Lessons.
The chapters reduce to nine working rules, one for each part of the book, and each traces back to the chapters named at its end.
- Judge a device in its use, not a material on its own. The same plastic meets the body differently in a bottle, a breathing hose and a blood bag, because the contact, the route, the duration and the patient differ (Chapters 1, 2 and 11).
- Evaluate what a material holds and what it becomes. In nearly all cases additives are not bound to the polymer, so they diffuse out. Water, oxygen and ozone cut the polymer's chains into molecules that the recipe never held, and heat speeds each cut (Chapters 3 to 5).
- Take the test article from the finished device as made, sterilized, aged and cleaned. Molding, sterilization, packaging and cleaning each change what a material releases, and, on this book's reading, accelerated aging of a packed device models neither ozone nor cleaning (Chapters 6 and 7).
- Expect a response wherever blood or tissue fluid wets a surface. Cells meet the protein coat that forms on the surface, not the material, so even a material that releases nothing draws a response (Chapters 8 to 10).
- Set the reporting line from the daily threshold before the analysis starts. The line decides which compounds are listed at all. Each listed compound becomes a dose set against its tolerable intake or threshold, and tests answer the effects that no threshold covers (Chapters 12 to 18).
- Test the gas that aged, assembled devices deliver. What leaves a part that never touches the patient can still reach the lungs, and the recalled foam's breakdown was found in use (Chapters 19 and 20).
- Keep the evaluation traceable in the design file, and reopen it when a change can alter what reaches the patient. A change of material, formulation, supplier or processing reopens the evaluation wherever the changed part reaches the patient, directly or through the gas, even within specification. A number that cannot be recomputed from the file is a broken link (Chapters 21 to 24).
- Choose a laboratory against the plan, and check its recognitions on the day. ISO/IEC 17025 accreditation shows a laboratory's competence for a method, GLP governs how a nonclinical safety study is planned, run, recorded and archived, and neither says the device is safe (Chapter 25).
- Build one evaluation, and cite the edition each jurisdiction accepts, with its date. As of November 2026, FDA recognizes ISO 10993-1:2025 in part and the EU cites it, while India's standard for the evaluation is the 2018 edition as modified in 2023 (Chapters 26 to 28).