Who makes it, and what is changing.

On three consecutive days in September 2026, FDA cleared three energy devices from three makers: a LigaSure vessel sealer for Medtronic's Hugo robot on 15 September, Olympus's SOLTIVE thulium fiber laser systems on 16 September and Ethicon's HARMONIC 1100 ultrasonic shears on 17 September. Each energy in this book has its own makers and its own public record, and both change from month to month. This part is dated October 2026. It sets out who makes each kind of device and what the regulator's record shows for it, then what is changing at the edge of the field and what is likely to stay hard.

23 — Who makes surgical energy

The makers, energy by energy.

+ The questionWhich companies make each kind of surgical energy device in October 2026, and what does the regulator's record show for each?

The record counts decisions rather than products

Between 1 January 2024 and 20 October 2026, FDA recorded decisions on 153 premarket notifications under the product code for electrosurgical devices. A product code is the three-letter code under which FDA files every record for one type of device. The electrosurgical code, GEI, holds more than surgical generators, because aesthetic RF devices, endovenous RF devices, electrodes and pencils share it. The 153 records were filed under 98 forms of applicant name, and one company can appear under several of them: Olympus, for example, under four name forms of three legal entities. One record can cover a generator together with many instruments. FDA publishes these data through its openFDA service and states that they are unvalidated.

Solved example: the surgical makers' share of the electrosurgical record

Eight makers whose surgical generators, sealers, pencils or instruments this chapter describes appear among the applicants. Erbe holds 6 of the 153 records, Stryker's two units 5, Covidien 3, Olympus 4 under four name forms, LivSmed 3, Intuitive 2, Ethicon 2 and Megadyne 1. Adding them gives 26 records. Dividing 26 by 153 gives about 0.17, so these makers hold about one record in six. The other 127 records belong to the remaining applicants, many of them makers of aesthetic RF devices. A count under one product code therefore measures what the code holds, and the part of it that belongs to surgical energy has to be worked out applicant by applicant.

The makers are therefore best followed energy by energy, through the records of their current products and what each company states about them. The record's words are kept throughout: a 510(k) is cleared, a De Novo request is granted, and a premarket approval, or PMA, is approved (Chapter 26). A figure, a count or a "first" that comes from a company is given as that company's statement.

62 — US records by energy, January 2024 to October 2026
2024 2025 2026 20 Oct 2026 Electrosurgery and sealing GEI Ethicon generator VIO 3n LigaSure for Hugo 153 records Ultrasonic and hybrid LFL · GEI THUNDERBEAT II HARMONIC 1100 aspirators and similar (LFL) LFL: 8 records, aspirators and similar devices Surgical lasers GEX SOLTIVE India 100+ applicants Microwave ablation NEY IntelliBlate 7 applicants Cryoablation GEH · De Novo ProSense, low-risk breast cancer GEH: skin, warts too Pulsed field ablation PMA 13 Dec 2023 dual-energy catheter (announced) Kardium Abbott PMA route first four premarket approvals Focused ultrasound, histotripsy 6 Oct 2023 Exablate Neuro, Parkinson's kidney kidney: no FDA decision announced Robotic systems SCV and others da Vinci 5 Versius Hugo SCV: created by Versius De Novo 510(k) PMA De Novo solid: decision date in FDA's record ring: date from a company or investor announcement dashed: request, no decision openFDA, queries of 20 and 21 Oct 2026; marks are the records the chapter names
ElectricalAcousticOpticalHeatColdFocal detail
Plate 62 — Each energy reaches the US by its own route. Generators, instruments, lasers and microwave systems are cleared by 510(k), pulsed field systems are approved through premarket approval, and a new kind of device arrives by De Novo. The marks are the records this chapter names; the notes at the right show how much more each product code holds.

Generators, sealers and instruments that pass current

Medtronic's annual report for its fiscal year to 24 April 2026 lists its electrosurgical products as the Valleylab FT10 and FX8 energy platforms and its LigaSure vessel-sealing instruments. Chapter 8 read the FT10's monopolar modes, and Chapter 9 its sealing. Two of its records in the period are a pencil and a sealer. FDA cleared a smoke-management pencil from the company's Covidien unit on 1 July 2024, and Medtronic states that the LigaSure sealer cleared for its Hugo robot on 15 September 2026 seals vessels up to and including 7 mm. Erbe holds the most records of the eight makers, and its generators also drive other makers' instruments. FDA cleared Erbe's VIO 3n generator with its accessories on 29 August 2025, and Chapter 25 reads its public summary. FDA also cleared Erbe's HybridTherm System on 5 August 2024, and Intuitive lists an Erbe generator, the VIO dV, among those that drive its Vessel Sealer Extend.

FDA cleared a generator from Johnson & Johnson's Ethicon, the ETHICON Total Energy System, on 17 January 2025. The company states that it has since renamed the generator DUALTO with no change to its intended use or regulatory status, and that DUALTO powers several energy modalities and runs any two instruments at once whose maximum power draw is 150 W or less. The same company also lists two other generators as separate products: GEN11, for its HARMONIC ultrasonic and ENSEAL bipolar devices, and a MEGADYNE generator. FDA cleared LivSmed's ArtiSeal, a vessel sealing system with its own generator, on 15 November 2024. Stryker's records since 2024 include PhotonBlade 3, an electrosurgical blade with smoke evacuation cleared on 21 October 2024, and a remanufactured LigaSure Maryland sealer from its reprocessing unit, cleared on 18 June 2026.

A maker missing from a recent record may still be selling

BOWA's latest generator record under the electrosurgical code is the ARC 400, cleared in 2020. KLS Martin's latest is a generator sold under another company's name, the Codman Electrosurgical Generator, cleared in 2019, and the latest records of CONMED, for its CleanSeal bipolar sealer, and of Apyx Medical, for its Renuvion helium-plasma handpieces, date from 2023. None of the four holds a record under these names and this code with a decision date from January 2024 to October 2026. A product cleared earlier can stay on the US market, and a company can file under another code or another legal name, so a quiet record shows only that nothing new was cleared there.

Ultrasonic instruments and lasers

Ethicon, Olympus and Medtronic each sell ultrasonic instruments of the kind Chapter 11 describes. FDA cleared Ethicon's HARMONIC 1100 Shears on 17 September 2026. FDA cleared Olympus's THUNDERBEAT II, with a transducer that Olympus calls cordless, on 7 October 2025, and its front-actuated THUNDERBEAT shears on 1 July 2026. Olympus's word means that the cable has moved to the base of the handle, not that the device runs on a battery (Chapter 11). Medtronic lists Sonicision 7, which it calls a cordless ultrasonic dissection system, in its 2026 annual report.

The product code for ultrasonic surgical instruments, LFL, holds eight 510(k)s decided between January 2024 and October 2026, and FDA still lists the code as unclassified as of October 2026 (Chapter 26). All eight are aspirators or similar devices: two each for Integra's CUSA Clarity and Stryker's Sonopet iQ, and one each for Misonix's neXus, Trice Medical's Tenex, Hunan Handlike and Shanghai Saints Sages. None of them comes from Ethicon, Medtronic or Olympus: the clearances of HARMONIC 1100 and of both THUNDERBEAT shears appear under the electrosurgical code, GEI, so a search under LFL alone would miss them.

The surgical-laser code, GEX, lists more than 100 applicants over the same period, most of them makers of aesthetic lasers. Among makers of surgical lasers, El.En holds seven records, Quanta System four, and Boston Scientific, Olympus and IPG Medical three each. In 2021 Boston Scientific bought Lumenis's surgical business, with its urology and ENT lasers, its fibers and the MOSES pulse technology of Chapter 12. Lumenis Be also holds three records under the code, but for aesthetic devices, because it kept Lumenis's aesthetic and vision businesses after that sale. FDA cleared Boston Scientific's Flexiva Pulse laser fibers on 9 July 2025 and its MOSES Raydar holmium laser system on 18 May 2026. FDA cleared Olympus's SOLTIVE thulium fiber laser systems three times, on 23 August 2024, 24 July 2026 and 16 September 2026.

Ablation by heat, cold, pulses and sound

Seven applicants hold the 510(k)s for microwave ablation decided between January 2024 and October 2026. Among them is Varian, whose IntelliBlate system FDA cleared on 19 July 2024. Medtronic's annual report lists the Emprint microwave system of Chapter 13. AngioDynamics states in its annual report that it discontinued its RF ablation product line as of 29 February 2024. It keeps its Solero microwave system, indicated for the ablation of soft tissue during open procedures, and its NanoKnife IRE system (Chapter 15). The cryosurgical code, GEH, also covers devices for skin and warts; its applicants in the period include IceCure, Boston Scientific and Erbe, and IceCure's ProSense holds the De Novo grant for low-risk breast cancer described in Chapter 14.

Pulsed field ablation first reached US patients through four premarket approvals, from Medtronic's PulseSelect on 13 December 2023 to Johnson & Johnson's Varipulse on 6 November 2024, with Boston Scientific's FARAPULSE and Medtronic's Sphere-9 between them, and the plate below dates each. Approvals continued after those four, and FDA approved further systems from Kardium and Abbott in 2025 (Chapter 15). Johnson & Johnson announced on 8 July 2026 that FDA had approved its Dual Energy THERMOCOOL SMARTTOUCH SF, which delivers RF and pulsed field energy through a single catheter. A maker's own figure gives a sense of the scale of use: Boston Scientific states in its 2025 annual report that more than 500,000 patients have been treated with FARAPULSE. A European marking can also come before a US approval. Medtronic announced on 23 January 2026 that its Affera Sphere-360 catheter had received CE marking in Europe, and as of October 2026 no FDA approval of it had been found. Chapter 24 follows Varipulse's first year in the US and its newer European version.

63 — Pulsed field ablation in the US, December 2023 to October 2026
2024 2025 2026 this guide DECIDED IN THE US Dual Energy THERMOCOOL SMARTTOUCH SF 8 Jul 2026 approved: RF and pulsed field (company announcement) 13 Dec 2023 PulseSelect, Medtronic (PMA) 30 Jan 2024 FARAPULSE, Boston Scientific (PMA) 24 Oct 2024 Sphere-9 and Affera, Medtronic: pulsed field and RF (PMA) 6 Nov 2024 Varipulse, Johnson & Johnson (PMA) 14 Feb 2025 Varipulse correction: FDA classed it the most serious type of recall 26 Jun 2025 Varipulse irrigation flow rate approved 27 Aug 2025 Globe, Kardium (PMA) 19 Dec 2025 Volt, Abbott (PMA) OUTSIDE THE US OR UNDER STUDY 23 Jan 2026 Sphere-360 CE marking (announced) Mar 2026 Varipulse Pro CE marking (company statement) 24 Apr 2026 PERSIGMA trial announced, up to 466 participants Boston Scientific's 2025 annual report: more than 500,000 patients treated with FARAPULSE maker's figure
ElectricalFocal detail
Plate 63 — Four pulsed field systems were approved in the US in less than a year, and about three months after the last of them FDA classed a correction to its instructions as the most serious type of recall. Systems from Kardium and Abbott followed in 2025. Newer catheters from Medtronic and Johnson & Johnson carried CE marking in Europe first and had no US approval at the dates shown.

HistoSonics' Edison, the histotripsy system of Chapter 16, holds a De Novo grant for liver tumors, and Chapter 16 follows its kidney request. The company's own announcements give a sense of its scale. HistoSonics announced on 7 August 2025 that a consortium of investors had agreed a management-led acquisition of a majority stake, valuing the company at about US$2.25 billion, and in June 2026 it announced financing at a valuation of US$3.75 billion. Among the makers of other ultrasound ablation systems, Insightec announced on 8 July 2025 that FDA had approved its Exablate Neuro for staged treatment of both sides of the brain in advanced Parkinson's disease. Profound Medical states that its TULSA-PRO is cleared by 510(k) and reports 78 installed systems at the end of 2025. EDAP states that FDA cleared its Focal One i System in November 2025.

Robots and the energy they carry

Intuitive announced FDA's clearance of its da Vinci 5 system on 14 March 2024 and states that the system integrates an electrosurgical unit and insufflation. Its E-200 generator holds a clearance dated 5 August 2024, and Intuitive states that SynchroSeal, the 5 mm sealer of Chapter 17, runs on the E-100 and the E-200. The robots that followed show the De Novo route of Chapter 26 at work, in which the first device of a kind creates a product code that later devices follow. FDA granted CMR Surgical's Versius a De Novo on 11 October 2024, creating the product code SCV for modular electromechanical surgical systems. Medtronic's Hugo was cleared by 510(k) under that code on 3 December 2025, with Versius as its predicate, and as of October 2026 its US indication remains urologic procedures (Chapter 24). FDA cleared Versius Plus on 16 December 2025, and CMR's investor Cambridge Innovation Capital reports that the clearance covers gallbladder removal.

On 12 October 2026 Stryker announced that it would invest about US$400 million in CMR Surgical, with an exclusive option to buy all of the company and the role of exclusive US sales agent for Versius, subject to regulatory approvals and closing conditions. SS Innovations, an Indian company, states that its 510(k) for the SSi Mantra robot was under FDA review for gallbladder removal as of October 2026. FDA asked for more information, the company responded on 5 October 2026, and it expects the review to be completed in the first quarter of 2027. Neither CMR's nor SS Innovations' announcements name an energy instrument or a generator.

India's makers and the record abroad

In FDA's open data, the only applicant with an Indian address under the surgical-laser code is Allengers Global Healthcare, of Derabassi in Punjab. It holds two clearances, for a holmium laser on 18 August 2023 and for a thulium fiber laser on 17 December 2024. The only Indian-address applicant ever recorded under the electrosurgical code is Ace Medical Devices of Pune, for resection and vaporization electrodes cleared in 2016, and as of October 2026 no Indian electrosurgical generator or vessel sealer appears in those data. India's Department of Pharmaceuticals publishes the approved applicants of its Production Linked Incentive scheme for the domestic manufacture of medical devices. Its list as on 6 November 2023 names one surgical-energy system: a thulium fiber laser, listed for Indovasive Private Limited together with laser probes and loop and cutting electrodes for prostate resection.

Medtronic launched its Valleylab FT10 electrosurgical generator and its FT10 vessel sealing generator in India on 30 September 2025 and described both as "USFDA-approved", although a device that reaches the US market by 510(k) is cleared. BioSpectrum India, a trade journal, reported on 9 June 2026 that Biorad Medisys of Pune had launched a 20 W thulium fiber laser for kidney stones, adding to its 40 W and 75 W units. As of October 2026, India's 2021 classification lists place general-purpose electrosurgical units in class C, and Chapter 27 sets out what that class requires.

Reading the field's public record

Before a design review or a competitive analysis, list the devices of the same energy from the regulator's database rather than from press releases: product code, record type, applicant as filed, decision date and the indication in the summary. Merge the name forms of one company, and check whether a count covers products or only records. Label every figure that comes from a company as the company's own, and give every status with the date on which it was checked.

+ What this chapter established
  • FDA's open data hold 153 premarket notifications under the electrosurgical product code from January 2024 to October 2026, and the eight surgical makers this chapter describes hold about one in six of them.
  • FDA cleared generators, sealers, ultrasonic instruments, laser systems or laser fibers from Medtronic, Ethicon, Olympus, Erbe, Intuitive and Boston Scientific in that period, while all eight records under the ultrasonic code, LFL, went to aspirators and similar devices, and the ultrasonic shears appear under the electrosurgical code.
  • Four pulsed field systems were approved in the US between December 2023 and November 2024 and two more in 2025, while AngioDynamics left RF ablation in 2024, an investor consortium's majority stake in HistoSonics was announced in 2025 and Stryker announced an investment in CMR Surgical in 2026.
  • India's makers appear in FDA's data with lasers and resection electrodes, and as of October 2026 no Indian electrosurgical generator or vessel sealer holds a US clearance there.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.