predicates: the VIO 3n differs from its predicates in software, interfaces, modes and maximum power, and the Megadyne generator in a lower maximum power among other changes. Neither summary says which difference made a new 510(k) necessary, so a reader learns where the threshold lies from the guidance, and the records show only that each maker submitted a new one.
After sale: reports, corrections and recalls
Clearance marks a date, and the record continues after it. FDA collects medical device reports of deaths, serious injuries and malfunctions and counts them by product code, and the counts can weigh in a classification decision: FDA's briefing for the 2022 panel on ultrasonic instruments counted the reports filed under LFL.
FDA's executive summary for the panel counted 46,673 medical device reports under product code LFL from 1 January 2002 to 31 December 2021: 44,354 malfunctions, 2,263 serious injuries and 56 deaths. The three add up to the total, since 44,354 and 2,263 make 46,617, and 56 more make 46,673. Malfunctions are 44,354 divided by 46,673, or 95 % of the reports. Serious injuries are 2,263 divided by 46,673, about 4.8 %, and deaths 56 divided by 46,673, about 0.12 %, or 1 report in about 830. Over the twenty years the reports average 46,673 divided by 20, about 2,300 a year. The summary gives no count of instruments in use or of procedures performed, so none of these figures is a rate of harm per operation. They describe what was reported, and 95 % of it was malfunctions.
When a marketed device puts patients at risk, its maker may correct it in place or remove it, and FDA classes the action on a scale of its own, separate from the device classes. FDA calls a Class I recall "the most serious type of recall", and such a recall need not remove anything. The Varipulse pulsed field ablation catheter shows such a recall. In February 2025 Biosense Webster, a Johnson & Johnson company, sent customers a correction updating the instructions for the catheter, after 4 of 132 patients in its US External Evaluation had a stroke or transient ischemic attack. That is about 3 %, against an expected rate of 1 % or less. FDA classed the correction as the most serious type although the correction left the catheter on the market, and its notice asked users to follow the updated instructions, including their indication for "symptomatic drug refractory paroxysmal atrial fibrillation" (Chapter 15 tells the pause that preceded it). Because the catheter is a PMA device, its later change of irrigation flow rate went back to FDA for approval, which FDA gave on 26 June 2025 (Chapter 22).
The return pad of Chapter 8 has its own entry. Megadyne began a correction of its MEGA 2000 and MEGA SOFT reusable return electrodes on 1 June 2023 after reports of burns, and FDA posted it as a Class I recall on 31 July 2024, covering 21,200 devices in the US, with 63 injuries reported and no deaths. Less serious actions fall in Class II. Olympus began one on 25 July 2025 for 155 units of its ESG-410 generator, which could show a system error and then restart, sometimes repeatedly, and FDA posted it on 19 August 2025 (Chapter 24). Each such action returns to the loop of Chapter 22, where the correction is itself a change to be assessed.
Write the intended use and the list of claims first, and only then classify the device: find the product code whose regulation describes it, its class and the route that class requires. Read the product code's page for third-party eligibility and the standards FDA lists against it. For each foreseeable change, record in advance the answer the 2017 flowcharts give, so that the change plan names the changes that need a new 510(k). Set up the procedures for reports and corrections before the first sale.
- In the US, a device type's regulation, product code and class set its route: electrosurgical generators and surgical lasers are class II by 510(k), ultrasonic instruments remain unclassified as of October 2026, and cardiac pulsed field ablation systems are class III by PMA.
- A new kind of device with no predicate can be classified by a De Novo grant, as Sonablate, Edison and ProSense were, and a later device can then follow by 510(k), as one did four weeks after Sonablate.
- Since 2 February 2026 the US quality system regulation incorporates ISO 13485:2016, under which a maker runs the design controls of Chapter 22, and section 524B adds cybersecurity obligations to submissions for cyber devices made on or after 29 March 2023.
- A change that could significantly affect safety or effectiveness needs a new 510(k), and after sale FDA counts medical device reports and classes corrections and recalls, as it classed the 2025 Varipulse correction among the most serious.