Fifty thousand cells a minute.
In March 1969, four researchers at the Los Alamos Scientific Laboratory in New Mexico described an instrument built around a stream of water fifty micrometers wide. Their report filled two pages of Science. Hamster cells had been stained so that the DNA in their nuclei glowed green, and they flowed down the stream and across the beam of an argon-ion laser running at about one watt. Each cell took 15 to 20 microseconds to cross the beam and gave one flash of that length. A photomultiplier, a detector for very faint light, caught the flashes, and a pulse-height analyzer sorted them by brightness, typically 50,000 a minute.
The result was a histogram with two peaks. Cells that had not yet copied their DNA made the first. Cells that had copied it, before or during division, made the second, and its brightness was 1.97 times that of the first, against the 2.00 that a doubling of DNA predicts. The authors set their result beside one made on mouse cells with a scanning microscope: 594 cells, measured over many hours. Theirs held about 50,000 cells and took a minute.
The instrument used no microscope and, in the paper's words, "a minimum of optics." It made no image of any cell. It recorded flashes and their brightness, and the analyzers built since have kept that kind of record, with many detectors in place of one.
The 1969 histogram could be checked against a simple expectation, because DNA doubles exactly and one stain reported it. Most of what cytometry is asked today is harder: which of a dozen kinds of cell a particle belongs to, whether a particle a tenth of a micrometer across is there at all, and whether a count is good enough to decide a diagnosis.