Thirteen rules distilled from the chapters, for people who design, verify, buy, regulate or write about flow cytometers and the tests run on them. Each names the chapter whose mechanism it rests on.
- Treat every number as a measurement of light, not of a particle. An event is a pulse that crossed a threshold: it may be one cell, two cells, debris or a swarm of vesicles. A particle that stayed below the threshold leaves no record at all (Chapters 1, 6, 9 and 19).
- Specify detection channel by channel, as detection efficiency Q and background B. In the published example of Chapter 8, background, not detector efficiency, set the dimmest resolvable signal. Single sensitivity figures on datasheets cannot be compared between manufacturers (Chapters 8 and 15).
- Never read forward scatter as size, or bead size as vesicle size. Scatter depends on refractive index and collection angle as much as on diameter; a 100-nanometer polystyrene bead scatters like a vesicle of about 175 nanometers (Chapters 4, 15 and 17).
- Fix the flow rate, the trigger and the threshold, and record them. A faster run widens the core and raises coincidence. The threshold decides which particles exist in the data (Chapters 6 and 9).
- Expect compensation to restore averages, never precision. Spillover spreading grows with the square root of the spilled light. Dim markers therefore belong in detectors that bright dyes on the same cells do not spread into (Chapters 10 and 14).
- Set each positive boundary on a control that contains everything except the specific binding. Fluorescence-minus-one controls and cells known to lack the marker do that. Unstained and isotype samples do not (Chapter 11).
- Keep the original file, and record what it cannot hold. Transforms, gates, matrices, denominators, laser delays and software versions decide what a number means. The standard file keeps only some of them (Chapters 12, 13 and 18).
- Count donors as replicates, and count enough events of the population reported. A million events from one sample are one observation. A rare population's precision depends on its own events: one hundred events give about 10 % (Chapters 12 and 22).
- Own the interfaces, and check them in every run. Failures in stream speed, laser delays, sample preparation and settings arise at the boundaries between subsystems. They look like biology rather than like errors (Chapters 7, 13 and 16).
- Standardize to make instruments alike; calibrate to make numbers mean something. Shared targets made beads agree within 5.5 % and cells within about 30 %. Calibrated units carry the assumptions of their models (Chapters 16 and 17).
- Prove that each vesicle event is one particle, and a vesicle. A dilution series, buffer and reagent controls and a detergent step come before any count, and every count needs its size window, brightness threshold and measured volume (Chapters 19 and 20).
- Validate for the intended use, and report the cells acquired with every rare-event result. A limit of 20 abnormal cells in ten million applies only to a sample that reached ten million cells (Chapters 21 and 22).
- Assemble the evidence once, then map it to each jurisdiction. The instrument, the reagents and a laboratory's own panel are classed differently in the US, the EU and India. A research-use label never makes clinical use lawful by itself (Chapter 23).