11 — Controls

Where a positive signal begins.

+ The questionEvery cell gives some light in every detector, stained or not. Which comparison decides that a cell carries a marker, and why do some widely used controls answer the wrong question?

Positive is a comparison, not a value

In 1998 an editorial on counting lymphocytes and CD34-positive cells carried the subtitle "time to let go!" Its target was the isotype control, then in widespread use in clinical cytometry: an antibody made to bind nothing on the cells, so that its signal stands for the test antibody's nonspecific binding.

No detector reads zero for an unlabeled cell. Cells glow faintly by themselves, which is called autofluorescence. The glow comes from flavins, small molecules made from vitamin B2, and from other molecules inside the cells. Light from the panel's other dyes spills into the detector and leaves its spread behind after compensation (Chapter 10). Antibodies also stick weakly where their target is absent: to Fc receptors on some white cells, to dead cells, and to other cells through simple stickiness (Chapter 5). The authors of a 2009 review grouped these sources of background into three kinds: autofluorescence, spectral overlap and undesirable antibody binding.

A cell is therefore never positive in an absolute sense. It is positive when its signal in one detector is higher than that of comparable cells that differ only in lacking the specific binding being tested. A control is the sample, or the population within a sample, that supplies that comparison. Each kind of control leaves out a different source of background, so the choice of control decides what the boundary between positive and negative actually measures.

29 — What a detector reads for a stained cell
ONE DETECTOR autofluorescence spillover spreading nonspecific binding specific signal STAINED CELL UNSTAINED autofluorescence only FMO + spillover spreading ISOTYPE another antibody's binding, no spread INTERNAL NEGATIVE same tube: all but specific binding, if cells are alike schematic; layer sizes illustrative
ParticlesFluorescenceReadoutFocal detail
Plate 29 — A detector's reading for a stained cell stacks autofluorescence, spillover spreading, nonspecific binding and the specific signal. Each control removes different layers, so the choice of control decides what the positive boundary measures.

The boundary matters most where a marker is dim. A bright marker separates from its negatives on almost any reasonable control. A dim one lies within a few multiples of the background, and moving the boundary by the width of the negative population can change the fraction of positive cells severalfold. Leukemic cells, whose markers are often abnormally dim, have to be read in that region, and so do vesicles, whose labels sit close to the background (Chapter 5).

The unstained cell is the wrong reference

The simplest control is a sample of the same cells with no stain at all. It measures autofluorescence and nothing else. In the 2001 review of Chapter 10, Mario Roederer explained why that is not enough for a multicolor panel. He wrote that on compensated data "any gate based on a completely unstained sample" would give wrong results, because at high intensities the negatives spread upward into the positive region. An unstained sample has no spillover, so it has no spillover spreading, and a boundary set on it is too low for the stained tube.

The same review gave the alternative a name. A fluorescence-minus-one control, or FMO, contains every antibody of the panel except the one whose positive boundary is being set. It carries all the spillover into that detector, with all its spread, and none of the specific signal. One manufacturer's staff wrote the 2006 guideline of Chapter 10. The guideline defined the FMO as a sample with all the conjugates "except one" and noted its limit: an FMO does not measure the background binding of the missing antibody itself, because that antibody is absent.

A boundary set on the wrong control

Suppose 99 % of unstained cells fall below 300 in the phycoerythrin detector, and a boundary is set there. In the FMO tube, where the other dyes are present but phycoerythrin is not, spillover spreading pushes the upper edge of the negatives higher, and 8 % of cells now lie above 300. Those cells carry no phycoerythrin, so in the fully stained tube the same 8 % appear positive for a marker they lack. A boundary set on the FMO at 1,200, below which 99 % of its cells fall, leaves only 1 % of the FMO's cells above it. The values are illustrative, chosen for this guide.

30 — A boundary set on the wrong control
UNSTAINED FMO FULL STAIN 0 300 1,200 10⁴ 10⁵ PE, LOG-LIKE DISPLAY illustrative values boundary on unstained: 99 % below boundary on FMO: 99 % below 8 % above 300: cells that carry no PE, called positive
ReadoutFocal detail
Plate 30 — A boundary set on unstained cells is too low for a stained tube: here 8 % of cells that carry no phycoerythrin lie above it because of spillover spreading. A boundary set on the fluorescence-minus-one control leaves only 1 % of such cells above it.

FMO controls are needed most for dim markers, for markers whose level varies continuously from cell to cell, and for detectors that receive heavy spread. They are needed least where positive and negative cells form clearly separated groups, which is why laboratories choose them channel by channel rather than for every marker of every panel.

Why isotype controls usually fail

Antibodies come in structural classes and subclasses, which differ in the part of the molecule that does not bind the target; an antibody's class and subclass are its isotype. An isotype control is an antibody of the same class and subclass as the test antibody, carrying the same dye, but made to bind a target that the cells do not have. The idea is that it binds nonspecifically as the test antibody does, so its signal shows how much of the test signal is background. The 2006 guideline stated the intent: such controls "are meant to account for nonspecific staining" of a given isotype conjugated to a given dye.

In practice the match is poor. Two antibodies of the same subclass differ in structure, in the number of dye molecules attached and in how they were purified, and each conjugate has its own background. The authors of the 1998 editorial reported that isotype controls do not stain the same number of events as the test antibody stains nonspecifically, and they argued that the widespread use of isotype controls should be questioned. The 2006 guideline added that, on their own, isotype controls "do not account for fluorescence spillover from other channels."

The evidence against isotype controls is a guideline and an editorial with observations, not a controlled study, and some reagent suppliers still recommend isotype controls from the same supplier. Their remaining use is narrow. The 2006 guideline describes one case: an isotype control in the channel of interest, together with all the panel's other antibodies. That combination takes care of spillover but still does not match the test antibody's own background, and it does not set a positive boundary by itself.

A percentage positive depends on the boundary

A result such as "12 % CD38-positive" is a property of the cells, the reagent and the boundary together. If the boundary was set on an unstained or isotype sample, the figure includes spillover spreading and nonspecific binding of unknown size. Without the control named, a percentage of dim positive cells cannot be compared with another laboratory's.

Biology often gives the best negative

The most faithful control is often inside the tube. Some cells in the same sample belong to a type known not to carry the marker, and they have experienced the same staining, the same spillover, the same antibody and the same instrument settings. T cells do not carry CD19, the B-cell protein of Chapter 5, so in a tube that stains B cells for CD19 the T cells give the negative directly. The 2006 guideline made the same point for stimulation assays, which measure how cells respond to a deliberate stimulus. In such an assay, the unstimulated sample "usually provides the best means" of telling positive from negative.

Clinical immunophenotyping reads markers this way. ICSH and ICCS are international bodies for standardization in hematology and for clinical cytometry, and their 2013 recommendations describe abnormal intensity, bright or dim, relative to the normal counterpart of the abnormal cells. The 2018 international guidelines for chronic lymphocytic leukemia describe three markers as characteristically dim compared with normal B cells: surface immunoglobulin, the antibody that a B cell carries on its own surface, and the proteins CD20 and CD79b. That judgment is easiest when normal B cells are present in the same sample. Chapter 21 shows how a report states such patterns. Dead cells, which bind many antibodies nonspecifically, are excluded beforehand with a viability dye (Chapter 12).

For vesicles, the first question is whether an event is a vesicle at all, and the controls change accordingly. The reporting framework for vesicle cytometry lists controls for this question. Buffer is the salt solution, held at a steady pH, in which the sample is diluted. The controls include buffer alone and buffer with the labeling reagents, because antibodies and dyes form small aggregates that cross the threshold like particles. They also include a detergent step that dissolves membranes, so that events which survive it cannot be vesicles. Chapter 20 takes these controls further.

Titration sets how far positive sits from negative

How much antibody to add is itself a control on the boundary. Too little leaves binding sites empty, so positive cells are dim and their brightness depends on how many cells are in the tube. Too much raises nonspecific binding, so the negatives brighten and spread. Between the two lies an amount at which positives are well separated from negatives, and titration finds that amount by staining the same cells with a series of antibody amounts and measuring the separation at each.

31 — Titration chooses separation, not brightness
10² 10³ 10⁴ 0 50 100 150 MEDIAN FLUORESCENCE (LOG) STAIN INDEX 1/32 1/16 1/8 1/4 1/2 1 2 ANTIBODY AMOUNT, × A REFERENCE (DOUBLING STEPS) POSITIVE MEDIAN STAIN INDEX, RIGHT AXIS NEGATIVE MEDIAN too little: dim, depends on cell number too much: negatives brighten best separation schematic
ReadoutFocal detail
Plate 31 — As antibody is added, positive cells brighten and plateau while negatives brighten at high amounts. The stain index peaks before the brightest staining, and that amount is the one to use.

The measure of separation is the stain index of Chapter 5, which rewards a narrow negative population as much as a bright positive one. The best concentration is therefore not always the one that makes positive cells brightest. A titration holds only for the reagent lot, cell type, staining volume and instrument it was done on. The spectral guideline of Chapter 10 advised that new lots of tandem dyes be titrated again, and a manufacturer's staff were among that guideline's authors.

Name the control for every boundary

In an assay specification, state for each marker which control sets its positive boundary, with acceptance limits for that control. The control may be an FMO, an internal negative population, an unstimulated sample or, for bright and well-separated markers, the stained tube itself. Record titration data for each reagent and repeat it after a change of lot, staining volume or cell type. For vesicles, include buffer-only, reagent-only and detergent controls in every run.

+ What this chapter established
  • Background comes from autofluorescence, spillover spreading and nonspecific binding, so a positive signal is always a comparison with a control.
  • An unstained sample sets the boundary too low for a stained tube; a fluorescence-minus-one control includes the spillover spreading and so accounts for it, though not for the missing antibody's own background.
  • Isotype controls rarely match the test antibody's background and do not account for spillover; cells known to lack the marker, in the same tube, are often the best negative.
  • Titration chooses the antibody amount that maximizes separation, not brightness, and must be repeated for new lots, especially of tandems.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.