+ Part VI · Defending the device
Cybersecurity.
On 28 August 2017 Abbott wrote to physicians that a firmware update was ready for several of its pacemakers, formerly made by St. Jude Medical. The update was meant to close vulnerabilities in the devices' radio link, weaknesses that an attacker could use. It could not be sent remotely. Each patient had to visit a clinic, where the update was sent through the wand of the programmer, the clinic's instrument for reading and setting an implanted pacemaker. The update took about three minutes, and the pacemaker ran in a backup mode while it was installed. Press reports of the FDA's communication put the number of affected devices in the US at about 465,000. Nothing in those pacemakers had failed. According to a cardiology society's summary of the FDA's communication, the flaws could allow intrusions, and some of those intrusions could affect how the device operated. The three chapters of this part deal with weaknesses of that kind: what threats a device faces, which defenses must be designed in before release, and how weaknesses found after release are judged and fixed.
An attacker is an input.
+ The questionHow can a device that meets every requirement still be made to harm a patient?
A pump that could not be patched
On 27 June 2019 the FDA warned patients and clinicians that certain Medtronic MiniMed insulin pumps were being recalled because an unauthorized person "could potentially connect wirelessly to a nearby MiniMed insulin pump and change the pump's settings." The recalled pumps were the MiniMed 508 and several models of the Paradigm series. They communicated by radio with remote controllers and glucose meters, and the protocol they used lacked proper authentication, the check that confirms who sent a message, so a pump could not tell its own patient's devices from an attacker's transmitter within range. Medtronic had identified about 4,000 patients in the US who were potentially using vulnerable pumps.
The remedy was unusual. The FDA said that Medtronic was "unable to adequately update" the pumps "with any software or patch," and the company offered patients replacement pumps with stronger built-in protection. The FDA said it was not aware of any confirmed patient harm. On the same day CISA, the US cybersecurity agency, published an advisory that described the weakness as improper access control. According to the advisory, only an attacker nearby could exploit it, doing so required high skill, and there was no known public exploit, meaning no published code or method built to use the weakness. The FDA posted its recall record for the MiniMed 508 in March 2020, and the recall was still open in October 2026.
Safety risk and security risk
The pumps met their requirements. They delivered insulin as programmed, alarmed as specified and passed their verification. What they lacked was a requirement that nobody had written in the form that mattered: refuse commands from anyone but the patient's own devices. Like every software fault in this guide, the fault was present on the day the pumps shipped; what differed was the input that reached it, which was not an unlucky combination of events but a deliberate one.
That difference changes how risk is judged. Safety risk management estimates how likely a sequence of events is, following the ISO 14971 chain of Chapter 4. An attacker is not a sequence of events with a probability; an attacker chooses the inputs, searches for the rare combination and repeats it at will. For that reason the FDA's 2026 guidance on cybersecurity in premarket submissions describes security risk assessment as non-probabilistic. Such an assessment judges a vulnerability by its exploitability, how easily an adversary could use it, and by the harm that would follow. The FDA also treats known vulnerabilities as reasonably foreseeable. Where a security weakness could lead to patient harm, the two assessments overlap, and the security control that blocks the attack is also a safety control.
In its 2019 advisory, the US cybersecurity agency scored the pump weakness 7.1, High, on version 3 of the Common Vulnerability Scoring System. The score comes with a vector, a string that records each factor behind it: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H. From left to right, its factors are these: the attack vector is adjacent, meaning within radio range rather than over the internet, and the attack complexity is high. The attacker needs no privileges, meaning no access rights on the device, and no action by the user. The scope is unchanged, meaning the effects of an attack stay within the component that has the weakness. The impact is low on confidentiality, the secrecy of data. It is high on integrity, whether data stay correct, and on availability, whether the device stays usable, because settings and delivery could be changed or stopped. In a 2018 advisory on the same pumps' remote controllers, two weaknesses had been scored 4.8 and 5.3, Medium. They scored lower because each affected only one of confidentiality or integrity, and one also needed user interaction. The score summarizes technical severity but says nothing about how many patients use the device or what a wrong dose of insulin does, so a maker adds its own assessment of the clinical harm alongside the score.
The attack surface
Every way data or commands can enter a device is part of its attack surface: radio and Bluetooth links, Wi-Fi and Ethernet, USB and serial ports, the connection to a cloud service, a service interface used only by technicians. Since 29 March 2023 a US law has set requirements for every premarket submission for what it calls a "cyber device": among other conditions, a device that includes software and can connect to the internet. Chapter 29 sets out those requirements. The FDA reads the condition of connecting to the internet broadly, "whether intentionally or unintentionally, through any means": Wi-Fi, cellular, Bluetooth, magnetic inductive links, USB, Ethernet and serial ports all count, and so does a USB connection used briefly for service.
A threat model lists those entry points, the assets behind them and the adversaries who might want those assets. The assets include the dose settings, the patient's data and the device's software itself. For each path, the threat model records what an attacker could do and what would stop the attack. In 2021 MITRE, a US non-profit research organization, and the Medical Device Innovation Consortium published a playbook for threat modeling medical devices, with funding from the FDA. The FDA's premarket cybersecurity guidance expects the threat model to be part of the design and to be updated as the architecture of Chapter 5 changes. The special controls set for IDx-DR in 2018 already required a cybersecurity vulnerability and management process, because the device's images and answers cross the internet between the client and the server.
Estimates that an attack is unlikely rest on assumptions about who would try, with what skill and for what gain, and those assumptions go out of date. Research, published exploits and automated tools lower the skill an attack needs. Judging a weakness by how easily it could be used, and by the harm that would follow, stays valid longer than judging it by how likely an attack seems today.
When security becomes safety
Some attacks threaten patients directly, as a changed insulin dose would. Others threaten data, and an attack on data can undermine trust and care. In January 2025 the FDA warned that the software of the Contec CMS8000 patient monitor could send patient data outside the care setting and contained a backdoor, a hidden way into the software that bypasses its normal access controls. The monitor was also sold under another name. The FDA advised hospitals that relied on the monitors' remote monitoring to unplug them and stop using them, and advised other users to use them for local monitoring only. The fix released in July 2025 removed the monitors' networking entirely: the remedy for the unsafe connection was to have no connection at all.
The pump and the monitor show two limits of device security. A defense that was not designed in may be impossible to add later, and for some weaknesses the safest response is to give up a function. Chapter 23 describes the defenses that have to be in the design from the start.
Start from the architecture diagram of Chapter 5 and mark every interface, including service and debug ports and cloud connections. For each, list what could be read, changed or stopped through it, by whom and with what effort, and the control that prevents it. Judge each path by exploitability and clinical harm rather than by an estimate of how likely an attack is, and update the model with every design change.
- MiniMed pumps recalled in 2019 accepted radio commands without proper authentication and could not be patched, so patients were offered replacements.
- Security risk is judged by exploitability and harm, not probability, because an attacker chooses the inputs; known vulnerabilities count as foreseeable.
- Every interface is part of the attack surface; US law and the FDA's broad reading make almost any connectable device a cyber device.
- A threat model built from the architecture names each path, its harm and its control, and some weaknesses can be removed only with the function.