+ Part VII · Cytometry as a test
From a measurement to a medical result.
A research measurement has to convince its author; a clinical test has to be right for a patient who will never see the data. Part VII follows cytometry into the clinic through the guide's second case, leukemia immunophenotyping. Chapter 21 shows how patterns of markers become part of a diagnosis and what a report contains. Chapter 22 shows how a laboratory proves that an assay performs as claimed, down to the rarest cells it reports, and Chapter 23 how the instrument, its reagents and a laboratory's own panel are regulated in the United States, the European Union and India. The last part of the central question, when a count can be trusted as a test result, is answered here.
When a result becomes a diagnosis.
+ The questionA clinical laboratory turns a tube of blood or marrow into a sentence in a patient's record. How do patterns of markers point to a disease, what does a report contain, and what can the cytometer not decide on its own?
What clinical cytometry is asked
The US classification for flow cytometric leukemia and lymphoma panels lists the conditions they help to diagnose: acute and chronic leukemias, non-Hodgkin lymphomas, myeloma, myelodysplastic syndromes and myeloproliferative neoplasms, in blood, bone marrow and lymph node. All of them are cancers or related disorders of blood cells and of the marrow cells that produce them. Beyond these, clinical laboratories use cytometry to count CD4 T cells in people with HIV and CD34 stem cells before transplantation. Laboratories also use cytometry to detect the missing surface proteins of paroxysmal nocturnal hemoglobinuria, to screen for inherited immune deficiencies and to look for leukemia left after treatment. Paroxysmal nocturnal hemoglobinuria is a rare disease in which blood cells lack a group of surface proteins and red cells break down.
In 2006 an international consensus meeting set out when diagnostic cytometry is indicated. A 2025 educational module from the College of American Pathologists restates the list for blood: abnormal cells seen under the microscope, an increase in lymphocytes that may be reactive or malignant, increased blasts, suspected relapse and several rarer situations. Blasts are the immature cells from which blood cells develop.
The module also states when cytometry is not indicated. Neutrophils are the most common white cells in blood. Normal mature neutrophils, red cells and platelets look the same whether or not they come from a clone, a population descended from a single cell, so an isolated rise in neutrophils, red cells or platelets is not a reason to order cytometry. Routine diagnostic panels characterize abnormal populations above about 1 % of cells. Detecting residual disease below 0.01 % is a separate, specialized test (Chapter 22).
A pattern, not a marker
No single marker identifies a leukemia. Blood cells develop along several branches, the lineages of Chapter 12. Normal blood and marrow contain cells at every stage of maturation in every lineage, and almost every marker found on leukemic cells is also found on some normal cell. What identifies a population is a combination: which markers it carries, how brightly, and in what pattern relative to its normal counterpart (Chapter 11).
The EuroFlow consortium of Chapter 10 built this idea into the design of its panels. Each eight-color tube combines backbone markers, which identify the main populations in the sample, with characterization markers, which describe them further. A panel of several tubes carries the backbone on the same dye in every tube. The population of interest then appears in the same place in each tube, and the tubes can be analyzed together (Chapter 16).
EuroFlow's entry point for suspected acute leukemia is a single tube, the acute leukemia orientation tube, and its one question is which lineage the leukemic cells belong to. The tube pairs markers found inside the cell, in its cytoplasm, with surface markers of immaturity and lineage. The internal markers are myeloperoxidase for the myeloid lineage, CD3 for T cells and CD79a for B cells, and the surface markers include CD34, CD45, CD19, CD7 and surface CD3. The myeloid lineage is the branch that produces red cells, platelets and the white cells other than lymphocytes.
The tube was refined over five design cycles on 385 acute leukemias and then tested prospectively on 483 further cases in eight laboratories. In the textbook reading, myeloperoxidase points to a myeloid leukemia, cytoplasmic CD3 with CD7 to a T-lineage one, and CD19 with cytoplasmic CD79a to a B-lineage one, while CD34 and dim CD45 mark the cells as immature. CD45 is carried by nearly all white cells, and immature ones carry less of it.
A worked case: chronic lymphocytic leukemia
Chronic lymphocytic leukemia shows how a pattern becomes a criterion. The 2018 international guidelines require at least 5 × 10⁹ clonal B lymphocytes per liter of blood, which is 5,000 per microliter, persisting for at least three months, and they require cytometry to show that the cells form a clone. The cells carry CD5, normally a T-cell marker, together with the B-cell markers CD19, CD20 and CD23. Their surface immunoglobulin, CD20 and CD79b are characteristically dim compared with normal B cells.
Surface immunoglobulin is the antibody that each B cell carries on its surface (Chapter 11). An antibody is a protein built from two heavy chains and two light chains, and cytometry shows clonality through the light chains of that surface antibody. Every B cell uses either a kappa or a lambda light chain, and a normal population contains both. A clone uses only one, because all its cells descend from one cell. A population of CD5-positive B cells that is all kappa or all lambda is therefore clonal, and the diagnosis rests on that measurement.
The thresholds draw the boundaries of the diagnosis. The same clonal cells below 5 × 10⁹ per liter define monoclonal B lymphocytosis when the patient has no enlarged lymph nodes or organs, no low blood counts and no symptoms. With enlarged nodes, the same cells suggest small lymphocytic lymphoma, which a biopsy confirms. In 2018 a European harmonization project agreed on six required markers: CD19, CD5, CD20, CD23, kappa and lambda. The project also agreed on recommended markers for borderline cases, and it validated the approach retrospectively on 14,643 cases from 13 centers, with more than 97 % concordance with the diagnostic approaches then in use.
Counting cells in absolute numbers
Some clinical results are counts per microliter rather than percentages. In its 2003 guidelines for CD4 counts, the US Centers for Disease Control and Prevention favored single-platform methods, in which a known number of fluorescent counting beads is added to a measured volume of blood. The alternative, dual-platform methods, multiply a cytometry percentage by a count from a separate hematology analyzer, and so they compound the errors of both instruments.
Suppose a tube receives 50,000 counting beads and 50 microliters of blood, so there are 1,000 beads for each microliter. After staining, the cytometer records 40,000 bead events and 20,000 CD4 T-cell events. Recording 40,000 of the 50,000 beads means that 40 microliters of blood were analyzed, so the count is 20,000 ÷ 40,000 × 1,000, or 500 CD4 T cells per microliter. The guidelines ask for at least 2,500 gated lymphocytes and accurate pipetting, because errors in the bead count and the blood volume pass straight into the result. The values are illustrative, for this guide.
The same principle counts CD34 stem cells. For that count, the US Pharmacopeia, which publishes official standards for medicines, sets a method that asks for at least 75,000 CD45-positive events and at least 100 CD34-positive cells. Because the cells and the beads are counted in the same run, the instrument's flow rate drops out of the result, and single-platform counting thereby solves, for cells, the volume problem that Chapter 17 found still open for vesicles.
Blasts can be reported as a percentage of all events, of CD45-positive cells, of nucleated cells or of a gated population. These percentages can differ widely in a marrow with many red-cell precursors or much debris. A report that does not state the denominator and the gating behind it cannot be compared with a microscope count or with the next report on the same patient.
The cytometer is one witness
A cytometry result is never a diagnosis on its own. The 2022 World Health Organization classification of myeloid neoplasms describes itself as predicated on integrating morphology, immunophenotype, and molecular and cytogenetic data. Morphology is the appearance of cells under the microscope, the immunophenotype is the pattern of markers that cytometry measures, and molecular and cytogenetic data describe the cells' genes and chromosomes. The classification also dropped the requirement of 20 % blasts for most acute myeloid leukemias that carry defining genetic abnormalities. A parallel International Consensus Classification, also of 2022, takes the same multiparameter approach. The US regulation for leukemia and lymphoma panels requires their labeling to say that results "should be interpreted by a pathologist or equivalent professional" together with other clinical and laboratory findings.
The 2013 recommendations of the international hematology and clinical cytometry societies describe the report itself. Interpretation is reserved for people who know malignant hematology, instrumentation, software and data analysis. Abnormal populations are quantified as percentages and described by their intensity, bright or dim, relative to their normal counterparts, and the findings are correlated with morphology. Counts such as lymphocyte subsets carry a comment that compares them with the reference range (Chapter 20). An Italian society's report format for acute leukemia lists 13 elements. They run from the specimen type and the sample quality, including viability, to the gating procedure, the blast phenotype and the panel to use later for residual disease.
For a panel developed for clinical use, state the intended use: the conditions, the specimen types, the population reported and its denominator. State also the limit of detection for an abnormal population and the interpretation statement that the labeling will carry. For a laboratory, fix a report template that names the specimen, the sample quality and viability, and the gating strategy. The template also gives each population with its denominator and its intensity relative to normal, absolute counts where relevant, and the interpreter's comment.
- Clinical cytometry helps diagnose leukemias, lymphomas and related neoplasms, counts CD4 and CD34 cells, and detects residual disease, but it is not indicated for isolated rises in mature blood cells.
- Lineage is read from patterns: EuroFlow's orientation tube pairs internal lineage markers with surface markers, refined on 385 leukemias and tested on 483.
- Chronic lymphocytic leukemia rests on CD5-positive B cells with a single light chain at 5 × 10⁹ per liter or more for at least three months.
- Results are integrated with morphology and genetics by a qualified interpreter, and every reported percentage needs its denominator.