Cold, wiring and the signal chain of a qubit.
In 1984 and 1985, John Clarke's group at the University of California, Berkeley, set out to show that a whole electrical circuit could behave like an atom. John Martinis was a graduate student in the group, and Michel Devoret a postdoctoral researcher. The first attempt was wired the way such experiments usually were, and in Martinis's words it "completely failed." The group then filtered every line, above all at microwave frequencies, and "the data made perfect sense." At 18 to 28 millikelvin, a few hundredths of a degree above absolute zero, a circuit absorbed microwaves in discrete steps. The paper appeared on 7 October 1985, forty years to the day before the Nobel Prize in Physics honored it. Part VI treats a quantum computer as that experiment already was, an instrument whose hardest problem is keeping the world out while signals go in and come back. Chapter 17 follows the cold, Chapter 18 the signal chain of one qubit, and Chapter 19 the machines built from ions, atoms, photons and spins.
Inside the refrigerator.
+ The questionWhy does a superconducting quantum computer need a refrigerator colder than space, and what does each stage do?
A chandelier of stages
A dilution refrigerator wired for qubits hangs as a stack of round metal plates, each colder than the one above. A 2019 study described such a system at ETH Zurich, built on a commercial Bluefors refrigerator. Its plates ran at about 35 K, 2.85 K, 0.88 K and 0.082 K, and the bottom plate at 6 millikelvin, about six thousandths of a degree above absolute zero. The quantum chip sits in a shielded can on the coldest plate. Every cable from the room-temperature electronics runs down through the plates, and each plate exists to remove heat that would otherwise reach the stage below. The plates also serve as shields, each blocking the infrared glow of the warmer plate above it.
The coldest plate is called the mixing chamber, after the process that cools it. Below about 0.87 K, a mixture of the two isotopes of helium, helium-3 and helium-4, separates into two layers, one rich in helium-3 and one dilute. Helium-3 atoms crossing from the rich layer into the dilute one absorb heat, much as evaporation cools a liquid, and this dilution gives the refrigerator its name. Helium-3 is pumped out of the dilute side at a warmer plate, called the still, and that keeps the atoms crossing. A mechanical cooler called a pulse tube precools everything to about 3 K, so the refrigerator needs no bath of liquid helium. Heinz London proposed the method in the early 1950s, and it was first made to work in 1964 at the Kamerlingh Onnes Laboratory in Leiden.
The cooling power at the bottom is tiny. The 2019 system could absorb about 19 microwatts at 20 millikelvin, about two hundred-thousandths of a watt. Bluefors lists more than 30 microwatts at 20 millikelvin for one of its current models, and more than 14 for a smaller one. Its largest platform, built for over a thousand qubits, is listed at more than 90 microwatts, split across three cooling units. Every design choice below the still is a decision about how to spend those microwatts.
Why so cold
A qubit at 5 GHz stores the energy of one microwave photon, and surroundings above absolute zero hold such photons, more of them the warmer they are. Their average number in each mode, a single pattern of oscillation at a given frequency, follows a simple formula from the physics of heat. The number falls steeply once the temperature drops below the photon's energy expressed as a temperature, which at 5 GHz is about 0.24 K. A stray photon absorbed by a qubit in 0 flips it to 1, so the qubit's surroundings must hold almost none.
The calculation divides 0.24 K by the temperature and raises e to that power. Subtracting 1 and dividing 1 by what remains gives the average number of photons in one 5 GHz mode. At room temperature, 300 K, the number is about 1,250. At 4 K it is about 16, and the cosmic microwave background, the 2.7 K glow of space, gives about 11. At 100 millikelvin it falls to 0.1, at 20 millikelvin to about 6 in a million, and at 10 millikelvin to about 4 in 10¹¹. A qubit at 10 millikelvin is therefore far quieter than one in deep space, by a factor of several hundred billion in stray photons.
The same arithmetic explains the 1985 filters. Cooling the chip is not enough if a cable carries photons down from a warmer stage. A line from room temperature delivers about 1,250 photons per mode at 5 GHz, and the only way to remove them is to absorb them along the way, at stages cold enough not to emit new ones. A line is anchored at a stage when it is clamped to that stage's plate so that it cools to the plate's temperature. The Berkeley experiment of 1985 worked at frequencies from 2 to 5.7 GHz. At the low end, a line anchored at 4 K would have carried dozens of photons per mode, against about five thousandths at the 18 millikelvin of the junction.
Every line brings heat and noise
Each cable that drives a qubit is a path for heat and for noise. The 2019 study measured the heat that one stainless-steel drive line conducts into each plate: about 45 milliwatts into the 35 K plate, 1 milliwatt into the 3 K plate, and about 13 billionths of a watt into the mixing chamber. Adding 66 lines raised the mixing chamber from 6.1 to 8.4 millikelvin. Pulses sent through the lines add heat of their own wherever they are absorbed.
Noise is removed with attenuators, components that absorb most of a signal and turn it into heat at their own stage. In the 2019 design, 20 decibels of attenuation, a factor of 100 in power, sat at each of three stages: the 3 K plate, the 0.08 K plate and the mixing chamber. The qubit therefore receives about a millionth of the power sent from the room, together with the leftover noise. The result was about 0.003 thermal photons at the chip, close to the design's target of a few thousandths. Putting all the attenuation at the 3 K plate would have left more than ten photons, because an attenuator cannot cool a signal below its own temperature: it replaces the photons it absorbs with photons of its own.
Filters complete the job. Absorbing filters block the infrared light that leaks down cables and breaks the electron pairs of the superconductor. In a 2011 study, Martinis's group found that a second light-tight shield around the chip improved its circuits about tenfold. Magnetic shields protect components that steer the readout signal. As in 1985, every gap that lets in warm light or stray fields shows up as qubits that do not behave as expected. In a June 2026 preprint, a team reported that better infrared shielding cut the rate at which stray energy disturbed its qubits by more than ten-thousandfold. The qubits were left at an effective temperature of about 17 millikelvin, measured from the qubits themselves.
A refrigerator's base temperature says little by itself. What reaches the qubits depends on how every line is attenuated, filtered and anchored at each stage, and a single badly anchored cable can warm the qubits it serves. Ask what the effective temperature of the qubits is, measured from their own excitation, not what the thermometer on the plate reads.
The wiring bottleneck
The cables also run out of room. By the ETH team's count, a 50-qubit processor needs 124 radio-frequency lines. The ETH team estimated that their refrigerator's heat budget would allow about 150 qubits if three times as many lines could be fitted, but there was no room for them. Space, not heat, was the binding limit. In 2023 IBM reported that its 1,121-qubit Condor processor used more than a mile of flat, dense cryogenic wiring inside a single refrigerator, a company figure. Cryogenic means built to work at very low temperatures. A year earlier IBM had shown Goldeneye, a concept refrigerator of 6.7 tonnes with about 1.7 cubic meters of cold space and room for six cooling units. IBM described it as a proof of concept rather than a production system.
One answer is to move the electronics into the cold. By Google's 2019 estimate, ordinary control electronics use about a watt per qubit, far more than a 3 K plate can remove. A controller placed in the cold must therefore be hundreds of times more frugal, and the chip Google showed that year ran at 3 K on less than 2 milliwatts. The authors of a 2026 preprint argue that the limit is now shifting to the 3 K stage itself, where the readout amplifiers sit. Each amplifier of the usual kind draws several milliwatts at the operating point its maker rates it for. Another 2026 preprint describes a machine large enough for RSA-2048. Its author took the estimate of about 900,000 physical qubits from a 2025 preprint by Craig Gidney of Google Quantum AI, assumed about 10,000 qubits per refrigerator and arrived at roughly 90 refrigerators working together.
For a system design, count the lines each qubit needs for drive, readout and tuning, and multiply by the passive heat per line at each stage. Add the heat of the pulses absorbed in the attenuators and of the amplifiers at 3 K. Compare each stage's total with its cooling power, with margin. The stage that runs out first, not the base temperature, sets how many qubits one refrigerator can hold.
- A dilution refrigerator cools a superconducting chip to about 10 millikelvin through a stack of plates at roughly 35 K, 3 K, 0.9 K and 0.1 K, with only about 20 to 90 microwatts of cooling at the bottom.
- At 5 GHz a qubit's surroundings hold about 16 thermal photons per mode at 4 K and about 4 in 10¹¹ at 10 millikelvin, which is why the cold is needed.
- Attenuators at several cold stages and absorbing filters remove the noise every cable brings down, leaving about 0.003 photons at the chip in one 2019 design, while attenuation at a single warm stage cannot.
- Wiring space, heat at 3 K and amplifier power limit how many qubits one refrigerator holds, and one 2026 analysis puts an RSA-2048 machine at about 90 refrigerators.