15 — Pulsed fields

Pulsed fields: IRE and PFA.

+ The questionHow can kilovolt pulses kill cells without heating them, and why would anyone want an ablation that does not use heat?

The problem heat cannot solve

On 13 December 2023, FDA approved Medtronic's PulseSelect, the first system in the US that treats atrial fibrillation, a disorder of the heart's rhythm, with strong electric pulses instead of heat or cold. Such a system answers a problem that heat cannot solve, because every energy so far kills by temperature, and temperature does not discriminate. Three targets show the problem: a tumor wrapped around the portal vein, a prostate cancer beside the nerves and sphincter that control erection and continence, and the wall of the left atrium lying against the esophagus. In each, the cells to be killed sit next to structures that must survive, and heat destroys both alike. The collagen framework of a vessel or duct survives many other insults, but the temperatures that ablation reaches destroy it along with the cells around it.

Pulsed electric fields offer a different mechanism, in which very short, very strong electric pulses open holes in cell membranes. If the holes are large and numerous enough, the cell cannot repair them and dies, while the scaffold of collagen and elastin between the cells is left intact, because it has no membrane to open. In tumor ablation the technique is called irreversible electroporation, or IRE, and in cardiology it is called pulsed field ablation, or PFA.

A voltage across the membrane

Chapter 3 described a cell as a conducting interior wrapped in an insulating membrane. In an external electric field, charge piles up on that membrane, positive on one side and negative on the other. Within about a microsecond, the voltage across the membrane settles at a value that depends on the field, the cell's size and the position on the cell. For a spherical cell, the voltage is largest at the two poles that face the field, where it equals 1.5 times the field strength times the cell's radius. It falls away from the poles and is zero at the equator, midway between them.

Solved example: the voltage across a small cell's membrane

A spherical cell with a radius of 5 µm sits in a field of 400 V/cm. In meters, the field is 40,000 V per meter and the radius is 0.000005 m, five millionths of a meter. The voltage at each pole is 1.5 times the field times the radius, and 1.5 times 40,000 V/m times 0.000005 m gives 0.30 V. The voltage rises in step with the field, so 600 V/cm gives 0.45 V and 1,000 V/cm gives 0.75 V. Membranes are reported to open pores once the voltage exceeds about 0.5 to 1 V, and reaching 0.5 V at this cell's poles takes 0.5 V divided by 1.5 and by 0.000005 m, about 67,000 V per meter, or about 670 V/cm. A cell with twice the radius reaches 0.5 V at half that field, so a larger cell crosses the threshold first.

43 — A cell in an electric field
UNIFORM APPLIED FIELD equator: 0 V peak pores open where the voltage peaks membrane voltage at the poles = 1.5 × field × radius for a cell of radius 5 µm: 400 V/cm → 0.30 V 600 V/cm → 0.45 V 1,000 V/cm → 0.75 V pores open above about 0.5–1 V a bigger cell reaches the threshold at a lower field: heart muscle cells are larger than esophageal smooth-muscle cells (proposed basis of PFA selectivity)
ElectricalFocal detail
Plate 43 — An external field drives charge onto the cell membrane until the voltage across it is about 1.5 times the field times the cell radius, largest at the poles. Above roughly half a volt to a volt the membrane opens pores. Because the voltage grows with radius, large cells cross the threshold at lower fields than small ones.

Membranes are reported to open pores once the voltage across them exceeds about 0.5 to 1 V. Above that threshold, the membrane's lipid layer rearranges into nanometer-scale pores. After a brief, mild exposure the pores reseal and the cell survives: this is reversible electroporation, which is used to push drugs or DNA into cells. After stronger or more numerous pulses the damage is beyond repair, and the cell dies over the following hours. Because the voltage at the poles grows with the cell's radius, a larger cell reaches the threshold at a lower field than a smaller one.

Thresholds and the shape of the zone

The field around a pair of needle electrodes is strongest between and immediately around them and falls off with distance, so the ablation zone is the region where the field exceeds the irreversible threshold for that tissue. That threshold differs from one tissue to another. In a 2023 laboratory study funded by Medtronic, Kos and colleagues compiled the mean irreversible thresholds that earlier studies had measured in living tissue: 539 V/cm for liver, 538 for kidney, 585 for brain, 700 for prostate and 1,050 for bone. Three of the study's authors were Medtronic employees. Below the irreversible threshold lies the reversible one, at which pores open and then reseal, and a 2005 modeling study by Davalos and colleagues used about 360 V/cm for it in liver. The thresholds fall as more and longer pulses are delivered.

Clinical settings keep the field well above these values. In 2016, Martin and colleagues called for a standard way of delivering IRE energy in locally advanced pancreatic cancer, and they report that surgeons typically apply 1,400 to 2,000 V for each centimeter between electrodes. The electrodes are usually placed no more than 2 cm apart, so that the field across the target stays above threshold with margin. The first author was a paid consultant for AngioDynamics, the maker of the NanoKnife IRE system, when the study was published.

44 — The field around two electrodes
about 15 mm about 360 V/cm: reversible, cells recover about 540 V/cm: irreversible (liver) vessel wall scaffold survives; no heat sink the vessel passes through undamaged irreversible thresholds, mean (V/cm): liver 539 kidney 538 brain 585 prostate 700 bone 1,050 clinical IRE, pancreas: 1,400–2,000 V per cm of electrode spacing, electrodes usually no more than 2 cm apart
ElectricalFocal detail
Plate 44 — The field is strongest between and around the electrodes and falls off with distance, so the ablation zone is the region above the irreversible threshold. Because the effect is not thermal, flowing blood does not carry the effect away, and the collagen scaffold of vessels and ducts inside the zone survives.

Because the mechanism is electrical rather than thermal, flowing blood does not carry the effect away. There is no heat sink of the kind Chapter 2 described, and the zone reaches right up to the wall of a vessel crossing it. Animal studies report that vessels, bile ducts and bronchi within the zone keep their structure. The mechanism predicts this, because their framework of collagen and elastin has no membrane to open. That preservation is the reason to use IRE, because it lets IRE treat tissue that heat-based ablation cannot safely reach.

Every pulse still deposits some heat

IRE pulses carry a lot of energy: one maker's manual for its IRE generator lists outputs of 500 to 3,000 V and up to 50 A, in pulses of 20 to 100 µs carrying up to about 15 J each. Every pulse deposits some Joule heat (Chapter 3). Near the electrodes, after many pulses, tissue can heat enough to add thermal injury. Non-thermal describes the mechanism that kills most of the zone, not an absence of heating.

Pulses and the heart

Pulses of thousands of volts stimulate every nerve and muscle they reach and can trigger the heart, so IRE needs two precautions. The pulses are synchronized with the ECG, the trace of the heart's electrical activity, and delivered in the refractory period just after each heartbeat, when the heart cannot be triggered. The generator whose manual is cited above uses this synchronized mode by default and also offers fixed rates of 90 pulses a minute, for soft tissue below the waist, and 240, for the prostate. And because the pulses make skeletal muscle contract violently, patients are treated under general anesthesia with complete neuromuscular block, in which drugs stop the skeletal muscles from contracting.

45 — Monophasic and biphasic pulses
ECG PULSE TRAINS IRE monophasic PFA biphasic delivered in the refractory period (ECG-synchronized) pulse where the heart cannot be triggered 20–100 µs, up to 3,000 V (manufacturer data) strong muscle contraction: general anesthesia with full neuromuscular block biphasic pulses stimulate muscle less schematic; each row on its own time scale
ElectricalReadoutFocal detail
Plate 45 — Kilovolt pulses stimulate nerve and muscle and can trigger the heart. IRE systems time their pulses to the heart's refractory period and need full muscle relaxation under anesthesia. Cardiac PFA uses short biphasic pulses, which stimulate muscle less.

Cardiac PFA addresses the same problem differently, with short pulses of a different shape. Each pulse can make nearby skeletal muscle twitch, so pulse shape matters in practice. Pulses that reverse direction, called biphasic, appear to reduce this. In a 2022 preclinical study, Bi and colleagues found less muscle contraction with biphasic pulses than with long monophasic pulses, which drive current in one direction only. One of the authors worked for a company, Tianjin Intelligent Health Medical. The pulses remain strong. One present-day system is Boston Scientific's FARAPULSE, and its maker states that it offers outputs of 1.8 to 2 kV in applications of 2.5 seconds each.

Pulsed field ablation for atrial fibrillation

Pulsed fields are being adopted rapidly in cardiology. Atrial fibrillation is treated by creating lines of scar around the pulmonary veins, and for decades that was done with RF heat or with cold. Both carry a small risk of injuring the esophagus, which lies against the back of the left atrium, and the phrenic nerve, which runs near the right-sided veins. The phrenic nerve drives the diaphragm, and an injured one can leave the diaphragm paralyzed, a phrenic palsy. Pulsed fields can spare such structures if heart muscle dies at a lower field than they do.

Heart muscle has a comparatively low irreversible threshold, below that of every tissue in the list above, although the list includes neither the esophagus nor the phrenic nerve. In the same Medtronic-funded study that compiled that list, Kos and colleagues measured a median of 416 V/cm in three human donor hearts, kept perfused outside the body and treated with Medtronic's biphasic pulses. A proposed explanation for the selectivity of PFA is the size of the cells. The solved example showed that a cell with a larger radius reaches the pore-opening voltage at a lower field, and heart muscle cells have a larger radius than the smooth-muscle cells of the esophagus, so heart muscle crosses the threshold first.

After PulseSelect, FDA approved Boston Scientific's FARAPULSE on 30 January 2024, Medtronic's Affera Sphere-9 on 24 October 2024 and Johnson & Johnson's Varipulse on 6 November 2024. Sphere-9 delivers both pulsed field and RF energy. In January 2025, Johnson & Johnson paused US cases with Varipulse after neurovascular events, which affect the blood supply of the brain, and in February it announced a limited restart with updated instructions for use (Chapter 26). Later in 2025, FDA approved two more pulsed field systems, Kardium's Globe on 27 August and Abbott's Volt on 19 December.

MANIFEST-17K is a retrospective registry, compiled from the records of patients already treated, and many of its authors report ties to Boston Scientific. In it, researchers at 106 centers in Europe and Israel reported on 17,642 patients treated with one system, Boston Scientific's FARAPULSE. They recorded no esophageal complications and no persistent phrenic nerve palsy, and major complications in 0.98 % of patients.

Measures of adoption depend on what is counted. A market-intelligence firm modeled the purchasing data of more than 4,500 US facilities and estimated that PFA accounted for 69 % of spending on atrial fibrillation ablation catheters in 2025 and 78 % in early 2026, but for 48 % and 57 % of the catheters bought. A physician survey projected that PFA would be used in 49 % of procedures in 2025. All three figures can be accurate, because they measure different things.

IRE for tumors has moved more slowly, and its US change of December 2024 concerns its indication. An indication is the use a maker claims for its device, and a clearance or approval covers the indication on the label (Chapter 2). AngioDynamics' NanoKnife had long been cleared by 510(k) for the surgical ablation of soft tissue. On 9 December 2024, AngioDynamics announced a clearance that extended that indication to name prostate tissue. NanoKnife's US indication names the tissue rather than a disease, a distinction that matters in reading any device's label.

In practice

Treat the field map as the design of a pulsed-field device. The electrode geometry, the pulse amplitude, width, number and polarity, and the tissue's threshold together decide where the zone ends, and the heating near the electrodes decides whether the zone stays purely non-thermal. Verify the system by measuring zone size and thermal effects across the range of pulse settings, and not only the generator's electrical output.

+ What this chapter established
  • At a cell's poles, an external field induces a membrane voltage of about 1.5 times the field times the radius, pores open above about 0.5 to 1 V, and after enough pulses the cell cannot repair them.
  • Irreversible thresholds are several hundred volts per centimeter and vary by tissue, and the zone is wherever the field exceeds the threshold, with no heat sink to carry the effect away.
  • In animal studies, the collagen scaffold of vessels and ducts survives, which is the reason to use pulsed fields, although every pulse still deposits some heat near the electrodes.
  • IRE pulses are timed to the heartbeat and given under complete neuromuscular block, while cardiac PFA uses biphasic pulses and has been adopted rapidly since FDA first approved a PFA system in December 2023.