The RF generator.
+ The questionHow can one box seal a vessel at 20 Ω and sustain an arc into a load of two kilohms through the same output?
The load problem
The impedance an electrosurgical generator drives spans a hundredfold range, from about 20 Ω in a sealer's jaws to 2 kΩ or more across an arc. No designer of power supplies would choose such a load. The load is the tissue as the generator sees it, and its impedance depends on the tissue, the electrode and the contact area, and above all on what the energy is doing to the tissue at that instant. A vessel clamped wet and well compressed in a sealer's jaws looks like about 20 Ω, and tissue under a cutting electrode like about 300 Ω. A coagulation arc across a gap, or tissue that has dried, can be 2 kΩ or more. An arc's impedance does not hold steady even for a single cycle of the carrier, which lasts a few microseconds, because it changes as the arc forms and goes out within that cycle.
Manufacturers therefore specify each mode at a rated load, the impedance at which they define its power and waveform. Medtronic's rated loads for the Valleylab FT10 are 20 Ω for sealing, 100 Ω for bipolar and soft coagulation, 300 Ω for monopolar cut and 500 Ω for fulguration and spray. A rated load fixes only one point, while in use the tissue moves the load far from it. FDA's electrosurgical guidance gives 100 to 2,000 Ω, a twentyfold span, as an example of the loads across which a monopolar mode's power is plotted. A single generator must therefore cover a wide range of load with every mode it offers.
Because power is voltage squared over resistance, the RMS voltage is the square root of the product of power and resistance, and the current is voltage over resistance. Sealing at 350 W into 20 Ω needs the square root of 7,000, about 84 V, and about 4.2 A. Cutting at 300 W into 300 Ω needs 300 V, the square root of 90,000, and 1.0 A. Coagulation at 120 W into 500 Ω needs about 245 V and 0.49 A, with a peak of about 1.47 kV at a crest factor of 6. Spray at 80 W into 2,000 Ω needs 400 V and 0.2 A, with a peak of 3.2 kV at a crest factor of 8. The coagulation and spray figures are illustrative assumptions, not a particular generator's specification. Sealing therefore needs the most current, about 4.2 A, and spray the highest peak voltage, about 3.2 kV.
The four modes show the difficulty. The output must deliver more than four amperes in one mode and more than three kilovolts peak in another. The powers of the four modes differ by a factor of only about four, but their load resistances differ by a factor of a hundred, so what makes an electrosurgical output stage hard to design is the range of load, far more than the power.
The power chain
Published patents describe how a generator turns mains power into that output in stages. Covidien's patent granted in 2017 describes the core of the chain, and research designs follow the same structure. The list below combines them with earlier designs.
- The mains input filters and rectifies the incoming supply.
- A DC supply is isolated from the mains and produces a steady internal voltage.
- A DC-DC converter sets the amplitude of the output by adjusting that voltage. It is typically a buck converter, which steps a voltage down.
- An inverter switches the DC at the carrier frequency, usually 300 to 500 kHz, and creates the burst patterns of the coagulation modes by switching on and off. It is often a full H-bridge, four power transistors that can drive current through the load in either direction.
- An output network filters the switched waveform into a sine in many designs, sometimes using a resonant circuit. The Covidien design has no such network and drives the load with a square wave.
- An isolation transformer passes the output across the barrier between the internal circuits and the patient and sets the voltage ratio.
- Blocking capacitors in each output lead stop direct and low-frequency current, as Chapter 8 described.
Splitting the job this way lets each stage do one thing well. The converter controls the amplitude, which is what closed-loop control needs to adjust quickly. The inverter controls frequency and timing, and it can stay at a fixed frequency at which the output network is efficient. The transformer and the internal supply voltage together bound the voltage and current that the stage can reach, and choosing them to span both ends of the load range is the central trade-off of the design.
The earliest solid-state generators used the same ideas in a simpler form. Valleylab's patent filed in 1971 describes coagulation produced near the resonant frequency of an output circuit, at about 400 kHz, with 4,000 V across the output when no load is connected. It describes cutting produced off resonance, at a lower peak voltage.
Holding the output: three limits
A generator cannot deliver its set power at every load, because its power stage has a maximum current and a maximum voltage. Into a very low impedance it reaches the current limit first, and at that fixed current the power rises in proportion to the load. Into a very high impedance it reaches the voltage limit first, and at that fixed voltage the power falls as the load rises, in inverse proportion to it. Only between the two can the generator hold the power the surgeon set, so every electrosurgical mode has three regions: current-limited, constant power and voltage-limited.
The number on the dial is the power the generator delivers at the rated load and across its constant-power band. Outside that band it delivers less, sometimes much less. Two generators set to the same power can deliver different power to the same tissue, which is one reason clinical results do not always transfer between generators.
The closed loop works to hold the output in whichever region the tissue puts it. It measures the voltage and current at the output, computes the power and the impedance, and adjusts the converter, often thousands or hundreds of thousands of times a second. Medtronic states that the FT10's tissue-sensing technology monitors impedance 434,000 times a second. Holding the power steady is difficult, because an arc can change the load faster than a slow loop can react, and control methods for exactly this problem are an active research topic.
FDA's guidance asks for the resulting curves, and so does IEC 60601-2-2, the international standard for the safety of high-frequency surgical equipment (Chapter 25). FDA's electrosurgical guidance asks for graphs of output power at the maximum setting and at half of it, over the range of expected loads, derived from measured data rather than from theory. It also asks for the output waveform of each mode at its rated load, with its amplitude, frequency, duty cycle and crest factor, and it asks for the maximum output voltage in the labeling, the information supplied with the device.
Duty cycle and high current
A generator also has thermal limits of its own, so its output cannot stay on all the time. Medtronic states that the FT10 can work at a duty cycle of 25 %, 10 seconds active and 30 seconds inactive, in any mode for 4 hours. This activation duty cycle, the share of time the output is switched on, is a different quantity from the waveform duty cycle of Chapter 8, the share of each cycle in which the carrier is on. The two can differ widely: pure cut's waveform runs at 100 %, while the FT10's activation rating of 25 % covers every mode, pure cut included.
The activation duty cycle matters to the patient as well, because current that flows for a long time heats the tissue under the return electrode (Chapter 8). IEC 60601-2-2 therefore adds rules for outputs that deliver a lot of current for a long time. It defines a heating factor, current squared multiplied by time, and it treats a monopolar mode as a high current mode when its intended use, at maximum current and maximum activation duty cycle, gives more than 30 A²s in any 60 seconds. For the cut of the solved example, at 1.0 A, 30 A²s is 30 seconds of activation within one minute. Such a mode brings stricter requirements for the return electrode. The standard's 2023 amendment also asks for heating under the return electrode to be considered in the risk analysis for any mode whose intended use has an activation duty cycle above 45 %, even below 30 A²s.
Read a generator's datasheet mode by mode, against the tissue the device will meet. For each mode, find the rated load, the maximum power and the load at which the power-versus-load curve leaves its flat region. Find the carrier frequency and burst repetition rate, and the peak voltage and crest factor, which govern arcing, coupling and insulation stress. Find also the declared high-frequency leakage current, which Chapter 20 explains, the alarm rule of the return-pad monitor from Chapter 8, and any duty-cycle limit. Then compare the rated load with the impedance of the tissue the device will actually meet. A mode rated at 300 Ω and used mostly on tissue at 1,000 Ω may run in its voltage-limited region, at less power than its setting, and the power-versus-load curve shows whether it does.
- A generator must drive loads from about 20 Ω to 2 kΩ or more, and that hundredfold range makes its output stage harder to design than its power does.
- The power chain runs from the mains input through a DC supply, a converter that sets the amplitude and a high-frequency inverter, and in many designs an output network, to an isolation transformer and blocking capacitors.
- Every mode has current-limited, constant-power and voltage-limited regions, and the closed loop holds the output within them as the tissue changes.
- FDA's guidance asks a 510(k) to include power-versus-load curves at the maximum and half settings and each mode's waveform at its rated load, and IEC 60601-2-2 adds rules for outputs that deliver high current for a long time.
Let's Talk