09 — Bipolar and vessel sealing

Bipolar and vessel sealing.

+ The questionA sealer cannot see inside its own jaws. How does it know when a seal is complete?

Bipolar: the circuit in the instrument

Bipolar forceps held on a vessel for too long dry the tissue until it chars and sticks to their tips, and when the forceps open, the vessel may tear. Held too briefly, they leave the vessel open. Both failures have one cause: with conventional bipolar forceps the surgeon decides how long to apply energy. A bipolar instrument carries both electrodes in its tips, usually the two blades of a forceps. Current flows from one tip to the other through only the tissue held between them, so no return pad is needed and almost no current wanders through the rest of the patient.

A generator offers bipolar modes alongside the monopolar modes of Chapter 8, and the bipolar modes work at low voltages. Medtronic's specification for its Valleylab FT10 generator lists the bipolar modes at 70 W into a rated load of 100 Ω (Chapter 10). Their maximum peak voltages run from 184 to 530 V depending on the mode, a fraction of the kilovolt peaks of monopolar coagulation in Chapter 8. Conventional bipolar forceps coagulate small vessels and bleeding points precisely. For vessels of 3 mm or more, however, conventional bipolar coagulation was not reliable enough to replace a tie, the suture knotted around a vessel to close it.

Advanced bipolar: pressure, current and a closed loop

The advanced bipolar vessel sealer changed that. Valleylab's LigaSure is one such sealer, and FDA cleared it on 28 August 1998. As the introduction explained, a device is cleared when FDA finds it substantially equivalent to one already legally on the US market, its predicate device. The maker makes that case in a premarket notification called a 510(k) (Chapter 26).

A later 510(k) summary for LigaSure describes the mechanism in regulatory language. Collagen and elastin in the tissue are re-formed by heat and pressure to fuse the vessel walls into a permanent seal, while a microprocessor monitors the tissue, ends energy delivery and allows the seal to cool before signaling that the seal is complete. LigaSure is indicated for vessels up to and including 7 mm in diameter, which means that sealing such vessels is the use its maker claims for it (Chapter 2). Many advanced bipolar sealers share the 7 mm limit, and some robotic sealers are limited to 5 mm.

Three things distinguish a sealer from bipolar forceps, and the first is that the jaws apply high, controlled pressure, squeezing the vessel flat so that its two walls touch along their whole width and blood is pushed out of the jaw. The generator delivers high current at low voltage: Medtronic rates the FT10's sealing mode at 20 Ω, with up to 350 W, a maximum current of 5.5 A and a peak voltage of at most 244 V. Into wet, clamped tissue the voltage stays far lower, so the current heats the tissue without arcing. And the generator runs a closed loop, measuring the tissue continuously and deciding for itself when to stop.

25 — Inside a sealing jaw
high pressure pushes the walls together current flows only between the jaws: no pad two walls become one band collagen and elastin fuse thermal spread: about 1.3–2.1 mm in one study by the manufacturer's staff
ElectricalHeatFocal detail
Plate 25 — An advanced bipolar sealer squeezes a vessel flat, passes current only through the tissue between its jaws, and heats it until the collagen and elastin of the two walls fuse into one band. Heat escaping sideways is the thermal spread, about 1.3–2.1 mm in one study by the manufacturer's staff.

The seal itself is the collagen fusion of Chapter 1. Heat unwinds the helices of the collagen in the two opposed walls, and as the tissue cools under pressure, their collagen and elastin entangle into a single translucent band. Heat that escapes sideways from the jaw is the lateral thermal spread of Chapter 2. In a 2025 study funded by Medtronic, Chambers and two other Medtronic employees report a spread in living tissue of about 1.3 to 2.1 mm for Medtronic's current LigaSure sealers, with seal times of about 2 to 4 seconds.

Impedance tells the generator what is happening

Inside the jaw the generator has no camera and no thermometer in the tissue, so all it has is the voltage and current at its own output, and from those it computes the tissue's electrical impedance, many thousands of times a second. Medtronic states that the FT10's tissue-sensing technology monitors changes in tissue impedance 434,000 times a second.

The impedance follows the state of the tissue closely. As the tissue warms, its conductivity rises, so the impedance first falls. Below 42 °C, where it has been measured, the rise is about 2 % per degree, although a seal takes the tissue far above that temperature (Chapter 2). As the tissue approaches 100 °C, its water begins to boil off, the tissue that remains conducts worse and worse, and the impedance turns upward and then climbs steeply. When the climb shows that the water has largely gone, heating can do no more for the seal, and further energy would only char the tissue, so the generator stops, waits while the seal cools in the closed jaws, and sounds a tone.

26 — Impedance and temperature through a seal
HEAT DRY STOP COOL UNDER PRESSURE 40 60 80 100 °C low high Ω 0 s 1 s 2 s 3 s 4 s TIME tissue temperature end-point threshold energy off warming raises conductivity about 2 % per °C (measured below 42 °C), so impedance first falls drying makes it climb stop here: the water has gone schematic: shapes from published descriptions, not a device record
HeatReadoutFocal detail
Plate 26 — Tissue conducts better as it warms and worse as it dries. A sealer watches the impedance fall and then climb, stops the energy when the climb shows the water has gone, and holds pressure while the seal cools. It never measures the seal itself; it infers it from impedance.

Different devices apply the same idea in different ways. Some regulate on impedance directly. Ethicon's EnSeal uses a jaw material whose conductivity falls sharply as it warms, so each part of the jaw limits its own temperature, and Olympus's PK system pulses its bipolar energy. All of them manage the same physics, because each must deliver enough energy to fuse the collagen and stop before the tissue dries into char.

Insight

A sealer never measures the seal. It measures impedance and infers the seal from the shape of the impedance curve, using an algorithm tuned on bench and animal tests. That is why the strength of a seal must be demonstrated by testing rather than by reading the generator's display. It is also why a seal may not be as strong as the tone suggests when it is made outside the conditions the algorithm was tuned for, on a larger vessel or in unusual tissue.

How strong is a seal?

The bench test for a seal measures its burst pressure. A sealed vessel is filled with liquid and the pressure is raised until the seal fails, and the pressure at that moment is the burst pressure. The pressure in the arteries peaks with each heartbeat, and this systolic pressure is normally about 120 mmHg. A widely used acceptance level for a seal is 360 mmHg, three times systolic pressure. LigaSure's maker uses 360 mmHg as its lower specification limit, and published studies often compare against that level. The level is an industry convention and not a regulatory requirement, because FDA's guidance for bipolar vessel sealers sets out how to test burst pressure but gives no pass value.

Measured burst pressures are usually far above 360 mmHg when a device is used within its indication. Okhunov and colleagues published an independent bench comparison of five devices in 2018. The authors were at the University of California, Irvine, and Duke University, and they declared no competing financial interests. Three of the devices were bipolar sealers: LigaSure, EnSeal and Caiman. The other two were ultrasonic devices, the Harmonic ACE+ and the Harmonic ACE+7. On arteries of 5.1 to 7 mm, the mean burst pressures of the five ranged from about 570 to 1,740 mmHg, the lowest from the Harmonic ACE+, which is indicated for vessels only up to 5 mm. A mean can hide weak seals, though. Even on these vessels, 4 of 10 EnSeal seals and 2 of 10 Harmonic ACE+ seals burst below 300 mmHg, the study's threshold for failure.

The same study also tested all five devices on arteries of 7.1 to 9 mm. Four of them lost about 46 to 60 % of their mean burst pressure there. LigaSure fell from 981 to 530 mmHg beyond its 7 mm indication, EnSeal and the Harmonic ACE+7 both fell from 1,165 to 467 mmHg, and the Harmonic ACE+ fell to 254 mmHg, below the 360 mmHg convention, on vessels beyond its 5 mm indication. Caiman held about 1,680 mmHg, so one device kept its strength while four lost about half of it or more. Beyond the indicated vessel size, then, a seal's strength varies from device to device, and the maker's evidence no longer covers it.

27 — Seal strength against vessel size
0 500 1,000 1,500 2,000 mean burst pressure (mmHg) arteries 5.1–7 mm arteries 7.1–9 mm 1,740 1,676 Caiman 571 254 Harmonic ACE+ 1,165 467 Harmonic ACE+7 1,165 467 EnSeal 981 530 LigaSure indicated to 5 mm 3 × systolic: industry convention, not an FDA requirement normal systolic pressure LigaSure, 981 to 530 mmHg: about 46 % weaker; still above 360 independent bench comparison (Okhunov 2018, Table 3); Caiman 1,740: proximal jaw, distal jaw 1,600
HeatFocal detail
Plate 27 — Burst pressure is the bench test for a seal. In one independent comparison, five devices sealed 5.1–7 mm arteries at mean burst pressures of about 570 to 1,740 mmHg. On 7.1–9 mm arteries four of them lost about 46 to 60 % of their mean burst pressure, and the fifth, also a bipolar sealer, held its strength. The 360 mmHg line is an industry convention; FDA's guidance sets the test method but no pass value.
Solved example: what a seal loses beyond its indication

A normal systolic pressure of about 120 mmHg, multiplied by three, gives the acceptance level of 360 mmHg. On arteries of 5.1 to 7 mm, within its 7 mm indication, LigaSure's seals in Okhunov and colleagues' 2018 comparison had a mean burst pressure of 981 mmHg. That is about eight times systolic pressure and about 2.7 times the acceptance level. On arteries of 7.1 to 9 mm the mean fell to 530 mmHg, a fall of 451 mmHg. Dividing 451 by 981 gives about 0.46, so the seals lost about 46 % of their strength. A mean of 530 mmHg is still about 4.4 times systolic pressure and still lies above the 360 mmHg convention, but the margin over that convention fell from about 2.7 times to about 1.5 times on vessels the device was not indicated for.

A burst pressure depends on its test conditions

Burst pressure depends on the vessel's size, type and collagen content, on whether the vessel was tested fresh, and on how the pressure was applied. Numbers from different studies, or from a manufacturer's own study against a competitor, are rarely comparable. A fair comparison needs the same vessels, the same method and an independent laboratory.

FDA issued its guidance for bipolar vessel sealers in 2016, and it describes the evidence FDA expects before such a device reaches the market. Burst pressure is tested for each mode against the predicate device, across the vessel sizes and types claimed, and thermal spread is measured by histology, as Chapter 2 describes. A chronic animal study follows at least five animals for at least three weeks to show that the seals stay closed as the tissue heals. It is run under good laboratory practice, the rules FDA sets for the conduct of such studies. And the indication must state the maximum vessel size, measured in its rounded, perfused state in a subject with normal blood pressure.

In practice

Treat the design or evaluation of a sealer as three controlled variables and one inferred outcome. Jaw pressure, energy delivery and end-point detection are designed and verified, while seal strength is inferred and must be demonstrated. Test each variable at its limits: the lowest jaw force, the largest indicated vessel and the thickest tissue. A verification plan built that way tests the inference as well as the hardware.

+ What this chapter established
  • Bipolar instruments pass current only through the tissue between their tips, at low voltage and with no return pad.
  • Advanced sealers add high jaw pressure, high current at low voltage and a closed loop that ends energy delivery by itself, and many are indicated for vessels up to 7 mm.
  • The generator infers the seal from impedance, which falls as the tissue warms and climbs steeply as it dries.
  • Burst pressure is the bench test for a seal, its widely used acceptance level of 360 mmHg is an industry convention rather than an FDA requirement, and in one comparison four of five devices lost about half their seal strength or more on vessels beyond their indicated size.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.