Surgical lasers.
+ The questionChapter 5 showed that the wavelength decides where light is absorbed. What does it take to make light of one wavelength, and how does that choice shape a laser's whole design?
How a laser makes light
A photon of exactly the right energy passes an excited atom, and in that instant it can trigger the atom to emit a second photon identical to the first, with the same wavelength, the same phase and the same direction. This is stimulated emission, and each such event turns one photon into two. Atoms and molecules hold energy in discrete levels, and they absorb or emit light when they move between them. Normally most of them are in their lowest level, so light passing through a material is more likely to be absorbed than to trigger emission, and a laser has to reverse that balance.
The material that amplifies the light is called the gain medium. A pump lifts its atoms into an excited state faster than they can fall back, until more of them are in the upper laser level than in the level below it. This state is a population inversion. The pump may be a flashlamp, a diode laser or an electrical discharge. Mirrors at either end of the gain medium send the light back and forth through it, so the light is amplified on every pass. One mirror lets a small fraction through, and that fraction is the beam. Lasing begins when the gain on each pass exceeds the losses.
How the pump is made largely sets how much of the mains power ends up as light. A holmium laser pumped by flashlamps converts roughly 1 to 2 % of its electrical input into light, because the lamp's broad spectrum overlaps the absorption of holmium only slightly, and the rest becomes heat that water cooling must remove. A thulium fiber laser pumped by diodes whose output matches thulium's absorption converts about 12 % and can be air-cooled. In carbon dioxide lasers an electrical discharge through the gas does the pumping, and they reach 10 to 20 %.
The difference shows on the nameplate. The manufacturer of one 120 W holmium system lists a single-phase supply of 200 to 240 V at up to 46 A and a weight of 240 to 260 kg, depending on the version. A 60 W thulium fiber laser from another manufacturer runs from 100 to 240 V at 1,200 VA and weighs about 40 kg, about a sixth as much. The optical output divided by the electrical input comes to about 1.1 to 1.3 % for the holmium system and at least 5 % for the thulium laser, which is consistent with the published efficiencies.
Lasers and their jobs
Chapter 5's absorption curves decide what each laser is for.
The CO₂ laser emits at 10.6 µm, and tissue absorbs that light within about 16 µm, so the laser vaporizes a thin layer precisely, with little coagulation beneath it. It is used for cutting and vaporizing soft tissue, and in airway surgery, where precision near delicate structures matters. Tissue absorbs the light of the Ho:YAG laser, at 2.1 µm, within about 0.5 mm, and the light of the thulium fiber laser, at 1.94 µm, within about 0.1 mm. Both lasers therefore cut and coagulate together, thulium in a shallower layer. Both also work well under water, which is why they are used to treat stones and to enucleate an enlarged prostate.
Water also absorbs the light of the 1,470 nm diode laser within about half a millimeter, and this laser is used, for example, to close varicose veins from inside with a radial fiber. The light of the 980 nm diode and of the Nd:YAG laser at 1,064 nm is weakly absorbed and scatters over millimeters, so these lasers coagulate deeply. The 532 nm green laser is made with a KTP or LBO crystal. Hemoglobin absorbs its light and water lets it pass, which suits the vaporization of a blood-rich prostate under irrigation.
Getting the light there
The wavelength also decides how light reaches the tissue. Silica optical fibers carry light from the visible to about 2 µm with little loss, so every laser above except CO₂ can be delivered through a flexible fiber small enough to pass down an endoscope. The size of the fiber's core matters too. Holmium lasers need fibers with a core of 200 µm or larger, because their beam quality limits how tightly they can be focused into a fiber, while thulium fiber lasers couple into fibers as small as 50 to 150 µm. The GreenLight XPS system for the prostate uses the MoXy fiber, a side-firing fiber with a 750 µm core that directs the beam sideways at the tissue.
Silica absorbs light at 10.6 µm strongly, so it cannot carry a high-power CO₂ beam. CO₂ lasers are therefore delivered by an articulated arm of mirrors, or through hollow waveguides made for the purpose.
Pulses, bubbles and stones
Chapter 5 showed that the length of a pulse sets how far its heat spreads and how high its peak power climbs, and lasers for lithotripsy, the breaking of stones, use both effects. A holmium pulse delivers about 1 J in roughly 300 µs, a peak of about 3.3 kW. A thulium fiber laser works at a peak of about 500 W, so the same joule takes about 2 ms, and the laser compensates with far higher repetition rates. The manufacturer of one current thulium system states that it offers pulse energies from 0.025 to 6 J at up to 2,400 Hz.
Under water, a laser meets an obstacle before it reaches the stone, because the water itself absorbs the light. The front of each pulse therefore boils a vapor bubble at the fiber tip, and the light that reaches the stone is the light that passes through that bubble. Since 2017, pulse-shaping modes have exploited the bubble by firing a short, low-energy sub-pulse to open it and then the main pulse through it, so more of the energy reaches the stone. The manufacturer of one such mode claims 50 % less stone movement and a 33 % higher ablation rate.
Figures such as the claimed 50 % less stone movement come from manufacturers' own testing, under conditions they chose. They describe the mode under those conditions and are not a measure of clinical benefit in a particular patient or stone type. Independent comparisons, with the same stones, fibers and settings, are the evidence to look for.
Pulse settings then decide how a stone breaks. Two published sources give the settings: a 2017 review by Aldoukhi and colleagues and a 2018 doctoral thesis by Hardy on thulium fiber laser lithotripsy. The review's settings are for a 120 W Lumenis holmium laser, and one of its authors, Ghani, consulted for Lumenis and Boston Scientific when it was published. For dusting, between them, the two sources give low pulse energies within about 0.2 to 0.5 J at 40 to 80 Hz, which erode the stone into fine particles that can be washed out. For fragmentation, they give higher energies within about 0.5 to 2 J at 5 to 30 Hz, which break the stone into pieces to be removed with a basket. Longer pulses, lower energies and smaller fibers all reduce the recoil that pushes a stone away from the fiber.
Between them, the review and the thesis above give settings within about 0.5 to 2 J at 5 to 30 Hz for fragmentation and about 0.2 to 0.5 J at 40 to 80 Hz for dusting. The average power is the energy of each pulse multiplied by the number of pulses each second. One setting inside each range shows the difference. A fragmentation setting of 1 J at 20 Hz gives 1 J times 20, or 20 W, and a dusting setting of 0.5 J at 80 Hz gives 0.5 J times 80, or 40 W. The peak power is the energy of one pulse divided by the time it lasts. If the fragmentation pulses are holmium pulses of 1 J lasting roughly 300 µs, or 0.0003 seconds, each one peaks at about 3,300 W, about 170 times the 20 W average. At these two settings, dusting therefore runs at twice the average power of fragmentation with pulses of half the energy, and the average power of a setting says little about the peak each pulse reaches.
A laser's wavelength decides what absorbs it, how deep the energy goes, which fibers can carry it, and, through the pump that makes it, how much mains power it draws and how it must be cooled. Choosing a wavelength is choosing most of the system.
Treat the wavelength as the first requirement of a laser system, and derive most of the rest from it. Carry the pump's efficiency into the mains supply, the cooling and the weight, because a flashlamp-pumped holmium laser turns less than 2 % of its input into light and needs water cooling, while a diode-pumped thulium fiber laser can be cooled by air. Check that silica fiber can carry the wavelength, and which core sizes the beam can enter. Specify each pulse setting by its energy, rate and duration together, because the average power follows from the energy and the rate, and the peak power from the energy and the duration.
- A laser pumps a gain medium into population inversion and amplifies stimulated emission between mirrors, and the way it is pumped sets its efficiency: under 2 % for flashlamp-pumped holmium, about 12 % for thulium fiber and 10 to 20 % for CO₂.
- The wavelength decides the job: CO₂ for precise vaporization, holmium and thulium for cutting under water and for stones, 980 nm diode and Nd:YAG for deep coagulation, 1,470 nm diode for vein closure, 532 nm for blood-rich tissue.
- Silica fibers carry every surgical wavelength up to about 2 µm, while CO₂ needs mirrors or hollow waveguides.
- Under water each pulse must first boil a vapor channel, and pulse shaping and pulse settings decide how much energy reaches a stone and whether it dusts or fragments.