Why the lasers cross the stream at different places.
+ The questionHow does an instrument with several lasers choose and shape them, and know which flash of light belongs to which particle?
Why lasers
A cytometer needs a light source with three properties. Its light must focus to a spot between about 5 and 22 micrometers high, and it must deliver enough power to make a dim particle glow measurably in a microsecond. It must also have a single, narrow color, so that a filter can block it completely and pass the fluorescence beside it. A laser does all three. Early instruments used mercury arc lamps, and some still use light-emitting diodes for particular purposes, but the laser has been the standard source since the 1970s.
The colors in use come in families set by the dyes they excite. A 2015 review of near-infrared lasers in cytometry lists the laser lines in use. They run from ultraviolet at 355 nanometers through violet diodes at 405 to blue at 488, a line originally produced by argon-ion lasers. Green and yellow lines at 532 to 561 nanometers suit phycoerythrin and the red fluorescent proteins, labels that cells can be engineered to make themselves. Red lines run from 633 to 640 nanometers, and near-infrared diodes from 660 to 730. Some current instruments go further into the infrared. One manufacturer states that its current five-laser analyzer carries lines at 355, 405, 488, 561 and 637 nanometers, at powers between 60 and 200 milliwatts.
Matching lines to dyes
A laser excites a dye only if the laser's wavelength falls within the dye's absorption band, the range of colors that the dye absorbs (Chapter 3). Each dye can therefore be paired only with the lasers it absorbs. Phycoerythrin absorbs strongly in the green and yellow, so it is excited well by a 561-nanometer laser and less well, though still usefully, by the blue line at 488. Allophycocyanin needs a red laser. The polymer dyes of Chapter 5 made bright labels available for the violet and ultraviolet lasers.
Adding lasers is one of the most effective ways to add colors. Dyes excited by different lasers can emit at similar wavelengths and still be told apart, because their light arrives at the detectors at a different moment and through a different collection path, as the rest of this chapter shows. A cytometer with five lasers can therefore use many more dyes, with less overlap between them, than one with a single laser and the same number of detectors.
Shaping the spot
The beam is focused to an ellipse rather than a circle. Its short dimension lies along the direction of flow and sets how long each particle is illuminated, and therefore how close behind one another two particles can pass and still be resolved as two (Chapter 6). Its long dimension lies across the stream and gives tolerance: a particle slightly off the axis of the core still receives nearly the full intensity. Manufacturers quote spots of 22 by 66 micrometers on an older analyzer, 9 by 65 on a newer one and 5 by 80 on another. Some instruments shape the beam so that its intensity is flat across the center rather than Gaussian. The design intent is to make the illumination less sensitive to where a particle crosses the beam, but this guide found no published comparison that measured the benefit.
The mean intensity of a focused beam is its power divided by the area of the spot. Twenty milliwatts focused into an ellipse 9 by 65 micrometers gives about 4 × 10⁷ watts per square meter, roughly forty thousand times the intensity of sunlight at the Earth's surface. The 1997 study of Chapter 3 described fluorescein, PE and APC as considerably saturated and bleached in standard cytometer conditions. Its clearest case, APC, gave only about eleven times more light for fourteen times more power, at a mean intensity of about 10⁹ watts per square meter. On these figures, typical analyzers run one to two orders of magnitude below that intensity. A 50-milliwatt source in a 5-by-80-micrometer spot comes within a factor of ten of it, and sources of 100 to 200 milliwatts focused to a similar spot approach it. Fluorescein and PE lost signal within microseconds in the same study, so a lower intensity does not guarantee a linear response. The powers quoted by manufacturers are measured at the laser, so the intensity at the stream is lower still. The comparison is derived for this guide.
The intensity of the spot also bears on how much laser power to use. Jet-in-air sorters move particles faster than analyzers, so each particle spends less time in the beam, and the sorting guide of Chapter 6 notes that "higher laser powers are generally required to illuminate the cell with the same number of photons." For dyes that already saturate, however, extra power adds little signal, and it can bleach a dye partway through the beam. The best power for a measurement therefore depends on the dye, the flow speed and the spot.
One stream, several beams
An instrument with several lasers could focus them all on the same spot. Most do not. They place the beams at separate points a short distance apart along the stream, so that each particle passes through one laser, then the next. The light produced at each point is collected by its own optics and sent to its own set of detectors. A dye excited only by the red laser then produces a signal only at the red point, even if its emission overlaps with that of a dye excited at the blue point.
Separating the beams in space separates them in time. A particle crosses the first beam, which usually triggers the event, and reaches the next one microseconds or tens of microseconds later. The electronics must assign the second pulse to the same event by waiting a fixed interval, the laser delay, before looking for it. The delay depends on the distance between the beams and on the speed of the stream, which in turn depends on the sheath pressure. The delay must therefore be measured with beads and checked as part of routine setup.
A wrong delay does not stop the instrument working. It makes the second laser's detectors integrate the wrong part of the pulse, or the tail of the previous particle. A note from one core facility, a laboratory that runs shared instruments, records a delay corrected from −93 to −91 microseconds. Before the correction, a population that should have been almost entirely positive for two markers read 89.4 % double-positive; after it, 97.6 %. Two microseconds is a large part of a pulse at analyzer speeds. The same note records that the affected channel's fluorescence intensity read about 11 % lower before the correction, enough to push part of the population out of the positive gate.
Whether a fluorescence value belongs to the right particle depends on the delay setting. Yet the file standard FCS 3.1 has no keyword for laser delays, windows or sheath pressure. Some instruments write these settings into their own keywords, in units that the standard does not define. A data set exchanged between laboratories can therefore lose the one setting that decides whether its multi-laser values are valid.
For an instrument specification, state each laser's wavelength, its power at the interrogation point rather than at the laser, the spot dimensions and profile, and the spacing between beams. For daily operation, include verification of the laser delays with beads in the setup procedure, and record the delays with the data. Repeat the check after any change of sheath pressure, nozzle or flow cell. The reporting standard MIFlowCyt asks for the power measured "at the intersection of the light source beam with particles" where possible.
- Lasers give a narrow, intense, focusable color; current instruments use lines from about 355 to beyond 730 nanometers at tens to a couple of hundred milliwatts.
- Each laser excites only the dyes that absorb at its line, and adding lasers adds colors with less overlap than adding detectors to one laser.
- The spot is an ellipse, short along the flow and long across it; typical analyzers run one to two orders of magnitude below the intensity at which APC clearly saturated in a 1997 study, though some dyes lose signal at lower intensities.
- Separate beams along the stream are separated in time, and a laser delay assigns each pulse to its event; a delay off by two microseconds can turn 97.6 % into 89.4 %.
Let's Talk