The frontier.
+ The questionWhat is changing in surgical energy, and what is likely to stay hard?
Non-thermal ablation goes mainstream
In January 2025, about two months after FDA approved it, Johnson & Johnson paused US cases with its Varipulse pulsed field catheter (Chapter 15). The correction that followed came after 4 of the 132 patients in the catheter's US evaluation had a stroke or a transient ischemic attack, about 3 % where 1 % or less was expected. FDA classed the correction as the most serious type of recall (Chapter 26 gives FDA's record). The case for pulsed fields rests on sparing the esophagus and the phrenic nerve, which heat and cold can injure, but the first months of one approved system showed that a new energy can bring a risk of its own.
The correction left the catheter on the market. Johnson & Johnson announced on 14 July 2025 that FDA had approved an updated irrigation flow rate for the Varipulse platform, and by November 2025 the company stated that more than 25,000 Varipulse procedures had been performed worldwide. The catheter's newer version reached Europe first. Johnson & Johnson states that Varipulse Pro received CE marking in March 2026, and it launched the catheter in Europe on 7 April 2026. Varipulse Pro had no US approval as of October 2026. On 24 April 2026 Johnson & Johnson announced PERSIGMA, a randomized trial of Varipulse Pro against Boston Scientific's FARAPULSE in persistent atrial fibrillation, with up to 466 participants at 50 sites.
In its release of 14 July 2025, Johnson & Johnson gives a neurovascular adverse event rate for Varipulse of less than 0.5 % and states that more than 10,000 procedures had been performed worldwide. The company also states that the rate was calculated from internal customer-reported data and complaints rather than from a prospective clinical investigation, and calls it preliminary and pending publication. A rate built from complaints counts only the events that someone reports, while the 4 of 132 came from a defined group of patients in an evaluation. The two figures measure different things, and neither shows by itself whether the risk has fallen.
One published safety record covers 17,642 patients, far more than the 132 of the Varipulse evaluation. Ekanem and colleagues published it in 2024 as MANIFEST-17K, a retrospective survey of the use of Boston Scientific's FARAPULSE at 106 centers in Europe and Israel after the system reached the European market. Its cases date mostly from early 2022 to June 2023, before FDA approved the system. It had no external funding, and many of its authors report ties to the company. Its findings bear on the case for pulsed fields made above: it found no esophageal complications and no persistent palsy of the phrenic nerve (Chapter 15).
MANIFEST-17K counted 173 major complications among 17,642 patients. Dividing 173 by 17,642 gives about 0.0098, and multiplying by 1,000 gives about 9.8 major complications per thousand patients, the 0.98 % the registry reports. Transient palsy of the phrenic nerve, in 11 patients, works out the same way at about 0.6 per thousand, and esophageal complications, of which there were none, at zero. The denominator decides how much one patient moves a rate. One patient more or less changes the registry's rate by 1,000 divided by 17,642, about 0.06 per thousand, while in the Varipulse evaluation of 132 patients one patient changes it by 1,000 divided by 132, about 7.6 per thousand. Its four strokes or transient ischemic attacks come to about 30 per thousand, against an expected 10 or fewer. The two rates cannot rank the catheters, because they count different events in different kinds of study; the arithmetic shows that a rate from 132 patients rests on far fewer cases than a rate from 17,642.
In tumor ablation, histotripsy received its first US authorization, a De Novo grant for liver tumors, on 6 October 2023. The evidence behind it came from HistoSonics' own single-arm trial in 44 patients (Chapter 16). HistoSonics has since sought an indication beyond the liver. It announced a De Novo request for kidney tumors on 11 May 2026, and no decision had been announced as of October 2026. In June 2026 HistoSonics itself still described the kidney request as pending, with use outside the liver "limited to investigational use". In Europe, HistoSonics announced CE marking of its Edison system for liver tumors on 13 July 2026, and Chapter 16 follows both steps. Investors had backed the company before both steps: an investor consortium's purchase of a majority stake had been announced in August 2025 (Chapter 23). Elsewhere, a clearance announced in December 2024 extended one IRE system's indication to name prostate tissue, and cryoablation, an older energy, gained a US indication in low-risk breast cancer through a De Novo granted on 3 October 2025 (Chapter 14).
What pulsed fields and histotripsy share is the idea of Chapter 6: for some tasks, the harm that heat does to the collagen scaffold of vessels and ducts matters more than how well it kills cells. Both energies now hold US authorizations. Histotripsy holds its grant for the liver, and four pulsed field systems were approved in the US between December 2023 and November 2024, with two more in 2025 (Chapter 15). The question for the next few years is how far the idea extends beyond the heart and the liver, and what evidence each new indication will need.
The rules and standards that these devices must meet changed in the same years. The plate marks those that changed between 2023 and 2026: the US quality regulation in force since 2 February 2026, a 2025 edition of the ultrasonic standard IEC 61847, an April 2026 amendment to the laser standard IEC 60601-2-22, and FDA's May 2026 database entry for edition 6.1 of the electrosurgical standard. Chapters 25 to 27 give each with its date and what it changes.
Robotic energy
Robotic surgery is changing how energy is controlled as much as where it is delivered. Intuitive's E-200 generator holds a US clearance dated August 2024, and Intuitive states that it works with its da Vinci systems with the energy settings controlled from the surgeon console, and that it is also available as a backup and standalone unit. The company also describes force feedback on some da Vinci 5 instruments, which applied up to 43 % less force to tissue in its preclinical trials.
Medtronic's Hugo is newer to the US: it was cleared there on 3 December 2025 for urologic procedures, and as of October 2026 those remained its only US indication. On 3 June 2026 Medtronic announced that it had submitted 510(k)s to extend Hugo to general surgery, including hernia repair, and to gynecology, and as of October 2026 both were pending. Medtronic states that the LigaSure sealer cleared for Hugo on 15 September 2026 is designed for use only with its Valleylab FT10 generator. Hugo's 510(k) summary shows that generator on the robot's tower, so the robot's sealer runs on a generator that the robot itself carries (Chapter 21).
As the robot takes over more of what the surgeon's hand once did, the questions of Chapter 7 grow: which controller decides when energy flows, how the robot and the generator are verified together, and how reused wristed instruments keep their insulation. Chapter 21 treats the first as an interface between products, and Chapter 7 gives the insulation failures of robotic instruments.
Instruments, lasers and power electronics
In urology, thulium fiber lasers now compete directly with holmium for stones and prostate work. The engineering difference is efficiency: a diode-pumped thulium fiber laser turns about 12 % of its input into light and can be air-cooled, against under 2 % for a flashlamp-pumped holmium laser that needs water cooling (Chapter 12). The difference shows in the mains supply each needs. Olympus specifies a supply of 100 to 240 V at 1,200 VA for its 60 W air-cooled SOLTIVE thulium systems, while Boston Scientific lists up to 46 A at 200 to 240 V for its 120 W Pulse 120H holmium system (Chapter 12). FDA cleared new thulium fiber systems from Olympus in 2024 and 2026 and from Allengers, an Indian maker, in December 2024 (Chapter 23). Holmium systems continue to develop as well, and FDA cleared Boston Scientific's MOSES Raydar holmium laser system in May 2026 (Chapter 23).
Instruments change in smaller steps. Olympus states that THUNDERBEAT II, cleared on 7 October 2025, adds an ultrasonic-only mode to its hybrid jaw and moves the transducer's cable to the base of the handle (Chapter 11). The generators that drive such instruments are controlled by software today, and they carry the software and cybersecurity obligations of Chapters 20 and 26. Faults in such generators reach the recall record as well. In July 2025 Olympus began a recall of 155 ESG-410 electrosurgical generators that could show a system error and then restart, sometimes repeatedly, and FDA classed it as Class II (Chapter 26).
What stays hard
Some problems in this guide are not going away, because they come from physics rather than from technology.
The effect is still inferred rather than measured. A sealer infers a seal from impedance, an RF ablation generator infers a zone from temperature or impedance, and the margin of an ablation is judged afterward on a scan. Direct, real-time measurement of the tissue effect remains rare. MRI thermometry during focused ultrasound, the bubble cloud of histotripsy and the visible ice ball come closest, and even the ice ball's visible edge marks 0 °C, with the lethal edge farther in (Chapter 14).
Energy still escapes out of view. Insulation still fails in reusable instruments, more often in robotic ones, and capacitive coupling is a property of any shaft in a cannula (Chapters 7 and 18). Monitoring helps, and the physics stays the same.
Tissue still varies. Perfusion, water content, collagen content and large vessels change what a given output does. Bench tests in unperfused tissue leave out the heat sink of Chapter 2, so they show the most favorable case.
Evidence is still hard to compare. Many performance studies are written or funded by manufacturers, conditions differ between studies, and each device generation changes the answer. The newest energies show the range. The trial behind histotripsy's liver authorization was its maker's own, the 17,642-patient registry had no external funding but many authors with ties to the maker, and the randomized comparison of two pulsed field catheters announced in April 2026 was launched by one of the two makers. Independent, standardized comparisons remain the scarcest evidence in the field. The words for regulatory status carry a hazard of their own, and Chapter 26 sets out how a request, a clearance, a grant, an approval and a certificate differ.
Treat every claim about a new energy, indication or rate as dated evidence before it enters a design review, a plan or a regulatory file. Record who made it and in which document, whether it is an announcement, a submission, a decision or a result, and its date; for a rate, record the count, the denominator and how the cases were found. Check the status against the regulator's own record before the claim is reused, and date it "as of" the day it was checked.
- The Varipulse correction of February 2025, after strokes or transient ischemic attacks in about 3 % of the patients in the catheter's US evaluation, showed that a new pulsed field system can bring a risk of its own.
- As of October 2026, histotripsy is authorized in the US for liver tumors only and no FDA decision on its kidney request has been announced, while one IRE system's indication names prostate tissue and cryoablation has gained an indication in low-risk breast cancer.
- Intuitive states that its generator's energy settings are controlled from the surgeon console, and as of October 2026 Hugo remains cleared in the US for urologic procedures only, while thulium fiber lasers turn more of their input into light and generators carry software and cybersecurity obligations.
- The effect is still inferred rather than measured, energy still escapes out of view, tissue still varies, and independent comparisons remain the scarcest evidence in the field.