04 — Scatter

What scattered light can and cannot tell you.

+ The questionWhy is forward scatter a poor measure of size, and why does a vesicle of 100 nanometers scatter almost no light at all?

Light the particle itself redirects

A polystyrene bead 100 nanometers across sends a side-scatter detector about as much light as a vesicle of about 175 nanometers. The detector is working correctly: scatter depends on more than size.

Fluorescence needs a dye. Scatter needs nothing but the particle. A material's refractive index measures how much it slows light compared with a vacuum. Any object whose index differs from that of the surrounding fluid deflects some of the light that strikes it. The deflected light keeps the wavelength of the light that arrived, and a cytometer collects it at two places.

Forward scatter is the light deflected through small angles and collected by a lens or detector placed opposite the laser. The angles run from about one degree to roughly fifteen, depending on the instrument. The direct laser beam is far brighter than this light, so a small opaque strip called an obscuration bar blocks the beam itself and lets only the slightly deflected light through. Side scatter is light deflected roughly at right angles to the beam. It is collected by the same lens that gathers fluorescence, a high-aperture lens that accepts light over a wide cone, and it reaches its own detector through a filter at the laser's wavelength.

08 — Where scattered light goes
laser focusing lens FORWARD SCATTER small angles interrogation point stream, seen end-on obscuration bar blocks the beam, and with it the smallest angles side collection lens, at about 90° dichroic SIDE SCATTER laser wavelength FLUORESCENCE behind bandpass filters
ParticlesFluidExcitation lightFluorescenceScattered lightFocal detail
Plate 08 — Forward scatter is collected at small angles around the beam, behind a bar that blocks the laser itself; side scatter and fluorescence are collected at about 90 degrees by a separate lens. Each path sees different light from the same particle.

Because every particle scatters, scatter is the usual trigger that decides what becomes an event (Chapter 1), and the familiar first plot of almost every analysis is forward scatter against side scatter. On that plot, blood separates into three clouds that correspond, more or less, to lymphocytes, monocytes and granulocytes. Granulocytes are white cells whose interior holds many small granules. The plot is the origin of two habits of speech: forward scatter as "size" and side scatter as "granularity."

What forward scatter is called, and what it is

Neither habit is a measurement. A 2011 study tested how well optical signals track the volume of cells, and it found that the area of the forward-scatter pulse "performed poorly" as an indicator of volume. The authors listed the reasons: forward scatter is influenced by the difference in refractive index between particle and fluid, by light-absorbing substances inside the particle, and by the optical design of the forward-scatter path. In some cell lines side scatter tracked volume better than forward scatter did. For particles comparable in size to the wavelength of light, the relation is not even monotonic: scattered light can rise and fall as size increases, because light waves scattered from different parts of the particle interfere.

A specification sheet that quotes a forward-scatter "resolution" of 1 micrometer describes how well the instrument separates beads of known size from one another. The figure is not a claim that forward scatter measures diameter, and the size it implies holds only for particles made of the same material as the beads.

"Size" and "granularity" are labels, not quantities

Forward scatter depends on size, refractive index, shape, absorption and the instrument's collection angles. Side scatter depends on internal structure and on size as well. Both are useful for separating populations whose differences are already known, and neither measures a physical dimension unless the instrument has been calibrated with a model of the particle's optical properties (Chapter 17).

Refractive index decides as much as size

Scatter depends on the contrast in refractive index between particle and medium. Water at 488 nanometers has an index of about 1.33. Most calibration beads are made of polystyrene, which has an index of about 1.61 at that wavelength, and silica has about 1.44.

Living cells are close to water: a 2016 review gives about 1.35 for most normal cells, with higher values for some cancer cells and for red cells. Extracellular vesicles are also close to water, though further from it than most cells. Published estimates for vesicles run from about 1.37 to 1.45, and values near 1.39 to 1.40 are used most often. Plasma, the liquid part of blood, also carries lipoproteins: particles of fat and protein that transport fats around the body. Lipoprotein particles outnumber vesicles in plasma and have higher indices, near 1.47 in one study.

The contrast matters enormously. At 100 nanometers, a sphere of polystyrene scatters about seventeen times as much light as a vesicle of the same diameter, and at 1 micrometer about ten times as much. A 2010 review noted that a gold particle of 200 nanometers scatters 27 times more than a polystyrene sphere of similar size, which in turn scatters 15 times more than a vesicle. Even a small change matters: if the index assumed for a vesicle falls from 1.40 to 1.36, its expected scatter falls about sixfold.

09 — Same size, different scatter
0 5 10 15 RELATIVE SIDE SCATTER (VESICLE OF n = 1.40 = 1) polystyrene n ≈ 1.61 silica n ≈ 1.44–1.46 vesicle n = 1.40 vesicle n = 1.36 17× about 17 times a vesicle of the same size 2.6–3.7× 1 the reference 0.16 about one sixth, from a small change of index all particles 100 nm, in water (n = 1.333); side scatter at 488 nm, derived (Mie)
ParticlesScattered lightFocal detail
Plate 09 — Scatter depends on refractive index as much as size. At the same 100-nanometer diameter, polystyrene scatters about 17 times as much as a vesicle of index 1.40, so a bead size is not a vesicle size.

The sixth power

For particles much smaller than the wavelength of light, scattering follows a simple law worked out by the physicist Lord Rayleigh. The scattered light grows with the sixth power of the diameter and falls with the fourth power of the wavelength, and a factor set by the refractive-index contrast multiplies the result. A 2010 review of vesicle detection put the consequence plainly: "If a microvesicle is only 10-fold smaller than another microvesicle, the scattering cross-section and thus the scattered amount of light decreases 10⁶-fold."

How fast scatter falls

Under Rayleigh's law, halving the diameter divides scattered light by 2⁶ = 64. Vesicles of 100 to 200 nanometers are at the edge of that law in blue light, and for them the full theory of scattering by spheres, called Mie theory, applies instead. This guide made a Mie calculation for vesicles of index 1.40 in water at 488 nanometers, and it gives a factor of 53 between 50 and 100 nanometers rather than 64. The same calculation gives a 1-micrometer polystyrene bead about 300,000 times the total scatter of a 100-nanometer vesicle, and a sphere the size of a lymphocyte about ten million times. A 2014 study reported the same order of magnitude: the smallest and largest vesicles in one sample typically differ 25-fold in size and 10,000,000-fold in scattered light.

The steep fall of scatter with size is why a cytometer that sees every white cell with ease can miss most vesicles entirely. A cell scatters millions of times more light in total than a 100-nanometer vesicle of low refractive index. On a conventional instrument, such a vesicle sends so few photons that the signal is indistinguishable from the light that the optics, the fluid and the detector produce with no particle present. A 2010 survey put the lower detection limit of commercial cytometers at 300 to 500 nanometers for polystyrene beads, and because vesicles scatter far less than polystyrene, the smallest vesicle detectable on the same instruments was larger still.

10 — Scatter falls with the sixth power
0.1 10 1,000 100,000 10,000,000 50 100 200 500 1,000 DIAMETER, nm (LOG SCALE) SCATTERING CROSS-SECTION, nm² (LOG SCALE) detection floor (schematic) Rayleigh, d⁶ polystyrene vesicle, n = 1.40 50 → 100 nm: 53× 53-fold for a doubling of diameter derived (Mie), 488 nm, in water schematic between the marked values
Scattered lightFocal detail
Plate 10 — Below a few hundred nanometers, scatter falls roughly with the sixth power of diameter: halving a vesicle from 100 to 50 nanometers cuts its scatter about 53-fold. Below a size set by each instrument, vesicles drop beneath its detection floor.

Seeing small particles with side scatter

The same physics points to the remedies. A particle much smaller than the wavelength scatters light over a wide range of angles, so the side-scatter path, with its large collecting lens, gathers more of that light than the narrow forward path, which must also contend with glare around the obscuration bar. The Mie calculation above took side scatter as 45 to 135 degrees and forward scatter as 1.5 to 15 degrees. On those angles, a 100-nanometer vesicle sends about nineteen times more light to the side than forward. A shorter wavelength helps through the fourth-power term. Moving from blue light at 488 nanometers to violet at 405 roughly doubles the scatter of a small particle, which is why some instruments offer violet side scatter for small particles.

In 2020 a group showed how far an instrument can be pushed. The group rebuilt a common clinical analyzer step by step: a laser ten times more powerful, a photomultiplier in place of the forward-scatter photodiode, a redesigned obscuration bar, a pinhole to reject stray light and a narrower sample stream. The changes improved the separation of particles from noise about 30-fold on side scatter. With them, the estimated detection limits for vesicles of index 1.40 were 246 nanometers on forward scatter and 91 nanometers on side scatter.

Beads cannot substitute for a calculation of that kind, made with the refractive index of vesicles. Because polystyrene scatters so much more than a vesicle, the 100-nanometer bead of this chapter's opening gives the side scatter of a vesicle of about 175 nanometers. In 2012 a widely used bead mixture for setting size gates was tested on one instrument, and it selected single vesicles between 800 and 2,400 nanometers across. The reporting framework for vesicle cytometry was published in 2020, and it concludes that "a gating strategy based on polystyrene beads alone is not a sound standardization methodology."

Specify scatter by angle, wavelength and model

Specify an instrument meant for small particles by its scatter detection limit for a stated material and refractive index, at a stated wavelength and collection angle, rather than by a bead size alone. A method that reports particle size from scatter should state the optical model it used, the refractive index it assumed and the instrument's collection angles, as Chapter 17 describes.

+ What this chapter established
  • Scatter is laser light deflected by the particle itself: forward scatter at small angles, side scatter near right angles; it needs no label and usually triggers events.
  • Forward scatter is not size and side scatter is not granularity: both depend on refractive index, shape, absorption and the instrument's optics.
  • Refractive index matters as much as diameter: at 100 nanometers, polystyrene scatters about 17 times more than a vesicle of the same size.
  • Below a few hundred nanometers scatter falls roughly with the sixth power of diameter, so small vesicles vanish into the noise unless the optics are built and calibrated for them.