Proving an assay does what it claims.
+ The questionAn instrument can be verified with beads, but a clinical result depends on cells that have no certified reference. How does a laboratory show that a flow assay is precise, sensitive and stable enough, and how many cells does a rare-event result need?
Validation without a reference cell
Most clinical tests are anchored to reference materials: a solution of known glucose, a serum of known hormone concentration. Cytometry has almost none for the quantities it reports. The international hematology and clinical cytometry societies published recommendations in five parts in 2013 (Chapter 21). The recommendations state the problem plainly: stable reference preparations "are unavailable for the cell-based measurements," and tracing a cell measurement back to a defined molecular quantity is not yet feasible. A laboratory must therefore prove its assay with patient samples, stabilized cells and careful design rather than with certified standards. That proof is validation, which shows that an assay serves its clinical or scientific purpose, just as verification shows that an instrument meets its specification (Chapter 15).
The Clinical and Laboratory Standards Institute, which writes standards for clinical laboratories, published the most complete guide to validation on 27 October 2021. It is H62, Validation of Assays Performed by Flow Cytometry, a first edition of 234 pages. Its governing idea is fit for purpose: "the level of validation should be tailored to the intended use of the data." The standard adds that this is "not an endorsement for inadequate validation." The institute noted that before H62 there were no official guidance documents for validating flow assays, and this guide found no FDA guidance specific to them.
In the United States the legal minimum is set by the CLIA regulations of Chapter 18. A laboratory using an unmodified cleared test verifies its accuracy, precision and reportable range, the span of results the test can report. It also confirms the reference intervals for its own patients. A reference interval is another name for the reference range of Chapter 20. For a test it developed or modified, the laboratory must establish accuracy, precision, analytical sensitivity and specificity, reportable range and reference intervals itself.
What kind of result
What has to be validated depends on what the assay reports. H62 and the 2013 recommendations sort flow assays into categories. A quantitative assay with true reference standards reports absolute values, which cytometry rarely achieves. A quasi-quantitative assay reports numbers that are proportional to the quantity but have no calibration standard, and a CD4 count is the usual example. A qualitative assay reports a category, such as the presence of an abnormal population with a given phenotype, and leukemia immunophenotyping belongs here. When the FDA authorized the first leukemia panels, it described them as qualitative assays.
The category decides the validation menu. A quantitative or quasi-quantitative assay needs precision across its range, linearity, limits of detection and quantitation, and reference intervals. A qualitative assay needs agreement with a reference diagnosis, precision of the call, and the smallest abnormal population it can reliably detect. For leukemia and lymphoma panels, the FDA sets special controls, requirements specific to that type of device (Chapter 23). They express that smallest population as limits of blank, detection and quantitation. The limit of blank is the highest result expected from a sample without the target. The limit of detection is the smallest amount that can be told apart from it, and the limit of quantitation the smallest that can be measured with acceptable precision and accuracy. Both kinds of assay need specimen and reagent stability, carryover and the effect of reagent lots.
Precision and agreement
Precision is measured with replicates, but each replicate is already the average of thousands of cells. The analytical part of the 2013 recommendations therefore limited replicates to no more than six, "since each 'measurement' actually is derived from measurements of tens of thousands of cells." A 2024 teaching workshop summarized the targets of the series' fifth part, on assay performance: a coefficient of variation below 10 % for most results and below 20 % for populations of about one in a thousand or rarer. The US special controls for leukemia and lymphoma panels go further for manufacturers, requiring precision on clinical samples at a minimum of three sites, at least two of them external, and reproducibility across at least three reagent lots.
Agreement is where the category matters most. The manufacturer of the first authorized leukemia panels ran the study behind them, and the FDA reviewed it. Two blinded experts read 279 specimens from four sites, and their readings were compared with each site's final diagnosis. Their sensitivity was 82 and 86 %, and their specificity 94 and 93 %. Sensitivity here is the share of patients with the disease whom a reader called positive, and specificity the share of patients without it whom a reader called negative.
In a later submission for a newer instrument, the same manufacturer reported 100 % agreement with its predicate instrument, the cleared instrument it was compared with, across 512 specimens stained with the manufacturer's later ten-color panels (Chapter 23). Both figures are correct, and they answer different questions. The first measures how often the panel's reading matches the patient's eventual diagnosis, and the second how often two instruments give the same reading.
Rare events and the limits of detection
Residual-disease testing pushes validation to its limit. The 2018 guidelines of Chapter 21 define undetectable disease in chronic lymphocytic leukemia as fewer than one leukemic cell in 10,000 leukocytes, or white blood cells. Myeloma is a cancer of plasma cells, the cells that make antibodies, and it grows in the bone marrow (Chapter 18). For myeloma, the EuroFlow consortium described next-generation flow in a 2017 study: two eight-color tubes, bulk lysis to acquire at least 10 million cells per sample, and automated gating against a database of normal marrows. Bulk lysis removes the red cells from a large volume of sample, which leaves many more white cells to measure.
The myeloma study defined the limit of detection as 20 abnormal plasma cells among at least 10⁷ events, 2 in a million or lower, and the lower limit of quantitation as 50, 5 in a million or lower. In 110 marrows from patients in deep response, 47 % were positive by the new method against 34 % by conventional eight-color cytometry.
The limits are counts, so they depend on how many cells were actually acquired. In the same study the median sample yielded 10.4 million cells, and 7 of 110 fell below 7 million. For B-cell precursor leukemia, a leukemia of immature B cells, a 2017 study compared flow cytometry with a PCR test, which copies and counts a DNA sequence found only in the leukemic cells. The study found flow cytometry at least as sensitive, reaching 1 in 100,000 "if sufficient cells are measured," which meant more than 4 million. In a 2022 analysis of acute myeloid leukemia, the target of 500,000 CD45-positive events was reached in 158 of 261 patients.
Counts of rare cells follow Poisson statistics (Chapter 12). If a sample truly contains enough abnormal cells to give 20 events on average, a single measurement shows 20 or more only about 53 % of the time. A 95 % chance of reaching 20 needs an expected count of about 28, and seeing at least one abnormal cell with 95 % probability needs an expected count of 3. The 20- and 50-event conventions are rules for how many events make a population, not statistical detection limits in the sense used for chemistry tests. The probabilities are derived for this guide.
Regulators have taken note of residual disease as evidence. In a 2020 guidance on its use in drug development, the FDA asked that an assay detect at least tenfold below the clinical decision threshold. In April 2024 an FDA advisory committee voted 12 to 0 that residual disease could support accelerated approval of myeloma drugs, a route that approves a drug on an early measure likely to predict its benefit. In January 2026 the FDA issued a draft guidance that describes that early measure as the proportion of patients in complete response whose bone marrow is negative for residual disease by flow cytometry or sequencing, at a threshold of at least 1 in 100,000 cells. As of October 2026 this guide found no final version, and the only residual-disease device it found authorized in the United States was a sequencing test. Flow cytometric residual-disease assays run as laboratory-developed tests, which a laboratory designs and runs itself.
A residual-disease result reported as negative at 1 in 100,000 is negative at that level only if enough cells were acquired, the sample was fresh enough and the controls passed. A sample that yields a tenth of the target cells has a detection limit ten times higher, and a report that omits the number of cells analyzed hides that change.
Stability, lots and quality over time
A validated assay stays valid only within its tested conditions. Heparin and EDTA are anticoagulants, which keep blood from clotting. The 2013 recommendations on preanalytical issues, the handling of a sample before analysis, describe blood in heparin as optimally stable for 48 to 72 hours and in EDTA for up to 48 hours at room temperature. They note that some markers degrade faster with time than others: CD138 and CD16 more than CD45 or CD64. CD138 marks plasma cells, so in a myeloma residual-disease test the age of the marrow sample bears directly on the result. Stability is validated by testing at least five specimens of each type and anticoagulant over time, up to the oldest sample the laboratory will accept.
Reagent lots are a recurring change. Under the analytical recommendations, a new lot or any major hardware change calls for a full compensation procedure, and tandem dyes need values specific to each lot (Chapter 10). Over time, laboratories hold quality with daily instrument checks (Chapter 16), external quality assessment and accreditation. External quality assessment compares laboratories in the same way as the proficiency tests of Chapter 16, and the 2013 recommendations make it mandatory at least three times a year. Accreditation is to ISO 15189:2022, the international standard for medical laboratories. Under the arrangement among accreditation bodies, accreditation to the 2012 edition has not been recognized since the transition ended in December 2025. One UK scheme runs 11 accredited and 6 pilot programs for cytometry, from stem-cell counts to residual disease. List-mode files and analysis files are kept for at least two years (Chapter 18).
State the assay category, then list the parameters it needs, each with acceptance criteria. They include precision with replicates and sites, agreement with a stated reference, and limits of detection and quantitation as event counts and cell numbers. They also include specimen and reagent stability, carryover, lot-to-lot reproducibility and reference intervals. Use H62 and the 2013 recommendations as the checklist. For rare-event assays, report the number of cells acquired and the resulting detection limit with every result.
- Cytometry lacks certified reference cells, so assays are validated with patient samples, stabilized cells and design; CLSI H62 of 2021 sets a fit-for-purpose approach.
- The assay category, quantitative, quasi-quantitative or qualitative, decides what must be shown; agreement with a diagnosis differs from agreement between instruments.
- Residual-disease limits are counts: 20 abnormal cells in 10⁷ for myeloma, which only samples that actually reach 10⁷ cells can claim.
- Stability, reagent lots, external quality assessment and accreditation keep a validated assay valid over time.
Let's Talk