16 — Stability

Keeping one instrument the same, and two alike.

+ The questionA cytometer's lasers age, its filters fade and its detectors drift. How is one instrument kept the same from day to day, and how far can laboratories go toward making different instruments give the same numbers?

Every instrument drifts

A cytometer that met its specification on the day it was installed will not meet it unchanged a year later. Laser output and alignment drift, filter coatings age (Chapter 8), detectors lose efficiency, and the fluidics wear. Each change shifts the numbers a sample produces, and Chapter 13 showed that most shifts look like biology rather than like faults. Stability therefore has to be measured with something that does not change, on every day the instrument is used.

That something is a set of quality-control beads: plastic particles dyed with fixed amounts of fluorescent dye, run before samples to check that each channel reads where it should, with the spread it should have. One manufacturer's system uses beads at three levels, and the manufacturer explains the choice. In its bright beads, counting statistics contribute negligibly to the spread. In its mid beads they contribute a large part, and in its dim beads the background contributes significantly. Spread has three sources: the beads' own variation, the counting statistics of Chapter 8 and the background. Because the three levels expose them separately, a system of this kind can report relative values of Q and B for each channel, along with detector voltages and laser delays.

Each day's results are plotted against time, often on a Levey–Jennings chart with limits drawn around the expected value, so that a slow trend becomes visible before it crosses a limit. This guide found no peer-reviewed data set that follows many instruments over years. The published evidence on drift comes mostly from the daily logs of standardization programs, such as the one described below.

Choosing the detector voltage

Before an instrument can be kept the same, its detector settings have to be chosen. The voltage on a photomultiplier sets its gain. Too low, and dim signals are lost in the electronic noise of the amplifier and converter. Too high, and bright signals run off the top of the scale. Between the two lies a range in which raising the gain no longer improves the spread of a dim population, because counting statistics, not electronics, now dominate.

44 — Choosing the detector voltage
bright beads run off the top 300 400 500 600 700 800 900 999 rCV of the second-dimmest bead peak (schematic) DETECTOR VOLTAGE, STEPS OF 50 electronic noise dominates counting statistics dominate chosen voltage schematic; EuroFlow procedure with eight-peak beads
ReadoutFocal detail
Plate 44 — As gain rises, a dim bead's spread falls out of electronic noise until counting statistics dominate and the curve flattens. EuroFlow chose the voltage at the beginning of that plateau; more gain only loses room at the top of the scale.

The EuroFlow consortium found that range with beads of eight brightness levels. On a reference instrument, the second-dimmest bead population was measured at voltages from 300 to 999 volts in steps of 50 volts, and its robust CV was calculated at each. The chosen voltage was the one "at the beginning of the plateau phase of the curve." Manufacturers' systems use related rules, such as placing a dim bead at ten times the spread of the electronic noise. One core-facility protocol aims for electronic noise to contribute only 10 to 20 % of the measured variance. Above the plateau, extra gain buys nothing at the bottom of the scale and loses room at the top.

Target values make instruments alike

Keeping one instrument stable is not enough when results from many laboratories must mean the same thing. EuroFlow addressed that problem for the leukemia panels of Chapter 21, and in 2012 it described its instrument settings, the product of six years of joint experiments. Its laboratories used instruments with three lasers at about 405, 488 and 633 nanometers and a fixed set of eight dyes. The filter configurations were identical, apart from exceptions that the description lists.

The method turned one instrument's settings into targets for all. After the voltages were chosen on the reference instrument, the brightest bead peak was measured in every channel. Those values became the target values, which every other instrument had to reach by adjusting its own voltages. One master lot of beads was used throughout. Each day, an instrument had to pass two checks: its bead values had to lie within 15 % of target, and the spread of its brightest peak had to be below 4 % on the blue and violet channels and below 6 % on the red channels and the PE–cyanine 7 channel.

45 — Target values make instruments alike
BEFORE A B B reads low AFTER target ±15 % A B bead intensity, log scale brightest-peak CV < 4 % (blue, violet), < 6 % (red, PE-Cy7) beads across 8 instruments: < 5.5 % cells across 11 labs: about 30 % beads test instruments; cells test everything else 0 % 10 % 20 % 30 % EuroFlow 2012 and quality rounds 2010–2013
ParticlesReadoutFocal detail
Plate 45 — EuroFlow set every instrument so that one bead lot read its target values, within ±15 % each day. Bead values then agreed within 5.5 % across eight instruments, while cell intensities across laboratories still varied by about 30 %.

The approach worked on beads. Across the eight instruments of the 2012 evaluation, the variation of the brightest peak's value was "systematically lower than 5.5%." The instruments were not identical, but they had been made to read the same reference material the same way, and that property can be checked every day.

How far standardization reaches

Beads are the easy part. Between 2010 and 2013, EuroFlow ran four annual rounds of quality assessment on 123 blood samples from healthy donors, measured on 14 instruments in 11 laboratories in nine countries with uniform gating. The average median fluorescence in each round ranged from 86 to 125 % of the overall median, and for 7 of 11 markers the variation stayed below 30 %. The authors who reported the rounds described the scheme as a complement to established proficiency tests, not a replacement. In a proficiency test, an outside organizer sends the same samples to many laboratories and compares their results.

A 2019 study transferred the method to 10 clinical laboratories in Switzerland, six with instruments of one manufacturer and four with another's. Over three rounds, the average variation across 11 markers was 30.3 %, 32.9 % and 27.8 %, comparable with EuroFlow's own laboratories. Median fluorescence did not differ significantly between the two makes. By the third round, 89 % of participants had at least nine in ten of their measurements within the acceptance criteria.

Why beads agree within 5.5 % and cells within about 30 %

A bead is the same particle in every laboratory, measured with the same dye, so its value tests only the instrument. A cell population measured in a quality round carries everything else as well: blood from different donors, reagents from different vials and lots, staining by different people, and gating by different analysts. The bead figure is the floor set by the instruments. The cell figure stays near 30 % even after full standardization, and it is the realistic spread for intensities compared between sites. The comparison of the two published figures is made for this guide.

Without a shared scheme, beads alone spread more widely than EuroFlow's 5.5 %. The 2012 study of Chapter 8, which covered 133 cytometers, found that cross-calibration with hard-dyed beads, which carry dye through their whole volume, varied by typically 20 % or more, with variation between instrument models often greater than within one. The 2018 study of Chapter 8 found that detection efficiency varied between analyzers of the same model by about a factor of two in some channels and by more than an order of magnitude in others. Its authors included staff of a cytometer manufacturer and of an analysis-software company. No shared settings exist for vesicles, and Chapters 17 and 20 show how calibration in physical units narrows a spread that is far larger still.

What standardization buys is that populations land in the same place. A pattern of markers read in one laboratory can then be compared with another laboratory's and with reference databases of normal and malignant cells. EuroFlow's panels were designed for that automated comparison.

Daily quality control covers the instrument, and only the instrument

Bead checks show that the instrument reads beads as it did yesterday. They do not detect a degraded tandem dye in a reagent vial, a wrong compensation matrix, a sample prepared differently, or a gate moved by hand. Quality control of the instrument is necessary for a trustworthy result and is not sufficient for one.

Treat targets, limits and logs as controlled records

For an instrument used in a regulated or multi-site setting, define how detector voltages are chosen, the bead material and lot that carries the targets, the daily acceptance limits for value and spread, and what happens when a limit is crossed. Keep the charts as records, review trends at fixed intervals, and repeat the voltage and target procedure after any service that touches lasers, filters or detectors. Join an external quality assessment, a scheme of regular proficiency tests, for each panel in routine use.

+ What this chapter established
  • Lasers, filters and detectors drift, so stability is measured daily with dyed beads and tracked on charts, with Q and B where the bead system allows.
  • Detector voltages are chosen at the start of the plateau where a dim population's spread stops improving with gain.
  • EuroFlow made instruments alike by setting each to target bead values with one bead lot, reaching variation below 5.5 % on beads across eight instruments.
  • Cell intensities still varied by about 30 % between standardized laboratories, enough for populations to land in the same place, not at identical values.