27 — The evidence a software device files

Documents, validation and clinical evidence.

+ The questionWhat does each regulator ask to see before a software or AI device is sold, and what do the validation rules ask of the sites and makers that use software under good manufacturing practice?

Two levels of documentation

The FDA's guidance on premarket submissions for device software has been final since 14 June 2023. It replaced a guidance from 2005 and applies to every device with software, AI or not. In it, the FDA defines a documentation level: the amount of software evidence a submission must contain, which depends on what a failure could do. The level is Enhanced where a failure or latent flaw in the software "could present a hazardous situation with a probable risk of death or serious injury," judged before any risk control is counted. Otherwise the level is Basic.

Most of the file is the same at both levels: the evaluation of the level itself, a description of the software, the risk management file, the software requirements, an architecture diagram, the version history and the list of unresolved anomalies. At the Basic level, the testing evidence is a summary of testing and the protocol and report of system-level testing. The Enhanced level adds the detailed design specification, the configuration management and maintenance plans, and the protocols and reports of unit and integration testing. A maker can describe its development practices by declaring conformity to the clauses of IEC 62304 that the guidance names for its level, fewer at Basic than at Enhanced. The FDA has recognized the standard in full since 2019, but the guidance does not ask for a declaration to the whole standard, because the two schemes classify software differently (Chapter 4). Chapters 4 and 7 traced where each of these records comes from, and the guidance lists which of them the maker sends to the FDA.

53 — Basic and Enhanced documentation
could a failure or latent flaw present a hazardous situation with a probable risk of death or serious injury, judged before risk controls? the level is set by the worst case, not the design's defenses no yes BASIC ENHANCED SAME AT BOTH LEVELS documentation level evaluation · software description risk management file · software requirements architecture diagram · version history · unresolved anomalies SOFTWARE DESIGN SPECIFICATION — in the submission DEVELOPMENT AND MAINTENANCE PRACTICES summary, or IEC 62304 declaration + configuration management and maintenance plans TESTING summary + system-level protocol and report + unit and integration protocols and reports FDA guidance on device software content, final 14 June 2023 (replaced the 2005 guidance)
SpecificationCodeReadoutFocal detail
Plate 53 — The FDA sets software documentation at Basic or Enhanced by what a failure could do before any risk control counts. Most of the file is the same at both levels; Enhanced adds the design specification, development plans and unit and integration test records.

A submission draws its records from the maker's quality system. The FDA's quality rule for devices is now called the Quality Management System Regulation. Since 2 February 2026 it has incorporated the international standard ISO 13485:2016 by reference, which makes the standard's requirements part of the rule, and the agency's inspectors have stopped using their old inspection technique. The surgical energy guide in this library treats IEC 62304 and the documentation levels as items on a generator maker's compliance list. This chapter has followed them from the side of the software.

What the FDA asks of AI

For AI, the FDA's main document is a draft guidance on lifecycle management and marketing submissions for AI-enabled device software functions. The FDA published it in January 2025, and as of October 2026 it was still a draft. In it, the FDA proposes the sections a submission should contain beyond the general software file: a description of the device and its user interface, labeling, risk assessment, data management, the model's description and development, validation, plans for monitoring performance in the field, cybersecurity, and a public summary.

The draft's expectations read like Parts III and IV of this guide. Test data should be independent of training data, for example drawn from different clinical sites, and sequestered from the developers. Validation data should represent the intended population, and relying on a single collection site is generally not appropriate. As an example of what may be appropriate, the draft gives at least three geographically diverse US sites. Performance should be reported across subgroups such as sex, age, race, ethnicity, disease severity, site and acquisition equipment, at each operating point. The draft suggests, but does not require, a model card: a short structured summary of the model, its data and its performance, placed in the labeling. The final guidance on change control plans, the subject of Chapter 21, completes the FDA's set of documents for AI devices.

The EU and India

In the EU, the maker puts a software device's evidence into its technical documentation, which a notified body reviews. For clinical evidence, the coordination group's guidance of March 2020 sets out three components. The first is a valid clinical association: the software's output is associated with the clinical condition it targets. The second is technical performance: the software generates its intended output accurately and reliably from its input data. The third is clinical performance: the output is clinically relevant in accordance with the intended purpose. The maker keeps all three current through post-market evaluation. The guidance predates the recent growth in AI devices and does not mention AI.

Neither IEC 62304 nor the security standard of Chapter 23 is a harmonized standard under the EU's device regulation. A harmonized standard is one whose reference the EU has published for a regulation, and a device that meets it is presumed to meet the requirements the standard covers. Neither standard appears in the EU's list of harmonized standards for the device regulation as consolidated on 17 June 2026. A maker can still use them as the state of the art, the generally accepted current practice that the regulation requires makers to take into account, but meeting them gives no presumption of conformity.

In its final guidance of July 2026, India's regulator uses the same three components. For AI, it adds requirements of its own on datasets: the composition of training, validation and test data, including demographic distribution, geographic origin and clinical diversity; whether the data are real-world, from public databases or synthetic; and a justification where models were trained or validated outside India. A maker that shows performance on non-Indian data should supplement it with validation in representative Indian populations. The guidance lists IEC 62304, IEC 81001-5-1, ISO 14971 and ISO 13485 among the standards that may apply.

A draft guidance signals direction while the final one governs

Three drafts describe where regulators are heading: the FDA's AI-enabled device guidance, the FDA's draft on change control plans for all devices, and the EU's draft manufacturing annex on AI. Makers sensibly design toward them. A regulator still judges a submission against the final documents in force on the day it is filed, so the file needs to show which version of each document the maker followed.

Standards in motion

Standards change more slowly than guidance, and their status is easy to misreport. IEC 62304's first edition dates from 2006, with an amendment in 2015. A second edition, widened to health software in general, has been in preparation for years. Its status can be read directly from the international project record.

Solved example: reading a standard's stage code

ISO and IEC track every project with a stage code. Stage 30 is the committee stage, where drafts circulate among experts; 40 is the enquiry, when a draft goes to national bodies for a vote; 50 is the approval of a final draft; and 60.60 is publication. A national standards body mirrors the international project record. In that copy, the second edition of IEC 62304 was at stage 30.20 from 7 August 2026, with the next milestone due on 27 November 2026; stage 30.20 means that a ballot on a committee draft has begun. A consultancy that follows the committee reported almost 1,500 comments on the first committee draft and expected the final draft no earlier than 2028. A vendor's web page, by contrast, stated that the final draft stage had begun on 22 May 2026 and that publication was scheduled for 12 August 2026. A standard cannot be at stage 30 and stage 50 at once, and the project record is the authority on its stage. In October 2026 the 2015 consolidated first edition is the one in force. Anything said about the second edition describes a draft, including the proposed replacement of the three safety classes by two levels of process rigor that the German association VDE reported.

54 — Where IEC 62304's second edition stands
20 PREPARATORY 30 COMMITTEE drafts among experts 40 ENQUIRY national bodies vote 50 APPROVAL (FINAL DRAFT) final draft vote 60.60 PUBLISHED project record: 30.20 committee draft ballot, from 7 Aug 2026; next milestone 27 Nov 2026 a standard cannot be at stage 30 and 50 at once a vendor's web page: final draft from 22 May 2026, publication 12 Aug 2026 (contradicted) consultancy: final draft not before 2028; IEC forecast publication Oct 2028 IN FORCE IN OCTOBER 2026 IEC 62304 edition 1.1 (2006 + amendment 2015), recognized by the FDA in full almost 1,500 comments on the first committee draft (2025) draft proposals (e.g. two process-rigor levels replacing classes A, B, C) may change
SpecificationFaultReadoutFocal detail
Plate 54 — Stage codes make a standard's status checkable. In October 2026 the second edition of IEC 62304 was at committee draft stage 30.20, whatever one vendor's page said, and edition 1.1 remained the one in force.

Validation rules for those who use software

Sites and makers that use software, rather than sell it, follow a parallel set of rules. In the US, the FDA's rule on electronic records and signatures, 21 CFR Part 11, has applied since 1997. The FDA finalized its guidance on computer software assurance in September 2025 and revised it on 3 February 2026 under a new title that refers to quality management system software. The guidance sets out the risk-based testing of Chapter 9 for software used in production and in the quality system, and it does not cover the device software itself.

In the EU, a revised Annex 11 on computerized systems and a new Annex 22 on AI went out for consultation from July to October 2025. Both are annexes to the EU's guidelines on good manufacturing practice, or GMP, which is the set of rules on how regulated products such as medicines are made and controlled. Annex 22 is the draft described in Chapter 19. On 8 October 2026 the EU's list of GMP guidelines still contained the Annex 11 of 2011 and no Annex 22.

Build the submission from the records, release by release

Determine the documentation level before design starts, and keep each required record, from requirements to unresolved anomalies, current with every release, so that preparing a submission means exporting records rather than starting a project. For AI, keep the data management records, the sequestered test sets and the subgroup results the FDA, EU and Indian documents ask for, and record which version of each guidance and standard the file follows.

+ What this chapter established
  • The FDA asks for Basic or Enhanced software documentation, depending on whether a software failure could present a probable risk of death or serious injury before risk controls are counted.
  • The FDA's AI guidance is still a draft; it asks for test data that are independent, sequestered and representative, with results by subgroup, and it suggests a model card.
  • The EU and India ask for evidence of valid clinical association, technical performance and clinical performance, and India adds validation in Indian populations for AI.
  • IEC 62304's second edition is a committee draft; Part 11, the FDA's computer software assurance guidance and the EU's Annex 11 of 2011 govern software used under GMP, while a revised Annex 11 and a new Annex 22 are still drafts.