Sound in tissue.
+ The questionCurrent heats tissue because it pushes ions through it. How can a vibration heat tissue, cut it or tear it apart?
Sound is a traveling pressure wave
The tip of an ultrasonic instrument is a metal blade that vibrates along its length tens of thousands of times a second, and the tissue pressed against it heats and can part. Where the current of Chapter 3 moves ions through tissue, a vibration moves the tissue itself, because sound is a mechanical wave. A vibrating surface pushes on the material next to it and compresses it, and the compressed region pushes on the next in turn. The disturbance travels outward as a wave of alternating compression and rarefaction, regions where the pressure rises above and falls below its resting value, while each bit of material only moves back and forth about its place.
In soft tissue the wave travels at about 1,540 meters per second, slightly faster than in water at 1,481 and far faster than in air at 343, both at room temperature. The wavelength is the speed divided by the frequency. At 1 MHz, a typical therapeutic frequency, the wavelength in tissue is 1.54 mm, and at 4 MHz it is 0.39 mm. These lengths matter, because both the finest detail a beam of sound can resolve and the smallest spot it can be focused to are of the order of one wavelength.
Surgery uses sound in two quite different regimes. Ultrasonic instruments vibrate a metal tip at 23 to 55.5 kHz, and they act on the tissue that touches the tip: the tip rubs, shears and churns that tissue at its face. Focused ultrasound works at 0.7 to 4 MHz and sends a wave through intact skin and tissue to act at a point centimeters deep. The first regime depends on friction and cavitation at a surface, while the second depends on how waves travel, reflect and are absorbed. This chapter builds both.
What happens at a boundary
Each material has an acoustic impedance, its density multiplied by its speed of sound, and sound crosses a boundary well only when the impedances on the two sides match. Acoustic impedance is measured in rayls, kilograms per square meter per second, so water, at about 1,000 kilograms per cubic meter and 1,481 m/s, has about 1.48 million rayls, or megarayls. Soft tissue has about 1.63 megarayls, a close match to water. Cortical bone is about 7, more than four times the value for tissue, though reported values vary widely. Air is about 0.0004, four thousand times lower than tissue. The fraction of the sound's intensity reflected at a boundary is the square of a ratio: the difference between the two impedances divided by their sum. Between tissue and water the fraction is 0.2 %, and between tissue and bone it is about 39 %. Between tissue and air it is 99.9 %, so essentially all the sound bounces back.
Three practical consequences follow. A transducer turns an electrical drive into sound, and it must be coupled to the skin with gel or water, because even a thin layer of air would reflect almost all of its energy. A beam aimed through gas in the bowel or lung is blocked, and a beam aimed through bone is partly reflected and partly absorbed, so it heats the bone. Focused ultrasound therefore needs an acoustic window, a path to its target free of gas and bone. Inside an ultrasonic instrument, the vibration stays in the metal waveguide that carries it to the tip, because the gas around the waveguide is so poorly matched to the metal that almost nothing leaks out.
Absorption turns sound into heat
As sound travels through tissue, it loses some of its energy in every cycle to internal friction between moving molecules and structures, and the lost energy appears as heat, while a little more of it is scattered out of the beam. Together, absorption and scattering attenuate sound in soft tissue by about 0.54 decibels per centimeter for every megahertz of frequency, and absorption accounts for most of the loss. A loss of 3 dB halves the intensity, so at 1 MHz the intensity halves in about 5.6 cm of tissue, and at 4 MHz in about 1.4 cm.
The attenuation and the wavelength together set the central trade of focused ultrasound. A lower frequency reaches deeper, but its longer wavelength focuses to a larger spot, while a higher frequency focuses more finely but loses its energy sooner. A system that treats the prostate from inside the rectum works a few centimeters from its target and can use 4 MHz, while the histotripsy system that treats liver tumors through the abdominal wall uses 700 kHz. Chapter 16 describes both systems.
A target lies 5.6 cm deep, the depth at which a 1 MHz beam has lost half its intensity. Tissue attenuates sound by about 0.54 dB per centimeter for every megahertz, so at 4 MHz the loss is four times as fast, about 2.2 dB per centimeter, and the intensity halves every 1.4 cm. Over the 5.6 cm to the target, the 4 MHz beam therefore halves four times, to a half, a quarter, an eighth and finally a sixteenth of its starting intensity. That leaves about 6 % of the 4 MHz beam at the target, against about 50 % of the 1 MHz beam. The 4 MHz wavelength in tissue is 0.39 mm against 1.54 mm, a quarter as long. At this depth the 4 MHz beam can therefore focus to a spot about a quarter as wide. Since 6 % is an eighth of 50 %, an eighth as much sound arrives there as at 1 MHz, but a spot a quarter as wide has a sixteenth of the area, so the intensity at the focus is about twice as high. At 5.6 cm the finer focus therefore more than makes up for the greater loss.
Friction at a vibrating tip
Each vibration of an ultrasonic blade has an excursion of 50 to 100 micrometers, and a movement that small is still fast, because it repeats tens of thousands of times a second. If 100 µm is the full stroke at 55.5 kHz, the tip reaches a peak speed of about 17 meters per second and a peak acceleration of about six million meters per second squared, more than 600,000 times the acceleration of gravity.
The blade shears and rubs the tissue pressed against it at high speed, so friction at the interface and internal losses in the deformed tissue turn the mechanical energy into heat, concentrated where the tissue touches the blade. As in Chapter 1, the proteins unfold and form a sticky coagulum that can seal small vessels, and with enough pressure and energy the tissue parts along the blade. No current flows through the patient, because the transducer is driven electrically inside the handpiece and only vibration leaves it. Chapter 11 follows the whole instrument.
Cavitation
The rarefaction half of each acoustic cycle pulls on the liquid in tissue, and if the pull is strong enough, small gas bubbles can grow in it. The growth and motion of such bubbles in a sound field is cavitation, and it takes two forms. In stable cavitation, a bubble oscillates gently in size in step with the wave, swelling a little in each rarefaction and shrinking in each compression. In inertial cavitation, a bubble grows to many times its size, within one cycle or over several, and then collapses violently. The collapse concentrates the bubble's energy into a tiny volume and launches a shock wave, and it is the collapse that damages tissue.
Two surgical tools use cavitation on purpose. An ultrasonic aspirator vibrates a hollow tip at 23 or 36 kHz and creates cavitation just ahead of the tip's rim. Together with direct mechanical cutting, the cavitation fragments tissue with a high water content and weak cohesion, while it spares collagen-rich structures such as vessels, nerves and ducts. Vieira-Coimbra and Ferreira describe this selectivity in a prospective case series of 2025 on the laparoscopic excision of endometriosis with the CUSA aspirator. Integra LifeSciences, which makes the CUSA, supported the publication.
The second tool, histotripsy, generates inertial cavitation deep inside the body with focused pulses (Chapter 16). Whether a pulse starts cavitation depends on whether it pulls hard enough on the liquid, and Vlaisavljevich and colleagues measured that threshold in a 2015 laboratory study. For single-cycle pulses, cavitation starts even without a seed bubble once the peak negative pressure exceeds about 25 to 30 megapascals. The study found this threshold nearly the same in excised tissues of different stiffness and in tissue-like gels, at frequencies from 345 kHz to 3 MHz, so one narrow band of pressure describes the onset across all of them. Three of its authors declared financial interests or other relationships with HistoSonics, which makes histotripsy systems.
Cavitation damages tissue mechanically, with little heating, so a device that uses it can be mistaken for a gentle one. The energy still reaches the tissue, but as pressure rather than heat. Whether a structure survives depends on its mechanical strength, which is why collagen-rich vessels survive and an organ's parenchyma does not. Parenchyma is the soft, water-rich working tissue of an organ.
Sound can act on tissue in three ways, and different devices use each one. When tissue absorbs sound along its path, the sound heats it, which is how thermal focused ultrasound works. When a vibrating surface rubs against tissue, friction heats the tissue at that surface, which is how the ultrasonic shear works. When the negative half of each cycle pulls hard enough on the liquid in tissue, bubbles grow and tear the tissue mechanically, which is how the aspirator and histotripsy work.
Trace the path from the transducer to the target before judging a focused-ultrasound design or its test. List every boundary on that path, because a thin layer of air reflects almost all of the sound, and bone reflects about 39 % of it and absorbs part of the rest, which heats the bone. Work out the loss over the target's depth at the chosen frequency, at about 0.54 dB per centimeter for every megahertz, and set it against the wavelength, which limits how small the focus can be. For a device that relies on cavitation, specify the peak negative pressure it reaches as well as the temperature, because cavitation damages tissue mechanically with little heating.
- Sound travels through soft tissue at about 1,540 m/s, so its wavelength at 1 MHz is about 1.5 mm, and the wavelength limits how finely a beam can focus.
- A boundary reflects more sound the larger the mismatch in acoustic impedance across it: 0.2 % between tissue and water, about 39 % at bone and 99.9 % at air.
- Tissue attenuates sound by about 0.54 dB per centimeter for every megahertz, so a higher frequency focuses more finely but reaches less deep.
- A vibrating tip heats tissue by friction at its surface, while cavitation tears tissue mechanically and, with single-cycle pulses, starts without a seed bubble above a peak negative pressure of about 25 to 30 MPa.
Let's Talk