How a flash of light becomes a number.
+ The questionHow does a pulse lasting a few microseconds become a height, an area and a width, and who decides which pulses count as events at all?
The shape of a pulse
As a particle moves through the laser spot, the light it sends to a detector rises, peaks and falls. The detector turns that light into a current that follows the same course, and the electronics see a pulse: a bump in the signal lasting from under a microsecond in a fast sorter to a few microseconds in an analyzer. The pulse's shape is set by the profile of the beam along the direction of flow and by the size of the particle, because a large cell takes longer to enter and leave the beam than a small one.
Three numbers are taken from each pulse. One manufacturer's technical bulletin defines them simply: the height is the peak intensity of the pulse, the width is the time the particle takes to pass through the interrogation point, and the area is "calculated by integration and represents the whole pulse." Area grows with the total light collected during the passage, height with the brightest moment, and width with the time spent in the beam. Most software attaches the suffixes -A, -H and -W to these values, but FCS 3.1 does not define them; they are naming habits. MIFlowCyt requires a report to say which pulse measurement it used.
Telling one particle from two
The three measurements together reveal clumps. Two cells stuck together, or passing one directly behind the other, produce a pulse that is longer than one cell's and carries about twice its area. The pulse's height rises much less, so area and height no longer keep the ratio that single cells show. On a plot of area against height, single cells fall along a diagonal line and pairs fall above it. A plot of width against height does the same, with pairs at greater width. Drawing a gate on such a plot to keep only single particles is called doublet discrimination, and it is a routine first step of analysis.
Doublet discrimination has limits. The sorting guide of Chapter 6 notes that pulse processing reduces doublets but "can never completely eliminate" them. The method also works only when a particle is comparable to or larger than the height of the beam. Two small particles side by side across the stream, or several vesicles inside the beam at once, produce a pulse that looks like one particle's. A doublet of two cells that is not removed can look like a single cell positive for both their markers. Immunophenotyping, which identifies kinds of cell by the combinations of proteins they carry, looks for exactly such populations.
The threshold decides what exists
Most of the time, nothing is passing the beam, and the detectors register only background. The electronics need a rule for deciding when a particle has arrived. The rule is the trigger and threshold: one chosen detector, and a level that its signal must exceed (Chapter 1). When the signal exceeds that level, the electronics open a window, measure the pulse in every detector and record an event. Everything below the threshold is discarded and leaves no record.
The choice of trigger shapes the data as much as any other setting. A threshold set too high discards real but dim particles, and one set too low fills the file with noise events and increases the overlap of pulses, which leads to more electronic aborts (Chapter 6). Through its $TR keyword, FCS 3.1 can record one trigger detector and its threshold, in channel numbers, the instrument's own arbitrary scale. More elaborate triggers, such as an event declared when either of two detectors crosses its level, have no standard representation.
For vesicles, the choice of trigger is decisive. Their scatter is so weak that a scatter trigger set above the noise misses most of them. Triggering on fluorescence instead declares an event when a labeled particle glows, and it changes the count dramatically. In a 2016 study of plasma vesicles, fluorescence triggering raised the number detected 15-fold for vesicles from red cells, 40-fold for those labeled with annexin V and 75-fold for those from platelets. Annexin V is a protein that binds phosphatidylserine, a membrane lipid that some vesicles expose on their outer surface.
MIFlowCyt asks for the threshold to be stated, and its extension for vesicle cytometry asks for the trigger channel and threshold, preferably in calibrated units rather than in channel numbers. A scatter threshold can then be given as a cross-section in square nanometers, a measure of how much light a particle scatters into the detector, and a fluorescence threshold as a level in calibrated molecules (Chapter 17).
From current to digits
Modern cytometers are described as digital, but the sorting guide states plainly: "A fully digital flow cytometer does not exist. All cytometers, including sorters, are hybrid analog–digital systems." The detector's current is converted to a voltage and amplified by analog electronics. An analog-to-digital converter then samples the voltage tens of millions of times a second, typically at 10 to 100 megahertz according to the same guide. Software computes height, area and width from the samples.
A converter with n bits distinguishes 2ⁿ levels: 16 bits give 65,536. The number of decades is the span from the smallest step to the largest value, counted in powers of ten: 16 bits span about 4.8 decades, 18 bits about 5.4 and 22 bits about 6.6. One manufacturer quotes "22-bit 6.5 log decades," consistent with that arithmetic. But a linear converter has very few steps at the bottom of its range: in an 18-bit range, the first decade, from 1 to 10, contains only nine integer steps. A pulse lasting 3 microseconds yields about 30 samples at 10 megahertz and about 180 at 60 megahertz. The figures are derived for this guide.
The bit depth printed in a specification does not always describe the converter. Some instruments combine two 16-bit converters per channel to produce data with an effective 23-bit range. An area stored as a 32-bit number is a sum of many samples, not a 32-bit conversion. What matters to the measurement is the noise and resolution at the bottom of the scale, where dim particles fall.
Log amplifiers and logarithmic displays
Older instruments did not digitize a linear signal. They passed it through logarithmic amplifiers that compressed four decades of signal into 1,024 channel numbers, which made bright and dim populations visible on the same axis. Analog log amplifiers were never perfect: one manufacturer's patent notes that their transfer function "can contain 'ripples' that may distort the input signal." In the 2018 comparison of 23 instruments in Chapter 8, the study's estimates of Q and B were less precise on instruments with log amplifiers than on linear ones: 85 % of their standard errors were below 10 %, against 90 % below 3 % on linear instruments. The comparison's authors included staff of a cytometer manufacturer and of an analysis-software company.
Linear digitization removed the hardware but not the habit of viewing data on a logarithmic axis, and the habit causes a problem once data are compensated. Compensation subtracts one dye's light from another dye's detector, and for cells that carry none of a dye it leaves values near zero and below zero, which a logarithmic axis cannot show (Chapter 10). The 2006 paper of Chapter 8 described how log displays "often show a peak above the actual mean or median of the population with a pileup of events on the baseline," a distortion known as the log artifact. The paper's remedy is the logicle, or biexponential, display, which is linear near zero and logarithmic far from it. Chapter 12 shows that this display is a choice that can change how a plot looks without changing the data.
Height and area of the same pulse can rank particles differently. The threshold decides which particles exist at all, and the transform, the scale on which the data are displayed, decides how populations look. A number reported without these three facts cannot be reproduced on another instrument or in another laboratory.
In an instrument specification, state the sampling rate, the converter resolution, the electronic noise and the baseline stability in units of the detector signal. Test linearity down to the dimmest populations the instrument is meant to measure. In an assay, fix the trigger channel, threshold and pulse measurements, and record them. For small particles, express the threshold in calibrated units so that it means the same on another instrument.
- Each pulse yields a height, an area and a width; the suffixes are conventions, and a report must say which measurement it used.
- Area against height separates single particles from doublets when particles are comparable to or larger than the beam height, but never perfectly.
- The trigger and threshold decide what becomes an event; for plasma vesicles, fluorescence triggering found 15 to 75 times more particles than scatter triggering.
- Cytometers are hybrid analog–digital systems; linear converters, which have few steps at the bottom of the scale, replaced log amplifiers, and logicle displays, linear near zero, show the compensated values near and below zero that logarithmic axes cannot.
Let's Talk