20 — Vesicles counted

How many vesicles, and how big.

+ The questionA vesicle count means something only with its size window, its brightness threshold and its volume. What do calibrated cytometers report today, how far do laboratories agree, and how do their numbers compare with other methods?

A count is a window

The question "how many vesicles are in this plasma?" has no single answer. Chapter 19 showed reported concentrations spanning more than seven orders of magnitude. Methods that count every particle above about 70 to 100 nanometers give values near 10¹⁰ per milliliter, and their counts include lipoproteins. A calibrated cytometer reports about 10⁷ per milliliter for each cell type, because it counts only vesicles that carry a marker of that cell type, measure between 150 and 1,000 nanometers and shine above a stated brightness. Both numbers can be right, because they count different things.

Every vesicle count therefore needs three limits stated with it. The first is the size window: the smallest and largest diameter included, in calibrated scatter, with the refractive index that the conversion assumed (Chapter 17). The second is the dimmest label included, in calibrated fluorescence units. The third is the analyzed volume, measured rather than assumed. Without these three limits, the count describes the instrument as much as the sample.

54 — A count is a window
all particles > 70–100 nm ≈ 10¹⁰ per mL, lipoproteins included conventional cytometer > about 300 nm calibrated small-particle cytometer 150–1,000 nm + label threshold ≈ 10⁷ per mL per cell type 30 100 300 1,000 DIAMETER (nm, LOG) NUMBER (LOG) schematic distribution the lower limit decides the count EVERY COUNT NEEDS size window calibrated scatter, index assumed brightness threshold calibrated units measured volume
ParticlesReadoutFocal detail
Plate 54 — Methods that count every particle above about 70 to 100 nanometers find about 10¹⁰ per milliliter, lipoproteins included; a calibrated cytometer counting labeled vesicles from 150 to 1,000 nanometers finds about 10⁷ per cell type. Both are right about different windows.
Why the window dominates the count

Suppose vesicle numbers rise steeply toward small sizes, so that each halving of the lower size limit multiplies the count by about eight. A method that counts from 75 nanometers would then find about eight times as many vesicles as one that counts from 150, and about sixty-four times as many as one that starts at 300, in the same sample. This steepness is an illustration chosen for this guide, and real distributions differ from sample to sample and are poorly known at the smallest sizes. The point holds for any distribution as steep as those Chapter 19 described: moving the lower limit changes the count more than most biological differences do.

Reporting what was measured

The community's answer is a reporting framework called MIFlowCyt-EV. A working group of the societies for vesicles, cytometry and thrombosis published it in 2020. It extends the MIFlowCyt standard of Chapter 18 with what vesicle measurements need. In its own words, it "does not prescribe specific protocols": it asks that whatever was done be reported in a form that others can compare. The international guidelines MISEV2023, published in 2024, point to it for flow cytometry.

The framework's items begin with the sample before analysis and its preparation, and they include the controls of Chapters 11 and 19: buffer, reagent, unstained, isotype, single-stained, procedural, serial dilution and detergent controls. They also cover the instrument's calibration: the trigger channel and threshold, the flow rate, and the calibration of fluorescence and scatter. Results are reported as a calibrated detection range, as a concentration in the starting material and as brightness, preferably in standard units. The framework also states that a gating strategy based on polystyrene beads alone is not a sound basis for standardization. Chapter 4 reached the same conclusion from the history of bead mixtures.

How far laboratories agree

Comparisons between laboratories show calibration closing a very wide gap. A 2023 review summarizes two workshops of the international thrombosis society, in 2009 and 2010: they used a bead mixture to set size gates, and the counts proved comparable only among instruments of the same type. In 2017 a workshop of 44 laboratories found that counts of platelet vesicles in its two samples varied by 37 and 28 % between laboratories. The maker of the workshop's beads supplied them and employed two of its authors.

In 2018 the same society sent samples to 33 participants with 46 instruments: with gates set by bead diameter, counts varied by 139 % between them, and with gates set in vesicle diameter, worked out from scatter models, they varied by 81 %. That figure is for a gate of 1,200 to 3,000 nanometers, which 31 of the 46 instruments could detect. The largest comparison so far was published in 2025, by authors who included staff of the company that made the calibration beads used and a cofounder of a calibration company. They calibrated flow rate, fluorescence and scatter on each of 25 cytometers in 18 laboratories. The spread of counts between laboratories fell from 67 % before calibration to 25 to 31 % after it. The study measured that spread as a median absolute deviation, which shows how far typical counts lie from the middle count.

55 — How far laboratories agree
SPREAD BETWEEN LABORATORIES (%) 0 50 100 150 standardized clinical cell intensities (Ch. 16) 2009–10 bead gates: comparable only within an instrument type sample 1 28 % sample 2 37 % bead gates 139 % vesicle gates 1,200–3,000 nm 81 % uncalibrated 67 % calibrated 2017 workshop 2018, 46 instruments 2025, 25 cytometers, 18 labs 25–31 % calibrated: comparable for the larger, brighter vesicles measures differ by study: CV or median absolute deviation
ReadoutFocal detail
Plate 55 — Calibration closed a wide gap. Bead gates gave 139 % variation between 46 instruments in 2018; vesicle-diameter gates 81 %; full calibration of 25 cytometers in 2025 cut the median absolute deviation from 67 % to 25–31 %.

That spread is of the same order as the variation of about 30 % that fully standardized clinical laboratories show in the brightness of cells (Chapter 16). The two figures are not directly comparable, because one describes counts and the other, a coefficient of variation, describes intensities. The 2025 figure also applies only to vesicles inside the gates that the compared instruments shared. When the more sensitive half of the instruments were compared, the shared gate began at about 113 to 133 nanometers. When 18 to 19 of the 25 were compared, it began at about 240 nanometers, with higher fluorescence thresholds. Calibration has made vesicle counts comparable between laboratories for the larger, brighter fraction of vesicles, not for vesicles as a whole.

Reference ranges

Comparable counts made reference ranges possible. A reference range is the span of values found in healthy people, against which a patient's value is read. The authors of the 2025 study measured plasma from 224 healthy donors on one calibrated instrument. They counted vesicles between 150 and 1,000 nanometers and above stated fluorescence thresholds, and they described the results as the first reference ranges of cell-type-specific vesicles in human plasma. The median concentrations were about 20 million per milliliter for vesicles from red cells, 18 million for vesicles from white cells and 34 million for vesicles from platelets.

In 2023 the same group had described a plasma vesicle test sample. It is a ready-to-use material that resembles the particles of plasma and stays stable for 12 months at −80 °C, so that laboratories can check their whole procedure on a shared sample. The missing piece is a reference material whose size and concentration are traceable. A European project set targets for one: sizes of 50 to 1,000 nanometers, refractive indices of 1.37 to 1.42 and concentrations of 10⁹ to 10¹² per milliliter, and its candidates range from hollow organosilica beads and liposomes to recombinant vesicles. As of October 2026 this guide found no diagnostic test based on vesicle cytometry cleared, granted or approved by the FDA; the measurement is a research tool on its way to becoming a clinical one.

Other methods see other vesicles

Comparisons with other methods belong here, because they too are comparisons of windows. The 2014 study of Chapter 19 measured one pooled sample of urine vesicles by five methods: electron microscopy, resistive pulse sensing, particle tracking and two cytometers, one of them built for small particles. Its authors wrote that each technique "gave a different size distribution and a different concentration for the same vesicle sample." The concentrations ranged from 10⁴ to 10¹² per milliliter, and the authors concluded that the differences were "primarily caused by differences between the minimum detectable vesicle sizes." The size distribution peaked at 45 nanometers by electron microscopy, 75 by resistive pulse sensing and 95 by particle tracking. For the last two, the peak lay just above the method's own lower limit.

56 — Other methods see other vesicles
TEM: peak 45 nm, limit ≈ 1 nm RPS: peak 75 nm, limit 70–100 NTA: peak 95 nm, limit 70–90 dedicated FC: limit 150–190 conventional FC: limit 270–600 the peak sits near its own limit 10 30 100 300 1,000 DIAMETER (nm, LOG) one pooled urine sample, 2014 study; distributions schematic dashed boxes: lower detection limits
ParticlesFocal detail
Plate 56 — Measured by five methods, one vesicle sample gave five size distributions and concentrations from 10⁴ to 10¹² per milliliter. Where a method's lower limit lay inside the vesicle range, its distribution peaked just above that limit: the differences are windows, not errors.

The authors judged the cytometer built for small particles the most accurate at sizing reference beads, but they expected it to be less accurate on vesicles, because the refractive index of vesicles varies. They named its strengths as speed and the ability to measure several fluorescent labels on each particle. Particle tracking, the method behind most high concentration estimates, counts lipoproteins and other particles along with vesicles. No method sees every vesicle: each sees those above its own threshold, and the cytometer is the one that can say which cell they came from.

Concentrations from different windows cannot be compared

A count from particle tracking, a count from a conventional cytometer and a count from a calibrated small-particle cytometer describe different subsets of the same sample. Comparing them compares methods, not samples. So does comparing a published value with a laboratory's own value without matching the size window, the label threshold and the volume calibration.

Report every vesicle count with its window

Follow MIFlowCyt-EV for every report. State the size window in calibrated scatter with the refractive index assumed, and the fluorescence threshold in calibrated units. Give the measured analyzed volume and the dilution range in which counts were proportional, and report the results of the buffer, reagent and detergent controls. Run a shared test sample where one exists, and compare counts with other laboratories or methods only within a matched window.

+ What this chapter established
  • A vesicle count is defined by its size window, brightness threshold and measured volume; without them it describes the instrument as much as the sample.
  • MIFlowCyt-EV asks for controls, trigger, flow rate and calibrations in standard units, and judges gates based on polystyrene beads alone unsound as a basis for standardization.
  • Calibration of flow rate, fluorescence and scatter cut the spread between laboratories from 67 % to 25 to 31 %, for the larger and brighter vesicles.
  • Reference ranges for plasma vesicles by cell type, which their authors described as the first, appeared in 2025, and other methods report different counts because they see different windows.