02 — One at a time

Why measure one particle at a time.

+ The questionA fluorometer measures the glow of a whole tube of cells at once. What does measuring every particle separately reveal that an average hides?

What an average cannot say

The simplest way to measure how much of a fluorescent label a sample of cells has taken up is to put the whole suspension in a small clear container called a cuvette, shine light through it and measure the total glow. A fluorometer does exactly that, and it is fast, cheap and precise. It cannot, however, answer the question that cytometry exists to answer, because it reports one number for millions of cells.

Suppose that 12 % of the white cells in a sample carry about 50,000 molecules of a dye each and the rest carry none. The average over all cells is 6,000 molecules per cell. A second sample in which every cell carries exactly 6,000 molecules gives the same total glow, the same average and the same reading on the fluorometer. Biologically the two samples could not be more different: one contains a distinct population of strongly labeled cells, the other a uniform population of weakly labeled ones. Only a measurement of each cell separately tells them apart.

03 — Same average, different samples
IN BULK: ONE NUMBER ONE PARTICLE AT A TIME SAMPLE A 12 % of cells at 50,000 molecules, the rest none laser detector 6,000 average per cell 88 % dark 12 % bright invisible to any average average SAMPLE B every cell at 6,000 molecules laser detector 6,000 average per cell 0 1,000 10,000 100,000 all cells 6,000 MOLECULES PER CELL schematic histograms
ParticlesExcitation lightFluorescenceReadoutFocal detail
Plate 03 — A bulk measurement reports only the average. Two samples with the same average can differ completely: one where 12 % of cells are bright and the rest dark, one where every cell is moderately bright. Measuring each particle separately tells them apart.
Three things an average destroys

Measuring each particle separately preserves three kinds of information that a bulk measurement averages away. The first is the distribution: whether a population is one group or several, and how wide each is. The second is rare events: a population that makes up 0.001 % of a sample, one cell in 100,000, changes a bulk signal by less than its noise, but it can be counted if enough cells are measured one by one. The third is correlation: whether the same particle that is bright in one color is also bright in another. Correlation is how cytometry separates helper T cells, a kind of lymphocyte that carries both CD3 and CD4, from monocytes, another kind of white cell that carries some CD4 but no CD3.

Those three properties are why cytometry became one of the standard tools of immunology and of hematology, the study of blood. Blood is a mixture of dozens of cell types, and so is bone marrow, the tissue inside bones where blood cells are made. Many of these cell types look alike under a light microscope and differ mainly in the proteins on their surfaces. A disease such as leukemia shows up as an unusual population: sometimes a large one at diagnosis, and sometimes, after treatment, a very small one hidden among normal cells. A method that measures every cell, and several properties of each, is the natural instrument for finding either.

Counting particles in a stream

Measuring one particle at a time began with counting. In 1949 Wallace Coulter filed a patent for a device that counted particles suspended in a conducting fluid as they passed one at a time through a small opening carrying an electric current. The patent was granted in 1953. Each particle displaced some of the fluid and briefly changed the resistance of the opening, so each produced an electrical pulse whose size grew with the particle's volume. The Coulter principle is electrical rather than optical, and it is still used in the hematology analyzers that count red and white cells in routine blood tests.

Optical counting of cells in a flowing stream was explored from the 1930s onward. In 1953 a paper described a way of confining the cells to the center of the stream by surrounding them with a faster-moving sheath of clean fluid. This hydrodynamic focusing is the arrangement that almost every flow cytometer still uses to put particles in single file at a fixed point in a light beam (Chapter 6).

From absorption to fluorescence

In 1965 Louis Kamentsky and colleagues described an instrument that measured how much ultraviolet light individual cells absorbed, at rates above 500 cells per second. The absorption served as an estimate of each cell's nucleic acid content, its DNA and RNA. In the same year Mack Fulwyler at Los Alamos described the first instrument that sorted cells one at a time by a measurement of each. It measured each cell's volume by the Coulter principle and then broke the stream into droplets, using a method that Richard Sweet had just published for ink-jet recording. The instrument charged the droplets that contained the cells it wanted and deflected them into a separate container. Neither instrument measured fluorescence.

Fluorescence arrived at the end of the decade. Wolfgang Dittrich and Wolfgang Göhde in Germany filed a patent in December 1968. In 1969 they published a method of measuring the fluorescence of single cells in a flowing suspension, and an instrument based on that method was sold commercially soon afterward. Also in 1969, the Los Alamos group of this guide's opening story reported measuring fluorescence from cells excited by a laser, and Leonard Herzenberg's group at Stanford published the first sorting of cells according to their fluorescence. In 1972 the Stanford group described a laser-based instrument in a paper titled "Fluorescence Activated Cell Sorting." The medical-technology company Becton Dickinson brought commercial sorters to market in the early 1970s. The name "flow cytometry" came into general use in the second half of the decade.

04 — From counting cells to fifty colors
1950 1960 1970 1980 1990 2000 2010 2020 1949 Coulter patent filed (granted 1953): electrical counting 1953 hydrodynamic focusing: a faster sheath centers the cells 1965 absorption measured at over 500 cells/s 1965 cells sorted by volume into droplets 1968–69 fluorescence measured in flow 1969 first sorting by fluorescence (Stanford) measure and keep: light decides which cell is sorted 1972 laser-based cell sorter 1975 monoclonal antibodies 2013 first commercial full-spectrum instrument COLORS MEASURED ON ONE CELL 2 colors at the start 12 17 40 50
ReadoutFocal detail
Plate 04 — Flow measurement grew in steps: electrical counting, a focused stream, fluorescence and sorting, monoclonal antibodies, then more colors on each cell until full-spectrum instruments separated dozens of dyes mathematically.

The decisive change came from biology rather than engineering. In 1975 Georges Köhler and César Milstein described how to make monoclonal antibodies, identical antibodies produced without limit by one cloned line of cells. A monoclonal antibody binds to one site on one protein, and with a fluorescent dye attached it labels every cell that carries that protein. From the late 1970s onward, panels of such antibodies turned the cytometer from an instrument that measured size and DNA content into one that could identify cell types by the proteins they display.

More colors at once

Each additional color measured on the same cell adds a dimension in which populations can be separated. The Stanford group recalled in 2002 that they had started with two colors, fluorescein and rhodamine, and by then had a twelve-color instrument. A 2004 review described seventeen-color cytometry. In 2020 a published panel used forty colors on a full-spectrum instrument, and two of its three authors worked for the instrument's manufacturer. In 2024 another published panel used fifty. It was developed on instruments from two makers, and two of its five authors worked for one of them.

The growth did not come from adding more of the same detectors. Fluorescent dyes emit over broad, overlapping ranges of wavelength, so the light of one dye reaches the detectors meant for others even when only two dyes are used, and with every added color the instrument has more to separate mathematically (Chapter 10). Full-spectrum or spectral instruments rely on that mathematical separation for every dye. Instead of one detector behind one filter for each dye, they spread the light from each cell across an array of dozens of detectors and work out the contribution of every dye from the shape of the whole spectrum. The maker of the first commercial full-spectrum cytometer announced it in 2012 and launched it in 2013. Chapter 14 compares the two architectures as a systems engineer would.

More colors are not free

Every added color brings more light from other dyes into each detector and more uncertainty into every value calculated from it. A forty-color panel is possible because of careful choice of dyes, detectors and controls, not simply because the instrument has enough detectors. A panel with fewer colors, chosen well, often resolves a given question better than a larger one chosen badly.

What stayed the same

Across seventy years the principle has not changed. A cytometer brings particles one at a time to a fixed point, measures what each one does to a beam of light or a current, and records numbers for each. Every generation has added detail to what is measured about each particle: volume, then absorption, then scatter and fluorescence, then many colors and whole spectra. None has changed the fact that the measurement is of a signal from a particle, not of the particle.

Sorting rests on the same idea. A sorter makes the same measurements as an analyzer and then acts on them while the particle is still in the stream. It charges a droplet that contains a chosen particle, so that an electric field deflects the droplet into a collection tube. This guide treats sorting in Chapter 24, as an extension of the analyzer rather than as a separate subject.

Ask what question needs single-particle data

A bulk measurement is cheaper, faster and more precise for a quantity that is uniform across a sample, such as the total concentration of a soluble protein. Single-particle measurement earns its cost when the answer depends on a distribution, a rare population or a combination of markers on the same particle. Stating which of the three a measurement needs is the first step in specifying an instrument or designing an assay, because each one sets different demands on throughput, sensitivity and the number of colors.

+ What this chapter established
  • A bulk measurement reports one average and cannot distinguish a small bright population from a uniformly dim one; measuring each particle can.
  • Single-particle data preserve distributions, rare events and correlations between markers on the same particle.
  • Flow measurement grew from Coulter's electrical counting, with a patent filed in 1949, through absorption and volume sorting in 1965 to fluorescence between 1968 and 1972, monoclonal antibodies in 1975 and full-spectrum instruments in 2013.
  • Every generation added detail to the signal; none changed the principle that the cytometer measures a signal from each particle, one particle at a time.