02 — How heat moves

How heat moves.

+ The questionIf energy is deposited in one small place, why does the damage spread beyond it, and what carries the heat away?

Heat flows downhill, slowly

The damage an energy device causes can go on growing for a few seconds after activation ends, although no more energy is going in. Chapter 1 treated each point of tissue as if it kept whatever heat it received, and tissue does not. Heat flows from hotter tissue to cooler tissue, at a rate set by the temperature difference and by the tissue's thermal conductivity. Soft tissue conducts heat about as well as water does, about 0.5 watts per meter per degree for liver against 0.6 for water at room temperature, which makes tissue a poor conductor. Copper conducts heat about 800 times better.

How fast a change of temperature spreads depends on more than conductivity. It depends on the conductivity divided by the amount of heat that each unit of volume can store, and that ratio is the thermal diffusivity. For water and most soft tissues it is about 0.14 square millimeters per second. The number looks abstract, but it answers the practical question of how far heat travels in a given time.

The square-root rule

A pulse of heat deposited at a point does not travel outward at a steady speed, the way sound does. It diffuses, and diffusion covers distance in proportion to the square root of time, so a convenient estimate of how far heat has spread after a given time is the diffusion length: the square root of four times the diffusivity times the time.

Solved example: how far heat spreads in a given time

The diffusivity of soft tissue is taken as 0.14 square millimeters per second. After a tenth of a second, four times the diffusivity times the time is 0.056 square millimeters, and its square root is about 0.24 mm. After one second the product is 0.56 square millimeters and the distance about 0.75 mm, and after ten seconds the product is 5.6 and the distance about 2.4 mm. After a minute the product is 33.6 and the distance about 5.8 mm. From a tenth of a second to ten seconds the time grows a hundredfold, but the distance grows only tenfold, because ten is the square root of a hundred.

04 — How far heat spreads in a given time
0.01 0.1 1 10 1 ms 10 ms 0.1 s 1 s 10 s 1 min mm TIME (LOG SCALE) seal cycle in one maker's study, 2–4 s 0.024 mm 0.24 mm 0.75 mm 5.8 mm diffusion length = √(4 × diffusivity × time) diffusivity 0.14 mm²/s: water and most soft tissue 10 s: 2.4 mm 1 s 10 s 0.1 s 1 mm to scale
HeatFocal detail
Plate 04 — Heat spreads a distance that grows with the square root of time: about 0.24 mm in a tenth of a second, 0.75 mm in one second and 2.4 mm in ten. A hundred times longer reaches only ten times farther, so energy delivered quickly stays close to where it is put, and energy delivered slowly spreads.

This one relationship explains a great deal of surgical practice. Energy delivered in a fraction of a second stays within a fraction of a millimeter of where it was put, because heat has had no time to move, while energy delivered over ten seconds has had time to reach tissue two or three millimeters away. To make a narrow effect, such as a clean incision or a precise laser crater, a device must therefore deliver its energy quickly. To make a broad effect, such as coagulating a wide area or ablating a tumor, it can take its time and let conduction do part of the work.

The same rule sets how far a vessel seal's damage spreads. In a 2025 study, Chambers and colleagues measured the seals made by Medtronic's current vessel sealers in excised and in living tissue. All three authors were Medtronic employees, and Medtronic funded the work, so the figures are a manufacturer's own. They report seal times of about 2 to 4 seconds, and over that time the square-root rule gives a diffusion length of roughly one to one and a half millimeters. Measured by histology in living tissue, the thermal spread beyond the jaws was about 1.3 to 2.1 mm, the same order as the rule predicts (Chapter 9).

Heat also keeps moving after the energy stops, which is why the damage can grow after activation ends. A hot jaw or blade goes on conducting heat into whatever it touches, and a heated zone goes on spreading until its temperature falls. Chapter 11 shows why this matters most for ultrasonic instruments.

Blood carries heat away

Conduction is not the only way heat leaves a heated zone. Capillaries run through living tissue, and blood entering them at 37 °C picks up heat and carries it away. The standard description is the bioheat equation that Harry Pennes published in 1948, and in words it is a simple heat balance: the temperature of a small volume of tissue rises by the heat the source deposits in it, minus the heat conducted to its neighbors, minus the heat carried off by blood. Metabolic heat is also in the equation, but during surgery it is negligible.

05 — Where deposited heat goes
capillary stored heat: temperature rises heat deposited by the source conducted to neighbors metabolic heat: negligible here carried away by blood varies about a hundredfold between tissues temperature rise = deposited − conducted away − carried off by blood Pennes' bioheat equation (1948), in words 1 10 100 1,000 blood perfusion, mL/min per 100 g 4 resting muscle 95 liver about 400 kidney
HeatFocal detail
Plate 05 — The bioheat balance in words: a small volume warms by what the source deposits, minus what conducts to its neighbors, minus what blood carries away. Perfusion differs about a hundredfold between resting muscle and kidney, so the same device heats different organs differently.

The blood term varies more than any other. Physiologists call the flow of blood through a tissue its perfusion. Resting skeletal muscle receives about 4 milliliters of blood per minute for every 100 grams of tissue, the liver about 95 and the kidney about 400. At the same power, the same device therefore heats kidney less than muscle, and any model or bench test that ignores perfusion will overestimate heating in well-perfused organs. FDA's guidance for electrosurgical devices recommends thermal testing in several tissue types for a device meant for soft tissue in general, and perfusion is one more reason that tests in dead, unperfused tissue need careful interpretation.

Heating also changes the balance as it proceeds. Inside a coagulated zone the blood in the capillaries clots and stops flowing, so the cooling term vanishes there and heating accelerates. The tissue's properties shift too. Its electrical conductivity rises by about 2 % for every degree, which matters for any energy delivered as current, and its thermal properties change as it dries. Berjano's 2006 review of models of RF ablation notes that the 2 % figure comes from measurements made below 42 °C.

The heat sink

Capillaries cool tissue a little everywhere, but a large vessel cools it a lot in one place. Blood flowing through a vein several millimeters wide holds the vessel wall close to 37 °C against all but intense heating, and the tissue immediately around it stays cooler than tissue farther away. Surgeons and engineers call this the heat sink effect.

06 — The heat sink beside a vessel
probe heated zone large vein wall near 37 °C cells beside the wall survive pig liver, RF: heat sink at 73 % of veins wider than 3 mm, none narrower; viable cells around every vein wider than 5 mm (Lu 2002) patients: incomplete treatment or recurrence in 48 % of tumors touching vessels of 3 mm or more, 7 % of the others (Lu 2003) plan view, schematic
HeatFocal detail
Plate 06 — Flowing blood holds a vessel wall near 37 °C, so the heated zone bends away from it and a cuff of cells beside the vessel survives. In pig liver the effect appeared around most veins wider than 3 mm and none narrower; in patients, tumors next to such vessels failed treatment far more often.

The effect is strongest in ablation, where the goal is to kill every cell in a volume. Lu and colleagues at the University of California, Los Angeles, measured it in 2002, in an animal study of RF ablation in the livers of 10 pigs. Of the veins wider than 3 mm, 73 % deformed the ablation zone, while none of the smaller veins did, and viable cells remained around every vein wider than 5 mm. In 2003 Lu and colleagues reported a clinical series of 105 liver tumors treated with RF ablation. Treatment was incomplete, or the tumor recurred, in 48 % of the tumors that touched a vessel of 3 mm or more, against 7 % of those that did not. Both studies divided the vessels at 3 mm, so together they support the "3 mm rule" of ablation practice, that a vessel 3 mm or more across beside a tumor acts as a heat sink.

The heat sink can also help. It protects large vessels that run near a heated zone, which is one reason tumors next to major vessels can be ablated at all. It is also one more reason a vessel sealer clamps its vessel shut, because with the flow stopped inside the jaws the vessel's own blood can no longer carry the heat away.

A clean zone on the bench can hide a heat sink

Most bench tests of thermal devices use excised tissue with no blood flow. Such a test shows the zone a device makes when no blood carries heat away, which is usually larger than the zone in living tissue, and it shows nothing of what happens next to a perfused vessel. Chapter 13 returns to the heat sink in ablation, where microwave energy reduces it but does not remove it.

Lateral thermal spread

The practical measure of all of this is lateral thermal spread: how far thermal damage extends beyond the tissue the device was meant to affect. It is measured after the fact, by cutting the tissue, staining it and measuring the damaged zone under magnification. A device's indication is the use its maker claims for it, and a general soft-tissue indication covers soft tissue in general rather than one procedure or organ. For a device with such an indication, FDA's electrosurgical guidance recommends exactly this measurement: the length, width and depth of the thermal damage zone in at least three tissues, such as liver, kidney and muscle, measured in triplicate at the minimum, default and maximum settings.

Spread depends on everything in this chapter: how much energy the device delivers, how fast, into which tissue, with how much blood flow, and how long the device and the tissue stay hot afterward. It is the single number that best summarizes whether a device put its effect where the surgeon intended, and Chapters 9, 17 and 22 return to it for vessel sealers.

In practice

When comparing figures for thermal spread, check the conditions under which each was measured. Spread measured in excised tissue after a short activation says little about a long activation in perfused tissue. The comparisons that mean something use the same tissue, the same vessel sizes, the same settings and activation times, and histology rather than surface appearance.

+ What this chapter established
  • Soft tissue conducts heat about as well as water, and its thermal diffusivity is about 0.14 square millimeters per second.
  • Heat spreads a distance that grows with the square root of time, about 0.24 mm in a tenth of a second, 0.75 mm in one second and 2.4 mm in ten, so energy delivered quickly stays close to where it was put.
  • Blood carries heat away, perfusion varies about a hundredfold between resting muscle and kidney, and large vessels act as heat sinks.
  • Lateral thermal spread is measured by histology and is the practical summary of where a device's heat actually went.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.