20 — Watching a model in the field

When the true answer arrives late.

+ The questionHow can a maker, or a hospital, tell a model is still working when the right answer arrives months later, or never?

Truth arrives late

A model can be wrong for months before anyone can know. When a screening program tells a patient to come back in a year, nobody checks whether that answer was right until the patient returns, if the patient returns at all. When the program refers a patient, the specialist confirms or overturns the answer weeks later, and the result is kept in another clinic's records. For most outputs of most diagnostic models, the reference standard of Chapter 12 is never applied at all in routine use. The direct measure of performance is agreement with the right answer, and that agreement is available late, partially and only for some patients.

The early months of the COVID-19 pandemic showed how quickly the conditions of use can change in the meantime. A study of 24 hospitals in four US health systems found that daily alerts from the sepsis prediction model of Chapter 13 rose by 43 percent while the number of patients in the hospitals fell by 35 percent, ahead of the surge in COVID admissions. The patients had changed, and the model's inputs had changed with them. Michigan Medicine was one of the four systems, and the university's account of the study states that the rise in alerts there led to the alerts being paused. Nobody needed to wait for confirmed diagnoses to see that something had shifted, because the change was visible in the number of alerts the model itself produced.

What can be watched at once

Monitoring therefore starts with what is available immediately. The inputs can be checked as they arrive: the equipment and software versions in the image headers of Chapter 18, measures of image quality, patient age and the other characteristics the device records. The outputs can be counted: the share of positive results, the share of declined cases and the distribution of scores. None of these measures shows directly whether the model is right. Each can show within days that the conditions the model was validated under no longer hold, and that finding is the trigger for a closer investigation.

Solved example: the positive rate as an early warning

Suppose a screening device with 87.4 percent sensitivity and 89.5 percent specificity runs at a site where 10 percent of patients have the disease. The device should return a positive result for about 18.2 percent of patients: 8.7 percentage points come from the diseased patients it detects, and 9.5 from false alarms among the healthy. At 200 screens a week, that is about 36 positives a week, and chance alone makes the weekly count vary around that figure with a typical spread of about 5. If a camera fault lowers specificity to 80 percent, the expected positive rate rises to about 26.7 percent, or about 53 positives a week, which is more than three typical spreads above the baseline. A simple chart of weekly positives, with its alarm line set three typical spreads above the expected count, would flag the change within a few weeks, months before referral outcomes could confirm that the extra positives were false.

39 — The positive rate as an early warning
0 10 20 30 40 50 60 70 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 WEEK POSITIVE RESULTS PER WEEK (200 SCREENS) expected 36 (18.2%) after a camera fault: expected 53 (26.7%) baseline + 3 spreads = 51 camera fault (specificity 89.5% → 80%) flagged within a few weeks schematic: expected levels from the worked example; real weeks scatter around them DRIFT THE COUNTS CAN MISS discrimination (AUC) maintained calibration predicted risk observed rate time → predicted risk overstates schematic, after a 9-year study of kidney-injury models
FaultReadoutFocal detail
Plate 39 — At 200 screens a week, a fall in specificity from 89.5 to 80 percent lifts expected positives from 36 to 53, more than three spreads above baseline, so a weekly chart flags it long before referral outcomes arrive. Calibration can drift while the area under the curve holds.

Some drift does not show in counts of inputs and outputs. A 2017 study at US Veterans Affairs hospitals followed seven models that predict acute kidney injury, a sudden loss of kidney function, over the nine years after the period their development data came from. The seven included machine-learning models and regression models, the traditional statistical kind that adds up weighted risk factors; the word has nothing to do with the regression tests of Chapter 7. All kept their discrimination, the ability to rank patients by risk. What drifted was their calibration, the agreement between the risks a model predicts and the rates that actually occur.

Every model came to overstate the risk as the rate of kidney injury changed, and the overprediction grew steadily for the regression models while staying small for the random forest, a model that combines many decision trees, and for the neural network. If a hospital had used one of the regression models with a fixed threshold on predicted risk, the model would have alerted more and more often for the same patients, while its area under the curve stayed stable throughout.

A stable area under the curve can hide a drifting threshold

A model's discrimination can hold while its calibration drifts. Monitoring that tracks only the area under the curve can miss a drift that moves every decision made at a fixed threshold, and that kind of drift changes alert rates and referrals.

Closing the loop with outcomes

The direct measures come later. A site can collect the specialist findings for referred patients and match them to the device's outputs, which gives a running estimate of the positive predictive value. It can also send a random sample of negative results for a reference reading, which is the only way to estimate missed cases before patients return with them. The maker uses complaints, adverse event reports and service records in its postmarket surveillance, the system through which it reviews information about a device once the device is on the market. In their joint statement of 2024, the radiology societies recommend re-evaluating each AI tool on updated local data "at specified intervals, but at least annually," and after any new version. They note that "naturally occurring data drift will cause AI model performance to degrade over time," and they ask institutions to define in advance how a problem will be escalated and resolved.

40 — Two clocks of a monitoring plan
WATCHED CONTINUOUSLY: DAYS inputs: equipment and software versions, image quality, patient age outputs: positive rate, declined rate, score distribution CLOSED ON A SCHEDULE: WEEKS TO MONTHS reference reading of a random sample of negatives specialist findings matched to positives: predictive value complaints, adverse events, service records → maker's postmarket surveillance baseline from the site validation (Chapter 19) limit crossed? investigate restrict use suspend change the device (Chapter 21) named reviewer at the site named reviewer at the maker the only early estimate of missed cases radiology societies (2024): re-evaluate on local data at least annually and after any new version
SpecificationDataReadoutFocal detail
Plate 40 — Monitoring runs on two clocks: inputs and output rates show within days that conditions have shifted, while reference reads of sampled negatives and matched outcomes measure accuracy over months. Named people on both sides act when a limit is crossed.

Who watches

Responsibility for monitoring is shared, and its division is not yet settled. The maker has postmarket obligations everywhere a device is sold, and the last of the good machine-learning practice principles states the expectation that deployed models are monitored for performance and that the risks of retraining are managed. The hospital sees the inputs, the workflow and the outcomes, and only the hospital can link the device's outputs to its own patients' later records. In September 2025 the FDA asked for public comment on how the real-world performance of AI-enabled devices should be measured and evaluated, with questions about metrics, monitoring methods, data sources and the triggers for a response. The FDA described its request as proposing no policy, and no follow-up had been published by October 2026. Under the EU's AI Act, hospitals that deploy high-risk AI will take on monitoring duties, on the timetable set out in Chapter 29.

What a monitoring plan needs is clear even where the division of labor is not. It needs a baseline from the site validation of Chapter 19, measures of inputs and outputs watched continuously, a reference sample and outcome matching on a schedule, thresholds that trigger investigation, and named people on each side who act when a threshold is crossed. When monitoring shows that performance has moved, the response is a change to the device, its labeling or its use, and Chapter 21 sets out how a model can be changed without losing the evidence behind it.

Write the monitoring plan before go-live

Define the input checks, the output and declined-case rates to chart, their baseline from the local validation and the limits that trigger investigation. Schedule a reference reading of a random sample of negatives and the matching of positives to specialist findings, assign who on the maker's and the site's side reviews each, and state what happens when a limit is crossed: investigation, restricted use, suspension or retraining.

+ What this chapter established
  • For most outputs of a diagnostic model the right answer arrives late or never, so accuracy cannot be monitored directly in real time.
  • Checks on inputs, output rates and declined-case rates can show within days that conditions have shifted, as the rise in sepsis alerts did early in the pandemic.
  • Calibration can drift while discrimination holds, so monitoring must watch decisions at the threshold, not only the area under the curve.
  • Reference readings of sampled negatives, outcome matching and agreed triggers close the loop, and the maker and the site share the work.