§ Glossary

Glossary.

Terms are defined as they are used in this guide.

510(k)
A US premarket notification showing that a device is substantially equivalent to a legally marketed device; a cleared 510(k) allows marketing.
Abort (electronic)
An event discarded by the electronics because a second pulse arrived while the first was still being processed.
Analog-to-digital converter
The circuit that samples a detector's voltage many millions of times a second and turns it into numbers.
Analyte-specific reagent
In the US, a single reagent such as an antibody sold for laboratories to build their own tests, without claims of performance.
Antibodies bound per cell
A calibrated unit giving the number of labeled antibodies bound to a cell, based on a biological standard such as CD4 on T cells.
Antigen density
The number of copies of a protein on a cell, which sets how many antibodies, and how much dye, a cell can carry.
Area, height and width
The three numbers taken from each pulse: the integrated signal, its peak and its duration.
Autofluorescence
The faint fluorescence of unlabeled cells, from molecules such as flavins inside them.
Avalanche photodiode
A semiconductor detector with high efficiency in the red and near-infrared and modest internal gain.
Backbone markers
Markers placed on the same dyes in every tube of a panel so that the population of interest lands in the same place in each tube.
Bandpass filter
A filter that passes a stated band of wavelengths; 530/30 passes 515 to 545 nanometers.
Calibration
Relating an instrument's arbitrary units to physical units, such as molecules, square nanometers or volume, through a reference material and a model.
Carryover
The fraction of one sample that appears in the next sample run on the same instrument.
CD (cluster of differentiation)
The international numbering of cell-surface proteins recognized by antibodies, such as CD4 or CD19.
CLIA
The US Clinical Laboratory Improvement Amendments, which regulate clinical laboratories, including the tests they develop themselves.
Clonality
Descent of a population of cells from one cell; for B cells, shown by a single light chain, kappa or lambda.
Clustering
Grouping events by similarity with an algorithm instead of hand-drawn gates.
Coefficient of variation (CV)
The standard deviation of a population divided by its mean, as a percentage; a robust CV uses percentiles instead.
Coincidence
Two or more particles in the laser beam within one processing window, recorded as one event or aborted.
Compensation
Subtracting each dye's spillover from the other detectors, using values measured on single-stain controls.
Control
A sample, or a population within a sample, that supplies the comparison on which a positive boundary is set.
Core stream
The thin thread of sample at the center of the sheath flow that carries particles in single file through the beam.
De Novo
The US route for a new type of low-to-moderate-risk device with no predicate; a granted request creates a new classification.
Dichroic mirror
A mirror that reflects light on one side of a stated wavelength and transmits the rest, dividing collected light by color.
Dimmest resolvable signal
The smallest signal whose population still separates from the negatives by two standard deviations each; 4 × (1 ÷ Q + background spread).
Doublet
Two cells passing together and recorded as one event; often removed by comparing pulse area with height.
Drop delay
In a sorter, the time between a cell crossing the laser and its arrival at the point where the jet breaks into drops.
Dynamic range
The span from the dimmest to the brightest signal measured within a stated deviation from linearity.
ERF
Equivalent reference fluorophore: the number of molecules of a stated reference dye in solution giving the same signal as one bead.
Event
A pulse that crossed the threshold and was recorded as one row in the data file.
Extracellular vesicle
A particle released from cells, delimited by a lipid bilayer, that cannot replicate on its own; most are smaller than 200 nanometers.
Fc receptor
A protein on some white cells that holds antibodies by their constant tail, causing binding unrelated to the target.
FCS
Flow Cytometry Standard, the file format for cytometry data; version 3.1 stores keywords and list-mode data and requires the data to be stored uncompensated.
Fit for purpose
Validation tailored to the intended use of the data, as set out in CLSI H62.
Flow cell
The quartz cuvette in which the laser meets the stream in most analyzers.
Fluorescence-minus-one (FMO) control
A sample stained with every antibody of a panel except one, used to set that marker's positive boundary.
Forward scatter
Light scattered by a particle at small angles around the laser's direction.
Full-spectrum (spectral) cytometer
An instrument that records each particle's emission across many detectors and separates dyes by unmixing.
Gate
A boundary drawn on a plot that selects events for the next step of analysis.
Gating hierarchy
The sequence of gates, from time and doublets through viability and lineage to subsets, behind every reported percentage.
Gating-ML
A standard for describing gates and transforms exactly, outside the FCS file.
Hydrodynamic focusing
Narrowing the sample into a thin core inside a faster sheath flow so that particles cross the beam one at a time.
Imaging flow cytometer
An instrument that records a picture of each cell as it flows.
In-house device
In the EU, a test made and used within a health institution under Article 5(5) of the IVDR.
Interface
A boundary between two parts of a system, with parameters both sides must agree on.
Interrogation point
The fixed point at which the laser crosses the stream and each particle is measured.
Intrinsic brightness
A dye's molar absorption coefficient multiplied by its quantum yield.
Isotype control
An antibody of the same class and dye as the test antibody but against an absent target; rarely a good match for its background.
IVDR
Regulation (EU) 2017/746 on in vitro diagnostic medical devices, applicable since 26 May 2022.
Jet-in-air
A sorter design in which the laser meets the free stream after it leaves the nozzle.
Laboratory-developed test
A test made and used within one laboratory; in the US regulated under CLIA.
Laser delay
The interval the electronics wait between a particle's pulse at one laser and its pulse at the next.
Limit of detection
The smallest amount an assay detects with stated reliability; in rare-event cytometry set by convention as 20 events in the final gate, a cluster-size rule rather than a statistical limit.
Linearity
The proportionality of an instrument's output to its input over a stated range.
List mode
Data stored as one row per event and one column per parameter.
Logicle
A display transform that is linear near zero and logarithmic at high values, so compensated data near and below zero can be shown.
Lower limit of quantitation
The smallest amount measured with acceptable precision; in rare-event cytometry often set as 50 events.
Mass cytometry
Cytometry in which antibodies carry metal isotopes that are counted by a mass spectrometer after the cell is vaporized.
MESF
Molecules of equivalent soluble fluorochrome: the number of free dye molecules in solution giving the same fluorescence as a particle.
Mie theory
The physics of light scattering by spheres of any size, used to relate scatter to diameter and refractive index.
MIFlowCyt
The minimum information standard for reporting a flow cytometry experiment, published in 2008.
MIFlowCyt-EV
The extension of MIFlowCyt for vesicle measurements, published in 2020.
Molar absorption coefficient
How strongly a molecule absorbs light of a given wavelength; also called the extinction coefficient.
Numerical aperture
A measure of the width of the cone of light a lens accepts, which sets the fraction of emitted light collected.
Photomultiplier tube
A detector in which one photoelectron is multiplied about ten million times by a chain of electrodes.
Pseudoreplication
Treating measurements that are not independent, such as events from one sample, as replicates.
Pulse
The rise and fall of a detector's signal as a particle crosses the beam.
Q and B
A channel's detection efficiency, in photoelectrons per unit of dye, and its background; together they set the dimmest resolvable signal.
Qualitative, quasi-quantitative and quantitative assays
Assays that report a category, a number without a calibration standard, or an absolute value traceable to standards.
Quantum yield
The fraction of absorbed photons that a molecule emits as fluorescence.
Reference material
A material whose value is known with a stated uncertainty, used for calibration.
Refractive index
How much a material slows light compared with a vacuum; scatter depends on the contrast between particle and medium.
Residual disease (MRD)
Leukemic or myeloma cells remaining after treatment, measured by flow cytometry at frequencies down to about 2 in a million.
Sheath
The clean fluid that surrounds and narrows the sample stream.
Side scatter
Light scattered by a particle at wide angles, collected by the same lens as fluorescence.
Signature
The pattern of one dye's light across all the detectors of a spectral instrument.
Silicon photomultiplier
An array of tiny photodiodes, each firing on a single photon, with high gain.
Single-platform count
An absolute count made in one tube on one instrument relative to a known number of counting beads.
Single-stain control
A sample stained with one dye of a panel, used to measure that dye's spillover or spectral signature.
Spillover
The fraction of a dye's signal in its own detector that appears in another dye's detector.
Spillover spreading
The widening of compensated populations caused by the counting noise of spilled light.
Stain index
The difference between positive and negative medians divided by twice the spread of the negatives.
Standardization
Making instruments read the same reference material alike, without assigning physical units.
Stokes shift
The difference between the wavelengths at which a dye absorbs and emits most strongly.
Swarm detection
Many particles below the detection limit, illuminated together, recorded as one event.
Tandem dye
Two dyes joined so that the first passes its energy to the second, which emits at a longer wavelength.
Target values
Bead intensities that every instrument in a standardized network is adjusted to reach.
Threshold
The level a chosen detector's signal must exceed for the electronics to record an event.
Titration
Staining with a series of antibody amounts to find the one that best separates positive from negative cells.
Trigger
The detector whose signal is compared with the threshold to decide that a particle has arrived.
Unmixing
Finding the amount of each dye whose combined signatures best fit a particle's readings across many detectors.
Validation
Showing that an assay serves its intended use, with precision, sensitivity, agreement and stability established.
Verification
Showing that an instrument or design meets its specification.
Viability dye
A dye excluded by intact membranes that marks dead cells so they can be removed from analysis.