The threat changes while the code does not.
+ The questionOnce a device ships, how are new weaknesses in its code and its components found, judged and fixed?
Eleven flaws in someone else's code
In July 2019 eleven vulnerabilities were disclosed in IPnet, a network stack: the part of an operating system that sends and receives data over a network. Six of them were serious enough to let an attacker on the network run code of the attacker's own choosing. Armis, the security company that found them, named them URGENT/11. It reported that the affected code had been part of the VxWorks real-time operating system since version 6.5, and of several other operating systems that had used the same stack. On 1 October 2019 the FDA warned patients, clinicians and makers that medical devices and hospital networks using those operating systems could be affected. The FDA named six operating systems from six suppliers. It said that device makers' notices so far included an imaging system, an infusion pump and an anesthesia machine, and that the agency knew of no adverse events.
No device maker had written IPnet. The stack reached their products inside operating systems that the makers had chosen and treated as SOUP under the rules of Chapter 6. The code in those devices had not changed since release. What changed was public knowledge about the code, and with that knowledge the risk. URGENT/11 shows the general pattern of security after release: the threat changes while the code does not, and most of the code at issue came from someone else.
Matching the list to the news
When a vulnerability is published, the first task is finding out which products contain the affected component, and in which version. The software bill of materials of Chapter 6 exists for that search. A maker that keeps a machine-readable SBOM for every released version can search all of them for the component within hours; a maker that must ask each development team, or inspect old build files, may take weeks. In its 2026 premarket guidance, the FDA expects makers to monitor sources such as NIST's National Vulnerability Database and to assess the known vulnerabilities in each component. It also expects them to give particular weight to those listed in CISA's catalog of vulnerabilities already being exploited in real attacks.
Whether the component is present is only the first question. A device may contain the vulnerable component without using the vulnerable function, or without exposing that function to any interface an attacker can reach. The maker records its answer for each vulnerability and each product as a VEX statement, short for Vulnerability Exploitability eXchange. In 2021 NTIA, the US Commerce Department's telecommunications agency, described a VEX as an assertion about the status of a vulnerability in a specific product, with four values: not affected, affected, fixed, or under investigation.
Vulnerabilities were also disclosed in another widely used network stack, made by Treck. According to a trade report, by June 2020 the infusion pump maker B. Braun had received 24 patches from Treck for its Outlook 400ES pump and had determined that 20 of them did not apply. The 4 that applied were 17 percent of the 24. Each of the 20 needed a reasoned statement of why the pump was not affected; each of the 4 needed an assessment of exploitability and harm, and a fix. The amount of that work grows with the number of components in the bill of materials. A code-audit vendor reported a mean of 1,180 open-source components per application (Chapter 6). If a product carries that many components, and 1 percent of them receive a new advisory in a year, about 12 advisories arrive for that product each year. At B. Braun's ratio, about 2 of the 12 would need a fix, and every one would need a written conclusion. Without an SBOM, the first step for each advisory is the slowest: finding out whether the component is present at all.
Judging what a weakness means for patients
A vulnerability that applies is judged on two scales. The first is technical severity, usually expressed with the Common Vulnerability Scoring System, the system behind the MiniMed score read in Chapter 22. The system's fourth version was published on 1 November 2023 and added a supplemental Safety metric. That metric records whether exploitation could cause injury, using the consequence categories of IEC 61508, the standard for functional safety: the part of safety that depends on electrical, electronic and programmable electronic systems working correctly. The specification states explicitly that supplemental metrics do not change the score. Under an FDA contract, MITRE wrote a rubric for applying the scoring system to medical devices. In October 2020 the FDA qualified the rubric as a medical device development tool, which means the agency has evaluated the tool and accepts that it works as intended within a stated context of use.
The second scale is clinical. In its postmarket cybersecurity guidance of December 2016, the FDA divides risks into controlled and uncontrolled by their residual risk, the risk that remains after controls are in place. A risk is controlled where the residual risk of patient harm from a vulnerability is acceptable, and uncontrolled where it is not.
For an uncontrolled risk, the agency states that it does not intend to enforce its usual reporting rules for corrections if four conditions hold. The first is that no serious adverse events or deaths are known. The other three concern the maker: it must tell its customers within 30 days of learning of the risk, with interim measures and a plan; fix the risk, validate the fix and distribute it within 60 days; and take part in an information-sharing organization for the sector. Since 29 March 2023 US law has required a maker that submits a cyber device to the FDA to make updates and patches available, with timing that depends on the vulnerability. Known unacceptable vulnerabilities must be fixed on a "reasonably justified regular cycle," and critical ones that could cause uncontrolled risks "as soon as possible out of cycle."
Coordinated disclosure
Many device vulnerabilities are found by people outside the company: academic researchers, security firms, hospital engineers. Outside researchers reported the MiniMed flaws, and a security firm found URGENT/11. Coordinated vulnerability disclosure is the practice by which they report a weakness privately to the maker, the maker acknowledges and investigates it, and the two agree when the details are published, ideally with a fix or mitigation available. A government coordinator often works between the two. The US government's coordinator, now part of CISA, published the MiniMed advisories of 2018 and 2019 after researchers reported the flaws and Medtronic analyzed which models shared them.
Two international standards describe the maker's side. ISO/IEC 29147 covers receiving and publishing reports, and ISO/IEC 30111 covers handling them inside the company; both were flagged for revision in 2025. Since March 2023 US law has required a maker that submits a cyber device to the FDA to include a plan for monitoring and addressing vulnerabilities after release, "including coordinated vulnerability disclosure and related procedures."
The hospital's share
A fix protects no patient until it is installed, and for many devices installation happens in the hospital. Each update has to be scheduled around clinical use and tested against the hospital's own configuration where the integration of Chapter 18 could be affected. It is then applied unit by unit, sometimes by the maker's service engineers and sometimes by the hospital's clinical engineering staff. In December 2020 Armis, which had found URGENT/11, estimated that 97 percent of the affected devices in operational technology had not been patched. Operational technology means the computers that control machines and physical equipment rather than handle office data. The estimate covers far more than medical devices, but it shows how slowly fixes reach equipment that is not a personal computer.
When a maker publishes a patch, the open risk moves from the maker's work queue to the hospital's. Until a unit is updated, its protection depends on what surrounds it: restricted access, disabled services, monitoring and network segmentation, which keeps the device on a separate part of the network. Hospitals need the maker's SBOM and VEX statements to know which units are exposed, and makers need to know which versions hospitals have installed in order to see which patients remain at risk.
The EU's medical device guidance on cybersecurity was issued in 2019 and revised in 2020. It describes security as a joint responsibility of makers, integrators, operators and users, while noting that only makers carry the legal duties. It also states that a device should not depend on security controls in its environment. In practice, the defense of a device in the field is shared: the maker designs the device to be patched, watches for new weaknesses and issues fixes; the hospital keeps an inventory of what it runs, installs fixes and protects what it cannot yet fix.
Keep the SBOM of every released version searchable, monitor vulnerability databases and the catalog of exploited vulnerabilities daily, and record a VEX statement for each match. Judge each applicable vulnerability by exploitability and clinical harm, set a target date in the regular cycle or out of it, publish a disclosure policy with a contact address, and tell hospitals which versions are affected and what to do until they can patch.
- URGENT/11 put eleven vulnerabilities into devices through an operating system's network stack that no device maker had written.
- With a searchable SBOM a maker finds affected products within hours, and a VEX statement records whether each one is actually exposed.
- Severity scores measure the weakness; FDA guidance ties its relief from reporting to fix deadlines for uncontrolled risks, and US law requires makers of cyber devices to make patches available on a regular cycle for known unacceptable vulnerabilities, or out of cycle for critical ones.
- Researchers report through coordinated disclosure, and a fix protects patients only once it is installed, often in the hospital; maker and hospital share the defense.
Let's Talk