+ Part II · Light meets matter
How a particle produces the light a cytometer measures.
A dye excited by blue light at 488 nanometers glows green, not blue. That shift in color lets a filter pass a faint glow while it blocks a far brighter laser. Fluorescence is one of the two ways a particle sends light back. The other is scatter: the particle itself deflects light, by an amount that depends on its size, its refractive index and the angle from which it is watched. Chapter 3 explains fluorescence and Chapter 4 explains scatter. Chapter 5 shows how antibodies carry dyes to particular proteins on a cell or a vesicle, and what decides how bright the result looks.
Why a molecule glows a different color.
+ The questionWhy does a dye give back light of a longer wavelength than the light it absorbed, why does that difference make fluorescence measurable, and why does one dye's light reach the detectors meant for others?
Light in, different light out
Light carries its energy in small packets called photons, and the more energy a photon carries, the shorter its wavelength. A fluorescent molecule absorbs a photon and emits one a few billionths of a second later. The emitted photon almost always has less energy than the absorbed one, which means a longer wavelength: a dye excited by blue light at 488 nanometers glows green, and one excited by red light at 640 nanometers glows in the deep red.
The reason lies in what happens between absorption and emission. Absorbing a photon lifts one of the molecule's electrons into a higher energy state, and usually into a vibrating version of that state with some energy to spare. The molecule sheds that spare energy as heat to the surrounding water within a tiny fraction of the time it spends excited. It then falls from the bottom of the excited state back to its lowest state and emits a photon, which carries less energy than the photon that arrived. A rule named for the chemist Michael Kasha states the regularity: molecules of this kind emit light with appreciable yield only from their lowest excited state, whichever higher state the absorbed photon reached.
The gap between the peak of absorption and the peak of emission for the same transition is the Stokes shift. The International Union of Pure and Applied Chemistry sets the standard terms of chemistry. It defines the Stokes shift as the difference between the band maxima of absorption and fluorescence "arising from the same electronic transition." A band is the range of wavelengths over which a molecule absorbs or emits, and its maximum is the band's peak. The shift is usually expressed in frequency units rather than in nanometers, because a gap of 10 nanometers in the red corresponds to less energy than a gap of 10 nanometers in the blue.
The Stokes shift is what makes fluorescence measurable at all in a cytometer. The laser delivers far more light to the particle than any dye gives back, and much of that light is scattered toward the detectors at the laser's own wavelength. Because the fluorescence comes out at longer wavelengths, an optical filter can block the laser light and pass the dye's glow. Without that difference in color, the faint emission would be lost in the far brighter laser light.
How bright is a dye
Two properties of a molecule decide how much light it can give back. The molar absorption coefficient measures how strongly the molecule absorbs light of a given wavelength, and it is often still called the extinction coefficient. The quantum yield is the fraction of absorbed photons that come back out as fluorescence; the rest of the energy is lost as heat or through other processes. Fluorescein was the dye of the earliest two-color instruments. A technical report of the International Union of Pure and Applied Chemistry gives its quantum yield as about 0.93 in alkaline solution, so in that solution nine of every ten photons that fluorescein absorbs return as light.
The product of the two properties is the usual measure of a dye's intrinsic brightness. The authors of a 2005 review of fluorescent proteins defined brightness in their tables as the "product of extinction coefficient and quantum yield," and the convention is common throughout fluorescence work. The two protein dyes most used in cytometry are R-phycoerythrin, or PE, and allophycocyanin, or APC.
For PE, a 1982 paper and one manufacturer's handbook agree on an absorption coefficient of 1,960,000 and a quantum yield of 0.82. For APC the handbook gives 700,000 and 0.68 for the whole molecule, which weighs about 104,000 daltons; a dalton is a unit of molecular mass, about the mass of one hydrogen atom. The 1982 paper gives the same quantum yield and a coefficient of 232,000 for one of the molecule's three identical units, each of 33,000 daltons. Three times that coefficient is about 700,000, the handbook's figure for the whole molecule.
PE: 1,960,000 × 0.82 ≈ 1.6 million. APC: 700,000 × 0.68 ≈ 476,000. The comparison is between whole molecules, as they are attached to antibodies, and on intrinsic brightness PE is about 3.4 times brighter than APC. The units of the absorption coefficient, liters per mole per centimeter, are the same for both and cancel in the ratio. The comparison assumes each dye is excited at its own absorption peak and that every emitted photon is counted, which no cytometer achieves.
Why the ranking does not survive the instrument
The 2005 review also warned that perceived brightness depends on much more: the wavelength and intensity of the illumination, the spectra of the filters and mirrors, and the sensitivity of the detector to the emitted color. A cytometer imposes all three. It excites a dye with whichever laser line it has, meaning the single wavelength of one of its lasers, and that line may sit well away from the dye's absorption peak. The filter in front of the dye's detector passes only part of the dye's emission. The detector converts photons to electrons with an efficiency that varies with wavelength, and background light from other sources sets the floor against which the signal must be seen.
The effect can be large. One manufacturer's application note reports measurements made on its own instruments with the same antibody against CD4, labeled once with PE and once with APC. The separation of stained from unstained cells differed between the two dyes by only about 10 %, against the threefold difference in intrinsic brightness. The comparison mixes many factors, from laser power to the number of dye molecules on each antibody, and that mixing is the point. The current community guidelines for cytometry in immunology make the same point in a figure caption: "fluorochrome brightness can be instrument-specific."
A table of dyes ranked by intrinsic brightness is a starting point for panel design, not an answer. What matters in a cytometer is how well a dye separates labeled from unlabeled particles on that instrument, with its lasers, filters and detectors, and that has to be measured on the instrument.
A nanosecond inside a microsecond
A dye molecule typically stays excited for a few nanoseconds before it emits, while a cell spends about a microsecond or a few in the laser beam. In principle each molecule could therefore be excited and emit hundreds of times during one pass, and more laser power would give proportionally more light.
In practice the light does not grow in proportion. At the high intensities in a focused laser spot, some molecules cross into a long-lived "dark" state from which they cannot be excited again for a while, and some are destroyed altogether, which is called photobleaching. A 1997 study measured fluorescein, PE and APC on cells and found them "considerably saturated and bleached in standard flow cytometric conditions." For APC excited with red light, the study raised the laser power about fourteenfold, from 35 to 510 milliwatts, and the maximum fluorescence rose only about elevenfold. Long illumination at high power caused serious bleaching.
Saturation matters to an engineer for two reasons. Once a dye saturates, doubling laser power does not double the signal, so a more powerful laser adds less signal than its rated power suggests. And because the shortfall depends on the dye, the intensity of the beam and how long the particle spends in it, two instruments with different lasers or flow speeds can rank the same dyes differently.
Broad colors that overlap
Fluorescence spectra are broad. A typical dye emits over a range of many tens of nanometers, with a long tail toward longer wavelengths, because the emitting molecule can land in any of many vibrational levels of its lowest state, the ground state. The Stokes shift of the protein dyes is small, about 13 nanometers for PE and 10 for APC. Cytometers therefore excite them away from their absorption peaks and collect their emission with filters placed to one side of the laser line.
The breadth of the spectra has a consequence that shapes the rest of the guide. When several dyes are used on the same particle, the tail of one dye's emission falls inside the filter meant for another. That overlap is called spillover. Better filters cannot remove it without throwing away most of each dye's light, and spillover is the reason a cytometer measuring many colors must separate the dyes mathematically (Chapter 10).
When designing a panel or specifying an instrument, measure how well each candidate dye separates stained from unstained particles on the instrument itself, with the antibody it will actually carry. Do not rely on tables of intrinsic brightness or on another instrument's results. For an instrument specification, state sensitivity per detector with the laser line and filter, because a single figure for "sensitivity" cannot describe how each dye will look.
- A fluorescent molecule absorbs a photon, loses some energy as heat, and emits a photon of longer wavelength; that Stokes shift lets filters separate the glow from the laser light.
- Intrinsic brightness is absorption coefficient times quantum yield: PE is about 3.4 times APC by that measure.
- Measured brightness depends on the laser line, filters, detector and background, and on one manufacturer's analyzers PE and APC differed by only about 10 %; brightness is instrument-specific.
- At high laser intensities dyes saturate and bleach, so their light stops growing in proportion to power, and their broad, overlapping spectra cause spillover between detectors.