Four hours of training.

The people holding the camera had attested that they had never taken a picture of the inside of an eye. They worked in ten primary-care offices spread across the United States, and each had been given one standardized training session of four hours, with no refresher afterward. Their task was to photograph the retinas of adults with diabetes enrolled at those offices: two pictures of each eye on a Topcon NW400, one centered on the optic disc, where the optic nerve leaves the eye, and one on the fovea, the small pit at the center of sharp vision. About three participants in four needed no drops to widen the pupil. The other quarter were dilated and photographed again.

The pictures went to a program that had been locked before the first participant arrived: its decision rules were fixed and did not change during the trial. Within seconds it returned one of three answers: more than mild diabetic retinopathy detected, not detected, or image quality insufficient. Diabetic retinopathy is damage to the blood vessels of the retina caused by diabetes, and it can threaten sight. Nobody at the clinic interpreted the photographs. Between January and July 2017, 900 people enrolled.

Every participant who completed the study was photographed a second time, by photographers certified by the Wisconsin Fundus Photograph Reading Center, with a far more demanding protocol: four overlapping stereo pairs of each retina, and a cross-sectional scan by optical coherence tomography of the macula, the central area of the retina. Three experienced readers graded the photographs without seeing the program's answer, and their majority decided which participants the trial would count as having the disease; the macular scans were graded separately. With input from the FDA on its endpoints, the trial's designers had fixed its test before it began. They planned the number of participants around an expected sensitivity of 85 percent and specificity of 82.5 percent, and the program would pass only if the lower end of each result's confidence interval cleared a floor: 75 percent for sensitivity and 77.5 percent for specificity (Chapter 13). Sensitivity is the share of participants with the disease that the program flags; specificity is the share without the disease that it clears.

Of the 892 participants who completed the study, 819 had reference images that could be graded and a usable answer from the program. On the reference images, 198 of the 819 had more than mild disease, and the program flagged 173 of them, or 87.4 percent. It cleared 556 of the other 621, or 89.5 percent. Among participants whose reference images could be graded, it gave a usable answer for 96.1 percent. The developer, IDx, funded the trial. The trial's first author was Michael Abràmoff, an ophthalmologist at the University of Iowa who had founded the company, and the paper discloses the funding and his financial ties to IDx. On 11 April 2018 the FDA granted the system a De Novo authorization, its route for a new kind of device of low or moderate risk with nothing comparable on the market. The FDA described it as the first device authorized to provide a screening decision "without the need for a clinician to also interpret the image or results."

The trial measured one locked program, run by those operators on that camera, for those patients. It said nothing about the same program in a clinic whose computers send images differently, after the camera is replaced or the model retrained, or when someone sets out to make it fail. A device that will be used for years on people who were never in its trial needs evidence of a different kind, and each later version needs more of it.