Focused ultrasound: HIFU and histotripsy.
+ The questionEvery energy so far reaches tissue through a needle, an instrument or an incision. How can sound destroy tissue deep inside a closed body without breaking the skin?
Bringing sound to a point
On 6 October 2023, FDA authorized HistoSonics' Edison system to destroy liver tumors by a non-thermal, mechanical process, in which pulses of focused sound sent in through the skin break the tumor's cells apart. The pulses can reach the tumor because sound travels through soft tissue with little reflection and moderate absorption, as Chapter 4 showed. A focused transducer uses both facts. Its surface is curved like the inside of a bowl, or made of many elements driven in a carefully timed sequence, so that sound leaving every part of it arrives at one point in phase. Spread over the large area of the transducer and the skin beneath it, the intensity is modest, but concentrated at the focus it is many times higher. That difference is what lets the sound destroy tissue at the focus while sparing the skin it crosses.
The focal zone is an elongated spot rather than a point, narrow across the beam and longer along it, and the wavelength and the geometry set its dimensions. Sonablate treats the prostate at 4 MHz, and FDA's decision summary for it describes a single lesion about 10 to 12 mm long and 1 to 2 mm wide, made at focal intensities of 1,200 to 2,000 W/cm². Larger volumes are treated by moving the focus and laying lesions side by side, under imaging.
Everything in the beam's path matters, and Chapter 4's reflection figures become rules for treatment. Gas in the bowel or lung reflects 99.9 % of the sound and blocks it, while bone reflects about two fifths, absorbs much of the rest and heats. Fat attenuates the sound. Perinephric fat surrounds the kidney, and in their 2024 review of HIFU, Turner and Cranston report a 2020 study in which that fat reduced the energy delivered from 58 % at a depth of 2 cm to 26 % at 5 cm. Treatment therefore needs the acoustic window of Chapter 4, a path of soft tissue free of gas and bone, and planning centers on finding one.
Thermal HIFU
In high-intensity focused ultrasound, or HIFU, the sound absorbed at the focus heats it during exposures that last seconds, and FDA's summary for Sonablate describes three seconds on and six off. Focal temperatures rise well above the 56 to 60 °C often quoted for tissue death within about a second, and Turner and Cranston report temperatures above 80 °C. Chapter 1 shows how much the time to tissue death at a given temperature varies from one tissue to another. Because the focus is small and heating is rapid, the lesion boundary is sharp, and tissue a few millimeters away is spared.
The advantages of a treatment without an incision are balanced by constraints. Each lesion is small, so large volumes take time. The heat is still heat, so perfusion and the heat sink of Chapter 2 apply, and nearby sensitive structures still need protection. The imaging that guides the treatment matters too, because systems guided by MRI can map temperature in the tissue during treatment, while ultrasound-guided systems see the anatomy in real time but have no direct thermometry.
In the US, FDA first approved a focused-ultrasound device on 30 June 1988, when it approved Sonocare's system for treating glaucoma by premarket approval, FDA's route for devices of the highest risk (Chapter 26). An MRI-guided system that treats uterine fibroids followed in 2004, also by premarket approval. For the prostate, FDA granted the Sonablate 450 a De Novo authorization on 9 October 2015. Chapter 14 introduced the De Novo route, which serves a new kind of device that has no predicate, no device already legally on the market to compare it with. With Sonablate's authorization, FDA created a new regulation for high-intensity ultrasound systems that ablate prostate tissue.
EDAP made a competing system called Ablatherm, and after Sonablate's authorization it withdrew its own De Novo request. EDAP filed a 510(k) instead, the route by which FDA clears a device by comparing it with a predicate, and named Sonablate as that predicate (Chapter 9). FDA cleared Ablatherm on 6 November 2015, and EDAP's later system Focal One received 510(k) clearance on 7 June 2018. The sequence shows how the US system works: the first device of a new kind creates a classification, and later devices of that kind follow it (Chapter 26). The international safety standard for such equipment is IEC 60601-2-62, which IEC first published in 2013. FDA's decision summary for Sonablate cites the diagnostic ultrasound standard, IEC 60601-2-37, instead.
Histotripsy: sound without heat
Histotripsy uses the same focused sound in a completely different way: instead of long exposures that let heat accumulate, it fires pulses lasting microseconds at duty cycles below 1 %, which means that the sound is on for less than a hundredth of the time. The peak negative pressure of each pulse is high enough to produce inertial cavitation at the focus. Chapter 4 described inertial cavitation as the violent growth and collapse of bubbles, and it put the threshold for single-cycle pulses at about 25 to 30 MPa of negative pressure, with no seed bubble already present. A cloud of bubbles forms, grows and collapses with each pulse, and the repeated collapses break the cells apart, after which the body clears the debris. With so little time on, the tissue barely warms.
FDA's decision summary for Sonablate describes HIFU exposures of three seconds on and six seconds off. Each cycle therefore lasts nine seconds, and the transducer is on for three of them, a third of the time, or about 33 %. A histotripsy system runs at a duty cycle below 1 %, so over the same nine seconds its pulses add up to less than a hundredth of nine seconds, under 0.09 seconds of sound. Three seconds divided by 0.09 seconds is about 33, so the HIFU transducer is on more than 33 times as long as the histotripsy transducer, which is silent for more than 99 % of the time. HIFU therefore keeps its sound on long enough for heat to build at the focus, while histotripsy leaves the focus almost no time in which to warm.
The mechanism has three useful consequences. There is no heat sink, because heat is not the agent. Collagen-rich structures such as vessels and bile ducts are more resistant to the mechanical stress than the cells around them, and in a 2026 review of preclinical studies, Yu and colleagues report that vessels larger than about 1 mm are preserved. And the bubble cloud is bright on ultrasound, so the treatment can be watched as it happens.
The same review describes Edison as operating at 700 kHz, with pulses shorter than 20 µs and duty cycles below 1 %. FDA's De Novo authorization of October 2023 covers the destruction of liver tumors by a non-thermal, mechanical process. The tumors it covers include unresectable ones, which surgery cannot remove. FDA's decision summary reports the results of HistoSonics' own trial, which supported the request: technical success was 95.5 %, in 42 of 44 tumors, and major complications occurred in 3 of 44 patients at 30 days. The trial treated tumors smaller than 3 cm, required an adequate acoustic window, and kept the main portal vein, main bile ducts, gallbladder, stomach and bowel out of the treatment volume.
HistoSonics announced on 11 May 2026 that it had submitted a De Novo request for kidney tumors, and as of October 2026 no FDA decision has been announced. On 13 July 2026, HistoSonics also announced CE marking for Edison in Europe, for the destruction of liver tumors. CE marking is the mark a manufacturer affixes before a device can be sold in the European Union, once its conformity assessment is complete (Chapter 27).
Press coverage of new energies often describes a new use as available before any regulator has authorized it. A De Novo request or a 510(k) submission is an application, and only FDA's decision makes a device legally marketable for that use. Histotripsy for liver is authorized in the US, while the kidney request had no announced decision as of October 2026.
HIFU and histotripsy can use similar transducers, focused through the same windows, and differ mainly in timing and peak pressure. Long exposures let heat build, and the focus coagulates. Microsecond pulses at a duty cycle under 1 % leave no time for heat to build but pull hard enough to cavitate, and the focus is fractionated mechanically. In principle, one transducer driven in two different ways becomes two different energies.
Treat the path to the focus as part of a focused-ultrasound device. Write the acoustic window into the requirements: the gas and bone the beam must avoid, the depth of fat it may cross, and how the user confirms a clear path before treatment. Specify each exposure by its timing and its peak pressure together, because, in principle, one transducer can coagulate or fractionate depending on both. State how the user watches the treatment, by MRI thermometry, by the bubble cloud on ultrasound or by the anatomy alone, and verify that view against lesions measured afterward. Before choosing a US route, check whether a device of the same kind has already created a classification, because a later device can follow it by 510(k), as Ablatherm followed Sonablate.
- A focused transducer concentrates sound at a small, elongated focus deep in the body while keeping the intensity low along the path, which must avoid gas and bone.
- Thermal HIFU heats the focus above about 60 to 80 °C in seconds, and because each lesion is small, volumes are treated lesion by lesion.
- Histotripsy fires microsecond pulses above the cavitation threshold at a duty cycle under 1 %, fractionating cells mechanically with little heat and no heat sink.
- Sonablate in 2015 and Edison in 2023 were authorized by De Novo, and later prostate systems followed by 510(k), while the kidney request that HistoSonics announced in May 2026 had no announced FDA decision as of October 2026.
Let's Talk