Site acceptance, local performance and computerized system validation.
+ The questionThe maker validated the software; why does the site have to validate it again, and how much is enough?
Two validations, two questions
A device that arrives at a hospital has already been validated by its maker, so a second validation can look redundant. It is not, because the two validations answer different questions. The maker's validation shows that the device meets its intended use in its specified environment, on the patients and inputs of its studies. The hospital's question is whether that evidence carries over to this hospital: its scanners and their settings, its interfaces, its patients and its way of working. Only the hospital has the data to answer that question.
Professional societies now say so plainly. In a joint statement published in January 2024, the radiology societies of the United States, Canada, Europe, Australia and New Zealand called testing on local data, with local systems and workflows, "essential", and recommended that each site "perform a statistically rigorous evaluation of performance on their own local data." Where such an evaluation is not feasible, the societies advised comparing the local data with the vendor's test data and, where the two differ, proceeding "with great caution, if at all."
A trial with its outputs hidden
In August 2020 a team at the Hospital for Sick Children in Toronto began running a deep-learning model on live clinical data with its outputs hidden from clinicians, an approach the team calls a silent trial. The model predicted, from ultrasound images of the kidneys, which children with swelling of the kidney would need surgery. Its development data covered 1,643 kidneys from 294 patients, and on a random 20 percent of those data, held out from training, the model's area under the curve had been 0.90. The silent trial covered 523 kidneys from 150 patients seen between August and December 2020, and there the area under the curve was 0.50, no better than chance.
The team traced the fall to three differences. The live patients were younger, the share of obstructed right kidneys was higher, and the images reached the model in a different form. The development data had been processed JPEG files, while the live data arrived as unprocessed PNG files that looked different to the model even after the same preprocessing. Adjusting for age and side barely helped, raising the area under the curve to 0.51. Reprocessing the live images to match the original processing raised it to 0.84 to 0.85. After the team retrained the model on the original and silent-trial data together, a second silent trial on 711 kidneys from 202 patients gave 0.91 to 0.92. The decisive fault was not clinical at all. It lay in the integration chain of Chapter 18, and it came to light only because the team ran the model on the hospital's own data before anyone acted on the model's outputs.
Acceptance and a local check
A site's own validation has two layers. The first is acceptance: confirming that the device is installed as specified, that each interface passes the right data, that results reach the right screen for the right patient, and that the fallback procedure works when the device is unavailable. The second is a local performance check: running the device on a sample of the site's own cases, silently or retrospectively, and comparing its outputs with a local reference. The societies suggest focusing that review on the predictive values that clinicians will see in practice, identifying the cases where the device would change care, and categorizing its false positives and negatives before deciding whether to deploy. They recommend repeating the review whenever the AI software or the equipment used with it changes.
A hospital wants evidence that a device's sensitivity on its patients is above 75 percent, and it expects the sensitivity to be about 87 percent, as in the device's studies. The hospital plans the check at one-sided 5 percent significance and 80 percent power. The check then has a 5 percent chance of wrongly concluding that sensitivity is above 75 percent when it is really 75 percent, and an 80 percent chance of showing that it is above 75 percent if it really is 87 percent. The standard calculation for comparing a proportion with a fixed value gives about 69 diseased cases. If 10 percent of the hospital's screened patients have the disease, the check needs about 690 consecutive cases. With 60 detections out of 69, 87 percent, the exact one-sided lower confidence bound is about 78 percent, above the target: at the same 5 percent level, the result rules out any sensitivity below that bound. A check run on a few dozen convenient cases cannot reach that conclusion, which is why the societies call for statistical rigor rather than a demonstration.
The results of the check become the baseline for the monitoring of Chapter 20: the distribution of inputs, the rate of declined cases, and the local sensitivity, specificity and predictive values against which later drift is measured.
Computerized system validation
Some sites must go further, because the rules they work under require it. Laboratories, blood services, organizations running clinical trials and pharmaceutical manufacturers work under good-practice rules, known collectively as GxP, and under these rules the user of any computerized system that affects product quality, patient safety or data integrity must validate that system. Many of these organizations follow ISPE's GAMP 5, the guide of Chapter 8, whose second edition appeared in July 2022. GAMP 5 is risk-based: the validation effort scales with the system's impact, complexity and novelty. Its authors describe it as placing patient safety and product quality ahead of compliance for its own sake, and as supporting iterative and agile methods.
GAMP 5 sorts software into categories: infrastructure software, such as operating systems and databases, and then a continuum from products used without configuration, through configured products, to custom software. The more a system is configured or customized for the site, the more verification the site must do itself; for a standard product, the user relies on the supplier's evidence after assessing the supplier. The work then follows a sequence of steps. The site writes user requirements that state what it needs, a risk assessment shows where a failure would matter, and the system is specified, configured and verified.
Verification has traditionally taken the form of installation, operational and performance qualification, which confirm in turn that the system is installed correctly, operates as specified and performs as needed in routine use. Increasingly it takes the form of risk-based testing, some of it scripted in advance and some unscripted (Chapter 9). A traceability matrix links each requirement to its tests, release into use is controlled, and the system is reviewed periodically for as long as it is used. Data integrity applies at every step: records must be attributable to the person who made them, complete and protected, with audit trails that log who changed what and when. The FDA has set requirements for such electronic records since 1997, in its rule on electronic records and signatures, 21 CFR Part 11.
With the supplier's documentation, a site can avoid repeating tests that do not depend on its environment. The documentation cannot cover the site's interfaces, configuration, data or patients. A site that accepts a package without assessing the supplier, or without local performance data, leaves its own question unanswered.
ISPE has extended the method to AI. The 2022 edition of GAMP 5 included an appendix on machine learning, which set out a life cycle for a machine-learning subsystem, and in July 2025 ISPE published a separate GAMP guide to AI, covering data governance, model testing, monitoring and change management for AI-enabled systems used in GxP work. In July 2025 the European Commission also put out for consultation a new annex on AI to the EU's good manufacturing practice rules. As drafted, the annex would admit only static, deterministic models in critical manufacturing uses: models that do not change during use and always give the same output for the same input.
The draft annex would not apply to generative AI, which produces text, images or speech rather than a score, or to large language models, which are trained on very large amounts of text to generate text. It would also require acceptance criteria at least as high as the performance of the process the model replaces. As of October 2026 the annex was still a draft. Its scope is medicines manufacturing, not medical devices, but its rules on test data read like a summary of Chapters 11 to 13.
The builder's standard and the user's
IEC 62304 and GAMP 5 describe the same software from opposite sides. IEC 62304 is written for the organization that builds device software, and under it the maker records how the device was made. GAMP 5 is written for the organization that uses a computerized system, and under it the user records evidence that the system does what that user needs, in that user's configuration. A laboratory that works under good-practice rules and installs an AI-enabled analyzer needs both: the maker's evidence, and its own validation built on it. So does a contract research organization, a company that runs clinical trials for sponsors, when it uses an AI reading tool in a trial. Makers can shorten the site's validation by supplying a validation package with their product: requirements and traceability, test scripts the site can re-run, and installation and operational qualification documents.
Before purchase, write the site's requirements and acceptance criteria, including the local performance check and its sample size, and ask the supplier for its validation package and its specified environment. Where GxP rules apply, scope the work with GAMP 5's categories and risk assessment; in every case, run the device silently or retrospectively on local cases before clinicians rely on it, and keep the results as the monitoring baseline.
- The maker's validation covers the intended use in a specified environment; only the site can show that the evidence carries over to its systems and patients.
- In a silent trial in Toronto, a model's area under the curve fell from 0.90 to 0.50, recovered to 0.84 to 0.85 once an image-format mismatch in the integration was fixed, and reached 0.91 to 0.92 after retraining.
- A local check needs enough diseased cases to support a conclusion, about 69 in the solved example to show a sensitivity above 75 percent, and its results become the monitoring baseline.
- Where GxP rules apply, sites validate AI systems with GAMP 5's risk-based method, which ISPE has extended to AI, and build on the supplier's evidence.