+ Part V · Into the hospital, and after

Deploying, watching and changing a model.

On 10 June 2021 the FDA cleared a new version of IDx-DR. The new version could receive images in DICOM, the format hospitals use for medical images, and it put its instructions on screen step by step. It let clinics configure file names and keep images locally, and it told operators during the exam whether an image was good enough to send. None of these changes altered what the program claimed to detect, and the FDA cleared the version without new clinical data. The changes were made to fit the device into the way clinics work. A year later the FDA cleared another version, in which the maker had replaced one of the models inside the analysis. The four chapters of this part follow a model from authorization into use: connecting it to a hospital's systems, checking it on the hospital's own patients, watching it once it runs, and changing it without losing the evidence. By the end of the part, the central question has its answer.

18 — Not plug-and-play

Fitting the software into a hospital.

+ The questionAn authorized AI device arrives at a hospital. What has to be connected, configured and agreed before it is used on the first patient?

The label on every image

Every image that a hospital's scanners and cameras produce in the DICOM format carries a header: a list of fields describing where the image came from. Each field is identified by a tag made of two four-digit codes. The field tagged 0008,0060 records the modality, such as CT or ophthalmic photography, and the field tagged 0008,0070 records the manufacturer of the equipment. Field 0008,1090 holds the model name, 0018,1020 the versions of the software running on the equipment, 0008,1030 a description of the study and 0010,0020 the patient's identifier. In DICOM a study is the set of images from one examination. The DICOM standard defines thousands of such fields. It is revised several times a year, and a 2026 release was current in October 2026.

The header is an AI device's first safeguard: the device can check it for each image before it analyzes the image. A device authorized for images from one camera can read the manufacturer and model fields and refuse anything else, and it can record the software version of the equipment that produced each image, so that a later change in performance can be traced to an update on the scanner. Many fields, however, are set by local protocols or typed by staff, and two hospitals can describe the same examination differently. A device that decides which images to analyze from their free-text descriptions will meet descriptions its makers never saw.

35 — What a DICOM header tells a device
image (pixel data) schematic DICOM HEADER 0008,0060 Modality e.g. CT, OP (ophthalmic photography) 0008,0070 Manufacturer equipment maker 0008,1090 Manufacturer's model name camera or scanner model 0018,1020 Software versions equipment software version 0008,1030 Study description free text: differs between hospitals 0010,0020 Patient ID local identifier typed by staff or set by local protocol: routing on it meets descriptions the makers never saw the device can check what it was tested on AI DEVICE: INPUT CHECK specified inputs (e.g. Topcon NW400) inside specification: analyze outside: refuse, with a message the user can act on logged with each result: trace later performance changes to an equipment update DICOM defines thousands of such fields; a 2026 release current in October 2026
SpecificationCodeDataFaultReadoutFocal detail
Plate 35 — The header of every DICOM image says which equipment and software produced it. A device can use it to refuse inputs outside its specification and to trace changes in performance to equipment updates, but free-text fields such as the study description vary from hospital to hospital.

Where the software sits

An AI device rarely works alone. A scanner sends each study to the picture archiving and communication system, or PACS, which stores the hospital's images and displays them to radiologists. An imaging model receives the study from the PACS and returns its results to the PACS, to the radiologist's worklist of studies waiting to be read, to the electronic health record or to a specialist's phone. The model may run on a server in the hospital or in the maker's cloud. Each of those connections is an interface with its own standard.

DICOM carries images and the results derived from them. HL7 version 2 carries orders and reports as messages; HL7, the standards body that publishes it, says that 95 percent of US healthcare organizations use it. FHIR is HL7's newer standard, built from web resources that can each be addressed individually. It has been at release 5 since March 2023. In July 2026 release 6 was in its second normative ballot, a formal vote on making its text a stable standard.

Standards for the AI step itself are younger. The IHE initiative writes profiles, which describe how vendors should combine existing standards for a task. It has published one profile for AI results, which defines how analysis results are encoded in DICOM objects and displayed, and one for AI workflow, which defines how a request for analysis is sent, managed and performed. In October 2026 both were still at trial implementation, the stage before final text. A hospital connecting several AI products therefore often builds part of each integration itself, or buys a platform that does that work.

Inputs inside and outside the specification

A device's authorization covers specified inputs. IDx-DR is indicated for use with the Topcon NW400, and the public summary of its 2021 clearance states a minimum image resolution of 22 pixels per degree of the camera's field of view. Inside those limits the trial's evidence applies; outside them it does not, whatever the images look like. When an input falls outside its specification, a device has two choices: it can refuse the input, with a message the user can act on, or it can analyze the input anyway and risk an answer the evidence never covered. IDx-DR refuses: it returns its "insufficient quality" output, and since 2021 it has also given feedback during the exam so that the operator can retake the image.

Inputs also change after installation. A hospital may change a scanner protocol to reduce dose, or a vendor may update the scanner's reconstruction software, which computes the image from the scanner's raw measurements. The PACS may begin compressing images to save storage, or a new interface may convert images to a different format on the way to the model. Such a change can be invisible to a person reading the images and still be visible to the model. That kind of change is the acquisition shift of Chapter 14. In a hospital study that Chapter 19 describes, a change of image format, more than any change in the patients, cut a model's performance from excellent to no better than chance.

A change on the hospital's side can change the device's inputs

A model's inputs can move outside the conditions it was tested under without any change to the device itself. A new scanner protocol, a reconstruction update or compression in the archive can each do so, and so can a format conversion in an interface engine, the software that translates and routes data between hospital systems. Agreeing in advance who notifies whom of such changes is part of installing the device.

Solved example: where the minutes go in stroke triage

A retrospective comparison with usual care tested the stroke-triage program authorized in 2018 on 44 cases with a confirmed blocked artery that the program had flagged correctly. With the program, the specialist was notified a median of 5.6 minutes after the CT angiogram; when notification came through the radiology report, the median was 51.5 minutes. A 2023 randomized trial tested the same company's software at four Houston stroke centers, in 243 treated patients. The time from arrival to the start of clot removal fell by an estimated 11.2 minutes from a median of 100, about 11 percent. The two studies measured different intervals in different hospitals, so their numbers cannot be subtracted. Together, though, they show that the model shortens one step and that the other steps decide how much of the gain reaches the patient: transferring the images, getting the alert to the right person, assembling the team and having the procedure room ready. The trial found no significant difference in patients' functional outcome, their level of disability after the stroke. That was an exploratory measure, one the trial was not designed to settle.

36 — Where the minutes go in stroke triage
CT angiogram images transferred AI analysis alert reaches specialist team assembles procedure room ready clot removal starts the link the software shortens the rest of the chain decides how much of the gain reaches the patient 2018 COMPARISON 44 cases with a confirmed blocked artery time from CT angiogram to specialist notified (median) through the radiology report: 51.5 min with the program: 5.6 min 0 10 20 30 40 50 60 min 2023 RANDOMIZED TRIAL 4 Houston stroke centers, 243 treated patients arrival to start of clot removal (median) usual care: 100 min with the software: estimated 11.2 min shorter 0 10 20 30 40 50 60 70 80 90 100 110 min functional outcome: no significant difference (exploratory) different intervals in different hospitals: the numbers cannot be subtracted
CodeDataReadoutFocal detail
Plate 36 — A triage model shortens one link in the chain from scan to treatment. In a 2018 comparison notification came at a median of 5.6 minutes against 51.5; in a 2023 trial the time to clot removal fell by an estimated 11.2 minutes from a median of 100. Transfer, the team and the procedure room decide how much of the gain reaches the patient.

Who answers for what

The maker answers for the device within its specified environment, and under the EU's medical device regulation it must state that environment: manufacturers "shall set out minimum requirements concerning hardware, IT networks characteristics and IT security measures" needed to run the software as intended. The hospital answers for the network, the interfaces, the other systems and the way the device is used. The integration lies between them, and neither party controls it alone.

The standard for that integration is IEC 80001-1. Its 2021 edition sets general requirements for how an organization applies risk management before, during and after it connects a medical device or health software to its IT infrastructure, covering safety, effectiveness and security. ISO has already marked the standard for revision and intends to replace it with a new standard in the 81001 series. An earlier technical report in the same family describes responsibility agreements: written statements of what each party does across the life of the connection. The parties include the hospital, the IT supplier and the device maker. The questions such an agreement must settle are concrete: who tests the interfaces, who approves changes on each side, who is told when a scanner or the archive is updated, and what clinicians do when the AI is unavailable.

Treat installation as a project with its own risk file

Before the first patient, list every interface the device depends on and test each with real local data, and confirm that the local images and their headers fall inside the device's specification. Measure the time from acquisition to result against the clinical need. Define what clinicians do when the device is down or declines an input, and sign a responsibility agreement that names who notifies whom of changes on either side.

+ What this chapter established
  • A device can read the DICOM header of every image to identify the equipment and software behind it, refuse inputs outside its specification and trace changes.
  • AI devices depend on interfaces to PACS, health records and cloud services; the IHE profiles for AI were still trial implementations in October 2026.
  • A model's evidence covers only inputs inside its specification, and changes on the hospital's side can move inputs outside it without any change to the device.
  • The maker states the required environment, the hospital manages the connection under IEC 80001-1, and a responsibility agreement divides the rest.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.