Tissue as a conductor.
+ The questionWhy does current heat tissue at all, and why must surgical current alternate hundreds of thousands of times a second?
Tissue conducts with ions
In 1891 the French physiologist Jacques-Arsène d'Arsonval found that the response of muscle to alternating current faded as he raised the frequency and had all but disappeared at about 10 kHz, the highest frequency his equipment could reach. The reason lies in how tissue carries current, which also explains why current heats tissue at all. A metal conducts because some of its electrons are free to move, but tissue has no free electrons. It conducts because it is salt water: sodium, potassium and chloride ions dissolved in the fluid inside and between cells carry charge when a voltage pushes them. Anything that changes how many ions there are, or how freely they can move, therefore changes how well a tissue conducts.
Conductivity measures how readily a material carries current, in siemens per meter, and it differs widely between tissues. Gabriel and colleagues published parametric models of the electrical properties of tissues in 1996, and at 500 kHz, a typical electrosurgical frequency, those models give blood a conductivity of about 0.75 siemens per meter, muscle 0.45, kidney 0.23 and liver 0.15. Fat holds little water and conducts at about 0.025, eighteen times less than muscle, and dry skin at about 0.004. Physiological saline at body temperature conducts at about 1.9, roughly thirteen times better than liver. Measurements on living liver are reported to be higher than those on excised tissue, so a fair range for liver is 0.1 to 0.4 S/m, depending on the source and the condition of the tissue.
These values explain where current goes inside a patient, because current offered a choice of paths divides in proportion to how well each path conducts, so it crowds into wet, lean tissue, blood and saline, and avoids fat and dry skin. A pool of saline irrigation around an electrode can carry current away from the tissue it was meant to heat.
How current becomes heat
Current heats any resistor it flows through, and school physics gives the power as the square of the current times the resistance, or equally as the square of the voltage divided by the resistance. Inside tissue it is more useful to think about each small volume and the current density in it, which is the current crossing each unit of area. The heat released in each cubic millimeter is the square of the local current density divided by the local conductivity. That is Joule heating, and it is how every RF device in this guide heats tissue.
Joule heating has a consequence that looks like a contradiction: for a given current density a poor conductor heats more, while for a given electric field a good conductor heats more. Which view applies depends on whether the current or the voltage is held fixed, and Chapter 10 shows that a generator switches between the two as the tissue changes.
Current density decides where the heat appears
The single most important idea in electrosurgery is that heating follows current density, and current density depends on geometry. Current leaving a small electrode must spread out through an ever larger area of tissue. Around a small sphere, the area the current crosses grows with the square of the distance from the center, so the current density falls with the square of the distance. Heating follows the square of the current density, so it falls with the fourth power of the distance: tissue twice as far away heats sixteen times more slowly.
A current of 0.5 A flows out of a small spherical electrode into muscle, whose conductivity is taken as 0.4 siemens per meter. At 1 mm from the center the current crosses the surface of a sphere 1 mm in radius, about 12.6 square millimeters, so the current density is about 4 A per square centimeter. The square of that density divided by the conductivity gives a heating power of about 4 kW in every cubic centimeter. Tissue needs about 3.8 J to warm one cubic centimeter by one degree, so before any heat conducts away the tissue there heats at about 1,000 °C per second. At 2 mm the sphere's area is four times larger, the current density a quarter and the heating a sixteenth, about 65 °C per second. At 4 mm it is about 4 °C per second, and at 10 mm about 0.1 °C per second.
The fall-off is so steep that heat appears only where current is concentrated. At the tip of an electrosurgical pencil, a fraction of a square millimeter carries the whole current, and tissue there boils in milliseconds. The same current leaves the patient through a return pad, a large electrode placed on the skin to carry it back to the generator. Spread over the pad's tens of square centimeters, the current warms the skin by a few degrees. Nothing in the circuit is different except area.
The same principle explains most of the burns that Chapter 18 describes. Any place where current is forced through a small area acts as an electrode: the edge of a return pad that is peeling, a pinhole in an instrument's insulation, the point where a shaft touches bowel. The current heats each of these small areas in the same way, whether or not the surgeon intended it to pass there.
Cells are tiny capacitors
So far tissue has been treated as a uniform conductor, but on the scale of cells it is not one, because each cell is wrapped in a membrane only a few nanometers thick that is an excellent electrical insulator, with conducting fluid on both sides. A thin insulator between two conductors is a capacitor, and so each cell membrane behaves as one.
A capacitor blocks steady current but passes alternating current, and it passes more the faster the current alternates. At low frequency the membranes block current, which then flows only through the fluid between cells. As the frequency rises, the membranes pass more and more of it, and current flows through the cells as well as around them. Gabriel and colleagues' models show the change directly. Between 100 kHz and 1 MHz, muscle's conductivity rises from about 0.36 to 0.50 S/m, because more of the current passes through the cells. Permittivity measures how much charge the membranes store, as a multiple of what empty space would store in the same field, and over the same range muscle's permittivity falls from about 8,000 to 1,800.
Why the current must alternate so fast
The capacitor picture also explains why surgical current alternates at hundreds of kilohertz. A nerve or muscle cell fires when the voltage across its membrane changes by enough, for long enough. Direct current pushes one way all the time, and mains current at 50 or 60 Hz pushes one way for milliseconds at a time, which in both cases is long enough to charge membranes fully and to fire nerves and contract muscles, including the heart.
At a higher frequency, each half-cycle becomes shorter than the time a membrane needs to charge and a nerve needs to respond: nerve cells need a stimulus lasting about 0.1 to 10 milliseconds, while at 400 kHz one half-cycle lasts 1.25 microseconds. The current still flows and still heats, but it no longer stimulates.
That is the fading that d'Arsonval saw as he raised the frequency. When he applied high-frequency current to himself at a high level, he felt nothing but warmth, and at a demonstration in July 1893 a commission recorded about 0.8 A passing through its members' arms. Modern generators work far above the 10 kHz his equipment reached. IEC 60601-2-2 is the safety standard for electrosurgical equipment, and it defines the high-frequency range of such generators as below 5 MHz and generally above 200 kHz. Most of them run between 300 and 500 kHz, near the low end of that range, for a reason given at the end of this section.
When an electrode arcs, the arc conducts better in one direction than the other and partly rectifies the current, which creates a small low-frequency component that can stimulate muscle. Generators place capacitors in series with their outputs to block direct and low-frequency current, and these capacitors limit the effect but do not remove it entirely (Chapter 8).
There is also a reason not to go much higher. The current that flows through a capacitor rises in proportion to frequency, and every cable, instrument shaft and patient forms small, unintended capacitors with whatever lies near it, which engineers call stray capacitance. At higher frequencies more current leaks away through those unintended paths, an effect Chapter 18 quantifies. The working band of a few hundred kilohertz is therefore a compromise between stimulation at the low end and stray coupling at the high end.
Ask two questions of any RF device. First, where is the current density highest, and is that where the heating is wanted? Second, what else in the circuit offers current a small area or a capacitive path, including the cables, the instrument shafts and the patient's contact with the table? Both questions appear in the risk analysis that every electrosurgical device must carry, though in different words.
- Tissue conducts through dissolved ions, and at 500 kHz its conductivity ranges from about 0.75 S/m for blood down to 0.025 S/m for fat and less for dry skin.
- The heat released in each small volume of tissue is the square of the current density divided by the conductivity, and near a small electrode it falls sixteenfold each time the distance doubles.
- Cell membranes act as capacitors, so at high frequency current passes through the cells as well as around them, and each half-cycle is too brief to fire a nerve.
- IEC 60601-2-2 defines the high-frequency range of surgical generators as below 5 MHz and generally above 200 kHz, and most run at 300 to 500 kHz, a compromise between nerve stimulation at the low end and stray coupling at the high end.
Let's Talk