What the file keeps, and what it drops.
+ The questionWhat does the standard data file record about how its numbers were made, what is lost on the way to a report, and when does the file itself become the record that matters?
Three segments and a list of keywords
Cytometry data have had a standard format for four decades. The first Flow Cytometry Standard, FCS 1.0, was adopted in 1984, and revisions followed in 1990 and 1997. Version 3.1 was published in 2010, and a version 3.2 followed in 2021. As of October 2026, the data-standards page of the International Society for Advancement of Cytometry lists the full specification of 3.1 but not of 3.2, so this guide describes 3.1. The format's stability is a large part of why files from different instruments and decades can still be opened by the same software.
An FCS file has three main segments. A short header says where the others begin. The text segment holds keywords, each a name and a value, such as the number of events, the number of parameters, and each parameter's name, range and detector settings. The data segment holds the measurements in list mode: one row per event and one column per parameter. Each parameter carries a short name for the detector, such as FL1-A, and may carry a long name for the axis label, such as "CD45 FITC Fluorescence - Area," so that the file can tell the detector from the marker it measured.
A file stores each measurement in 4 bytes when it uses 32-bit numbers. One million events with 24 parameters therefore need about 96 million bytes of data. A test for residual disease that acquires 10 million events produces a file of nearly a gigabyte. The header gives the start and end of each segment as byte positions in eight-digit fields, so once a segment ends beyond byte 99,999,999, its positions move into keywords in the text segment. The sizes are derived for this guide.
The data can be stored as unsigned integers, as 32-bit floating-point numbers or as 64-bit ones. The choice matters wherever values near and below zero occur, as they do in files that software exports or re-saves after compensation, even though the standard expects stored data to be uncompensated (Chapter 10). Such values survive in floating point, while unsigned integers cannot hold negative numbers at all.
What the file records
FCS 3.1 has keywords for much of the acquisition. For each parameter it can record the detector type, voltage, gain and filter, and the wavelength and power of the light that excited it. It can record one trigger channel and its threshold (Chapter 9), the volume analyzed, and the spillover matrix, which the standard calls "the only standardized way to specify compensation." The standard states that "the stored data is always uncompensated," so software applies compensation from the matrix, and only one matrix can be stored for each data set.
Three keywords added in version 3.1 matter for calibration, display and record-keeping. One carries a calibration factor per parameter that converts channel numbers into units such as MESF (Chapter 17); it is a single multiplier and is usually absent. A second recommends a display scale but recognizes only linear and logarithmic scales, so a logicle or arcsinh display cannot be recorded there. A third records the file's originality, from Original to DataModified. The standard advises that modifying data sets is bad practice and that "the original file should always be kept." The file can end with a checksum over the whole data set, but the checksum is optional, and software that skips it when writing the file fills the space with zeros.
What the file does not record
The list of what FCS 3.1 has no keyword for is as instructive as what it has. It does not record the name or version of the acquisition software, only the computer and operating system. It has no keyword for laser delays, processing window, sheath pressure, baseline restoration or detector temperature (Chapter 13), none for the daily quality-control results of Chapter 16, and none for any trigger more elaborate than one channel above one level. Manufacturers write some of these into their own keywords, in units the standard does not define.
The consequence is that a file alone cannot reproduce an acquisition. The file holds the measurements and much of the optical configuration, but several settings that decide whether those measurements are valid live only in the instrument, its log or its software. A data set exchanged between laboratories arrives without those settings unless they are documented separately.
What export and analysis drop
Further losses happen after acquisition. Some exporting software converts data to a linear scale, and the standard gives that as the reason for the display keyword. The reporting standard MIFlowCyt permits a spreadsheet export of list-mode data, and such an export keeps the numbers and loses every keyword: the detector settings, the matrix and the provenance. Gates and transforms with their parameters are not part of FCS 3.1 at all, apart from crude legacy region keywords. They are kept in the analysis software's workspace or in a separate standard, Gating-ML 2.0, which was published in 2015 and can describe logicle and arcsinh transforms and gate boundaries exactly. A compensation matrix edited at analysis time is also kept outside the file.
MIFlowCyt was published in 2008 and is still the community's reference. It asks for what the file cannot hold. A report "shall" state the denominator of every gate percentage, give the compensation matrix and say whether it is a spillover or a compensation matrix, and describe transforms by an open specification, while the software name and version are only recommended. Chapter 20 shows how MIFlowCyt's extension for vesicles, MIFlowCyt-EV, adds the flow rate and calibrations in standard units, and asks for the trigger channel and threshold preferably in calibrated units. MIFlowCyt also asks that a report provide its list-mode files or say how to obtain them. FlowRepository is a public archive where researchers deposit cytometry data. As of 6 October 2026 it had temporarily stopped accepting new experiments for lack of storage, while downloads continued.
A file can lose the spillover matrix, the scale, negative values and its originality status when it is exported to a spreadsheet, converted to integers, re-saved after editing or passed through software that rewrites keywords. It still opens normally. The numbers look the same, but the information needed to check them is gone.
The file as a record
In regulated settings the file becomes evidence. The US rule on electronic records, 21 CFR Part 11, applies to records kept under the FDA's own record requirements, such as those of a device manufacturer. It requires validated systems, accurate and complete copies, protected retrieval and "secure, computer-generated, time-stamped audit trails." Clinical laboratories in the United States keep their records instead under the Clinical Laboratory Improvement Amendments, CLIA, the federal rules for laboratories that test human samples. These rules require quality-control and patient test records, "including instrument printouts," to be kept for at least two years, and pathology reports for ten.
The file is also becoming the test object. EuroFlow runs an external quality assessment for residual disease in myeloma, a cancer of the antibody-producing cells of the bone marrow. The program sends participants FCS files rather than samples, so that laboratories are compared on analysis alone, with the variation of the instrument and the sample removed. For that comparison to mean anything, the files must arrive with their keywords intact.
Archive the original FCS files unchanged, with a checksum computed on arrival. Store the analysis workspace or Gating-ML file, the compensation matrix used, the transform parameters and the software version with each result. Record laser delays, sheath pressure, threshold and quality-control results in an instrument log linked to the files. Use MIFlowCyt as the checklist for reports, and set retention by the rules that apply: CLIA for a US clinical laboratory, and, for records kept under FDA requirements, the retention period of the governing FDA rule, with Part 11 controls on the electronic records.
- FCS files hold a header, keyword text and list-mode data; version 3.1 is the version this guide describes, and the standard requires data to be stored uncompensated, with at most one spillover matrix.
- The file records detector settings, excitation, one trigger, volume and an optional calibration factor, but not laser delays, sheath pressure, software version or quality control.
- Export and analysis can drop keywords, scales, negative values and gates; Gating-ML and MIFlowCyt cover what the file cannot hold.
- In regulated use the file is a record kept under Part 11 or CLIA, and EuroFlow already compares laboratories on supplied files.
Let's Talk