From requirement to evidence.
+ The questionHow does a team turn "seal vessels up to 7 mm" into requirements, tests and records that prove the device does it, and keep that proof true after every change?
From a claim to design inputs
The claim that a sealer seals vessels up to and including 7 mm cannot fail a test as it stands. It names no vessel type, no state in which the diameter is measured, no pressure the seal must hold and no number of seals to try. Valleylab's 510(k) summary for a LigaSure clearance of January 2005 describes the sealer as "capable of sealing vessels up to, and including, 7mm" (Chapter 9). Twenty-one years later Medtronic states the same limit for the LigaSure sealer of its Hugo robot (Chapter 23). FDA's guidance for bipolar vessel sealers of 15 August 2016 supplies one missing piece, asking the indication to state "a maximum vessel size measured in the rounded, perfused state in a normotensive subject", one with normal blood pressure. The vessel types, the pressure and the number of seals remain the team's work.
Design controls are the process that does that work. FDA's design control guidance of 11 March 1997 draws them as a waterfall. User needs become design inputs, the requirements the design must meet, and the design process turns these into design outputs. Verification checks each output against its input. Validation goes further: the guidance describes it as asking "whether devices produced in accordance with the design actually satisfy user needs and intended uses". FDA still lists the guidance as final as of October 2026, although it was written for the design-control section of the former US rule (Chapter 26). FDA's quality management system regulation replaced that rule on 2 February 2026 and rests on ISO 13485, the international standard for a device maker's quality management system (Chapter 26). ISO 13485 sets out the same steps, so the waterfall of 1997 still describes the work, and FDA's slides on the standard state that outputs must "meet the input requirements".
Requirements at four levels
A requirement is useful only if a test can fail it, and the claim becomes one first for the whole system. FDA's sealer guidance names the three system tests of Chapter 9. Burst pressure is measured for each mode and representative instrument, against the predicate, across the vessel sizes of the indication and the vessel types the team justifies. Thermal spread is measured by histology under magnification, two dimensions being "minimally acceptable". A chronic animal study follows the seals for at least three weeks in at least five animals, against success criteria "pre-defined and clinically justified", such as no vessel leaks and normal healing.
No single part makes the seal, so each system requirement is divided among the parts that produce it. The jaws must apply the controlled pressure of Chapter 9, and the instrument's rating must cover every peak the generator can deliver (Chapter 21). The generator must deliver its sealing mode and meet the essential performance of Chapter 20. For the FT10, Medtronic rates that mode at 20 Ω and up to 350 W, with a peak voltage of at most 244 V (Chapter 9). The software must infer the end of the seal from the impedance curve of Chapter 9, and the interfaces of Chapter 21 become requirements of their own. Each part can then be verified on its own, and the system test shows whether together they meet the claim.
Risk controls in their order, on one hazard
Risk management runs inside the design, because every control it chooses becomes a requirement with a test. As of October 2026, ISO lists ISO 14971:2019 as current, confirmed on 7 March 2025, and FDA's training slides quote its definitions. A hazard is a "potential source of harm", a hazardous situation the "circumstances in which people, property or environment is/are exposed to one or more hazards", and harm "injury or damage to health of people, or damage to property or environment". Every hazard of Chapters 18 and 19 enters the risk file in these terms. On the return pad of Chapter 8, the hazard is the monopolar current that leaves through the pad. The hazardous situation is a pad partly lifted during activation, so that the current crowds into the edge still in contact, and the harm is a burn there.
ISO 14971 ranks the ways to control a risk, and BSI gives the order in a white paper on the 2019 edition. BSI is a notified body, one of the organizations designated to assess devices under EU law (Chapter 27). In BSI's words, "The first and preferred option is to eliminate the risk", by inherently safe design; the second is protective measures; and the third, where those reduce the risk too little, is information for safety. UL Solutions, a certification body, gives barriers and alarms as examples of protective measures, and a written warning as an example of information for safety, in a summary of the 2019 edition published in 2020. On the lifting pad the three options fall on three parts of the system. The design is the pad's area, since Chapter 8 showed that an adult pad stays below the threshold of necrosis even with two thirds lifted, although the skin still in contact then heats nine times as fast. The protective measure is the split pad and its monitor, which stops the generator when the impedance between the halves rises. The information is the instruction not to cut or fold a pad.
Each control needs evidence of its own. The pad's area is tested by the return-electrode heating test, which allows the skin a rise of no more than 6 °C (Chapter 25). The monitor needs a study with each pad the labeling lists, and Intuitive's 2023 summary, for example, reports a contact-quality-monitor study and compatibility with an Erbe split pad (Chapter 21). Information works only if people act on it, and the knowledge test of Chapters 8 and 18 illustrates its weakness: 10 % of the 48 leaders thought a pad should be cut to fit a child, the opposite of the instruction not to cut or fold a pad. BSI adds that each control is verified once for its implementation and once for its effectiveness, and the remaining risk is then estimated again.
Verification, validation and use
Each level of test shows something the others cannot. Bench tests into resistors show that the generator delivers its declared output across the load range (Chapters 10 and 25), and say nothing about tissue. Burst and histology tests show what a seal does when it is tested, and only the chronic animal study shows whether it survives healing. Intuitive's summary for the 2024 clearance of its E-200 generator sorts its evidence the same way: bench verification of software, cybersecurity, EMC, electrical safety, system interface and instrument compatibility, then validation by simulated clinical use in pigs and by a human factors evaluation.
Use is the level that bench and animal tests miss. FDA's human factors guidance of 3 February 2016 defines a use error as a "user action or lack of action that was different from that expected by the manufacturer", and a critical task as one that, "if performed incorrectly or not performed at all, would or could cause serious harm". Formative evaluations test the user interface while it is still changing. Human factors validation testing comes at the end, on a user interface that "represents the final design", and the guidance states that "in general, the minimum number of participants should be 15" for each distinct user population. The Johner Institute, a consultancy, notes that IEC 62366-1 calls the same test a summative evaluation. On a robot one design has several set-ups to validate: Intuitive's 2023 summary reports human factors work on the E-200's integrated, backup and standalone set-ups, and Medtronic's Hugo summary lists a human factors validation, although neither gives results (Chapter 21).
One requirement traced from claim to label
Traceability links each claim to the requirements that serve it, the tests that verify them and the label line that limits it. FDA's sealer guidance builds the lowest link into its burst-pressure report: a table listing each vessel sealed with its burst pressure, vessel type, diameter and device sample number, so that each seal points to the instrument that made it. The top link is the label warning the guidance recommends: "DO NOT USE this instrument on vessels larger than __mm in diameter." The record that holds the chain is the design and development file, which ISO 13485 requires for each device type or family.
FDA's guidance sets no pass value, so the team chooses 360 mmHg, three times a systolic pressure of 120 mmHg, the industry convention of Chapter 9. The design input reads that seals of vessels up to 7 mm, measured rounded and perfused, hold at least 360 mmHg in each mode with each representative instrument. FDA sets no sample size either, so the team picks a confidence and a reliability and plans a test that allows no failure. A team that wants 95 % confidence that at least 90 % of seals hold, and treats each seal as an independent trial, reasons from the chance of a clean run. If only 90 % held, the chance that 28 seals in a row all passed would be 0.9 multiplied by itself 28 times, about 0.052, just above 5 %. For 29 seals it is about 0.047, below 5 %, so 29 seals without a failure are the smallest test that supports the requirement at that confidence. For 95 % of seals at 95 % confidence the count is 59, since 0.95 multiplied by itself 59 times is about 0.049, and for 90 % at 90 % confidence it is 22, since 0.9 multiplied by itself 22 times is about 0.098, below 10 %. The count also multiplies with the configurations the claim covers: a hypothetical sealer with one mode and two jaw sizes needs 29 seals of each size, 58 in all, with the predicate sealed beside them. The counts are the team's choices of confidence and reliability turned into seals, and a single failure leaves the requirement unproven under this plan.
Twenty-nine seals that held support the requirement for the vessels, sizes, modes and instruments sampled, and for nothing outside them. Chapter 9 showed what can happen outside them. In an independent bench comparison that Okhunov and colleagues published in 2018, four of five devices lost about half their seal strength or more on vessels larger than 7 mm. The label must therefore carry the limit the test carried.
The loop from the field
Proof made before launch must stay true afterward. A complaint is evidence of a possible failure to meet a specification, and FDA notes that ISO 13485 addresses the investigation of complaints and the corrective action that follows (Chapter 26). An investigation that finds a cause leads to a correction, a design change or both, and every change reopens the trace. The team finds the requirements the change touches, repeats the tests behind them, updates the risk file and the label, and assesses whether the change needs a new submission, by the US and EU criteria of Chapters 26 and 27.
The Varipulse pulsed field catheter shows the loop running. In February 2025 Johnson & Johnson's unit Biosense Webster sent a correction that updated the instructions for use, after 4 of 132 patients in the US evaluation had a stroke or a transient ischemic attack, a brief stroke-like episode (Chapters 15, 24 and 26). In ISO 14971's terms, the update was information for safety, the last of the three options. The catheter had reached the market by premarket approval, and FDA's records for that approval then show two later changes (Chapter 26). A label change to the recommended irrigation flow rate was approved on 26 June 2025. A software update to the TRUPULSE generator was approved on 23 December 2025 "to facilitate automatic delivery of irrigation at 30 mL/min upon initiation of energy delivery", which moved a control from the instructions into the design. FDA's records do not link the irrigation changes to the strokes, although Johnson & Johnson reported a neurovascular event rate from its complaint data in its release on the flow rate (Chapter 24).
A correction can come before its cause is known. When Megadyne corrected its reusable return electrodes in 2023 after reports of burns, it asked hospitals to confirm that staff follow the instructions for use and to post its visual aids, while it continued to evaluate the root cause (Chapter 26). Both corrections began as information for safety, and the trace then shows which requirement a later fix must change and which tests it reopens.
Keep one matrix per claim, one row per requirement, with an identifier in every cell, so that any test report, risk control or label line leads back to its claim. Fill the columns below for the 7 mm claim, then add rows for thermal spread, the chronic study and each interface. When a change touches a cell, rerun every test that depends on it before the change is released.
| Column | The 7 mm sealing claim, burst-pressure row |
|---|---|
| Claim | Seals vessels up to and including 7 mm |
| User need | Vessels up to 7 mm, sealed and divided, stay closed under arterial pressure |
| Design input | Seals up to 7 mm, measured rounded and perfused, hold at least 360 mmHg in each mode with each representative instrument |
| Design output | Jaw pressure, sealing mode, end-point algorithm, instructions for use |
| Verification | Zero-failure burst test, 29 seals per mode and instrument, beside the predicate |
| Validation | Chronic animal study, at least 3 weeks in at least 5 animals; human factors validation |
| Risk control | End-point detection in the generator; warning on vessel size |
| Label line | DO NOT USE this instrument on vessels larger than 7 mm in diameter |
- Design controls turn a claim into design inputs a test can fail, divided among the parts; verification shows that outputs meet inputs, and validation that the device meets its users' needs.
- FDA's sealer guidance sets no pass value or sample size, so the team chooses them: a zero-failure burst test at 95 % confidence and 90 % reliability needs 29 seals.
- ISO 14971 prefers inherently safe design, then protective measures, then information for safety, and on the lifting pad these are the pad's area, the monitor and the instructions, each with its own test.
- A trace from claim to test to label shows which tests a change must repeat, and the field feeds changes back, as Varipulse's instruction update was followed by new generator software.
Let's Talk