+ Part II · Access
The route to the tissue.
A 5 mm laparoscopic instrument must carry a conductor, its insulation, a jaw mechanism and often a cutting blade, and most of its shaft is out of the camera's view. Laparoscopic and robotic surgery add no new energy, but they change the conditions under which every energy works: what the surgeon can see and feel, how long and thin the instruments must be, where smoke and heat go, and who or what stands between the surgeon's foot and the generator. This part sets out those conditions before the instrument chapters meet them.
Open, laparoscopic, robotic.
+ The questionWhat changes for an energy device when the surgeon can no longer see or touch the tissue it acts on?
Open surgery: the reference case
An energy instrument in open surgery sits in the surgeon's hand a few centimeters from the tissue, and the whole of it is in view. The surgeon can feel the tissue, smoke drifts away into the room, and a burn in the wrong place is usually seen when it happens. Energy devices were designed for these conditions first. The apparatus that made electrosurgery routine was Bovie's generator, and Harvey Cushing first used it in open surgery, in the operation of the opening story, in the autumn of 1926.
Almost everything in open surgery gives the surgeon feedback that needs no thought. The color of the tissue shows that it has coagulated, the resistance it offers the instrument shows that it is being cut, and its smell and sound show when it is drying and charring. Minimally invasive surgery works through small openings in the body wall, and it removes most of that feedback, so energy devices have had to replace it.
Laparoscopy: working through a keyhole
In laparoscopic surgery the abdomen is inflated with carbon dioxide at 12 to 15 mmHg to lift the wall away from the organs. The instruments enter through cannulas, tubes that pass through the wall, and a camera on a rigid telescope shows the field on a monitor. Each of those facts changes how energy behaves. The view is partial, because the camera shows the working ends of the instruments and a few centimeters of their shafts, while the rest of each shaft lies inside the abdomen, out of view and within reach of the bowel.
The instruments are also long and thin, because they must reach the tissue through the cannulas. Cannulas come in a range of sizes, and one maker, Johnson & Johnson, sells them from 5 to 15 mm in diameter. Even at the small end of that range, a 5 mm instrument must carry a conductor, its insulation, a jaw mechanism and often a cutting blade. Its insulation is thin and rubs against the cannulas, and if the instrument is reusable, the insulation is cleaned and sterilized, or reprocessed, again and again. A long insulated conductor inside a metal cannula also forms a capacitor, which means that alternating current can cross the insulation to the metal without any contact, a transfer called capacitive coupling. Where the shaft is out of the camera's view, energy that leaves it through damaged insulation or by capacitive coupling goes unseen, and Chapter 18 is about exactly that.
The space is small and closed, so smoke from a few seconds of cutting fills the gas volume and obscures the camera, and the smoke has to be evacuated while the pressure is maintained. Heat has nowhere to go but into the tissue, because there is no room air to cool a hot jaw between activations, and a jaw that has just sealed a vessel can touch the bowel on its way to the next bite.
Motion is reversed as well: each instrument pivots at the abdominal wall, so its tip moves opposite to the hand. The surgeon also feels force poorly through a long shaft and the valve that keeps the gas in the cannula, so judging by feel alone how hard a jaw is squeezing, or when a seal is complete, becomes unreliable.
Robotic surgery: a controller in the loop
A surgical robot restores some of what laparoscopy removed: the surgeon sits at a console with a three-dimensional view, the instruments have wrists that bend inside the body, and the system scales the surgeon's motion and filters out tremor. The instruments still enter through ports in an abdomen inflated with gas, so every condition above still applies.
The robot also changes how the generator is switched on. In open and laparoscopic surgery, the surgeon's pedal or hand switch activates the generator directly, while in robotic surgery the command passes through the robot's own controller, which decides whether it reaches the generator. Some makers pair the generator with the robot. Intuitive, for example, describes its E-200 generator as integrated with its da Vinci systems, with the energy settings controlled from the surgeon's console, and its vessel sealers and its SynchroSeal instrument work with that generator. Energy delivery then depends on two control systems working together, so the risk analysis must cover both (Chapter 21).
A robot moves the instruments, but it does not decide by itself how much energy to deliver. On robotic systems the surgeon still commands the energy and a generator still controls it, and the same physics, hazards and tests apply as in laparoscopy. What the robot adds is another controller and another set of interfaces that must be verified.
Robotic instruments are reused a set number of times and have complex wristed shafts, and they bring an insulation problem of their own. In a prospective study published in 2011, Espada and colleagues at Mayo Clinic Arizona tested robotic and laparoscopic instruments for insulation failure. At a single test, they found failures in 32 % of the robotic instruments against 13 % of the laparoscopic ones. Among the instruments followed over time, 80 % of the robotic ones failed within ten uses, against 36 % of the laparoscopic ones. Chapter 18 returns to these numbers.
At the single test, the study found failed insulation on 25 of 78 robotic instruments and on 39 of 298 laparoscopic ones. Dividing each count by its total gives 32 % and 13 %. The figures over time rest on other totals. Of the 44 robotic instruments followed, 35 failed within ten uses, which is 80 %, and of the 189 laparoscopic instruments followed, 68 failed, which is 36 %. The first pair is the share found failed at one test, and the second is the share that failed while the instruments were followed, so the two pairs cannot be read as one rate rising with use. Within each pair, robotic instruments failed about 2.4 times as often as laparoscopic ones at one test, and about 2.2 times as often over time.
What the access route asks of energy devices
Taken together, minimally invasive access asks four things of an energy device that open surgery did not. The first is to do more than one job per instrument, because every exchange of instruments takes time and risks losing the view, so instruments that grasp, seal and cut in one device became central to laparoscopy. Advanced bipolar sealers with built-in blades, ultrasonic shears and hybrid instruments all answer this need. A hybrid instrument combines two energies in one jaw, such as bipolar current and ultrasonic vibration. Chapter 9 describes the sealers, and Chapter 11 the ultrasonic and hybrid instruments.
The second is to decide the end point, because when the surgeon cannot see the tissue blanch or feel it change, the generator must decide when a seal is finished. The answer is closed-loop control, in which the generator measures the tissue and stops by itself, and Chapters 9 and 10 describe how it works.
The third is to keep the energy inside the instrument: insulation, shielding and monitoring must stop current from leaving the shaft outside the camera's view. The fourth is to manage what the energy leaves behind. Smoke must be evacuated without losing the pressure that holds the abdomen open, and the heat retained in jaws and blades must be managed. For that heat, at least one current hybrid instrument now includes a thermal shield that slows the transfer of heat to the outer surface of its jaw.
Write the access route into the design inputs of a device for minimally invasive use from the start. List the shaft length and diameter, the insulation material and thickness, compatibility with metal cannulas and with the hybrid cannulas of metal and plastic that Chapter 18 describes, and smoke management. For robotic use, add the interface to the robot's controller and its activation path. Plan the capacitive-coupling test early, because FDA's electrosurgical guidance asks for one against a conductive cannula for monopolar instruments used in minimally invasive surgery.
- Minimally invasive access adds no energy but changes the conditions: a partial view, long thin instruments, a small closed space and reversed motion.
- Most of each instrument shaft is out of the camera's view and within reach of bowel, so energy that leaves it through damaged insulation or by capacitive coupling goes unseen.
- Robotic surgery restores dexterity and depth perception but places the robot's controller between the surgeon's pedal and the generator.
- The access route pushes energy devices toward multifunction instruments, closed-loop end points, containment of stray energy and management of smoke and residual heat.