+ Part I · The particle, the signal, the number
What a cytometer records, and why it records particles one at a time.
The 1969 instrument wrote one number for each cell, a brightness. A modern analyzer writes one number for every detector and every pulse measurement, and a result such as a percentage of CD4-positive cells lies several steps away from any of them. Chapter 1 separates the particle that crossed the laser, the light it sent to a detector and the number written into a file. Chapter 2 shows what measuring every particle on its own reveals that a measurement of the whole tube hides. Nothing here assumes more than school biology and physics.
What a flow cytometer actually measures.
+ The questionA report says that 12 % of the cells are CD4-positive. What did the instrument record to produce that number, and what did it never see?
A number with nothing visible behind it
A clinical report or a research figure states that 12 % of the cells in a blood sample carry a protein called CD4. CD4 is one of the numbered CD names that identify proteins on the surface of cells (Chapter 5). The report's statement sounds like an observation, as if someone had looked at the cells and counted. Nothing of the kind happened. The cells were first stained with an antibody that carries a fluorescent dye. An antibody is a protein of the immune system that binds to one particular target molecule, and a fluorescent dye is a molecule that absorbs light and gives some of it back in another color.
The stained cells were suspended in fluid and drawn into an instrument that pushed them, one at a time, through a focused laser beam. Each cell crossed the beam in a few microseconds. While it was there, it deflected some of the laser light, and its dye molecules gave some light back at a longer wavelength. Detectors behind lenses and filters turned those flashes into electrical pulses, and electronics turned each pulse into a few numbers. Software then decided which of those numbers belonged to the cells the report is about.
The instrument repeated that sequence for every particle it could detect, typically several thousand per second. One manufacturer specifies an analyzer of the kind found in clinical laboratories for up to 10,000 events per second. A stream moving at a few meters per second carries each particle through a beam about 5 to 22 micrometers high along the flow, so each crossing takes a microsecond or a few. The instrument took no image of any cell, touched none and identified none as a cell. The 12 % is a ratio of two counts, and each count is a count of recorded numbers that passed a series of decisions.
Three things that are easy to confuse
That sequence contains three different things, and this guide keeps them apart in every chapter.
The particle is the physical object: a lymphocyte, a red cell, a clump of two cells, a fragment of a dead cell, a calibration bead, a bubble, or a vesicle far smaller than any cell. Lymphocytes are one kind of white blood cell, and calibration beads are small spheres, usually of plastic, with a known size or brightness, made for setting up instruments. The instrument has no direct access to any of these particles.
The signal is the light the particle sent toward a detector during its few microseconds in the beam, and the electrical pulse the detector made from that light. The signal depends on the particle, but also on the laser's power, the angle and area of the light collected, the filters chosen, the detector's settings and the background light that arrives whatever the particle does.
The number is what the electronics computed from the pulse and wrote into a data file: the height of the pulse, its area or its width, in arbitrary units that mean nothing until a calibration converts them. Software then transforms these numbers for display and subtracts the overlap between dyes. The software also groups the numbers into populations by drawing boundaries called gates.
A cytometry result is trustworthy only to the extent that each step from particle to signal and from signal to number is known. The rest of the guide follows that chain. Part II covers how particles produce light, Part III how the instrument turns light into numbers, Part IV how numbers become populations, and Part V how the whole chain is engineered, calibrated and verified.
An event is not a cell
The data file does not contain cells. It contains events. Cytometry data files follow an international standard, the Flow Cytometry Standard, or FCS. Its version 3.1, which this guide describes, defines an event as "an ordered list of the cytometric measurements for one particle." A second standard, MIFlowCyt, sets the minimum information to report about a cytometry experiment. It calls an event "a unit of data representing one particle (e.g., a cell) as detected by flow cytometer." Both definitions say particle, and both give a cell only as an example.
The difference is not pedantic. Two cells that cross the beam stuck together, or one just behind the other, produce one pulse and one event. Two experienced operators wrote a guide to cell sorting, and they note that even with modern electronics "an event can still be 2 cells." Debris from dead cells, bits of membrane and particles in the reagents produce events too. One public example file, distributed with an open-source analysis package, holds the events recorded from stained, freeze-dried control blood cells. In that file, a "not debris" gate drawn by a human analyst removed 23 % of all recorded events before any analysis began, and a second gate for single cells removed a further 4 %.
For the smallest particles the difference becomes larger still. When vesicles far below the instrument's detection limit are present at high concentration, hundreds of them can be inside the beam at the same moment, and their combined light can cross the threshold as a single event. A 2012 study showed that one such "swarm" event could represent of the order of a hundred vesicles or more. Chapter 19 returns to swarms, because a swarm is one of the ways a cytometer can report a confident number about particles it cannot see.
What becomes an event at all is decided by a setting. The operator chooses one detector as the trigger and a level, the threshold, that its pulse must exceed. The trigger is often forward scatter, the light that a particle deflects through small angles in the direction of the laser beam (Chapter 4). Every pulse below the threshold is discarded and leaves no trace in the file.
FCS 3.1 provides an optional keyword, $TR, in which a file can record the trigger and threshold, and the standard states the rule: "When the threshold is exceeded, an event is declared." Some instruments record the threshold in their manufacturer's own keyword instead, and some do not record it at all. A count of events is therefore always a count of pulses that were bright enough in one chosen channel, the signal from one detector, and such a count depends on the sample and the settings together.
An event is a pulse that crossed a threshold. It may come from one cell, two cells, a fragment, or many vesicles at once, and a particle whose signal stayed below the threshold is not in the file at all. Any statement of the form "the sample contained N cells per microliter" depends on knowing which particles produced events, which did not, and what volume of sample was analyzed.
What the file does not contain
The numbers in an FCS file carry no physical units. A pulse area of 52,000 in a detector labeled for a green dye is not 52,000 of anything, because its value depends on the detector's voltage, the laser power and the optical path on that day and on that instrument. The standard provides an optional keyword, $PnCALIBRATION, for converting arbitrary units into a defined unit. One such unit is molecules of equivalent soluble fluorochrome, which expresses brightness as an equivalent number of free dye molecules (Chapter 17). Nothing requires the keyword, and most files do not carry it. Several properties that change the meaning of the numbers have no place in the standard at all: there is no keyword for the angle over which forward scatter was collected, and MIFlowCyt asks for that angle separately.
The file also leaves open what has already been done to the numbers. FCS 3.1 states that "the stored data is always uncompensated", meaning that the overlap between dyes has not yet been subtracted. Some instruments, however, mark the data in their files as compensated, in keywords of their own. A number on a plot axis has usually been transformed again for display. The value stored in the file, the value after compensation and the value displayed are three different numbers, and a report must say which it used.
A percentage of positive cells is the number of events inside one gate divided by the number inside another. MIFlowCyt requires the denominator to be stated: "percentage of events within the gate shall be provided specifically stating the denominator." For the CD4 example, the numerator is the events above a chosen boundary in the CD4 detector, and the denominator might be all lymphocytes, all white cells, or all events that passed the debris and single-cell gates. Each choice gives a different percentage from the same file. The boundary itself is a decision, made with the help of control samples (Chapter 11).
The vesicle problem
Most of the methods in this guide were developed for cells, which are typically 5 to 20 micrometers across and give bright, easily separated signals. This guide keeps returning to a much harder measurement. Extracellular vesicles are particles that cells release, enclosed by a lipid membrane and unable to replicate. A lipid membrane is a thin film of fat-like molecules, of the kind that encloses every cell. The International Society for Extracellular Vesicles, the field's scientific society, published its MISEV2023 guidelines in 2024. They describe small vesicles as "often" below 200 nanometers and note that measured size depends on the method. Vesicles are present in blood and other body fluids, and they are studied as possible markers of disease.
A vesicle of 100 nanometers has about a seventieth of the diameter of a lymphocyte 7 micrometers across. It scatters so little light that many instruments cannot distinguish it from the background. It carries only a few copies of any protein for an antibody to label, and it can arrive in such numbers that several vesicles are in the beam together. Measuring vesicles pushes every part of the chain from particle to number to its limit, and that is why vesicles serve as this guide's running case. The second case, in Part VII, is the diagnosis of leukemia, a cancer of the blood-forming cells. There the cells are large and bright, but the consequences of a wrong number are serious.
When a cytometry result will be compared, published or acted on, record with it the facts that make it interpretable. State which pulse measurement was used, height or area, and the trigger channel and threshold. Give the gates and the denominator of every percentage, say whether values were compensated and transformed, and name the instrument and its settings. MIFlowCyt lists these as minimum reporting requirements. In a regulated assay they become part of the fixed procedure, and in an instrument they become requirements on what the software must record.
- A cytometer pushes particles one at a time through a laser and records the light each one sends to its detectors, typically thousands per second, each in a few microseconds.
- The particle, the signal and the number are different things; every result depends on the chain that links them.
- An event is a pulse that crossed an operator-set threshold: it may be a cell, two cells, debris or a swarm of vesicles, and particles below the threshold leave no record.
- The file holds arbitrary units without physical meaning unless calibrated, and every percentage depends on its gates and its stated denominator.