+ Part III · Current through the patient
Electrosurgery.
A vessel sealer can close an artery up to 7 mm across, and nothing inside its jaws measures the seal: the generator infers it from the voltage and current at its own output. Electrosurgery was in surgical use by 1926, the year of the operation in the opening story, and from Bovie's cutting loop to today's sealer it rests on the rule of Chapter 3, that heat appears where current is densest. This part follows that current from the tip of a monopolar pencil, through the jaws of a sealer, to the generator that drives both, and shows how each is built so that the heat appears only where the surgeon intends.
Monopolar RF.
+ The questionIf current must cross the whole body to complete its circuit, how does it heat only the point the surgeon touches?
The circuit
The pad on a patient's thigh or back carries the same current as the electrode in the surgeon's hand, yet tissue at the tip boils within milliseconds while the skin under the pad warms by a few degrees. The difference comes from the geometry of the circuit alone. A generator drives high-frequency current through a lead to the active electrode, the pencil, hook or scissors in the surgeon's hand. The current enters the patient at the electrode's tip, spreads through the body and leaves through a large return electrode, the pad, which leads it back to the generator. The patient lies in the middle of the circuit, a conductor of irregular shape whose conductivity changes from one tissue to the next.
The same current flows through the tip and the pad, so the difference lies in the area each one offers it (Chapter 3). At the tip the current is squeezed into a fraction of a square millimeter, while at the pad it spreads over tens of square centimeters. Current density at the tip is therefore thousands of times higher, and because heating rises with the square of current density, the heat released in each cubic millimeter there is higher by a factor of millions.
The early history of electrosurgery is a history of learning where else the current could go. William Bovie filed a patent on his apparatus in 1928, and the apparatus it describes made both its cutting and its coagulating currents with spark gaps, as the opening story tells. In the same story, current from Bovie's early apparatus once reached Cushing's arm through a metal retractor. For decades afterward, generators were built with outputs referenced to earth. Current from such an output could return through any earthed contact on the patient, and where that contact was small, it burned. Valleylab's patent filed in 1971 names the electrodes of patient-monitoring equipment as one such path. It describes a generator built from solid-state components, with an isolated output that floats electrically and returns current only through its own return lead, "to prevent patient burns caused by alternate return current paths".
At the tip: from contact to arcs
What the current does at the tip depends mainly on the voltage. Below about 200 V peak, current flows by direct contact between the electrode and moist tissue. The tissue heats, its water boils away gradually, and it dries into a coagulum. This is the desiccation of Chapter 1, or soft coagulation, and it makes no sparks.
Above about 200 V peak, the first burst of heating makes a thin layer of steam between the electrode and the tissue. Steam does not conduct, but at these voltages the field across the thin layer is strong enough to ionize it, and current crosses the gap as tiny arcs. Each arc concentrates its energy into a very small spot, so the cells beneath it vaporize and burst. The electrode rides on its own envelope of steam and moves through the tissue along a line of bursting cells without touching it, which is how it cuts without being sharp. In a 2024 review of the physics of electrosurgery, Koninckx and colleagues put the onset of sparking at about 200 V peak. One of the authors is affiliated with BOWA-electronic, a generator maker. The Technology Committee of the American Society for Gastrointestinal Endoscopy states in a 2013 report that a continuous current with a peak above 200 V is often called a pure-cut current. Both figures put the change from contact heating to cutting at about 200 V peak.
At higher voltages still, in the kilovolts, interrupted bursts of current jump an air gap to the tissue surface. The arcs spread across the surface, dry and char it, and stop oozing from a broad area. This is fulguration, or spray coagulation, and it needs no contact at all.
The waveform decides the effect
A generator produces these effects by changing the shape of its output more than its power. For cutting it delivers a continuous sine wave of a few hundred volts. For coagulation it delivers the same carrier in short bursts, switched on for only a small part of each cycle, at a much higher peak voltage. The fraction of the time the output is on is the duty cycle. The RMS voltage is the steady voltage that would deliver the same power into the same load, and the ratio of the peak voltage to the RMS voltage is the crest factor. A continuous sine wave has a crest factor of 1.41. The lower the duty cycle, the higher the crest factor, and the higher the peak voltage must be to deliver the same power.
Two waveforms are to deliver the same average power into the same tissue: a continuous sine wave, and bursts of the same carrier switched on for 6.25 % of the time, one sixteenth. Power into a fixed load follows the square of the RMS voltage, so the two waveforms need the same RMS voltage. Switching the sine on for one sixteenth of the time divides its mean squared voltage by 16, so its RMS voltage falls to a quarter, the square root of one sixteenth. To restore it, the peak of each burst must be four times the peak of the continuous sine, and the crest factor rises by the same factor, from 1.41 to about 5.7. At equal power, a waveform that is on for one sixteenth of the time must therefore reach four times the peak voltage. Medtronic's specification for its Valleylab FT10 generator lists a typical crest factor of 5.4 for its fulguration mode, which runs at 6.25 %, close to the 5.7 computed here for ideal bursts.
Medtronic's specification sheet shows the same pattern across the FT10's monopolar modes, the preset outputs that the surgeon selects for each task: the lower a mode's duty cycle, the higher its crest factor and its peak voltage. The FT10's pure cut runs at a 100 % duty cycle, with a typical crest factor of 1.6 and a peak voltage of 1,287 V. Its blend runs at 50 %, with a crest factor of 2.2 and 2,178 V, and a third cut mode, which Medtronic names Valleylab, runs at 25 %, with 3.2 and 2,783 V. Its fulguration runs at 6.25 %, with 5.4 and 3,448 V, and its spray at 4.76 %, with 6.2 and 3,932 V.
The high peak voltage of the coagulation waveforms is what lets them arc across a gap, and it is also what makes them dangerous in the ways Part VII describes. A higher voltage breaks through damaged insulation more easily and drives more current through stray capacitance. The same power cuts at a low crest factor, and coagulates and fulgurates at a high one, so the effect on tissue follows the waveform far more than the number on the power dial.
Medtronic's table for the FT10 gives peak voltage, measured from zero to the top of the wave. A reseller's copy of the table for the older Valleylab Force FX gives peak-to-peak voltage, from the bottom of the wave to the top, and Koninckx and colleagues' 2024 review gives both. For a symmetrical wave the peak-to-peak figure is twice as large, so a "cut at 2,000 V" and a "cut at 1,000 V" can describe the same waveform. A comparison of two figures starts by establishing whether each source means peak or peak-to-peak.
The return pad
The return pad works only while its contact area stays large. If part of it peels away, the same current crowds into the remaining contact, and because heating rises with the square of current density, the skin under the edge that still touches heats fastest. A pad that has lost two thirds of its contact carries three times the current density, and each square centimeter of skin still in contact heats about nine times as fast.
Two engineering responses followed. The first is a test. The AAMI HF18 standard required that a return electrode carrying 700 mA for 60 seconds must not raise the temperature of the skin beneath it by more than 6 °C, and IEC 60601-2-2 sets the same 6 °C limit for its own test of return electrodes (Chapter 25). A MegaDyne patent cites the Emergency Care Research Institute for a threshold of necrosis under a return electrode of about 100 mA per square centimeter. An adult pad of about 130 cm² carrying 700 mA has a current density of about 5 mA per square centimeter, some twenty times below that threshold. With two thirds of it lifted, the same pad carries about 16 mA per square centimeter. That is still below the threshold, but each square centimeter in contact heats nine times as fast as in the test that allows a rise of 6 °C.
The second response is a monitor. A split pad has two halves, and the generator passes a small interrogating current from one half to the other through the skin. If the pad starts to lift, the impedance between the halves rises. Valleylab filed a patent on the idea in 1981, and it was granted in 1983. Its monitor uses a 2 mA signal and accepts an impedance between a fixed lower limit of 20 Ω and an adaptive upper limit, set at 120 % of the impedance it records once the pad is in place. When the impedance leaves that window, the generator stops. Later Valleylab generators use the same principle with other limits. The user's guide and service manual of the Force EZ give 5 to 135 Ω as the safe range and describe an alarm when the impedance rises more than 40 % above its starting value. Medtronic's 2023 specification for the FT10 states that its monitor measures contact resistance from 5 to 135 Ω, and gives no alarm rule.
The monitor protects only if the pad is used as designed. Pads come in sizes for neonates, infants, children and adults, and AORN, the association of perioperative nurses, states in a 2016 sample policy that they must not be cut or folded. Cutting reduces the area and can defeat the split-pad monitor. In 2011, 48 leaders of the Society of American Gastrointestinal and Endoscopic Surgeons took a knowledge test, and 10 % of them thought a pad should be cut to fit a child (Chapter 18 returns to the test). The same policy asks for the pad to go over a large, well-perfused muscle, as close to the operative site as practical, so that the current's path through the body is short.
Blocking capacitors
One more component sits in every monopolar output: a capacitor in series with each lead. Chapter 3 noted that an arc conducts slightly better in one direction than the other, so it partly rectifies the current and leaves a small direct-current component. A series capacitor passes the high-frequency current freely but blocks direct and low-frequency current, which could otherwise stimulate nerves and muscles. An ArthroCare patent application filed in 2010 points out a complication: the charge that arcing builds up on such a capacitor can itself discharge through tissue at a low frequency once the arc stops, at 100 Hz in the application's example. The size of the capacitor is therefore a design decision with consequences of its own.
Read a monopolar generator's datasheet for three things from this chapter. First, check the peak voltage and crest factor of each mode, because they govern arcing and coupling, and whether the voltage is quoted as peak or peak-to-peak. Second, check the rated load, because each waveform is specified at that load and nowhere else. Third, check the return-pad monitor's alarm rule, and whether the datasheet states one, because the rule decides how much pad lift the generator tolerates before it stops.
- Monopolar current enters through a small active electrode and leaves through a large return pad, and only the difference in area confines the heat to the tip.
- Below about 200 V peak, contact current desiccates tissue; above it, current arcs across a steam envelope and vaporizes cells, so the electrode cuts without being sharp.
- A lower duty cycle raises the crest factor, so at the same power it needs a higher peak voltage, and the waveform decides cut, blend, fulguration or spray more than the power setting does.
- Isolated outputs return current only through the generator's own return lead, a split-pad monitor stops the generator when the pad lifts, and blocking capacitors stop the direct and low-frequency current that could stimulate nerves and muscles.