06 — Six ways to deliver energy

Six ways to deliver energy.

+ The questionIf most energies end as heat in tissue, why does it matter what carries it there?

One frequency axis, two kinds of wave

Surgical energy reaches tissue at frequencies from 23 kHz to about 564 THz, a span of more than ten tenfold steps. A carrier is the current, vibration, field or light that brings energy to the tissue, and the previous chapters took the carriers one at a time. Placed on one frequency axis, the carriers fall in this order: ultrasonic instruments vibrate at 23 to 55.5 kHz, most RF generators run at 300 to 500 kHz, focused ultrasound works at 0.7 to 4 MHz and microwave ablation uses 915 MHz or 2.45 GHz. Surgical lasers emit infrared and visible light from about 28 THz, for the 10.6 µm of CO₂, to about 564 THz, for the 532 nm of KTP. Pulsed-field systems deliver bursts of microsecond pulses of current, which have no single carrier frequency but occupy the low end of the same range. That makes six carriers: the vibration of ultrasonic instruments, RF current, focused sound, microwaves, light and pulsed fields. They are the six ways of delivering energy that this chapter compares.

Solved example: the span of the carriers in tenfold steps

The lowest carrier is the vibration of an ultrasonic instrument at 23 kHz, and multiplying it by ten, step by step, places each of the other carriers on the axis. One tenfold step gives 230 kHz, just below the RF generators at 300 to 500 kHz. Two steps give 2.3 MHz, inside the band of focused ultrasound, and five steps give 2.3 GHz, beside the 2.45 GHz of microwave ablation. Nine steps give 23 THz, a little below the 28 THz of the CO₂ laser, and ten give 230 THz. The KTP laser's 564 THz lies between the tenth step and the eleventh, so the carriers span more than ten tenfold steps, a factor of about 25 billion between the lowest frequency and the highest.

The carriers on the axis are two kinds of wave. Acoustic carriers are mechanical, because in them matter moves back and forth. Electromagnetic carriers are currents, fields and light, in which charges and fields move with no bulk motion of the tissue. Ultrasonic shears and RF generators sit within about a factor of ten of each other in frequency, yet they share nothing in how they heat. Cryoablation has no place on the axis at all, because it carries no energy into the tissue and takes energy out instead.

17 — The carriers on one frequency axis
10 kHz 1 MHz 100 MHz 10 GHz 1 THz 100 THz frequency (log) ACOUSTIC: MECHANICAL WAVES ultrasonic shears and aspirators 23 kHz and 36 kHz aspirators 47 kHz Olympus 55.5 kHz Harmonic focused ultrasound: HIFU and histotripsy RF electrosurgery and RF ablation IEC range 915 MHz and 2.45 GHz: microwave ablation pulsed fields: microsecond pulses, not a carrier surgical lasers: CO₂ 10.6 µm to KTP 532 nm ELECTROMAGNETIC: CURRENTS, FIELDS AND LIGHT cryoablation: no carrier; it removes heat close in frequency, unrelated in physics
ElectricalAcousticOpticalColdFocal detail
Plate 17 — The carriers spread across more than ten tenfold steps of frequency in two kinds of wave. Acoustic carriers are mechanical vibrations; electromagnetic carriers are currents, fields and light. Ultrasonic shears and RF generators sit within about a factor of ten of each other, yet share nothing in how they heat. Cryoablation has no carrier at all.

Seven mechanisms

What matters to tissue is the mechanism by which a carrier deposits its energy, not the carrier's frequency. Every energy in this guide uses one of seven.

  1. Joule heating. Current flows through tissue and releases heat in each unit of its volume at a rate equal to the square of the current density divided by the conductivity, so the heat is greatest where the current is densest. Monopolar and bipolar electrosurgery, vessel sealing and RF ablation all work this way (Chapter 3).
  2. Friction and cavitation. A vibrating metal surface heats tissue by friction where the two touch and churns it by cavitation, which is how ultrasonic shears and aspirators work (Chapter 4).
  3. Photothermal absorption. A chromophore absorbs a laser's light within about one absorption depth of the surface, and the wavelength sets that depth (Chapter 5).
  4. Dielectric heating. In microwave ablation, an alternating electromagnetic field rotates water molecules and drives ions back and forth, and both motions dissipate energy as heat throughout a volume, with no conducting path to an electrode (Chapter 13).
  5. Focused acoustic absorption. In thermal focused ultrasound, sound converges on a small focal volume deep in the body and is absorbed there as heat (Chapter 16).
  6. Heat extraction. In cryoablation, a probe colder than tissue draws heat out, and ice grows outward from it (Chapter 14).
  7. Non-thermal injury. In irreversible electroporation and pulsed field ablation, strong electric pulses open pores in cell membranes, and in histotripsy, intense acoustic pulses tear cells apart by cavitation. The cells die without being heated much (Chapters 15 and 16).
18 — Seven ways energy acts on tissue
MECHANISM WHERE THE EFFECT APPEARS ENERGIES WHAT IT MEANS FOR THE DEVICE Joule heating in the tissue, where current density is highest monopolar, bipolar, sealing, RF ablation current needs a complete circuit through the patient Friction and cavitation at the blade–tissue interface ultrasonic shears and aspirators no current in the patient; the blade stays hot Photothermal absorption within about one absorption depth of the surface lasers the wavelength sets the depth; eye hazard Dielectric heating in a volume around the antenna microwave ablation no return pad; heats through dried tissue Focused absorption in a focal volume deep in tissue HIFU skin stays closed; needs an acoustic window Heat extraction at the probe, ice growing outward cryoablation no hemostasis; the ice shows on imaging Non-thermal membranes or cell structure IRE, PFA, histotripsy largely spares collagen scaffolds of vessels and ducts
ElectricalAcousticOpticalColdFocal detail
Plate 18 — Every energy in this guide acts through one of seven mechanisms. Five deposit heat, each in a different place; cryoablation removes it; the last group kills cells without relying on heat, which in preclinical studies largely spares the collagen of vessels and ducts.

The first five mechanisms deposit heat, but in five different places: in the tissue where current crowds, at the face of a blade, in a layer at the surface, in a volume around an antenna and at a point deep inside the body. Those places answer the part of this guide's central question that asks why surgery needs eleven kinds of energy. Each surgical task needs heat in a particular geometry: a line for an incision, a band across a clamped vessel, a surface layer for a precise ablation, a sphere several centimeters across for a tumor, a point behind intact skin. No single carrier can produce all of those geometries.

The seventh mechanism completes that answer. Heat that kills cells also destroys the collagen framework of vessels, bile ducts and nerves, and for a tumor wrapped around a major vessel, or for cardiac tissue next to the esophagus, destroying that framework is exactly the wrong effect. The non-thermal energies exist because the task is sometimes to kill cells while leaving their scaffold intact.

What the carrier decides for the device

The carrier also decides almost everything about the device that delivers it, so an engineer or founder who chooses an energy chooses a whole set of consequences at once.

Whether current flows through the patient. RF energies need a complete circuit through the patient, so they need either a return pad or a second electrode, and they bring every stray-current hazard of Part VII. Ultrasonic, laser and microwave energies need no circuit through the patient, which removes those hazards and brings others.

What the instrument must be. RF needs electrodes and insulation rated for kilovolts, and pulsed fields need electrodes that can withstand thousands of volts. Ultrasonic energy needs a titanium waveguide cut to exact half-wavelengths of its vibration. Lasers need optical fibers or, for CO₂, an articulated arm of mirrors. Microwaves need a coaxial cable and a cooled antenna, and cryoablation needs high-pressure gas lines or liquid nitrogen.

What the power electronics must do. An RF generator is an inverter with an isolation transformer, and an ultrasonic generator is a resonant driver that tracks a mechanical resonance. A laser needs a pump source and cooling, and a microwave generator is a gigahertz source. A pulsed-field generator stores energy and releases it as kilovolt pulses, while a cryoablation console controls gas flow rather than electrical power.

Which standards apply. Electrosurgical equipment is tested to IEC 60601-2-2, lasers to IEC 60601-2-22 and IEC 60825-1, and high-intensity therapeutic ultrasound to IEC 60601-2-62. Ultrasonic surgical systems have a standard for their output, IEC 61847, but no particular safety standard of their own was found for this guide. Chapter 25 takes up the standards in full.

The eleven energies at a glance

EnergyCarrierMechanismTypical taskChapter
Monopolar RFcurrent, 300–500 kHzJoule heating at a small electrodecut, coagulate, dissect planes8
Conventional bipolar RFcurrent between two tipsJoule heating between the tipscoagulate small vessels9
Advanced bipolar sealingcurrent through a clamped vesselJoule heating under pressureseal vessels up to 7 mm9
Ultrasonicvibration, 23–55.5 kHzfriction and cavitation at the tipseal and divide; fragment11
Laserlight, 532 nm to 10.6 µmphotothermal absorptioncut, vaporize, coagulate, break stones12
RF ablationcurrent, about 500 kHzJoule heating around a needledestroy tumors in place13
Microwave ablationfield, 915 MHz or 2.45 GHzdielectric heatingdestroy tumors in place13
Cryoablationnone; coldheat extractiondestroy tumors in place14
IRE and PFAkilovolt pulsesmembrane pores, non-thermaldestroy tissue, spare scaffold15
Thermal focused ultrasoundfocused sound, 0.7–4 MHzfocal absorptiondestroy tissue without incision16
Histotripsyfocused sound pulsesinertial cavitation, non-thermaldestroy tissue without incision16
Insight

Two devices can reach the same final temperature in the same tissue and still behave completely differently, because they get there through different geometry, speed and side effects. The carrier decides the geometry and the side effects, and the generator's control decides the speed. Each instrument chapter that follows explains its devices through those three properties.

In practice

Before comparing products, place each one on this table by its carrier, its mechanism and where it puts the heat. Two devices that put their effect in different places suit different steps of an operation, so many apparent disagreements about which device is "better" are settled once each one is placed. Chapter 17 does this for each step of one colectomy, an operation to remove part of the colon.

+ What this chapter established
  • The carriers span more than ten tenfold steps of frequency in two kinds of wave, acoustic and electromagnetic, and cryoablation has no carrier at all.
  • Every energy acts through one of seven mechanisms: five deposit heat, each in a different place, one removes heat, and one group kills cells without heating them much.
  • Surgery needs several energies because each task needs heat in its own geometry, and some energies are non-thermal because some tasks must spare the collagen scaffold of vessels, ducts and nerves.
  • The carrier fixes the instrument, the power electronics, the hazards and the standards a device must meet.