Section 524B, the AI Act and the deployer's duties.
+ The questionWhat do the cybersecurity and AI-specific rules add, what do they ask of the hospital that deploys the device, and how does one device meet two EU regulations at once?
Section 524B
On 29 December 2022 the US enacted an appropriations law, one of the laws that fund the federal government. Ninety days later, on 29 March 2023, a new section, 524B, took effect in the Food, Drug, and Cosmetic Act, the US law under which the FDA regulates devices. Section 524B applies to premarket submissions made on or after 29 March 2023 for any cyber device: one that includes software validated, installed or authorized by its maker, can connect to the internet, and has characteristics that could be vulnerable to cybersecurity threats.
A maker submitting such a device must do four things. It must provide a plan to monitor, identify and address vulnerabilities after release, including coordinated disclosure. It must design, develop and maintain processes giving reasonable assurance that the device and related systems are secure, and make updates and patches available on a regular cycle and, for critical vulnerabilities, out of cycle. It must provide a software bill of materials covering commercial, open-source and off-the-shelf components. And it must meet any further requirements the FDA sets by regulation.
The FDA gave makers a transition: until 1 October 2023 it worked with companies whose submissions lacked the new information, and after that date it could refuse to accept such submissions. The FDA revised its premarket cybersecurity guidance, the source of Chapters 22 to 24, on 3 February 2026. Two law and certification firms reviewed the revision and reported that it aligned the guidance with the new quality management system rule without adding technical expectations. In March 2026 a standards body announced that the FDA had recognized its consensus report on cybersecurity considerations unique to machine-learning devices.
One device, two EU regulations
In the EU, cybersecurity is part of the medical device regulation itself. The regulation's general safety and performance requirements state that software must be developed according to the state of the art, taking account of the development lifecycle, risk management including information security, verification and validation. They also require makers to set out the minimum requirements for hardware, IT networks and IT security measures, including protection against unauthorized access, needed to run the software as intended. Chapter 24 described the coordination group's cybersecurity guidance of 2019, revised in 2020, which states that a device should not rely on security controls in its operating environment. The EU's Cyber Resilience Act applies in full to most products with digital elements from 11 December 2027, and the duty of the makers of those products to report actively exploited vulnerabilities has applied since 11 September 2026, but the act excludes devices covered by the medical device and diagnostic regulations. For a hospital, that act therefore matters for the other software it buys, not for its medical devices.
For a device that uses AI, a second regulation applies as well. Under the AI Act, an AI system is high-risk if two conditions hold: it is a product covered by one of the EU laws listed in the act's first annex, or a safety component of such a product; and that law requires the product to be assessed by a third party. A safety component is a part that performs a safety function for a product, or whose failure endangers the health and safety of people or property. Medical devices are on that list, and almost all software that informs a clinical decision needs a notified body under Rule 11, so most medical device AI is high-risk.
The high-risk requirements cover risk management, data governance, technical documentation, automatic logging, information for those who deploy the system, human oversight, and accuracy, robustness and cybersecurity. The Digital Omnibus of July 2026 kept medical devices in that annex and moved the date from which these requirements apply to medical device AI to 2 August 2028. According to the consolidated text, the Omnibus also clarified that AI used solely for purposes other than safety does not qualify as a safety component.
The two regulations are applied in one assessment. A device that is also high-risk AI follows the conformity assessment of the medical device regulation, and within that assessment the notified body checks the AI Act's requirements. In June 2025 the device coordination group and the EU's AI board published joint guidance that describes one technical documentation, an integrated quality system and integrated risk management, rather than two parallel files.
Segmentation, firewalls and access control at the site are valuable layers of protection, and the EU's minimum IT requirements belong in the instructions for use. A device designed to be safe only behind those measures, though, depends for its security on an organization that did not design it and may change its network tomorrow.
The deployer's duties
The AI Act places duties on the organization that uses a high-risk system under its own authority, which the act calls the deployer. For an AI device used in a hospital, the deployer is the hospital. The deployer must use the system in accordance with its instructions and assign human oversight to people with the necessary competence, training and authority. Input data under its control must be relevant and sufficiently representative for the intended purpose. It must monitor the system's operation and suspend use where it sees a risk. It must inform the provider of risks without undue delay, and of serious incidents immediately.
Where the logs that the system generates automatically are under the deployer's control, it must keep them for at least six months. It must inform workers before the system is used in their workplace. The Omnibus also rewrote the act's article on AI literacy, the skills and understanding people need to use AI systems in an informed way and to be aware of their risks. Under the rewritten article, providers and deployers must take measures to support AI literacy, without having to guarantee any level of literacy in an individual.
The deployer's duties are the site practices of Chapters 18 to 20 made into law, almost line for line: integration within the specification, a local check that the inputs match the conditions of the evidence, monitoring with named people, and a channel back to the maker. For hospitals using AI that is a medical device, the deployer's duties set out above apply from 2 August 2028. The duty on AI literacy applies earlier: it sits in the act's first chapter, which has applied since 2 February 2025, and the Omnibus rewrote it without changing that date.
India
In its guidance of July 2026, India's regulator brings security into the same file as safety. Trade reports of the text state that the guidance asks makers to keep a bill of materials that includes an SBOM, covering third-party, open-source and commercial components, and to link it to the monitoring and mitigation of known vulnerabilities. By the same reports, the guidance treats cybersecurity as a risk area, asks that vulnerabilities be monitored after deployment, and asks that devices keep operating safely during incidents, including partial loss of connectivity, denial of service and loss of integrity. The guidance lists IEC 81001-5-1 among the standards that may apply, and Chapter 27 set out its rules on AI datasets.
Patient data are governed by the Digital Personal Data Protection Act of 2023 and its rules. The government formally published the rules in November 2025, and they take effect in phases. The rules on the data protection board applied at once. The registration of consent managers applies from 13 November 2026; consent managers are registered intermediaries through which people give, manage and withdraw their consent to the use of their data. Most obligations on those who process data, including security safeguards and breach notification, apply from 13 May 2027. The penalty for failing to keep reasonable security safeguards can reach 250 crore rupees, which is 2.5 billion rupees.
Counted from 8 October 2026, the next dates that apply to a software device or to the site that runs it are spread over two years. India's consent-manager rules apply in about one month, on 13 November 2026, and India's main data protection duties, including breach notification, apply in about seven months, on 13 May 2027. The Cyber Resilience Act applies in full on 11 December 2027, about 14 months away, to the hospital's software that is not a device. On 2 August 2028, about 22 months away, the AI Act's high-risk requirements for medical device AI apply, including the hospital's duties as deployer. A device authorized today and updated every quarter will go through about seven releases before the last of these dates, and each release will be assessed against the rules in force on its own release day.
For each device, list the rules that apply in each market and the date each takes effect: section 524B for cyber devices, the device regulation and the AI Act together in the EU, India's software guidance and data protection rules. Assign each duty to the maker or the site, write the site's duties into the responsibility agreement of Chapter 18, and review the list whenever a regulation changes its dates.
The four-hour operators, again
The operators of the opening scene had four hours of training and a locked program. What their trial showed in 2017 was narrow and solid: that program, on that camera, run by people like them, gave the right screening decision for adults like those enrolled often enough to clear performance thresholds agreed in advance. The trial could not show what the program would do in a clinic whose systems sent images differently, after a camera or model changed, or against someone trying to make it fail.
The same device filed in October 2026 would carry evidence for each of those gaps. Because its images and answers cross the internet, it is a cyber device, so its file would hold a threat model, an SBOM, security test results and a disclosure plan. Its software documentation would be at the level set by its potential for harm. Following the FDA's draft for AI, the maker would show test data drawn from several sites and kept from the developers, with results by subgroup and by camera, and would very likely include a change control plan for the retraining the model will need. In India the maker would be expected to add validation in Indian populations. In the EU, from August 2028, the notified body would assess the device's logging and human-oversight measures. The hospital that installed the device would test its interfaces, check its inputs against the specification, run it silently on local patients before relying on it, watch its positive and declined rates, and tell the maker when something changed.
None of that changes the fact that has shaped every chapter. Every way the software can fail was built in on the day it shipped or arrived with an update, and for the model the training data is part of what was built. The evidence that the software will be right for the next patient is a measurement of what was built, taken on patients chosen to resemble the next one. What keeps that evidence true after an update is the discipline of measuring again. Four hours were enough to take the pictures. Everything else in this guide is what it takes to keep trusting the answer.
- Since March 2023, a maker that submits a cyber device to the FDA must file a vulnerability plan, show secure processes with patching, and provide an SBOM.
- In the EU, a medical device that is high-risk AI goes through one assessment under both the device regulation and the AI Act, and the AI duties apply from 2 August 2028.
- Hospitals that deploy such AI become deployers under the AI Act, with duties for oversight, input data, monitoring and logs that mirror good site practice.
- Under India's 2026 guidance, SBOMs and cyber resilience belong in the device file, and India's data protection duties phase in through May 2027.