16 — Humans in the loop

The reader and the machine.

+ The questionWhen AI assists a clinician rather than deciding alone, what has to be measured: the model, the person, or the pair?

When the suggestion is wrong

An assistant that is right most of the time can still make an expert worse. In a study published in 2023, 27 radiologists read 40 test mammograms. Each mammogram came with a suggested category on the standard breast-imaging scale, which grades how likely a finding is to be cancer, and the readers were told that the suggestions came from an AI system. In fact the investigators had set the suggestions, and 12 of the 40 were deliberately wrong. When the suggestion was right, readers of every level of experience rated about 80 percent of mammograms correctly. When it was wrong, inexperienced readers rated 19.8 percent correctly, moderately experienced readers 24.8 percent and very experienced readers 45.5 percent.

The pattern has a name: automation bias, the tendency to accept a machine's output in place of one's own judgment. A 2012 systematic review found that it varies with a person's experience of the task and with trust and confidence, and that heavy workload, complex tasks and time pressure increase it. Automation bias turns the model's errors into the clinician's errors, and it is strongest among the people an assistant is often meant to help most, those with less experience.

31 — When the suggestion is wrong
INEXPERIENCED MODERATELY EXPERIENCED VERY EXPERIENCED 0% 20% 40% 60% 80% 100% SHARE OF MAMMOGRAMS RATED CORRECTLY correct suggestion wrong suggestion 79.7% 19.8% the model's error becomes the reader's 81.3% 24.8% 82.3% 45.5% Dratsch et al., Radiology 2023: 27 radiologists, 40 test mammograms the 'AI' suggestions were set by the investigators; 12 of 40 deliberately wrong
FaultReadoutFocal detail
Plate 31 — When a suggestion presented as AI was wrong, radiologists' accuracy fell from about 80 percent to 20 percent for inexperienced readers and 46 percent for very experienced ones. Automation bias turns a model's errors into the clinician's.
Solved example: how a model's errors pass through a reader

In the study's design, the suggestion was right in 28 of the 40 cases, 70 percent. An inexperienced reader was right 79.7 percent of the time with a correct suggestion and 19.8 percent of the time with a wrong one, so that reader's overall score is 0.7 times 79.7 plus 0.3 times 19.8, about 62 percent. For a very experienced reader, right 82.3 percent and 45.5 percent of the time, the same sum gives about 71 percent. Of the gap of about 9.5 points between the two readers, about 7.7 points come from the cases with a wrong suggestion: 0.3 times the 25.7-point difference in how often the two readers rated those cases correctly. The investigators chose the 30 percent rate of wrong suggestions to measure the effect, not to mimic any product. A better model would make wrong suggestions rarer, and each one may then be harder to spot.

Three roles for a model

Software that analyzes clinical data plays one of three roles, and each role changes what has to be measured. An autonomous device gives the answer itself: IDx-DR's output is the screening decision, and no clinician interprets the images first, so the device's own accuracy is what reaches the patient. An assistive device gives a reader extra information, such as marks on suspected lesions, and the reader decides; the performance that matters is the reader's with the device compared with the reader's without it. A triage or notification device flags cases that may need urgent attention and works alongside the usual workflow. The stroke-triage program of Chapter 12, which the FDA authorized in February 2018, is one. The FDA's decision summary describes it as a "notification-only, parallel workflow tool". The program alerts a specialist to a suspected blocked artery, while "trained radiologists read all images per standard of care, regardless of the performance" of the device.

The role also decides where automation bias can do harm. An autonomous device has no human to bias, but no human to catch its errors either. A parallel triage tool is designed so that its misses do not remove the usual reading by a radiologist, although its alerts may still change which images are read first. An assistive device places its output directly in front of the reader at the moment of decision, which is where the mammography study showed the risk. Clinical decision support is software that gives clinicians information or recommendations. The FDA last revised its guidance on such software in January 2026. Its criteria address the same risk: software that a clinician cannot independently review, or that supports a time-critical decision where the clinician is likely to rely on it without review, is a device software function.

32 — Three roles for a model
AUTONOMOUS IDx-DR model refer / rescreen patient the model's error reaches the patient directly ASSISTIVE detection aid model marks the reader also sees the image reader reader's decision where automation bias acts measure the reader with and without the aid PARALLEL TRIAGE stroke alert, 2018 model alert to specialist radiologist reads every image per standard of care report a miss does not remove the usual read
CodeDataReadoutFocal detail
Plate 32 — An autonomous device's accuracy reaches the patient directly; an assistive device's value is the reader's performance with it; a parallel triage tool adds alerts without removing the usual read. Each role changes what must be measured.

Measuring the pair

For an assistive device, the evidence comes from a reader study. The FDA describes the usual design in its guidance on computer-assisted detection in radiology, which it last revised in September 2022. Many readers each read many cases, with and without the device. In a "fully-crossed" design, every reader reads every case in both conditions, and the guidance states that such a design "offers the greatest statistical power for a given number of cases."

Under the guidance, reading sessions are separated "by at least four weeks to avoid memory bias," so that readers do not remember cases from an earlier session. The primary measure is usually the area under the ROC curve for readers with the device against readers without it, and the number of readers must be justified rather than fixed by rule. The guidance states that case sets may be enriched with diseased cases, but that enrichment can change how readers behave and biases predictive values, the problem of Chapter 13 in another form.

The good machine-learning practice principles name the target directly. The 2021 version stated that the focus should be on the performance of the human-AI team, and the international version of 2025 states that a device should be assessed with a focus on human-AI interaction in the intended use environment. A model's standalone figures remain necessary, because they show what the model does on its own, without a reader. For an assistive device, however, they do not show what patients receive.

The pair can perform worse than either partner

A strong model and a capable reader can work badly together if the reader follows the model's errors and overrides its correct calls. Only a reader study can show whether the pair beats the reader alone, and its cases must include the model's failures at a realistic rate. Standalone accuracy cannot answer that question.

Telling users what they need to know

Clinicians can resist a wrong suggestion only if they know when the model is likely to be wrong. In June 2024 the FDA, Health Canada and the MHRA published guiding principles on transparency for machine-learning devices. The principles state that makers should share the device's intended use, performance and limitations with the right users, at the right time and in a form designed for those users, and should also share the logic behind its outputs where that can be done. IDx-DR's labeling states the patients, the camera and the conditions under which performance was shown. It also states that an "insufficient quality" result after dilation may itself be a sign of disease that needs referral. Chapter 17 reads that kind of information as an outside reader would, and asks what it leaves out.

Design the display and measure the team

Decide whether the device is autonomous, assistive or parallel, and design the display for that role. Show the model's confidence or the evidence behind a suggestion where that helps review, and avoid showing a suggestion before the reader has looked at the case independently, because a suggestion seen first can anchor the reader's judgment. Measure the reader with and without the device in a study whose cases include the model's errors at a realistic rate, and train users on the failure modes the study finds.

+ What this chapter established
  • In a 2023 mammography study, wrong suggestions labeled as AI cut readers' accuracy from about 80 percent to between 20 and 46 percent.
  • Autonomous, assistive and parallel triage devices place the model differently, and each role changes what has to be measured.
  • Reader studies measure the clinician with and without the device, ideally with every reader reading every case in both conditions and with reading sessions at least four weeks apart.
  • Users can resist a wrong output only if they know when the model is likely to be wrong, and the 2024 transparency principles ask makers to share the device's intended use, performance and limitations with them.